TY - JOUR A1 - Ziemons, Karl A1 - Auffray, E. A1 - Barbier, R. A1 - Brandenburg, G. A1 - Bruyndonckx, P. T1 - The ClearPET™ project: Development of a 2nd generation high-performance small animal PET scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - Second generation high-performance PET scanners, called ClearPET™1, have been developed by working groups of the Crystal Clear Collaboration (CCC). High sensitivity and high spatial resolution for the ClearPET camera is achieved by using a phoswich arrangement combining two different types of lutetium-based scintillator materials: LSO from CTI and LuYAP:Ce from the CCC (ISTC project). In a first ClearPET prototype, phoswich arrangements of 8×8 crystals of 2×2×10 mm3 are coupled to multi-channel photomultiplier tubes (Hamamatsu R7600). A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the ring is 120 mm, the axial detector length is 110 mm.The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the exact pulse starting time, which is subsequently used for coincidence detection. The gantry allows rotation of the detector modules around the field of view. Preliminary data shows a correct identification of the crystal layer about (98±1)%. Typically the energy resolution is (23.3±0.5)% for the luyap layer and (15.4±0.4)% for the lso layer. early studies showed the timing resolution of 2 ns FWHM and 4.8 ns FWTM. the intrinsic spatial resolution ranges from 1.37 mm to 1.61 mm full-width of half-maximum (FWHM) with a mean of 1.48 mm FWHM. further improvements in image and energy resolution are expected when the system geometry is fully modeled. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 307 EP - 311 ER - TY - JOUR A1 - Yoshinobu, T. A1 - Iwasaki, H. A1 - Ui, Y. A1 - Furuichi, K. A1 - Ermelenko, Y. A1 - Mourzina, Y. A1 - Wagner, Torsten A1 - Näther, Niko A1 - Schöning, Michael Josef T1 - The light-addressable potentiometric sensor for multi-ion sensing and imaging JF - Methods. 37 (2005), H. 1 Y1 - 2005 SN - 1046-2023 SP - 99 EP - 102 ER - TY - JOUR A1 - Wagner, Torsten A1 - Yoshinobu, T. A1 - Rao, C. A1 - Otto, R. A1 - Schöning, Michael Josef T1 - „All-in-one“ solid-state device based on a lightaddressable potentiometric sensor platform JF - Transducers '05 : the 13th International Conference on Solid-State Sensors, Actuators and Microsystems ; Seoul, Korea, [June 5 - 9, 2005] ; digest of technical papers / sponsored by Korean Sensors Society. Technical co-sponsors: IEEE Electron Devices Society Y1 - 2005 SN - 0-7803-83995-6 N1 - International Conference on Solid State Sensors, Actuators and Microsystems ; (13, 2005, Seoul) SP - 1872 EP - 1875 PB - IEEE Operations Center CY - Piscataway, NJ ER - TY - JOUR A1 - Temiz Artmann, Aysegül A1 - Saklamaz, Ali A1 - Comlekci, Abdurrahman A1 - Caliskan, Sezer T1 - The beneficial effects of lipid-lowering drugs beyond lipid-lowering effects: A comparative study with pravastatin, atorvastatin, and fenofibrate in patients with type IIa and type IIb hyperlipidemia / Saklamaz, Ali ; Comlekci, Abdurrahman ; Temiz, Aysegu JF - Metabolism. 54 (2005), H. 5 Y1 - 2005 SN - 0026-0495 SP - 677 EP - 681 ER - TY - JOUR A1 - Temiz Artmann, Aysegül A1 - Resmi, Halil A1 - Akhunlar, Hülya A1 - Güner, Gül T1 - In vitro effects of high glucose concentrations on membrane protein sulfhydryl oxidation, G-actin and deformability of human erythrocytes. Resmi, Halil ; Akhunlar, Hülya ; Temiz Artmann, Aysegül ; Güner, Gül JF - Cell biochemistry and function. 23 (2005), H. 3 Y1 - 2005 SN - 0263-6484 SP - 163 EP - 168 ER - TY - JOUR A1 - Temiz Artmann, Aysegül A1 - Linder, Peter A1 - Kayser, Peter A1 - Digel, Ilya T1 - NMR in vitro effects on proliferation, apoptosis, and viability of human chondrocytes and osteoblasts JF - Methods and findings in Experimental and Clinical Pharmacology. 27 (2005), H. 6 Y1 - 2005 SN - 0379-0355 SP - 391 EP - 394 ER - TY - JOUR A1 - Temiz Artmann, Aysegül A1 - Kayatekin, Muammer B. A1 - Uysal, N. A1 - Resmi, H. T1 - Does antioxidant supplementation alter the effects of acute exercise on erythrocyte aggregation, deformability and endothelium adhesion in untrained rats? Kayatekin, Muammer B.; Uysal N.; Resmi, H.; Bediz, Seref C.; Temiz Artmann, A.; Genç, S.; Tugyan, K. JF - International Journal for Vitamin and Nutrition Research. 75 (2005), H. 4 Y1 - 2005 SN - 0300-9831 SP - 243 EP - 250 ER - TY - JOUR A1 - Temiz Artmann, Aysegül A1 - Akhisaroglu, M. A1 - Sercan, Z. A1 - Kayatekin, B. M. T1 - Adhesion of Erythrocytes to Endothelial Cells After Acute Exercise: Differences in Red Blood Cells from Juvenile and Adult Rats. Temiz Artmann, A.; Akhisaroglu, M.; Sercan, Z.; Kayatekin, BM.; Yorukoglu, K.; Kirkali, G. JF - Physiological Research (2005) Y1 - 2005 SN - 0862-8408 N1 - pre-press article ER - TY - JOUR A1 - Stulpe, Werner A1 - Bjelakovic, Igor T1 - The Projective Hilbert Space as a Classical Phase Space for Nonrelativistic Quantum Dynamics. Bjelakovic, Igor; Stulpe, Werner JF - International Journal of Theoretical Physics. 44 (2005), H. 11 Y1 - 2005 SN - 1572-9575 SP - 2041 EP - 2049 ER - TY - JOUR A1 - Streun, M. A1 - Christ, D. A1 - Hellendung, A. A1 - Larue, H. A1 - Ziemons, Karl A1 - Halling, H. T1 - Effects of crosstalk and gain nonuniformity using multichannel PMTs in the Clearpet® scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - The ClearPET® scanners developed by the Crystal Clear Collaboration use multichannel PMTs as photodetectors with scintillator pixels coupled individually to each channel. In order to localize an event each channel anode is connected to a comparator that triggers when the anode signal exceeds a common predefined threshold. Two major difficulties here are crosstalk of light and the gain nonuniformity of the PMT channels. Crosstalk can generate false triggering in channels adjacent to the actual event. On the one hand this can be suppressed by sufficiently increasing the threshold, but on the other hand a threshold too high can already prevent valid events on the lower gain channels from being detected. Finally, both effects restrict the dynamic range of pulse heights that can be processed. The requirements to the dynamic range are not low as the ClearPET® scanners detect the depth of interaction by phoswich pixels consisting of LSO and Lu0.7Y0.3AP, two scintillators with different light yields. We will present a model to estimate the achievable dynamic range and show solutions to increase it. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 402 EP - 405 ER -