TY - JOUR A1 - Kraff, Oliver A1 - Bitz, Andreas A1 - Breyer, Tobias A1 - Kruszona, Stefan A1 - Maderwald, Stefan A1 - Brote, Irina A1 - Gizewski, Elke R. A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - A transmit/receive radiofrequency array for imaging the carotid arteries at 7 Tesla: coil design and first in vivo results JF - Investigative Radiology N2 - Objective: To develop a transmit/receive radiofrequency (RF) array for magnetic resonance imaging (MRI) of the carotid arteries at 7 T. The prototype is characterized in numerical simulations and bench measurements, and the feasibility of plaque imaging at 7 T is demonstrated in first in vivo images. Materials and Methods: The RF phased array coil consists of 8 surface loop coils. To allow imaging of both sides of the neck, the RF array is divided into 2 coil clusters, each with 4 overlapping loop elements. For safety validation, numerical computations of the RF field distribution and the corresponding specific absorption rate were performed on the basis of a heterogeneous human body model. To validate the coil model, maps of the transmit B1+ field were compared between simulation and measurement. In vivo images of a healthy volunteer and a patient (ulcerating plaque and a 50% stenosis of the right internal carotid artery) were acquired using a 3-dimensional FLASH sequence with a high isotropic spatial resolution of 0.54 mm as well as using pulse-triggered proton density (PD)/T2-weighted turbo spin echo sequences. Results: Measurements of the S-parameters yielded a reflection and isolation of the coil elements of better than −18 and −13 dB, respectively. Measurements of the g-factor indicated good image quality for parallel imaging acceleration factors up to 2.4. A similar distribution and a very good match of the absolute values were found between the measured and simulated B1+ transmit RF field for the validation of the coil model. In vivo images revealed good signal excitation of both sides of the neck and a high vessel-to-background image contrast for the noncontrast-enhanced 3-dimensional FLASH sequence. Imaging at 7 T could depict the extent of stenosis, and revealed the disruption and ulcer of the plaque. Conclusions: This study demonstrates that 2 four-channel transmit/receive RF arrays for each side of the neck is a suitable concept for in vivo MRI of the carotid arteries at 7 Tesla. Further studies are needed to explore and exploit the full potential of 7 T high-field MRI for carotid atherosclerotic plaque imaging. Y1 - 2011 U6 - http://dx.doi.org/10.1097/RLI.0b013e318206cee4 SN - 1536-0210 VL - 46 IS - 4 SP - 246 EP - 254 PB - Wolters Kluwer CY - Köln ER - TY - JOUR A1 - Kobus, Thiele A1 - Bitz, Andreas A1 - Uden, Mark J. van A1 - Lagemaat, Miram W. A1 - Rothgang, Eva A1 - Orzada, Stephan A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - In vivo 31P MR spectroscopic imaging of the human prostate at 7 T: safety and feasibility JF - Magnetic Resonance in Medicine N2 - 31P MR spectroscopic imaging of the human prostate provides information about phosphorylated metabolites that could be used for prostate cancer characterization. The sensitivity of a magnetic field strength of 7 T might enable 3D 31P MR spectroscopic imaging with relevant spatial resolution in a clinically acceptable measurement time. To this end, a 31P endorectal coil was developed and combined with an eight-channel 1H body-array coil to relate metabolic information to anatomical location. An extensive safety validation was performed to evaluate the specific absorption rate, the radiofrequency field distribution, and the temperature distribution of both coils. This validation consisted of detailed Finite Integration Technique simulations, confirmed by MR thermometry and Burn:x-wiley:07403194:media:MRM24175:tex2gif-stack-1 measurements in a phantom and in vivo temperature measurements. The safety studies demonstrated that the presence of the 31P endorectal coil had no influence on the specific absorption rate levels and temperature distribution of the external eight-channel 1H array coil. To stay within a 10 g averaged local specific absorption rate of 10 W/kg, a maximum time-averaged input power of 33 W for the 1H array coil was allowed. For transmitting with