TY - JOUR A1 - Beer, S. A1 - Streun, M. A1 - Hombach, T. A1 - Buehler, J. A1 - Jahnke, S. A1 - Khodaverdi, M. A1 - Larue, H. A1 - Minwuyelet, S. A1 - Parl, C. A1 - Roeb, G. A1 - Schurr, U. A1 - Ziemons, Karl T1 - Design and initial performance of PlanTIS: a high-resolution positron emission tomograph for plants JF - Physics in Medicine and Biology N2 - Positron emitters such as 11C, 13N and 18F and their labelled compounds are widely used in clinical diagnosis and animal studies, but can also be used to study metabolic and physiological functions in plants dynamically and in vivo. A very particular tracer molecule is 11CO2 since it can be applied to a leaf as a gas. We have developed a Plant Tomographic Imaging System (PlanTIS), a high-resolution PET scanner for plant studies. Detectors, front-end electronics and data acquisition architecture of the scanner are based on the ClearPETâ„¢ system. The detectors consist of LSO and LuYAP crystals in phoswich configuration which are coupled to position-sensitive photomultiplier tubes. Signals are continuously sampled by free running ADCs, and data are stored in a list mode format. The detectors are arranged in a horizontal plane to allow the plants to be measured in the natural upright position. Two groups of four detector modules stand face-to-face and rotate around the field-of-view. This special system geometry requires dedicated image reconstruction and normalization procedures. We present the initial performance of the detector system and first phantom and plant measurements. Y1 - 2010 U6 - http://dx.doi.org/10.1088/0031-9155/55/3/006 SN - 1361-6560 VL - 55 IS - 3 SP - 635 EP - 646 PB - IOP CY - Bristol ER - TY - JOUR A1 - Streun, M. A1 - Beer, S. A1 - Hombach, T. A1 - Jahnke, S. A1 - Khodaverdi, M. A1 - Larue, H. A1 - Minwuyelet, S. A1 - Parl, C. A1 - Roeb, G. A1 - Schurr, U. A1 - Ziemons, Karl T1 - PlanTIS: A positron emission tomograph for imaging 11C transport in plants JF - 2007 IEEE Nuclear Science Symposium Conference Record, Vol. 6 N2 - Plant growth and transport processes are highly dynamic. They are characterized by plant-internal control processes and by strong interactions with the spatially and temporally varying environment. Analysis of structure- function relations of growth and transport in plants will strongly benefit from the development of non-invasive techniques. PlanTIS (Plant Tomographic Imaging System) is designed for non-destructive 3D-imaging of positron emitting radiotracers. It will permit functional analysis of the dynamics of carbon distribution in plants including bulky organs. It will be applicable for screening transport properties of plants to evaluate e.g. temperature adaptation of genetically modified plants. PlanTIS is a PET scanner dedicated to monitor the dynamics of the 11C distribution within a plant while or after assimilation of 11CO2. Front end electronics and data acquisition architecture of the scanner are based on the ClearPETTM system [1]. Four detector modules form one of two opposing detector blocks. Optionally, a hardware coincidence detection between the blocks can be applied. In general the scan duration is rather long (~ 1 hour) compared to the decay time of 11C (20 min). As a result the count rates can vary over a wide range and accurate dead time correction is necessary. Y1 - 2008 SN - 1082-3654 SP - 4110 EP - 4112 ER -