TY - JOUR A1 - Theysohn, Jens M. A1 - Kraff, Oliver A1 - Eilers, Kristina A1 - Andrade, Dorian A1 - Gerwig, Marcus A1 - Timmann, Dagmar A1 - Schmitt, Franz A1 - Ladd, Mark E. A1 - Ladd, Susanne C. A1 - Bitz, Andreas T1 - Vestibular effects of a 7 Tesla MRI examination compared to 1.5 T and 0 T in healthy volunteers JF - PLoS one N2 - Ultra-high-field MRI (7 Tesla (T) and above) elicits more temporary side-effects compared to 1.5 T and 3 T, e.g. dizziness or “postural instability” even after exiting the scanner. The current study aims to assess quantitatively vestibular performance before and after exposure to different MRI scenarios at 7 T, 1.5 T and 0 T. Sway path and body axis rotation (Unterberger's stepping test) were quantitatively recorded in a total of 46 volunteers before, 2 minutes after, and 15 minutes after different exposure scenarios: 7 T head MRI (n = 27), 7 T no RF (n = 22), 7 T only B₀ (n = 20), 7 T in & out B₀ (n = 20), 1.5 T no RF (n = 20), 0 T (n = 15). All exposure scenarios lasted 30 minutes except for brief one minute exposure in 7 T in & out B₀. Both measures were documented utilizing a 3D ultrasound system. During sway path evaluation, the experiment was repeated with eyes both open and closed. Sway paths for all long-lasting 7 T scenarios (normal, no RF, only B₀) with eyes closed were significantly prolonged 2 minutes after exiting the scanner, normalizing after 15 minutes. Brief exposure to 7 T B₀ or 30 minutes exposure to 1.5 T or 0 T did not show significant changes. End positions after Unterberger's stepping test were significantly changed counter-clockwise after all 7 T scenarios, including the brief in & out B₀ exposure. Shorter exposure resulted in a smaller alteration angle. In contrast to sway path, reversal of changes in body axis rotation was incomplete after 15 minutes. 1.5 T caused no rotational changes. The results show that exposure to the 7 Tesla static magnetic field causes only a temporary dysfunction or “over-compensation” of the vestibular system not measurable at 1.5 or 0 Tesla. Radiofrequency fields, gradient switching, and orthostatic dysregulation do not seem to play a role. Y1 - 2014 U6 - http://dx.doi.org/10.1371/journal.pone.0092104 SN - 1932-6203 VL - 9 IS - 3 PB - PLOS CY - San Francisco ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Weiland, Elisabeth A1 - Uden, Mark J. van A1 - Kobus, Thiele A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - (31) P MR spectroscopic imaging of the human prostate at 7 T: T1 relaxation times, Nuclear Overhauser Effect, and spectral characterization JF - Magnetic Resonance in Medicine Y1 - 2015 U6 - http://dx.doi.org/10.1002/mrm.25209 SN - 1522-2594 VL - 73 IS - 3 SP - 909 EP - 920 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Vos, E. K. A1 - Lagemaat, M. W. A1 - Barentsz, J. O. A1 - Fütterer, J. J. A1 - Zamecnik, P. A1 - Roozen, H. A1 - Orzada, S. A1 - Bitz, Andreas A1 - Maas, M. C. A1 - Scheenen, T. W. J. T1 - Image quality and cancer visibility of T2-weighted Magnetic Resonance Imaging of the prostate at 7 Tesla JF - European Radiology N2 - Objectives To assess the image quality of T2-weighted (T2w) magnetic resonance imaging of the prostate and the visibility of prostate cancer at 7 Tesla (T). Materials & methods Seventeen prostate cancer patients underwent T2w imaging at 7T with only an external transmit/receive array coil. Three radiologists independently scored images for image quality, visibility of anatomical structures, and presence of artefacts. Krippendorff’s alpha and weighted kappa statistics were used to assess inter-observer agreement. Visibility of prostate cancer lesions was assessed by directly linking the T2w images to the confirmed location of prostate cancer on histopathology. Results T2w imaging at 7T was achievable with ‘satisfactory’ (3/5) to ‘good’ (4/5) quality. Visibility of anatomical structures was predominantly scored as ‘satisfactory’ (3/5) and ‘good’ (4/5). If artefacts were present, they were mostly motion artefacts and, to a lesser extent, aliasing artefacts and noise. Krippendorff’s analysis revealed an α = 0.44 between three readers for the overall image quality scores. Clinically