TY - CHAP A1 - Schmitz, Annika A1 - Apandi, Shah Eiman Amzar Shah A1 - Spillner, Jan A1 - Hima, Flutura A1 - Behbahani, Mehdi ED - Digel, Ilya ED - Staat, Manfred ED - Trzewik, Jürgen ED - Sielemann, Stefanie ED - Erni, Daniel ED - Zylka, Waldemar T1 - Effect of different cannula positions in the pulmonary artery on blood flow and gas exchange using computational fluid dynamics analysis T2 - 4th YRA MedTech Symposium 2024 : February 1 / 2024 / FH Aachen N2 - Pulmonary arterial cannulation is a common and effective method for percutaneous mechanical circulatory support for concurrent right heart and respiratory failure [1]. However, limited data exists to what effect the positioning of the cannula has on the oxygen perfusion throughout the pulmonary artery (PA). This study aims to evaluate, using computational fluid dynamics (CFD), the effect of different cannula positions in the PA with respect to the oxygenation of the different branching vessels in order for an optimal cannula position to be determined. The four chosen different positions (see Fig. 1) of the cannulas are, in the lower part of the main pulmonary artery (MPA), in the MPA at the junction between the right pulmonary artery (RPA) and the left pulmonary artery (LPA), in the RPA at the first branch of the RPA and in the LPA at the first branch of the LPA. Y1 - 2024 SN - 978-3-940402-65-3 U6 - http://dx.doi.org/10.17185/duepublico/81475 SP - 29 EP - 30 PB - Universität Duisburg-Essen CY - Duisburg ER - TY - JOUR A1 - Thiebes, Anja Lena A1 - Klein, Sarah A1 - Zingsheim, Jonas A1 - Möller, Georg H. A1 - Gürzing, Stefanie A1 - Reddemann, Manuel A. A1 - Behbahani, Mehdi A1 - Cornelissen, Christian G. T1 - Effervescent atomizer as novel cell spray technology to decrease the gas-to-liquid ratio JF - pharmaceutics N2 - Cell spraying has become a feasible application method for cell therapy and tissue engineering approaches. Different devices have been used with varying success. Often, twin-fluid atomizers are used, which require a high gas velocity for optimal aerosolization characteristics. To decrease the amount and velocity of required air, a custom-made atomizer was designed based on the effervescent principle. Different designs were evaluated regarding spray characteristics and their influence on human adipose-derived mesenchymal stromal cells. The arithmetic mean diameters of the droplets were 15.4–33.5 µm with decreasing diameters for increasing gas-to-liquid ratios. The survival rate was >90% of the control for the lowest gas-to-liquid ratio. For higher ratios, cell survival decreased to approximately 50%. Further experiments were performed with the design, which had shown the highest survival rates. After seven days, no significant differences in metabolic activity were observed. The apoptosis rates were not influenced by aerosolization, while high gas-to-liquid ratios caused increased necrosis levels. Tri-lineage differentiation potential into adipocytes, chondrocytes, and osteoblasts was not negatively influenced by aerosolization. Thus, the effervescent aerosolization principle was proven suitable for cell applications requiring reduced amounts of supplied air. This is the first time an effervescent atomizer was used for cell processing. KW - tri-lineage differentiation KW - survival KW - twin-fluid atomizer KW - adipose-derived stromal cells (ASCs) KW - cell atomization KW - cell aerosolization Y1 - 2022 U6 - http://dx.doi.org/10.3390/pharmaceutics14112421 N1 - This article belongs to the Special Issue "Stromal, Stem, Signaling Cells: The Multiple Roles and Applications of Mesenchymal Cells" VL - 14 IS - 11 PB - MDPI CY - Basel ER - TY - JOUR A1 - Chloé, Radermacher A1 - Malyaran, Hanna A1 - Craveiro, Rogerio Bastos A1 - Peglow, Sarah A1 - Behbahani, Mehdi A1 - Pufe, Thomas A1 - Wolf, Michael A1 - Neuss, Sabine T1 - Mechanical loading on cementoblasts: a mini review JF - Osteologie N2 - Orthodontic treatments are concomitant with mechanical forces and thereby cause teeth movements. The applied forces are transmitted to the tooth root and the periodontal ligaments which is compressed on one side and tensed up on the other side. Indeed, strong forces can lead to tooth root resorption and the crown-to-tooth ratio is reduced with the potential for significant clinical impact. The cementum, which covers the tooth root, is a thin mineralized tissue of the periodontium that connects the periodontal ligament with the tooth and is build up by cementoblasts. The impact of tension and compression on these cells is investigated in several in vivo and in vitro studies demonstrating differences in protein expression and signaling