@article{WegmannHacklStaatetal.2016, author = {Wegmann, Kilian and Hackl, Michael and Staat, Manfred and Mayer, Katharina and M{\"u}ller, Lars-Peter}, title = {Double plate osteosynthesis of proximal ulna fractures: biomechanical and clinical results}, series = {Knee surgery, sports traumatology, arthroscopy}, volume = {Volume 24}, journal = {Knee surgery, sports traumatology, arthroscopy}, number = {Supplement 1}, publisher = {Springer}, address = {Berlin}, issn = {0942-2056}, doi = {10.1007/s00167-016-4079-0}, pages = {58 -- 59}, year = {2016}, abstract = {While plate fixation of proximal ulna fractures might lead to superior clinical results compared to tension band wiring, regular plates represent an established risk factor for wound complications. The olecranon double plates (Medartis, Basel, CH) might decrease complications related to the osteosynthesis because of their low profile and better anatomical fit. This study aimed to evaluate the biomechanical performance and clinical results of the olecranon double plates.}, language = {en} } @article{PookhalilAmoabedinyTabeshetal.2016, author = {Pookhalil, Ali and Amoabediny, Ghassem and Tabesh, Hadi and Behbahani, Mehdi and Mottaghy, Khosrow}, title = {A new approach for semiempirical modeling of mechanical blood trauma}, series = {The international journal of artificial organs}, volume = {39}, journal = {The international journal of artificial organs}, number = {4}, publisher = {Sage}, address = {London}, issn = {1724-6040}, doi = {10.5301/ijao.5000474}, pages = {171 -- 177}, year = {2016}, abstract = {Purpose Two semi-empirical models were recently published, both making use of existing literature data, but each taking into account different physical phenomena that trigger hemolysis. In the first model, hemoglobin (Hb) release is described as a permeation procedure across the membrane, assuming a shear stress-dependent process (sublethal model). The second model only accounts for hemoglobin release that is caused by cell membrane breakdown, which occurs when red blood cells (RBC) undergo mechanically induced shearing for a period longer than the threshold time (nonuniform threshold model). In this paper, we introduce a model that considers the hemolysis generated by both these possible phenomena. Methods Since hemolysis can possibly be caused by permeation of hemoglobin through the RBC functional membrane as well as by release of hemoglobin from RBC membrane breakdown, our proposed model combines both these models. An experimental setup consisting of a Couette device was utilized for validation of our proposed model. Results A comparison is presented between the damage index (DI) predicted by the proposed model vs. the sublethal model vs. the nonthreshold model and experimental datasets. This comparison covers a wide range of shear stress for both human and porcine blood. An appropriate agreement between the measured DI and the DI predicted by the present model was obtained. Conclusions The semiempirical hemolysis model introduced in this paper aims for significantly enhanced conformity with experimental data. Two phenomenological outcomes become possible with the proposed approach: an estimation of the average time after which cell membrane breakdown occurs under the applied conditions, and a prediction of the ratio between the phenomena involved in hemolysis.}, language = {en} } @inproceedings{BhattaraiStaat2016, author = {Bhattarai, Aroj and Staat, Manfred}, title = {Female pelvic floor dysfunction: progress weakening of the support system}, series = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, booktitle = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, editor = {Erni, Daniel}, publisher = {Universit{\"a}t Duisburg-Essen}, address = {Duisburg}, organization = {MedTech Symposium}, doi = {10.17185/duepublico/40821}, pages = {11 -- 12}, year = {2016}, abstract = {The structure of the female pelvic floor (PF) is an inter-related system of bony pelvis,muscles, pelvic organs, fascias, ligaments, and nerves with multiple functions. Mechanically, thepelvic organ support system are of two types: (I) supporting system of the levator ani (LA) muscle,and (II) the suspension system of the endopelvic fascia condensation [1], [2]. Significantdenervation injury to the pelvic