TY - JOUR A1 - Digel, Ilya A1 - Temiz Artmann, Aysegül A1 - Nishikawa, K. A1 - Cook, M. T1 - Bactericidal effects of plasma-generated cluster ions JF - Medical and Biological Engineering and Computing. 43 (2005), H. 6 Y1 - 2005 SN - 1741-0444 SP - 800 EP - 807 ER - TY - JOUR A1 - Maggakis-Kelemen, Christina A1 - Digel, Ilya A1 - Artmann, Gerhard T1 - Polystyrene sulfonate/Polyallylamine hydrochloride microcapsules as potential artificial red blood cells - improvement of capsule flexibility JF - Biomedizinische Technik. 50 (2005), H. Erg.-Bd. 1 Y1 - 2005 SP - 324 EP - 326 ER - TY - JOUR A1 - Khodaverdi, M. A1 - Chatziioannou, A. F. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. A1 - Pietrzyk, U. T1 - Investigation of different MicroCT scanner configurations by GEANT4 simulations JF - IEEE Transactions on Nuclear Science N2 - This study has been performed to design the combination of the new ClearPET (ClearPET is a trademark of the Crystal Clear Collaboration), a small animal positron emission tomography (PET) system, with a micro-computed tomography (microCT) scanner. The properties of different microCT systems have been determined by simulations based on GEANT4. We will demonstrate the influence of the detector material and the X-ray spectrum on the obtained contrast. Four different detector materials (selenium, cadmium zinc telluride, cesium iodide and gadolinium oxysulfide) and two X-ray spectra (a molybdenum and a tungsten source) have been considered. The spectra have also been modified by aluminum filters of varying thickness. The contrast between different tissue types (water, air, brain, bone and fat) has been simulated by using a suitable phantom. The results indicate the possibility to improve the image contrast in microCT by an optimized combination of the X-ray source and detector material. Y1 - 2005 SN - 0018-9499 VL - 52 IS - 1 SP - 188 EP - 192 ER - TY - GEN A1 - Achten, Richard A1 - Bauer, Andreas A1 - Bertram, Walter A1 - Cremer, Markus A1 - Daemen, Jos A1 - Dehnhardt, Markus A1 - Fleischer, Manfred A1 - Kirchner, Peter A1 - Leyendecker, Marco A1 - Pietrzyk, Uwe A1 - Schmitz, Jakob A1 - Ziemons, Karl A1 - Zilles, Karl T1 - Vorrichtung zum Halten eines lebenden Objektes bei physiologischen Messungen N2 - Die vorliegende Erfindung betrifft eine Vorrichtung zum Halten eines lebenden Objektes bei physiologischen Messungen mit einem Basiselement und Mitteln zum Arretieren des lebenden Objektes, die atraumatische Ohrenhalter sowie ein Mundstück aufweisen, das mit einer Aussparung für die Zähne des Lebewesens versehen ist, mit einem Masseanteil von wenigstens 95% an glasfaserverstärktem Kunststoff. Y1 - 2005 N1 - Patent Anmelder : Forschungszentrum Jülich GmbH SP - 1 EP - 17 ER - TY - JOUR A1 - Streun, M. A1 - Christ, D. A1 - Hellendung, A. A1 - Larue, H. A1 - Ziemons, Karl A1 - Halling, H. T1 - Effects of crosstalk and gain nonuniformity using multichannel PMTs in the Clearpet® scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - The ClearPET® scanners developed by the Crystal Clear Collaboration use multichannel PMTs as photodetectors with scintillator pixels coupled individually to each channel. In order to localize an event each channel anode is connected to a comparator that triggers when the anode signal exceeds a common predefined threshold. Two major difficulties here are crosstalk of light and the gain nonuniformity of the PMT channels. Crosstalk can generate false triggering in channels adjacent to the actual event. On the one hand this can be suppressed by sufficiently increasing the threshold, but on the other hand a threshold too high can already prevent valid events on the lower gain channels from being detected. Finally, both effects restrict the dynamic range of pulse heights that can be processed. The requirements to the dynamic range are not low as the ClearPET® scanners detect the depth of interaction by phoswich pixels consisting of LSO and Lu0.7Y0.3AP, two scintillators with different light yields. We will present a model to estimate the achievable dynamic range and show solutions to increase it. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 402 EP - 405 ER - TY - JOUR A1 - Ziemons, Karl A1 - Auffray, E. A1 - Barbier, R. A1 - Brandenburg, G. A1 - Bruyndonckx, P. T1 - The ClearPET™ project: Development of a 2nd generation high-performance small animal PET scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - Second generation high-performance PET scanners, called ClearPET™1, have been developed by working groups of the Crystal Clear Collaboration (CCC). High sensitivity and high spatial resolution for the ClearPET camera is achieved by using a phoswich arrangement combining two different types of lutetium-based scintillator materials: LSO from CTI and LuYAP:Ce from the CCC (ISTC project). In a first ClearPET prototype, phoswich arrangements of 8×8 crystals of 2×2×10 mm3 are coupled to multi-channel photomultiplier tubes (Hamamatsu R7600). A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the ring is 120 mm, the axial detector length is 110 mm.The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the exact pulse starting time, which is subsequently used for coincidence detection. The gantry allows rotation of the detector modules around the field of view. Preliminary data shows a correct identification of the crystal layer about (98±1)%. Typically the energy resolution is (23.3±0.5)% for the luyap layer and (15.4±0.4)% for the lso layer. early studies showed the timing resolution of 2 ns FWHM and 4.8 ns FWTM. the intrinsic spatial resolution ranges from 1.37 mm to 1.61 mm full-width of half-maximum (FWHM) with a mean of 1.48 mm FWHM. further improvements in image and energy resolution are expected when the system geometry is fully modeled. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 307 EP - 311 ER - TY - CHAP A1 - Braun, Carsten A1 - Boucke, Alexander A1 - Ballmann, Josef T1 - Numerical prediction of the wing deformation of a high speed transport aircraft type wind tunnel model by direct aeroelastic simulation T2 - Conference proceedings : CEAS/AIAA/DGLR International Forum on Aeroelasticity and Structural Dynamics IFASD 2005 : München, June 28 - July 1, 2005. DGLR-Bericht. 2005,04 Y1 - 2005 SN - 3-932182-43-X PB - DGLR CY - Bonn ER - TY - CHAP A1 - Braun, Carsten A1 - Boucke, Alexander A1 - Ballmann, Josef T1 - Numerical study of the influence of dynamic pressure and deflected ailerons on the deformation of a high speed wing model T2 - High performance computing in science and engineering '04. Transactions of the High Performance Computing Center Stuttgart (HLRS) 2004 / Eds.: Egon Krause ... Y1 - 2005 SN - 3-540-22943-4 SP - 225 EP - 236 PB - Springer CY - Berlin ER - TY - CHAP A1 - Müller, Karsten A1 - Rohlfing, Richard A1 - Hueben, Susanne T1 - Entwicklung und Umsetzung von Sanierungsstrategien am Beispiel der Stadt Rheine T2 - 6. Kölner Kanal-Kolloquium : 07. und 08. September 2005 im Maternushaus, Köln / hrsg. von Johannes Pinnekamp ... Aachener Schriften zur Stadtentwässerung. Bd. 6 Y1 - 2005 SN - 3-932-590-98-8 SP - 9/1 EP - 9/18 PB - Ges. zur Förderung der Siedlungswasserwirtschaft an der RWTH Aachen CY - Aachen ER - TY - THES A1 - Müller, Karsten T1 - Strategien zur Zustandserfassung von Kanalisationen Y1 - 2005 N1 - Aachen, Techn. Hochsch., Diss., 2005 ER -