pathways. In summary, osteogenic marker changes indicate that cyclic tensile forces support whereas static tension inhibits cementogenesis. Furthermore, cementogenesis experiences the same protein expression changes in static conditions as static tension, but cyclic compression leads to the exact opposite of cyclic tension. Consistent with marker expression changes, the singaling pathways of Wnt/ß-catenin and RANKL/OPG show that tissue compression leads to cementum degradation and tension forces to cementogenesis. However, the cementum, and in particular its cementoblasts, remain a research area which should be explored in more detail to understand the underlying mechanism of bone resorption and remodeling after orthodontic treatments. KW - Cementoblast KW - Compression KW - Tension KW - Mechanotransduction KW - Forces Y1 - 2022 U6 - http://dx.doi.org/10.1055/a-1826-0777 SN - 1019-1291 VL - 31 IS - 2 SP - 111 EP - 118 PB - Thieme CY - Stuttgart ER - TY - JOUR A1 - Lenz, Maximilian A1 - Kahmann, Stephanie Lucina A1 - Behbahani, Mehdi A1 - Pennig, Lenhard A1 - Hackl, Michael A1 - Leschinger, Tim A1 - Müller, Lars Peter A1 - Wegmann, Kilian T1 - Influence of rotator cuff preload on fracture configuration in proximal humerus fractures: a proof of concept for fracture simulation JF - Archives of Orthopaedic and Trauma Surgery N2 - Introduction In regard of surgical training, the reproducible simulation of life-like proximal humerus fractures in human cadaveric specimens is desirable. The aim of the present study was to develop a technique that allows simulation of realistic proximal humerus fractures and to analyse the influence of rotator cuff preload on the generated lesions in regards of fracture configuration. Materials and methods Ten cadaveric specimens (6 left, 4 right) were fractured using a custom-made drop-test bench, in two groups. Five specimens were fractured without rotator cuff preload, while the other five were fractured with the tendons of the rotator cuff preloaded with 2 kg each. The humeral shaft and the shortened scapula were potted. The humerus was positioned at 90° of abduction and 10° of internal rotation to simulate a fall on the elevated arm. In two specimens of each group, the emergence of the fractures was documented with high-speed video imaging. Pre-fracture radiographs were taken to evaluate the deltoid-tuberosity index as a measure of bone density. Post-fracture X-rays and CT scans were performed to define the exact fracture configurations. Neer’s classification was used to analyse the fractures. Results In all ten cadaveric specimens life-like proximal humerus fractures were achieved. Two III-part and three IV-part fractures resulted in each group. The preloading of the rotator cuff muscles had no further influence on the fracture configuration. High-speed videos of the fracture simulation revealed identical fracture mechanisms for both groups. We observed a two-step fracture mechanism, with initial impaction of the head segment against the glenoid followed by fracturing of the head and the tuberosities and then with further impaction of the shaft against the acromion, which lead to separation of the tuberosities. Conclusion A high energetic axial impulse can reliably induce realistic proximal humerus fractures in cadaveric specimens. The preload of the rotator cuff muscles had no influence on initial fracture configuration. Therefore, fracture simulation in the proximal humerus is less elaborate. Using the presented technique, pre-fractured specimens are available for real-life surgical education. KW - Proximal humerus fracture KW - Biomechanical simulation KW - Fracture configuration KW - Fracture simulation KW - Rotator cuff Y1 - 2022 U6 - http://dx.doi.org/10.1007/s00402-022-04471-9 SN - 1434-3916 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Malinowski, Daniel A1 - Fournier, Yvan A1 - Horbach, Andreas A1 - Frick, Michael A1 - Magliani, Mirko A1 - Kalverkamp, Sebastian A1 - Hildinger, Martin A1 - Spillner, Jan A1 - Behbahani, Mehdi A1 - Hima, Flutura T1 - Computational fluid dynamics analysis of endoluminal aortic perfusion JF - Perfusion N2 - Introduction: In peripheral percutaneous (VA) extracorporeal membrane oxygenation (ECMO) procedures the femoral arteries perfusion route has inherent disadvantages regarding poor upper body perfusion due to watershed. With the advent of new long flexible cannulas an advancement of the tip up to the ascending aorta has become feasible. To investigate the impact of such long endoluminal cannulas on upper body perfusion, a Computational Fluid Dynamics (CFD) study was performed considering different support levels and three cannula positions. Methods: An idealized literature-based- and a real patient proximal aortic geometry including an endoluminal cannula were constructed. The blood flow was considered continuous. Oxygen saturation was set to 80% for the blood coming from the heart and to 100% for the blood leaving the cannula. 