musculature, depolimerization of the collagen fibrils of the softvaginal hammock, cervical ring and ligaments during pregnancy and vaginal delivery weakens thenormal functions of the pelvic floor. Pelvic organ prolapse, incontinence, sexual dysfunction aresome of the dysfunctions which increases progressively with age and menopause due toweakened support system according to the Integral theory [3]. An improved 3D finite elementmodel of the female pelvic floor as shown in Fig. 1 is constructed that: (I) considers the realisticsupport of the organs to the pelvic side walls, (II) employs the improvement of our previous FEmodel [4], [5] along with the patient based geometries, (III) incorporates the realistic anatomy andboundary conditions of the endopelvic (pubocervical and rectovaginal) fascia, and (IV) considersvarying stiffness of the endopelvic fascia in the craniocaudal direction [3]. Several computationsare carried out on the presented computational model with healthy and damaged supportingtissues, and comparisons are made to understand the physiopathology of the female PF disorders.}, language = {en} } @inproceedings{JungStaat2016, author = {Jung, Alexander and Staat, Manfred}, title = {Computing olympic gold: Ski jumping as an example}, series = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, booktitle = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, editor = {Erni, Daniel}, publisher = {Universit{\"a}t Duisburg-Essen}, address = {Duisburg}, organization = {MedTech Symposium}, isbn = {978-3-940402-06-6}, doi = {10.17185/duepublico/40821}, pages = {54 -- 55}, year = {2016}, language = {en} } @inproceedings{KahmannHacklWegmannetal.2016, author = {Kahmann, Stephanie and Hackl, Michael and Wegmann, Kilian and M{\"u}ller, Lars-Peter and Staat, Manfred}, title = {Impact of a proximal radial shortening osteotomy on the distribution of forces and the stability of the elbow}, series = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, booktitle = {1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen}, editor = {Erni, Daniel}, publisher = {Universit{\"a}t Duisburg-Essen}, address = {Duisburg}, organization = {MedTech Symposium}, doi = {10.17185/duepublico/40821}, pages = {7 -- 8}, year = {2016}, abstract = {The human arm consists of the humerus (upper arm), the medial ulna and the lateral radius (forearm). The joint between the humerus and the ulna is called humeroulnar joint and the joint between the humerus and the radius is called humeroradial joint. Lateral and medial collateral ligaments stabilize the elbow. Statistically, 2.5 out of 10,000 people suffer from radial head fractures [1]. In these fractures the cartilage is often affected. Caused by the injured cartilage, degenerative diseases like posttraumatic arthrosis may occur. The resulting pain and reduced range of motion have an impact on the patient's quality of life. Until now, there has not been a treatment which allows typical loads in daily life activities and offers good long-term results. A new surgical approach was developed with the motivation to reduce the progress of the posttraumatic arthrosis. Here, the radius is shortened by 3 mm in the proximal part [2]. By this means, the load of the radius is intended to be reduced due to a load shift to the ulna. Since the radius is the most important stabilizer of the elbow it has to be confirmed that the stability is not affected. In the first test (Fig. 1 left), pressure distributions within the humeroulnar and humeroradial joints a native and a shortened radius were measured using resistive pressure sensors (I5076 and I5027, Tekscan, USA). The humerus was loaded axially in a tension testing machine (Z010, Zwick Roell, Germany) in 50 N steps up to 400 N. From the humerus the load is transmitted through both the radius and the ulna into the hand which is fixed on the ground. In the second test (Fig. 1 right), the joint stability was investigated using a digital image correlation system to measure the displacement of the ulna. Here, the humerus is fixed with a desired flexion angle and the unconstrained forearm lies on the ground. A rope connects the load actuator with a hook fixed in the ulna. A guide roller is used so that the rope pulls the ulna horizontally when a tensile load is applied. This creates a moment about the elbow joint with a maximum value of 7.5 Nm. Measurements were performed with varying flexion angles (0°, 30°, 60°, 90°, 120°). For both tests and each measurement, seven specimens were used. Student's t-test was employed to determine whether the mean values of the measurements in native specimen and operated specimens differ significantly.