significant cancer lesions in both peripheral zone and transition zone were visible at 7T. Conclusion T2w imaging with satisfactory to good quality can be routinely acquired, and cancer lesions were visible in patients with prostate cancer at 7T using only an external transmit/receive body array coil. Y1 - 2014 U6 - http://dx.doi.org/10.1007/s00330-014-3234-6 SN - 1432-1084 VL - 24 IS - 8 SP - 1950 EP - 1958 PB - Springer CY - Cham ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Breukels, Vincent A1 - Vos, Eline K. A1 - B., Adam A1 - Uden, Mark J. van A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Maas, Marnix C. A1 - Scheenen, Tom W. J. T1 - ¹H MR spectroscopic imaging of the prostate at 7T using spectral-spatial pulses JF - Magnetic Resonance in Medicine N2 - Purpose To assess the feasibility of prostate ¹H MR spectroscopic imaging (MRSI) using low-power spectral-spatial (SPSP) pulses at 7T, exploiting accurate spectral selection and spatial selectivity simultaneously. Methods A double spin-echo sequence was equipped with SPSP refocusing pulses with a spectral selectivity of 1 ppm. Three-dimensional prostate ¹H-MRSI at 7T was performed with the SPSP-MRSI sequence using an 8-channel transmit array coil and an endorectal receive coil in three patients with prostate cancer and in one healthy subject. No additional water or lipid suppression pulses were used. Results Prostate ¹H-MRSI could be obtained well within specific absorption rate (SAR) limits in a clinically feasible time (10 min). Next to the common citrate signals, the prostate spectra exhibited high spermine signals concealing creatine and sometimes also choline. Residual lipid signals were observed at the edges of the prostate because of limitations in spectral and spatial selectivity. Conclusion It is possible to perform prostate ¹H-MRSI at 7T with a SPSP-MRSI sequence while using separate transmit and receive coils. This low-SAR MRSI concept provides the opportunity to increase spatial resolution of MRSI within reasonable scan times. Y1 - 2016 U6 - http://dx.doi.org/10.1002/mrm.25569 SN - 1522-2594 VL - 75 IS - 3 SP - 933 EP - 945 PB - International Society for Magnetic Resonance in Medicine ER - TY - JOUR A1 - Noureddine, Yacine A1 - Bitz, Andreas A1 - Ladd, Mark E. A1 - Thürling, Markus A1 - Ladd, Susanne C. A1 - Schaefers, Gregor A1 - Kraff, Oliver T1 - Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study JF - Magnetic Resonance Materials in Physics, Biology and Medicine Y1 - 2015 U6 - http://dx.doi.org/10.1007/s10334-015-0499-y SN - 1352-8661 VL - 28 IS - 6 SP - 577 EP - 590 PB - Springer CY - Berlin ER - TY - JOUR A1 - Bank, Bart L. van de A1 - Orzada, Stephan A1 - Smits, Frits A1 - Lagemaat, Miriam W. A1 - Rodgers, Christopher T. A1 - Bitz, Andreas A1 - Scheenen, Tom W. J. T1 - Optimized (31) P MRS in the human brain at 7 T with a dedicated RF coil setup JF - NMR in Biomedicine Y1 - 2015 U6 - http://dx.doi.org/10.1002/nbm.3422 SN - 1099-1492 VL - 28 IS - 11 SP - 1570 EP - 1578 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Schmidt, K. A1 - Forkmann, K. A1 - Sinke, C. A1 - Gratz, M. A1 - Bitz, Andreas A1 - Bingel, U. T1 - The differential effect of trigeminal vs. peripheral pain stimulation on visual processing and memory encoding is influenced by pain-related fear JF - NeuroImage N2 - Compared to peripheral pain, trigeminal pain elicits higher levels of fear, which is assumed to enhance the interruptive effects of pain on concomitant cognitive processes. In this fMRI study we examined the behavioral and neural effects of trigeminal (forehead) and peripheral (hand) pain on visual processing and memory encoding. Cerebral activity was measured in 23 healthy subjects performing a visual categorization task that was immediately followed by a surprise recognition task. During the categorization task subjects received concomitant noxious electrical stimulation on the forehead or hand. Our data show that fear ratings were significantly higher for trigeminal pain. Categorization and recognition performance did not differ between pictures that were presented with trigeminal and peripheral pain. However, object categorization in the