TY - JOUR A1 - Bhattarai, Aroj A1 - Jabbari, Medisa A1 - Anding, Ralf A1 - Staat, Manfred T1 - Surgical treatment of vaginal vault prolapse using different prosthetic mesh implants: a finite element analysis JF - tm - Technisches Messen N2 - Particularly multiparous elderly women may suffer from vaginal vault prolapse after hysterectomy due to weak support from lax apical ligaments. A decreased amount of estrogen and progesterone in older age is assumed to remodel the collagen thereby reducing tissue stiffness. Sacrocolpopexy is either performed as open or laparoscopic surgery using prosthetic mesh implants to substitute lax ligaments. Y-shaped mesh models (DynaMesh, Gynemesh, and Ultrapro) are implanted in a 3D female pelvic floor finite element model in the extraperitoneal space from the vaginal cuff to the first sacral (S1) bone below promontory. Numerical simulations are conducted during Valsalva maneuver with weakened tissues modeled by reduced tissue stiffness. Tissues are modeled as incompressible, isotropic hyperelastic materials whereas the meshes are modeled either as orthotropic linear elastic or as isotropic hyperlastic materials. The positions of the vaginal cuff and the bladder base are calculated from the pubococcygeal line for female pelvic floor at rest, for prolapse and after repair using the three meshes. Due to mesh mechanics and mesh pore deformation along the loaded direction, the DynaMesh with regular rectangular mesh pores is found to provide better mechanical support to the organs than the Gynemesh and the Ultrapro with irregular hexagonal mesh pores. Insbesondere ältere, mehrgebährende Frauen leiden häufiger an einem Scheidenvorfall nach einer Hysterektomie aufgrund der schwachen Unterstützung durch laxe apikale Bänder. Es wird angenommen, dass eine verringerte Menge an Östrogen und Progesteron im höheren Alter das Kollagen umformt, wodurch die Gewebesteifigkeit reduziert wird. Die Sakrokolpopexie ist eine offene oder laparoskopische Operation, die mit prothetischen Netzimplantaten durchgeführt wird, um laxe Bänder zu ersetzen. Y-förmige Netzmodelle (DynaMesh, Gynemesh und Ultrapro) werden in einem 3D-Modell des weiblichen Beckenbodens im extraperitonealen Raum vom Vaginalstumpf bis zum Promontorium implantiert. Numerische Simulationen werden während des Valsalva-Manövers mit geschwächtem Gewebe durchgeführt, das durch eine reduzierte Gewebesteifigkeit modelliert wird. Die Gewebe werden als inkompressible, isotrop hyperelastische Materialien modelliert, während die Netze entweder als orthotrope linear elastische oder als isotrope hyperlastische Materialien modelliert werden. Die Positionen des Vaginalstumpfs, der Blase und der Harnröhrenachse werden anhand der Pubococcygeallinie aus der Ruhelage, für den Prolaps und nach der Reparatur unter Verwendung der drei Netze berechnet. Aufgrund der Netzmechanik und der Netzporenverformung bietet das DynaMesh mit regelmäßigen rechteckigen Netzporen eine bessere mechanische Unterstützung und eine Neupositionierung des Scheidengewölbes, der Blase und der Urethraachse als Gynemesh und Ultrapro mit unregelmäßigen hexagonalen Netzporen. Y1 - 2018 U6 - https://doi.org/10.1515/teme-2017-0115 SN - 2196-7113 VL - 85 IS - 5 SP - 331 EP - 342 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Peloni, Alessandro A1 - Dachwald, Bernd A1 - Ceriotti, Matteo T1 - Multiple near-earth asteroid rendezvous mission: Solar-sailing options JF - Advances in Space Research Y1 - 2017 U6 - https://doi.org/10.1016/j.asr.2017.10.017 SN - 0273-1177 IS - In Press, Corrected Proof PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Bhattarai, Aroj A1 - Staat, Manfred T1 - Modelling of Soft Connective Tissues to Investigate Female Pelvic Floor Dysfunctions JF - Computational and Mathematical Methods in Medicine N2 - After