@misc{LindelGreiserWaxmanetal.2012, author = {Lindel, Tomasz Dawid and Greiser, Andreas and Waxman, Patrick and Dietterle, Martin and Seifert, Frank and Fontius, Ulrich and Renz, Wolfgang and Dieringer, Matthias A. and Frauenrath, Tobias and Schulz-Menger, Jeanette and Niendorf, Thoralf and Ittermann, Bernd}, title = {Cardiac CINE MRI at 7 T using a transmit array}, series = {2012 ISMRM Annual Meeting Proceedings}, journal = {2012 ISMRM Annual Meeting Proceedings}, issn = {1545-4428}, year = {2012}, abstract = {With its need for high SNR and short acquisition times, Cardiac MRI (CMR) is an intriguing target application for ultrahigh field MRI. Due to the sheer size of the upper torso, however, the known RF issues of 7T MRI are also most prominent in CMR. Recent years brought substantial progress but the full potential of the ultrahigh field for CMR is yet to be exploited. Parallel transmission (pTx) is a promising approach in this context and several groups have already reported B1 shimming for 7T CMR. In such a static pTx application amplitudes and phases of all Tx channels are adjusted individually but otherwise imaging techniques established in current clinical practice 1.5 T and 3 T are applied. More advanced forms of pTx as spatially selective excitation (SSE) using Transmit SENSE promise additional benefits like faster imaging with reduced fields of view or improved SAR control. SSE requires the full dynamic capabilities of pTx, however, and for the majority of today's implemented pTx hardware the internal synchronization of the Tx array does not easily permit external triggering as needed for CMR. Here we report a software solution to this problem and demonstrate the feasibility of CINE CMR at 7 T using a Tx array.}, language = {en} }