presence of trigeminal pain was associated with stronger activity in task-relevant visual areas (lateral occipital complex, LOC), memory encoding areas (hippocampus and parahippocampus) and areas implicated in emotional processing (amygdala) compared to peripheral pain. Further, individual differences in neural activation between the trigeminal and the peripheral condition were positively related to differences in fear ratings between both conditions. Functional connectivity between amygdala and LOC was increased during trigeminal compared to peripheral painful stimulation. Fear-driven compensatory resource activation seems to be enhanced for trigeminal stimuli, presumably due to their exceptional biological relevance. Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.neuroimage.2016.03.026 SN - 1053-8119 VL - 134 SP - 386 EP - 395 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Chen, Bixia A1 - Schoemberg, Tobias A1 - Kraff, Oliver A1 - Dammann, Philipp A1 - Bitz, Andreas A1 - Schlamann, Marc A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Sure, Ulrich A1 - Wrede, Karsten H. T1 - Cranial fixation plates in cerebral magnetic resonance imaging: a 3 and 7 Tesla in vivo image quality study JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective This study assesses and quantifies impairment of postoperative magnetic resonance imaging (MRI) at 7 Tesla (T) after implantation of titanium cranial fixation plates (CFPs) for neurosurgical bone flap fixation. Materials and methods The study group comprised five patients who were intra-individually examined with 3 and 7 T MRI preoperatively and postoperatively (within 72 h/3 months) after implantation of CFPs. Acquired sequences included T₁-weighted magnetization-prepared rapid-acquisition gradient-echo (MPRAGE), T₂-weighted turbo-spin-echo (TSE) imaging, and susceptibility-weighted imaging (SWI). Two experienced neurosurgeons and a neuroradiologist rated image quality and the presence of artifacts in consensus reading. Results Minor artifacts occurred around the CFPs in MPRAGE and T2 TSE at both field strengths, with no significant differences between 3 and 7 T. In SWI, artifacts were accentuated in the early postoperative scans at both field strengths due to intracranial air and hemorrhagic remnants. After resorption, the brain tissue directly adjacent to skull bone could still be assessed. Image quality after 3 months was equal to the preoperative examinations at 3 and 7 T. Conclusion Image quality after CFP implantation was not significantly impaired in 7 T MRI, and artifacts were comparable to those in 3 T MRI. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s10334-016-0548-1 SN - 1352-8661 VL - 29 IS - 3 SP - 389 EP - 398 PB - Springer CY - Berlin ER - TY - JOUR A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - A method to approximate maximum local SAR in multichannel transmit MR systems without transmit phase information JF - Magnetic Resonance in Medicine N2 - Purpose To calculate local specific absorption rate (SAR) correctly, both the amplitude and phase of the signal in each transmit channel have to be known. In this work, we propose a method to derive a conservative upper bound for the local SAR, with a reasonable safety margin without knowledge of the transmit phases of the channels. Methods The proposed method uses virtual observation points (VOPs). Correction factors are calculated for each set of VOPs that prevent underestimation of local SAR when the VOPs are applied with the correct amplitudes but fixed phases. Results The proposed method proved to be superior to the worst-case calculation based on the maximum eigenvalue of the VOPs. The mean overestimation for six coil setups could be reduced, whereas no underestimation of the maximum local SAR occurred. In the best investigated case, the overestimation could be reduced from a factor of 3.3 to a factor of 1.7. Conclusion The upper bound for the local SAR calculated with the proposed method allows a fast estimation of the local SAR based on power measurements in the transmit channels and facilitates SAR monitoring in systems that do not have the capability to monitor transmit phases