menopause, decreased levels of estrogen and progesterone remodel the collagen of the soft tissues thereby reducing their stiffness. Stress urinary incontinence is associated with involuntary urine leakage due to pathological movement of the pelvic organs resulting from lax suspension system, fasciae, and ligaments. This study compares the changes in the orientation and position of the female pelvic organs due to weakened fasciae, ligaments, and their combined laxity. A mixture theory weighted by respective volume fraction of elastin-collagen fibre compound (5%), adipose tissue (85%), and smooth muscle (5%) is adopted to characterize the mechanical behaviour of the fascia. The load carrying response (other than the functional response to the pelvic organs) of each fascia component, pelvic organs, muscles, and ligaments are assumed to be isotropic, hyperelastic, and incompressible. Finite element simulations are conducted during Valsalva manoeuvre with weakened tissues modelled by reduced tissue stiffness. A significant dislocation of the urethrovesical junction is observed due to weakness of the fascia (13.89 mm) compared to the ligaments (5.47 mm). The dynamics of the pelvic floor observed in this study during Valsalva manoeuvre is associated with urethral-bladder hypermobility, greater levator plate angulation, and positive Q-tip test which are observed in incontinent females. Y1 - 2018 U6 - https://doi.org/10.1155/2018/9518076 SN - 1748-6718 VL - 2018 IS - Article ID 9518076 SP - 1 EP - 16 PB - Hindawi CY - New York, NY ER - TY - JOUR A1 - Werkhausen, Amelie A1 - Albracht, Kirsten A1 - Cronin, Neil J. A1 - Meier, Rahel A1 - Mojsen-Moeller, Jens A1 - Seynnes, Olivier R. T1 - Modulation of muscle-tendon interaction in the human triceps surae during an energy dissipation task JF - Journal of Experimental Biology Y1 - 2017 U6 - https://doi.org/10.1242/jeb.164111 SN - 0022-0949 VL - 220 IS - 22 SP - 4141 EP - 4149 ER - TY - JOUR A1 - Kotliar, Konstantin A1 - Hauser, Christine A1 - Ortner, Marion A1 - Muggenthaler, Claudia A1 - Diehl-Schmid, Janine A1 - Angermann, Susanne A1 - Hapfelmeier, Alexander A1 - Schmaderer, Christoph A1 - Grimmer, Timo T1 - Altered neurovascular coupling as measured by optical imaging: a biomarker for Alzheimer’s disease JF - Scientific Reports Y1 - 2017 U6 - https://doi.org/10.1038/s41598-017-13349-5 SN - 2045-2322 N1 - Article 12906 VL - 7 IS - 1 SP - 1 EP - 11 PB - Springer Nature CY - Cham ER - TY - JOUR A1 - Seifarth, Volker A1 - Grosse, Joachim O. A1 - Grossmann, Matthias A1 - Janke, Heinz Peter A1 - Arndt, Patrick A1 - Koch, Sabine A1 - Epple, Matthias A1 - Artmann, Gerhard A1 - Temiz Artmann, Aysegül T1 - Mechanical induction of bi-directional orientation of primary porcine bladder smooth muscle cells in tubular fibrin-poly(vinylidene fluoride) scaffolds for ureteral and urethral repair using cyclic and focal balloon catheter stimulation JF - Journal of Biomaterials Applications Y1 - 2017 U6 - https://doi.org/10.1177/0885328217723178 SN - 1530-8022 VL - 32 IS - 3 SP - 321 EP - 330 PB - Sage CY - London ER - TY - JOUR A1 - Albanna, Walid A1 - Lueke, Jan Niklas A1 - Sjapic, Volha A1 - Kotliar, Konstantin A1 - Hescheler, Jürgen A1 - Clusmann, Hans A1 - Sjapic, Sergej A1 - Alpdogan, Serdan A1 - Schneider, Toni A1 - Schubert, Gerrit Alexander A1 - Neumaier, Felix T1 - Electroretinographic Assessment of Inner Retinal Signaling in the Isolated and Superfused Murine Retina JF - Current Eye Research Y1 - 2017 U6 - https://doi.org/10.1080/02713683.2017.1339807 SN - 1460-2202 IS - Article in press SP - 1 EP - 9 PB - Taylor & Francis CY - London ER - TY - JOUR A1 - Tran, Linda A1 - Mottaghy, K. A1 - Arlt-Körfer, Sabine A1 - Waluga, Christian A1 - Behbahani, Mehdi T1 - An experimental study of shear-dependent human platelet adhesion and underlying protein-binding