@misc{GraesslRenzHezeletal.2012, author = {Gr{\"a}ßl, Andreas and Renz, Wolfgang and Hezel, Fabian and Frauenrath, Tobias and Pfeiffer, Harald and Hoffmann, Werner and Kellmann, Peter and Martin, Conrad and Niendorf, Thoralf}, title = {Design, evaluation and application of a modular 32 channel transmit/receive surface coil array for cardiac MRI at 7T}, series = {2012 ISMRM Annual Meeting Proceedings}, journal = {2012 ISMRM Annual Meeting Proceedings}, issn = {1545-4428}, year = {2012}, abstract = {Cardiac MR (CMR) at ultrahigh (≥7.0 T) fields is regarded as one of the most challenging MRI applications. At 7.0 T image quality is not always exclusively defined by signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Detrimental effects bear the potential to spoil the signal-to-noise (SNR) and contrast-to-noise (CNR) benefits of cardiac MR (CMR) at 7.0 T. B₁⁺-inhomogeneities and signal voids represent the main challenges. Various pioneering coil concepts have been proposed to tackle these issues, enabling cardiac MRI at 7.0 T. This includes a trend towards an ever larger number of transmit and receive channels. This approach affords multi-dimensional B₁⁺ modulations to improve B₁⁺ shimming performance and to enhance RF efficiency. Also, parallel imaging benefits from a high number of receive channels enabling two-dimensional acceleration. Realizing the limitations of existing coil designs tailored for UHF CMR and recognizing the opportunities of a many element TX/RX channel architecture this work proposes a modular, two dimensional 32-channel transmit and receive array using loop elements and examines its efficacy for enhanced B¹+ homogeneity and improved parallel imaging performance.}, language = {en} } @article{GraesslRenzHezeletal.2013, author = {Gr{\"a}ßl, Andreas and Renz, Wolfgang and Hezel, Fabian and Dieringer, Matthias A. and Winter, Lukas and {\"O}zerdem, Celal and Rieger, Jan and Kellmann, Peter and Santoro, Davide and Lindel, Tomasz D. and Frauenrath, Tobias and Pfeiffer, Harald and Niendorf, Thoralf}, title = {Modular 32-channel transceiver coil array for cardiac MRI at 7.0T}, series = {Magnetic Resonance in Medicine}, volume = {72}, journal = {Magnetic Resonance in Medicine}, number = {1}, publisher = {Wiley-Liss}, address = {New York}, issn = {1522-2594}, doi = {10.1002/mrm.24903}, pages = {276 -- 290}, year = {2013}, abstract = {Purpose To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. Methods The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Results Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. Conclusion The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T.}, language = {en} }