Y1 - 2016 U6 - http://dx.doi.org/10.1002/mrm.26398 SN - 1522-2594 VL - 78 IS - 2 SP - 805 EP - 811 PB - International Society for Magnetic Resonance in Medicine ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus temperature limits JF - Medical Physics Y1 - 2017 U6 - http://dx.doi.org/10.1002/mp.12034 N1 - This article is corrected by: Errata: Erratum: “RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus tissue temperature limits” [Med. Phys. 44(1), 143–157 (2017)] Volume 44, Issue 2, 772 VL - 44 IS - 1 SP - 143 EP - 157 ER - TY - JOUR A1 - Noureddine, Yacine A1 - Kraff, Oliver A1 - Ladd, Mark E. A1 - Wrede, Karsten H. A1 - Chen, Bixia A1 - Quick, Harald H. A1 - Schaefers, Gregor A1 - Bitz, Andreas T1 - In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla JF - Magnetic Resonance in Medicine Y1 - 2017 U6 - http://dx.doi.org/10.1002/mrm.26650 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - SAR Simulations & Safety JF - NeuroImage Y1 - 2017 U6 - http://dx.doi.org/10.1016/j.neuroimage.2017.03.035 SN - 1053-8119 IS - Epub ahead of print PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Lind, Thorsten Patric T1 - Rückzahlung von gewinnunabhängigen Ausschüttungen bei Auflösung einer stillen Gesellschaft : BGH, Versäumnisurteil vom 20.09.2016 - II ZR 120/15 (LG Berlin) : Anmerkung JF - Fachdienst Zivilrecht - LMK : kommentierte BGH-Rechtsprechung ; in Zsarb. mit der Neuen Juristischen Wochenschrift / Lindermaier-Möhring N2 - Zum Zweck der Kapitalanlage beteiligte sich der Bekl. an einer Publikumsgesellschaft. Diese war als (mehrgliedrige) atypische stille Gesellschaft organisiert (vgl. a. BGHZ 199, 104 = DNotZ 2014, 374 = NZG 2013, 1422 = DStR 2014, 45 Rn. 18). Der Gesellschaftsvertrag („GV“) sah für diejenigen Gesellschafter, die wie der Bekl. ihre Einlage in Form einer Einmaleinlage erbracht hatten, eine jährliche gewinnunabhängige Ausschüttung vor. Es sollte sich dabei ausdrücklich nicht um eine Garantieverzinsung handeln. Ende 2009 wurde die stille Gesellschaft durch Mehrheitsbeschluss der Stillen aufgelöst. Nach dem GV waren die Stillen im Falle ihres Ausscheidens sowie bei „Liquidation des Unternehmens“ des Geschäftsinhabers verhältnismäßig an dem jeweils seit ihrem Beitritt gebildeten Vermögen einschließlich der stillen Reserven sowie eines evtl. Geschäftswerts zu beteiligen (Auseinandersetzungswert). Den sich hiernach für den Bekl. auf seinem Kapitalkonto ergebenden Negativsaldo sollte dieser durch Erstattung der von ihm erhaltenen gewinnunabhängigen Auszahlungen ausgleichen. Anders als das LG als Berufungsinstanz bejahte der BGH auf Grundlage des GV einen Rückerstattungsanspruch der klagenden Geschäftsinhaberin. Y1 - 2016 PB - Beck CY - München ER - TY - JOUR A1 - Damm, Marc André A1 - Sauerborn, Markus A1 - Fend, Thomas A1 - Herrmann, Ulf T1 - Optimisation of a urea selective catalytic reduction system with a coated ceramic mixing element JF - Journal of ceramic science and technology Y1 - 2017 SN - 2190-9385 (Print) U6 - http://dx.doi.org/10.4416/JCST2016-00056 SN - 2190-9385 (Online) VL - 8 IS - 1 SP - 19 EP - 24 PB - Göller CY - Baden-Baden ER - TY - JOUR A1 - Lanzl, I. A1 - Kotliar, Konstantin T1 - Können Anti-VEGF-Injektionen Glaukom oder okuläre Hypertension verursachen? T1 - Can Anti-VEGF Injections Cause Glaucoma or Ocular Hypertension? JF - Klinische Monatsblätter für Augenheilkunde Y1 - 2017 U6 - http://dx.doi.org/10.1055/s-0043-101819 SN - 0023-2165 VL - 234 IS - 2 SP - 191 EP - 193 PB - Thieme CY - Stuttgart ER - TY - JOUR A1 - Enning, Manfred T1 - Vom "König der Landstraße" zur "Ameise 4.0" JF - ETR - Eisenbahntechnische Rundschau N2 - Auf Eisenbahn- und Logistikkonferenzen werden gelegentlich Videos von Blattschneiderameisen gezeigt, die effizient und selbstorganisiert große Mengen Material durch den Dschungel transportieren. Daran sollen wir uns ein Beispiel nehmen. Y1 - 2017 SN - 0013-2845 VL - 66 IS - 1/2 SP - 3 PB - DVV Media Group CY - Hamburg ER - TY - JOUR A1 - Enning, Manfred A1 - Pfaff, Raphael T1 - Güterwagen 4.0 - Der Güterwagen für das Internet der Dinge. Teil 1: Gesamtsystembetrachtung und grundlegendes Konzept JF - ETR - Eisenbahntechnische Rundschau Y1 - 2017 