@article{StreunChristHellendungetal.2005, author = {Streun, M. and Christ, D. and Hellendung, A. and Larue, H. and Ziemons, Karl and Halling, H.}, title = {Effects of crosstalk and gain nonuniformity using multichannel PMTs in the Clearpet® scanner}, series = {Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment}, volume = {537}, journal = {Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment}, number = {1-2}, isbn = {0168-9002}, pages = {402 -- 405}, year = {2005}, abstract = {The ClearPET® scanners developed by the Crystal Clear Collaboration use multichannel PMTs as photodetectors with scintillator pixels coupled individually to each channel. In order to localize an event each channel anode is connected to a comparator that triggers when the anode signal exceeds a common predefined threshold. Two major difficulties here are crosstalk of light and the gain nonuniformity of the PMT channels. Crosstalk can generate false triggering in channels adjacent to the actual event. On the one hand this can be suppressed by sufficiently increasing the threshold, but on the other hand a threshold too high can already prevent valid events on the lower gain channels from being detected. Finally, both effects restrict the dynamic range of pulse heights that can be processed. The requirements to the dynamic range are not low as the ClearPET® scanners detect the depth of interaction by phoswich pixels consisting of LSO and Lu0.7Y0.3AP, two scintillators with different light yields. We will present a model to estimate the achievable dynamic range and show solutions to increase it.}, language = {en} } @article{ChristHollendungLarueetal.2004, author = {Christ, D. and Hollendung, A. and Larue, H. and Parl, C. and Streun, M. and Weber, S. and Ziemons, Karl and Halling, H.}, title = {Homogenization of the MultiChannel PM gain by inserting light attenuating masks}, series = {2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4}, journal = {2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4}, issn = {1082-3654}, pages = {2382 -- 2385}, year = {2004}, abstract = {MultiChannel Photomultipliers (PM), like the R7600-00-M64 or R5900-00-M64 from Hamamatsu, are often chosen as photodetectors in high-resolution positron emission tomography (PET). A major problem of this PM is the nonuniform channel gain. In order to solve this problem, light attenuating masks were created. The aim of the masks is a homogenization of the output of all 64 channels using different hole sizes at the channel positions. The hole area, which is individually defined for the different channels, is inversely proportional to the channel gain. The measurements by inserting light attenuating masks improved a homogenization to a ratio of 1:1.2.}, language = {en} }