Feasibility of T₂-weighted turbo spin echo imaging of the human prostate at 7 tesla

  • Purpose To demonstrate that high quality T₂-weighted (T2w) turbo spin-echo (TSE) imaging of the complete prostate can be achieved routinely and within safety limits at 7 T, using an external transceive body array coil only. Methods Nine healthy volunteers and 12 prostate cancer patients were scanned on a 7 T whole-body system. Preparation consisted of B₀ and radiofrequency shimming and localized flip angle calibration. T₁ and T₂ relaxation times were measured and used to define the T2w-TSE protocol. T2w imaging was performed using a TSE sequence (pulse repetition time/echo time 3000–3640/71 ms) with prolonged excitation and refocusing pulses to reduce specific absorption rate. Results High quality T2w TSE imaging was performed in less than 2 min in all subjects. Tumors of patients with gold-standard tumor localization (MR-guided biopsy or prostatectomy) were well visualized on 7 T imaging (n = 3). The number of consecutive slices achievable within a 10-g averaged specific absorption rate limit of 10 W/kg was ≥28 in all subjects, sufficient for full prostate coverage with 3-mm slices in at least one direction. Conclusion High quality T2w TSE prostate imaging can be performed routinely and within specific absorption rate limits at 7 T with an external transceive body array.

Export metadata

Additional Services

Share in Twitter Search Google Scholar
Metadaten
Author:Marnix C. Maas, Eline K. Vos, Miriam W. Lagemaat, Andreas Bitz, Stephan Orzada, Thiele Kobus, Oliver Kraff, Stefan Maderwald, Mark E. Ladd, Tom W. J. Scheenen
DOI:https://doi.org/10.1002/mrm.24818
ISSN:1522-2594
Parent Title (English):Magnetic Resonance in Medicine
Publisher:Wiley-VCH
Place of publication:Weinheim
Document Type:Article
Language:English
Year of Completion:2014
Date of the Publication (Server):2017/04/03
Volume:71
Issue:5
First Page:1711
Last Page:1719
Link:https://doi.org/10.1002/mrm.24818
Zugriffsart:campus
Institutes:FH Aachen / Fachbereich Elektrotechnik und Informationstechnik
collections:Verlag / Wiley-VCH