SN - 0013-2845 VL - 66 IS - 1/2 SP - 12 EP - 16 PB - DVV Media Group CY - Hamburg ER - TY - JOUR A1 - Honarvarfard, Elham A1 - Gamella, Maria A1 - Channaveerappa, Devika A1 - Darie, Costel C. A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Katz, Evgeny T1 - Electrochemically Stimulated Insulin Release from a Modified Graphene–functionalized Carbon Fiber Electrode JF - Electroanalysis N2 - A graphene-functionalized carbon fiber electrode was modified with adsorbed polyethylenimine to introduce amino functionalities and then with trigonelline and 4-carboxyphenylboronic acid covalently bound to the amino groups. The trigonelline species containing quarterized pyridine groups produced positive charge on the electrode surface regardless of the pH value, while the phenylboronic acid species were neutral below pH 8 and negatively charged above pH 9 (note that their pKa=8.4). The total charge on the monolayer-modified electrode was positive at the neutral pH and negative at pH > 9. Note that 4-carboxyphenylboronic acid was attached to the electrode surface in molar excess to trigonelline, thus allowing the negative charge to dominate on the electrode surface at basic pH. Negatively charged fluorescent dye-labeled insulin (insulin-FITC) was loaded on the modified electrode surface at pH 7.0 due to its electrostatic attraction to the positively charged interface. The local pH in close vicinity to the electrode surface was increased to ca. 9–10 due to consumption of H+ ions upon electrochemical reduction of oxygen proceeding at the potential of −1.0 V (vs. Ag/AgCl) applied on the modified electrode. The process resulted in recharging of the electrode surface to the negative value due to the formation of the negative charge on the phenylboronic acid groups, thus resulting in the electrostatic repulsion of insulin-FITC and stimulating its release from the electrode surface. The insulin release was characterized by fluorescence spectroscopy (using the FITC-labeled insulin), by electrochemical measurements on an iridium oxide, IrOx, electrode and by mass spectrometry. The graphene-functionalized carbon fiber electrode demonstrated significant advantages in the signal-stimulated insulin release comparing with the carbon fiber electrode without the graphene species. Y1 - 2017 U6 - http://dx.doi.org/10.1002/elan.201700095 SN - 1521-4109 VL - 29 IS - 6 SP - 1543 EP - 1553 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Zange, Jochen A1 - Schopen, Kathrin A1 - Albracht, Kirsten A1 - Gerlach, Darius A. A1 - Frings-Meuthen, Petra A1 - Maffiuletti, Nicola A. A1 - Bloch, Wilhelm A1 - Rittweger, Jörn T1 - Using the Hephaistos orthotic device to study countermeasure effectiveness of neuromuscular electrical stimulation and dietary lupin protein supplementation, a randomised controlled trial JF - Plos one Y1 - 2017 U6 - http://dx.doi.org/10.1371/journal.pone.0171562 VL - 12 IS - 2 ER - TY - JOUR A1 - Kolditz, Melanie A1 - Albin, Thivaharan A1 - Brüggemann, Gert-Peter A1 - Abel, Dirk A1 - Albracht, Kirsten T1 - Robotergestütztes System für ein verbessertes neuromuskuläres Aufbautraining der Beinstrecker JF - at - Automatisierungstechnik N2 - Neuromuskuläres Aufbautraining der Beinstrecker ist ein wichtiger Bestandteil in der Rehabilitation und Prävention von Muskel-Skelett-Erkrankungen. Effektives Training erfordert hohe Muskelkräfte, die gleichzeitig hohe Belastungen von bereits geschädigten Strukturen bedeuten. Um trainingsinduzierte Schädigungen zu vermeiden, müssen diese Kräfte kontrolliert werden. Mit heutigen Trainingsgeräten können diese Ziele allerdings nicht erreicht werden. Für ein sicheres und effektives Training sollen durch den Einsatz der Robotik, Sensorik, eines Regelkreises sowie Muskel-Skelett-Modellen Belastungen am Zielgewebe direkt berechnet und kontrolliert werden. Auf Basis zweier Vorstudien zu möglichen Stellgrößen wird der Aufbau eines robotischen Systems vorgestellt, das sowohl für Forschungszwecke als auch zur Entwicklung neuartiger Trainingsgeräte verwendet werden kann. Y1 - 2016 U6 - http://dx.doi.org/10.1515/auto-2016-0044 SN - 2196-677X VL - 64 IS - 11 SP - 905 EP - 914 PB - De Gruyter CY - Berlin ER -