}, language = {en} } @book{IomdinaBauerKotliar2015, author = {Iomdina, E. N. and Bauer, S. M. and Kotliar, Konstantin}, title = {БИОМЕХАНИКА ГЛАЗА: ТЕОРЕТИЧЕСКИЕ АСПЕКТЫ И КЛИНИЧЕСКИЕ ПРИЛОЖЕНИЯ}, publisher = {Real Time}, address = {Moscow}, isbn = {978-5-903025-57-2}, pages = {208 Seiten}, year = {2015}, language = {ru} } @inproceedings{StaatDuong2016, author = {Staat, Manfred and Duong, Minh Tuan}, title = {Smoothed Finite Element Methods for Nonlinear Solid Mechanics Problems: 2D and 3D Case Studies}, series = {Proceedings of the National Science and Technology Conference on Mechanical - Transportation Engineering (NSCMET 2016), 13th October 2016, Hanoi, Vietnam, Vol.2}, booktitle = {Proceedings of the National Science and Technology Conference on Mechanical - Transportation Engineering (NSCMET 2016), 13th October 2016, Hanoi, Vietnam, Vol.2}, pages = {440 -- 445}, year = {2016}, abstract = {The Smoothed Finite Element Method (SFEM) is presented as an edge-based and a facebased techniques for 2D and 3D boundary value problems, respectively. SFEMs avoid shortcomings of the standard Finite Element Method (FEM) with lower order elements such as overly stiff behavior, poor stress solution, and locking effects. Based on the idea of averaging spatially the standard strain field of the FEM over so-called smoothing domains SFEM calculates the stiffness matrix for the same number of degrees of freedom (DOFs) as those of the FEM. However, the SFEMs significantly improve accuracy and convergence even for distorted meshes and/or nearly incompressible materials. Numerical results of the SFEMs for a cardiac tissue membrane (thin plate inflation) and an artery (tension of 3D tube) show clearly their advantageous properties in improving accuracy particularly for the distorted meshes and avoiding shear locking effects.}, language = {en} } @inproceedings{TranTranMatthiesetal.2016, author = {Tran, Ngoc Trinh and Tran, Thanh Ngoc and Matthies, Hermann G. and Stavroulakis, Georgios Eleftherios and Staat, Manfred}, title = {FEM Shakedown of uncertain structures by chance constrained programming}, series = {PAMM Proceedings in Applied Mathematics and Mechanics}, volume = {16}, booktitle = {PAMM Proceedings in Applied Mathematics and Mechanics}, number = {1}, issn = {1617-7061}, doi = {10.1002/pamm.201610346}, pages = {715 -- 716}, year = {2016}, language = {en} } @article{LeversStaatLaack2016, author = {Levers, A. and Staat, Manfred and Laack, Walter van}, title = {Analysis of the long-term effect of the MBST® nuclear magnetic resonance therapy on gonarthrosis}, series = {Orthopedic Practice}, volume = {47}, journal = {Orthopedic Practice}, number = {11}, pages = {521 -- 528}, year = {2016}, language = {en} } @article{NiedermeyerZhouDursunetal.2016, author = {Niedermeyer, Angela and Zhou, Bei and Dursun, G{\"o}zde and Temiz Artmann, Ayseg{\"u}l and Markert, Bernd}, title = {An examination of tissue engineered scaffolds in a bioreactor}, series = {Proceedings in Applied Mathematics and Mechanics PAMM}, volume = {16}, journal = {Proceedings in Applied Mathematics and Mechanics PAMM}, number = {1}, publisher = {Wiley-VCH}, address = {Weinheim}, issn = {1617-7061}, doi = {10.1002/pamm.201610038}, pages = {99 -- 100}, year = {2016}, abstract = {Replacement tissues, designed to fill in articular cartilage defects, should exhibit the same properties as the native material. The aim of this study is to foster the understanding of, firstly, the mechanical behavior of the material itself and, secondly, the influence of cultivation parameters on cell seeded implants as well as on cell migration into acellular implants. In this study, acellular cartilage replacement material is theoretically, numerically and experimentally investigated regarding its viscoelastic properties, where a phenomenological model for practical applications is developed. Furthermore, remodeling and cell migration are investigated.}, language = {en} }