mechanisms in a cylindrical Couette system JF - Biomedizinische Technik Y1 - 2017 U6 - https://doi.org/10.1515/bmt-2015-0034 SN - 0013-5585 VL - 62 IS - 4 SP - 383 EP - 392 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Horbach, Andreas A1 - Duong, Minh Tuan A1 - Staat, Manfred T1 - Modelling of compressible and orthotropic surgical mesh implants based on optical deformation measurement JF - Journal of the mechanical behavior of biomedical materials Y1 - 2017 U6 - https://doi.org/10.1016/j.jmbbm.2017.06.012 SN - 1751-6161 VL - 74 SP - 400 EP - 410 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Michael, Hackl A1 - Mayer, Katharina A1 - Weber, Mareike A1 - Staat, Manfred A1 - van Riet, Roger A1 - Burkhart, Klau Josef A1 - Müller, Lars Peter A1 - Wegmann, Kilian T1 - Plate osteosynthesis of proximal ulna fractures : a biomechanical micromotion analysis JF - The journal of hand surgery Y1 - 2017 U6 - https://doi.org/10.1016/j.jhsa.2017.05.014 SN - 0363-5023 VL - 42 IS - 10 SP - 834.e1 EP - 834.e7 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Ciritsis, Alexander A1 - Horbach, Andreas A1 - Staat, Manfred A1 - Kuhl, Christiane K. A1 - Kraemer, Nils Andreas T1 - Porosity and tissue integration of elastic mesh implants evaluated in vitro and in vivo JF - Journal of Biomedical Materials Research: Part B: Applied Biomaterials N2 - Purpose In vivo, a loss of mesh porosity triggers scar tissue formation and restricts functionality. The purpose of this study was to evaluate the properties and configuration changes as mesh deformation and mesh shrinkage of a soft mesh implant compared with a conventional stiff mesh implant in vitro and in a porcine model. Material and Methods Tensile tests and digital image correlation were used to determine the textile porosity for both mesh types in vitro. A group of three pigs each were treated with magnetic resonance imaging (MRI) visible conventional stiff polyvinylidene fluoride meshes (PVDF) or with soft thermoplastic polyurethane meshes (TPU) (FEG Textiltechnik mbH, Aachen, Germany), respectively. MRI was performed with a pneumoperitoneum at a pressure of 0 and 15 mmHg, which resulted in bulging of the abdomen. The mesh-induced signal voids were semiautomatically segmented and the mesh areas were determined. With the deformations assessed in both mesh types at both pressure conditions, the porosity change of the meshes after 8 weeks of ingrowth was calculated as an indicator of preserved elastic properties. The explanted specimens were examined histologically for the maturity of the scar (collagen I/III ratio). Results In TPU, the in vitro porosity increased constantly, in PVDF, a loss of porosity was observed under mild stresses. In vivo, the mean mesh areas of TPU were 206.8 cm2 (± 5.7 cm2) at 0 mmHg pneumoperitoneum and 274.6 cm2 (± 5.2 cm2) at 15 mmHg; for PVDF the mean areas were 205.5 cm2 (± 8.8 cm2) and 221.5 cm2 (± 11.8 cm2), respectively. The pneumoperitoneum-induced pressure increase resulted in a calculated porosity increase of 8.4% for TPU and of 1.2% for PVDF. The mean collagen I/III ratio was 8.7 (± 0.5) for TPU and 4.7 (± 0.7) for PVDF. Conclusion The elastic properties of TPU mesh implants result in improved tissue integration compared to conventional PVDF meshes, and they adapt more efficiently to the abdominal wall. