TY - JOUR A1 - Pourshahidi, Ali Mohammad A1 - Achtsnicht, Stefan A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim A2 - Offenhäusser, Andreas T1 - Frequency Mixing Magnetic Detection Setup Employing Permanent Ring Magnets as a Static Offset Field Source T2 - Sensors N2 - Frequency mixing magnetic detection (FMMD) has been explored for its applications in fields of magnetic biosensing, multiplex detection of magnetic nanoparticles (MNP) and the determination of core size distribution of MNP samples. Such applications rely on the application of a static offset magnetic field, which is generated traditionally with an electromagnet. Such a setup requires a current source, as well as passive or active cooling strategies, which directly sets a limitation based on the portability aspect that is desired for point of care (POC) monitoring applications. In this work, a measurement head is introduced that involves the utilization of two ring-shaped permanent magnets to generate a static offset magnetic field. A steel cylinder in the ring bores homogenizes the field. By variation of the distance between the ring magnets and of the thickness of the steel cylinder, the magnitude of the magnetic field at the sample position can be adjusted. Furthermore, the measurement setup is compared to the electromagnet offset module based on measured signals and temperature behavior. KW - magnetic sensors KW - biosensors KW - frequency mixing magnetic detection KW - magnetic nanoparticles Y1 - 2022 UR - https://opus.bibliothek.fh-aachen.de/opus4/frontdoor/index/index/docId/10457 SN - 1424-8220 VL - 22 IS - 22 PB - MDPI CY - Basel ER -