TY - JOUR A1 - Welden, Rene A1 - Jablonski, Melanie A1 - Wege, Christina A1 - Keusgen, Michael A1 - Wagner, Patrick Hermann A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Light-Addressable Actuator-Sensor Platform for Monitoring and Manipulation of pH Gradients in Microfluidics: A Case Study with the Enzyme Penicillinase JF - Biosensors N2 - The feasibility of light-addressed detection and manipulation of pH gradients inside an electrochemical microfluidic cell was studied. Local pH changes, induced by a light-addressable electrode (LAE), were detected using a light-addressable potentiometric sensor (LAPS) with different measurement modes representing an actuator-sensor system. Biosensor functionality was examined depending on locally induced pH gradients with the help of the model enzyme penicillinase, which had been immobilized in the microfluidic channel. The surface morphology of the LAE and enzyme-functionalized LAPS was studied by scanning electron microscopy. Furthermore, the penicillin sensitivity of the LAPS inside the microfluidic channel was determined with regard to the analyte’s pH influence on the enzymatic reaction rate. In a final experiment, the LAE-controlled pH inhibition of the enzyme activity was monitored by the LAPS. KW - microfluidics KW - enzyme kinetics KW - actuator-sensor system KW - light-addressable electrode KW - light-addressable potentiometric sensor Y1 - 2021 U6 - http://dx.doi.org/10.3390/bios11060171 SN - 2079-6374 N1 - This article belongs to the Special Issue "Selected Papers from the 1st International Electronic Conference on Biosensors (IECB 2020)" VL - 11 IS - 6 PB - MDPI CY - Basel ER - TY - JOUR A1 - Welden, Rene A1 - Nagamine Komesu, Cindy A. A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef A1 - Wagner, Torsten T1 - Photoelectrochemical enzymatic penicillin biosensor: A proof-of-concept experiment JF - Electrochemical Science Advances N2 - Photoelectrochemical (PEC) biosensors are a rather novel type of biosensors thatutilizelighttoprovideinformationaboutthecompositionofananalyte,enablinglight-controlled multi-analyte measurements. For enzymatic PEC biosensors,amperometric detection principles are already known in the literature. In con-trast, there is only a little information on H+-ion sensitive PEC biosensors. Inthis work, we demonstrate the detection of H+ions emerged by H+-generatingenzymes, exemplarily demonstrated with penicillinase as a model enzyme on atitanium dioxide photoanode. First, we describe the pH sensitivity of the sensorand study possible photoelectrocatalytic reactions with penicillin. Second, weshow the enzymatic PEC detection of penicillin. KW - enzymatic biosensor KW - penicillin KW - penicillinase KW - photoelectrochemistry KW - titanium dioxide photoanode Y1 - 2021 U6 - http://dx.doi.org/10.1002/elsa.202100131 SN - 2698-5977 N1 - Corresponding auhtor: Michael J. Schöning VL - 2 IS - 4 SP - 1 EP - 5 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Poghossian, Arshak A1 - Welden, Rene A1 - Buniatyan, Vahe V. A1 - Schöning, Michael Josef T1 - An Array of On-Chip Integrated, Individually Addressable Capacitive Field-Effect Sensors with Control Gate: Design and Modelling JF - Sensors N2 - The on-chip integration of multiple biochemical sensors based on field-effect electrolyte-insulator-semiconductor capacitors (EISCAP) is challenging due to technological difficulties in realization of electrically isolated EISCAPs on the same Si chip. In this work, we present a new simple design for an array of on-chip integrated, individually electrically addressable EISCAPs with an additional control gate (CG-EISCAP). The existence of the CG enables an addressable activation or deactivation of on-chip integrated individual CG-EISCAPs by simple electrical switching the CG of each sensor in various setups, and makes the new design capable for multianalyte detection without cross-talk effects between the sensors in the array. The new designed CG-EISCAP chip was modelled in so-called floating/short-circuited and floating/capacitively-coupled setups, and the corresponding electrical equivalent circuits were developed. In addition, the capacitance-voltage curves of the CG-EISCAP chip in different setups were simulated and compared with that of a single EISCAP sensor. Moreover, the sensitivity of the CG-EISCAP chip to surface potential changes induced by biochemical reactions was simulated and an impact of different parameters, such as gate voltage, insulator thickness and doping concentration in Si, on the sensitivity has been discussed. KW - equivalent circuit KW - multianalyte detection KW - control gate KW - on-chip integrated addressable EISCAP sensors KW - capacitive field-effect sensor Y1 - 2021 U6 - http://dx.doi.org/10.3390/s21186161 SN - 1424-8220 N1 - This article belongs to the Special Issue "Field-Effect Sensors: From pH Sensing to Biosensing" VL - 21 IS - 18 SP - 17 PB - MDPI CY - Basel ER -