50% and 90% venoarterial support levels from the total blood flow rate of 6 l/min were investigated for three different positions of the cannula in the aortic arch. Results: For both geometries, the placement of the cannula in the ascending aorta led to a superior oxygenation of all aortic blood vessels except for the left coronary artery. Cannula placements at the aortic arch and descending aorta could support supra-aortic arteries, but not the coronary arteries. All positions were able to support all branches with saturated blood at 90% flow volume. Conclusions: In accordance with clinical observations CFD analysis reveals, that retrograde advancement of a long endoluminal cannula can considerably improve the oxygenation of the upper body and lead to oxygen saturation distributions similar to those of a central cannulation. KW - computational fluid dynamics analysis KW - simulation KW - endoluminal KW - aortic perfusion KW - extracorporeal membrane oxygenation Y1 - 2022 U6 - http://dx.doi.org/10.1177/02676591221099809 SN - 1477-111X VL - 0 IS - 0 SP - 1 EP - 8 PB - Sage CY - London ER - TY - JOUR A1 - Mandekar, Swati A1 - Holland, Abigail A1 - Thielen, Moritz A1 - Behbahani, Mehdi A1 - Melnykowycz, Mark T1 - Advancing towards Ubiquitous EEG, Correlation of In-Ear EEG with Forehead EEG JF - Sensors N2 - Wearable EEG has gained popularity in recent years driven by promising uses outside of clinics and research. The ubiquitous application of continuous EEG requires unobtrusive form-factors that are easily acceptable by the end-users. In this progression, wearable EEG systems have been moving from full scalp to forehead and recently to the ear. The aim of this study is to demonstrate that emerging ear-EEG provides similar impedance and signal properties as established forehead EEG. EEG data using eyes-open and closed alpha paradigm were acquired from ten healthy subjects using generic earpieces fitted with three custom-made electrodes and a forehead electrode (at Fpx) after impedance analysis. Inter-subject variability in in-ear electrode impedance ranged from 20 kΩ to 25 kΩ at 10 Hz. Signal quality was comparable with an SNR of 6 for in-ear and 8 for forehead electrodes. Alpha attenuation was significant during the eyes-open condition in all in-ear electrodes, and it followed the structure of power spectral density plots of forehead electrodes, with the Pearson correlation coefficient of 0.92 between in-ear locations ELE (Left Ear Superior) and ERE (Right Ear Superior) and forehead locations, Fp1 and Fp2, respectively. The results indicate that in-ear EEG is an unobtrusive alternative in terms of impedance, signal properties and information content to established forehead EEG. KW - in-ear EEG KW - correlation KW - forehead EEG KW - impedance spectroscopy KW - biopotential electrodes Y1 - 2022 U6 - http://dx.doi.org/10.3390/s22041568 SN - 1424-8220 VL - 22 IS - 4 SP - 1 EP - 19 PB - MDPI CY - Basel ER - TY - CHAP A1 - Mandekar, Swati A1 - Jentsch, Lina A1 - Lutz, Kai A1 - Behbahani, Mehdi A1 - Melnykowycz, Mark T1 - Earable design analysis for sleep EEG measurements T2 - UbiComp '21 N2 - Conventional EEG devices cannot be used in everyday life and hence, past decade research has been focused on Ear-EEG for mobile, at-home monitoring for various applications ranging from emotion detection to sleep monitoring. As the area available for electrode contact in the ear is limited, the electrode size and location play a vital role for an Ear-EEG system. In this investigation, we present a quantitative study of ear-electrodes with two electrode sizes at different locations in a wet and dry configuration. Electrode impedance scales inversely with size and ranges from 450 kΩ to 1.29 MΩ for dry and from 22 kΩ to 42 kΩ for wet contact at 10 Hz. For any size, the location in the ear canal with the lowest impedance is ELE (Left Ear Superior), presumably due to increased contact pressure caused by the outer-ear anatomy. The results can be used to optimize signal pickup and SNR for specific applications. We demonstrate this by recording sleep spindles during sleep onset with high quality (5.27 μVrms). KW - EEG KW - sensors KW - Impedance Spectroscopy KW - Sleep EEG KW - biopotential electrodes Y1 - 2021 U6 - http://dx.doi.org/10.1145/3460418.3479328 N1 - UbiComp '21: Adjunct Proceedings of