TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Larue, H. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - A PET system based on data processing of free-running sampled pulses T2 - 2001 IEEE Nuclear Science Symposium Conference Record, Vol. 2 N2 - Within the developments for the Crystal Clear small animal PET project (CLEARPET) a dual head PET system has been established. The basic principle is the early digitization of the detector pulses by free running ADCs. The determination of the γ-energy and also the coincidence detection is performed by data processing of the sampled pulses on the host computer. Therefore a time mark is attached to each pulse identifying the current cycle of the 40 MHz sampling clock. In order to refine the time resolution the pulse starting time is interpolated from the samples of the pulse rise. The detector heads consist of multichannel PMTs with a single LSO scintillator crystal coupled to each channel. For each PMT only one ADC is required. The position of an event is obtained separately from trigger signals generated for each single channel. An FPGA is utilized for pulse buffering, generation of the time mark and for the data transfer to the host via a fast I/O-interface. Y1 - 2002 UR - https://opus.bibliothek.fh-aachen.de/opus4/frontdoor/index/index/docId/5059 SN - 1082-3654 SP - 693 EP - 694 ER -