TY - JOUR A1 - Frotscher, Ralf A1 - Staat, Manfred T1 - Stresses produced by different textile mesh implants in a tissue equivalent JF - BioNanoMaterials N2 - Two single-incision mini-slings used for treating urinary incontinence in women are compared with respect to the stresses they produce in their surrounding tissue. In an earlier paper we experimentally observed that these implants produce considerably different stress distributions in a muscle tissue equivalent. Here we perform 2D finite element analyses to compare the shear stresses and normal stresses in the tissue equivalent for the two meshes and to investigate their failure behavior. The results clearly show that the Gynecare TVT fails for increasing loads in a zipper-like manner because it gradually debonds from the surrounding tissue. Contrary to that, the tissue at the ends of the DynaMesh-SIS direct may rupture but only at higher loads. The simulation results are in good agreement with the experimental observations thus the computational model helps to interpret the experimental results and provides a tool for qualitative evaluation of mesh implants. Y1 - 2014 U6 - http://dx.doi.org/10.1515/bnm-2014-0003 SN - 2191-4672 (E-Journal); 2193-066X (E-Journal); 0011-8656 (Print); 1616-0177 (Print); 2193-0651 (Print) VL - 15 IS - 1-2 SP - 25 EP - 30 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Staat, Manfred A1 - Trenz, Eva A1 - Lohmann, Philipp A1 - Frotscher, Ralf A1 - Klinge, Uwe A1 - Tabaza, Ruth A1 - Kirschner-Hermanns, Ruth T1 - New measurements to compare soft tissue anchoring systems in pelvic floor surgery JF - Journal of Biomedical Materials Research Part B: Applied Biomaterials N2 - Suburethral slings as well as different meshes are widely used treating stress urinary incontinence and prolaps in women. With the development of MiniSlings and special meshes using less alloplastic material anchorage systems become more important to keep devices in place and to put some tension especially on the MiniSlings. To date, there are many different systems of MiniSlings of different companies on the market which differ in the structure of the used meshes and anchors. A new objective measurement method to compare different properties of MiniSling systems (mesh and anchor) is presented in this article. Ballistic gelatine acts as soft tissue surrogate. Significant differences in parameters like pull-out strength of anchors or shrinkage of meshes under loading conditions have been determined. The form and size of the anchors as well as the structural stability of the meshes are decisive for a proper integration. The tested anchorings sytems showed markedly different mechanical function at their respective load bearing capacity. As the stable fixation of the device in tissue is a prerequisite for a permanet reinforcement, the proposed test system permits further optimisation of anchor and mesh devices to improve the success of the surgical treatment Y1 - 2012 U6 - http://dx.doi.org/10.1002/jbm.b.32654 SN - 1552-4981 VL - 100B IS - 4 SP - 924 EP - 933 PB - Wiley CY - Hoboken, NJ ER -