the 2021 ACM International Joint Conference on Pervasive and Ubiquitous Computing and Proceedings of the 2021 ACM International Symposium on Wearable Computers, September 21–26, 2021, Virtual, USA SP - 171 EP - 175 ER - TY - JOUR A1 - Brockhaus, Moritz K. A1 - Behbahani, Mehdi A1 - Muris, Farina A1 - Jansen, Sebastian V. A1 - Schmitz- Rode, Thomas A1 - Steinseifer, Ulrich A1 - Clauser, Johanna C. T1 - In vitro thrombogenicity testing of pulsatile mechanical circulatory support systems: Design and proof-of-concept JF - Artificial Organs N2 - Thrombogenic complications are a main issue in mechanical circulatory support (MCS). There is no validated in vitro method available to quantitatively assess the thrombogenic performance of pulsatile MCS devices under realistic hemodynamic conditions. The aim of this study is to propose a method to evaluate the thrombogenic potential of new designs without the use of complex in-vivo trials. This study presents a novel in vitro method for reproducible thrombogenicity testing of pulsatile MCS systems using low molecular weight heparinized porcine blood. Blood parameters are continuously measured with full blood thromboelastometry (ROTEM; EXTEM, FIBTEM and a custom-made analysis HEPNATEM). Thrombus formation is optically observed after four hours of testing. The results of three experiments are presented each with two parallel loops. The area of thrombus formation inside the MCS device was reproducible. The implantation of a filter inside the loop catches embolizing thrombi without a measurable increase of platelet activation, allowing conclusions of the place of origin of thrombi inside the device. EXTEM and FIBTEM parameters such as clotting velocity (α) and maximum clot firmness (MCF) show a total decrease by around 6% with a characteristic kink after 180 minutes. HEPNATEM α and MCF rise within the first 180 minutes indicate a continuously increasing activation level of coagulation. After 180 minutes, the consumption of clotting factors prevails, resulting in a decrease of α and MCF. With the designed mock loop and the presented protocol we are able to identify thrombogenic hot spots inside a pulsatile pump and characterize their thrombogenic potential. Y1 - 2021 U6 - http://dx.doi.org/10.1111/aor.14046 SN - 1525-1594 VL - 45 IS - 12 SP - 1513 EP - 1521 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Kopp, Alexander A1 - Schunck, Laura A1 - Gosau, Martin A1 - Smeets, Ralf A1 - Burg, Simon A1 - Fuest, Sandra A1 - Kröger, Nadja A1 - Zinser, Max A1 - Krohn, Sebastian A1 - Behbahani, Mehdi A1 - Köpf, Marius A1 - Lauts, Lisa A1 - Rutkowski, Rico T1 - Influence of the casting concentration on the mechanical and optical properties of Fa/CaCl2-derived silk fibroin membranes JF - International Journal of Molecular Sciences N2 - In this study, we describe the manufacturing and characterization of silk fibroin membranes derived from the silkworm Bombyx mori. To date, the dissolution process used in this study has only been researched to a limited extent, although it entails various potential advantages, such as reduced expenses and the absence of toxic chemicals in comparison to other conventional techniques. Therefore, the aim of this study was to determine the influence of different fibroin concentrations on the process output and resulting membrane properties. Casted membranes were thus characterized with regard to their mechanical, structural and optical assets via tensile testing, SEM, light microscopy and spectrophotometry. Cytotoxicity was evaluated using BrdU, XTT, and LDH assays, followed by live–dead staining. The formic acid (FA) dissolution method was proven to be suitable for the manufacturing of transparent and mechanically stable membranes. The fibroin concentration affects both thickness and transparency of the membranes. The membranes did not exhibit any signs of cytotoxicity. When compared to other current scientific and technical benchmarks, the manufactured membranes displayed promising potential for various biomedical applications. Further research is nevertheless necessary to improve reproducible manufacturing, including a more uniform thickness, less impurity and physiological pH within the membranes. Y1 - 2020 U6 - http://dx.doi.org/10.3390/ijms21186704 SN - 1422-0067 N1 - Special issue: Optimization of Biomaterials for Reconstructive and Regenerative Medicine VL - 21 IS - 18 art. no. 6704 PB - MDPI CY - Basel ER - TY - THES A1 - Behbahani, Mehdi T1 - Modeling and Simulation of Shear-Dependent Platelet Reactions in Blood Vessels and Blood-Contacting Medical Devices Y1 - 2011 SN - 978-3-8439-0134-5 PB - Verlag Dr. Hut CY - München ER -