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 827–833, 2018. Y1 - 2018 U6 - https://doi.org/10.1002/jbm.b.33877 SN - 1552-4981 VL - 106 IS - 2 SP - 827 EP - 833 PB - Wiley CY - New York, NY ER - TY - JOUR A1 - Zange, Jochen A1 - Schopen, Kathrin A1 - Albracht, Kirsten A1 - Gerlach, Darius A. A1 - Frings-Meuthen, Petra A1 - Maffiuletti, Nicola A. A1 - Bloch, Wilhelm A1 - Rittweger, Jörn T1 - Using the Hephaistos orthotic device to study countermeasure effectiveness of neuromuscular electrical stimulation and dietary lupin protein supplementation, a randomised controlled trial JF - Plos one Y1 - 2017 U6 - https://doi.org/10.1371/journal.pone.0171562 VL - 12 IS - 2 ER - TY - JOUR A1 - Hackl, Michael A1 - Wegmann, Kilian A1 - Kahmann, Stephanie Lucina A1 - Heinze, Nicolai A1 - Staat, Manfred A1 - Neiss, Wolfram F. A1 - Scaal, Martin A1 - Müller, Lars P. T1 - Radial shortening osteotomy reduces radiocapitellar contact pressures while preserving valgus stability of the elbow JF - Knee Surgery, Sports Traumatology, Arthroscopy Y1 - 2017 U6 - https://doi.org/10.1007/s00167-017-4468-z SN - 1433-7347 VL - 25 IS - 7 SP - 2280 EP - 2288 PB - Springer CY - Berlin ER - TY - JOUR A1 - Albracht, Kirsten A1 - Arampatzis, Adamantios T1 - Influence of the mechanical properties of the muscle-tendon unit on force generation in runners with different running economy JF - Biological Cybernetics Y1 - 2006 U6 - https://doi.org/10.1007/s00422-006-0070-z SN - 1432-0770 VL - 95 IS - 1 SP - 87 EP - 96 ER - TY - JOUR A1 - Arampatzis, Adamantios A1 - Karamanidis, Kiros A1 - Albracht, Kirsten T1 - Adaptational responses of the human Achilles tendon by modulation of the applied cyclic strain magnitude JF - Journal of Experimental Biology Y1 - 2007 U6 - https://doi.org/10.1242/jeb.003814 SN - 0022-0949 VL - 210 IS - 15 SP - 2743 EP - 2753 ER - TY - JOUR A1 - Albracht, Kirsten A1 - Arampatzis, A. A1 - Baltzopoulos, V. T1 - Assessment of muscle volume and physiological cross-sectional area of the human triceps surae muscle in vivo JF - Journal of Biomechanics Y1 - 2008 U6 - https://doi.org/10.1016/j.jbiomech.2008.04.020 SN - 0021-9290 VL - 41 SP - 2211 EP - 2218 ER - TY - JOUR A1 - Arampatzis, Adamantios A1 - Karamanidis, Kiros A1 - Mademli, Lida A1 - Albracht, Kirsten T1 - Plasticity of the human tendon to short and long-term mechanical loading JF - Exercise and Sport Sciences Reviews Y1 - 2009 U6 - https://doi.org/10.1097/JES.0b013e31819c2e1d SN - 1538-3008 VL - 37 IS - 2 SP - 66 EP - 72 ER - TY - JOUR A1 - Aggeloussis, Nickos A1 - Giannakou, Erasmia A1 - Albracht, Kirsten A1 - Arampatzis, Adamantios T1 - Reproducibility of fascicle length and pennation angle of gastrocnemius medialis in human gait in vivo JF - Gait and Posture N2 - The purpose of the current study was to examine the reproducibility of fascicle length and pennation angle of gastrocnemius medialis while human walking. To the best of our knowledge, this is the first study of the reproducibility of fascicle length and pennation angle of gastrocnemius medialis in vivo during human gait. Twelve males performed 10 gait trials on a treadmill, in 2 separate days. B-mode ultrasonography, with the ultrasound probe firmly adjusted in the transverse and frontal planes using a special cast, was used to measure the fascicle length and the pennation angle of the gastrocnemius medialis (GM). A Vicon 624 system with three cameras operating at 120 Hz was also used to record the ankle and knee joint angles. The results showed that measurements of fascicle length and pennation angle showed high reproducibility during the gait cycle, both within the same day and between different days. Moreover, the root mean square differences between the repeated waveforms of both variables were very small, compared with their ranges (fascicle length: RMS = ∼3 mm, range: 38–63 mm; pennation angle: RMS = ∼1.5°, range: 22–32°). However, their reproducibility was lower compared to the joint angles. It was found that representative data have to be derived by a wide number of gait trials (fascicle length ∼six trials, pennation angle more than 10 trials), to assure the reliability of the fascicle length and pennation angle in human gait. Y1 - 2010 U6 - https://doi.org/10.1016/j.gaitpost.2009.08.249 SN - 0966-6362 VL - 31 IS - 1 SP - 73 EP - 77 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Arampatzis, Adamantios A1 - Peper, Andreas A1 - Bierbaum, Stefanie A1 - Albracht, Kirsten T1 - Plasticity of human Achilles tendon mechanical and morphological properties in response to cyclic strain JF - Journal of Biomechanics N2 - The purpose of the current study in combination with our previous published data (Arampatzis et al., 2007) was to examine the effects of a controlled modulation of strain magnitude and strain frequency applied to the Achilles tendon on the plasticity of tendon mechanical and morphological properties. Eleven male adults (23.9±2.2 yr) participated in the study. The participants exercised one leg at low magnitude tendon strain (2.97±0.47%), and the other leg at high tendon strain magnitude (4.72±1.08%) of similar frequency (0.5 Hz, 1 s loading, 1 s relaxation) and exercise volume (integral of the plantar flexion moment over time) for 14 weeks, 4 days per week, 5 sets per session. The exercise volume was similar to the intervention of our earlier study (0.17 Hz frequency; 3 s loading, 3 s relaxation) allowing a direct comparison of the results. Before and after the intervention ankle joint moment has been measured by a dynamometer, tendon–aponeurosis elongation by ultrasound and cross-sectional area of the Achilles tendon by magnet resonance images (MRI). We found a decrease in strain at a given tendon force, an increase in tendon–aponeurosis stiffness and tendon elastic modulus of the Achilles tendon only in the leg exercised at high strain magnitude. The cross-sectional area (CSA) of the Achilles tendon did not show any statistically significant (P>0.05) differences to the pre-exercise values in both legs. The results indicate a superior improvement in tendon properties (stiffness, elastic modulus and CSA) at the low frequency (0.17 Hz) compared to the high strain frequency (0.5 Hz) protocol. These findings provide evidence that the strain magnitude applied to the Achilles tendon should exceed the value, which occurs during habitual activities to trigger adaptational effects and that higher tendon strain duration per contraction leads to superior tendon adaptational responses. Y1 - 2010 U6 - https://doi.org/10.1016/j.jbiomech.2010.08.014 SN - 0021-9290 VL - 43 IS - 16 SP - 3073 EP - 3079 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Karamanidis, Kiros A1 - Albracht, Kirsten A1 - Braunstein, Bjoern A1 - Catala, Maria Moreno A1 - Goldmann, Jan-Peter A1 - Brüggemann, Gert-Peter T1 - Lower leg musculoskeletal geometry and sprint performance JF - Gait and Posture N2 - The purpose of this study was to investigate whether sprint performance is related to lower leg musculoskeletal geometry within a homogeneous group of highly trained 100-m sprinters. Using a cluster analysis, eighteen male sprinters were divided into two groups based on their personal best (fast: N = 11, 10.30 ± 0.07 s; slow: N = 7, 10.70 ± 0.08 s). Calf muscular fascicle arrangement and Achilles tendon moment arms (calculated by the gradient of tendon excursion versus ankle joint angle) were analyzed for each athlete using ultrasonography. Achilles tendon moment arm, foot and ankle skeletal geometry, fascicle arrangement as well as the ratio of fascicle length to Achilles tendon moment arm showed no significant (p > 0.05) correlation with sprint performance, nor were there any differences in the analyzed musculoskeletal parameters between the fast and slow sprinter group. Our findings provide evidence that differences in sprint ability in world-class athletes are not a result of differences in the geometrical design of the lower leg even when considering both skeletal and muscular components. Y1 - 2011 U6 - https://doi.org/10.1016/j.gaitpost.2011.03.009 SN - 0966-6362 VL - 34 IS - 1 SP - 138 EP - 141 PB - Elsevier CY - Amsterdam ER -