the 31P endorectal coil, our safety limit of less than 1°C temperature increase in vivo during a 15-min MR spectroscopic imaging experiment was reached at a time-averaged input power of 1.9 W. With this power setting, a second in vivo measurement was performed on a healthy volunteer. Using adiabatic excitation, 3D 31P MR spectroscopic imaging produced spectra from the entire prostate in 18 min with a spatial resolution of 4 cm3. The spectral resolution enabled the separate detection of phosphocholine, phosphoethanolamine, inorganic phosphate, and other metabolites that could play an important role in the characterization of prostate cancer. Y1 - 2012 U6 - http://dx.doi.org/10.1002/mrm.24175 SN - 1522-2594 VL - 68 IS - 6 SP - 1683 EP - 1695 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Klomp, D. W. J. A1 - Bitz, Andreas A1 - Heerschap, A. A1 - Scheenen, T. W. J. T1 - Proton spectroscopic imaging of the human prostate at 7 T JF - NMR in Biomedicine Y1 - 2009 U6 - http://dx.doi.org/10.1002/nbm.1360 SN - 1099-1492 VL - 22 IS - 5 SP - 495 EP - 501 ER - TY - JOUR A1 - Hansen, Volkert W. A1 - Bitz, Andreas A1 - Streckert, Joachim R. T1 - RF Exposure of Biological Systems in Radial Waveguides JF - IEEE Transactions on Electromagnetic Compatibility Y1 - 1999 U6 - http://dx.doi.org/10.1109/15.809852 SN - 1558-187X VL - 41 IS - 4 SP - 487 EP - 493 ER - TY - JOUR A1 - Franke, Helmut A1 - Streckert, Joachim A1 - Bitz, Andreas A1 - Goeke, Johannes A1 - Hansen, Volkert A1 - Ringelstein, E. Bernd A1 - Nattkämper, Heiner A1 - Galle, Hans-Joachim A1 - Stögbauer, Florian T1 - Effects of Universal Mobile Telecommunications System (UMTS) electromagnetic fields on the blood-brain barrier in vitro JF - Radiation Research Y1 - 2005 U6 - http://dx.doi.org/10.1667/RR3424.1 SN - 1938-5404 VL - 164 IS - 3 SP - 258 EP - 269 ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Schmidt, Simon A1 - Stelter, Jonathan K. A1 - Wittrich, Marco A1 - Quick, Harald H. A1 - Bitz, Andreas A1 - Ladd, Mark E. T1 - Performance and safety assessment of an integrated transmitarray for body imaging at 7 T under consideration of specificabsorption rate, tissue temperature, and thermal dose JF - NMR in Biomedicine N2 - In this study, the performance of an integrated body-imaging array for 7 T with 32 radiofrequency (RF) channels under consideration of local specific absorption rate (SAR), tissue temperature, and thermal dose limits was evaluated and the imaging performance was compared with a clinical 3 T body coil. Thirty-two transmit elements were placed in three rings between the bore liner and RF shield of the gradient coil. Slice-selective RF pulse optimizations for B1 shimming and spokes were performed for differently oriented slices in the body under consideration of realistic constraints for power and local SAR. To improve the B1+ homogeneity, safety assessments based on temperature and thermal dose were performed to possibly allow for higher input power for the pulse optimization than permissible with SAR limits. The results showed that using two spokes, the 7 T array outperformed the 3 T birdcage in all the considered regions of interest. However, a significantly higher SAR or lower duty cycle at 7 T is necessary in some cases to achieve similar B1+ homogeneity as at 3 T. The homogeneity in up to 50 cm-long coronal slices can particularly benefit from the high RF shim performance provided by the 32 RF channels. The thermal dose approach increases the allowable input power and the corresponding local SAR, in one example up to 100 W/kg, without limiting the exposure time necessary for an MR examination. In conclusion, the integrated antenna array at 7 T enables a clinical workflow for body imaging and comparable imaging performance to a conventional 3 T clinical body coil. KW - body imaging at 7 T MRI KW - thermal dose KW - tissue temperature KW - transmit antenna arrays Y1 - 2022 U6 - http://dx.doi.org/10.1002/nbm.4656 SN - 0952-3480 (Print) SN - 1099-1492 (Online) VL - 35 IS - 5 SP - 1 EP - 17 PB - Wiley ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Clemens, Markus A1 - Bitz, Andreas T1 - Safety of subjects during radiofrequency exposure in ultra-high-field magnetic resonance imaging JF - IEEE Letters on Electromagnetic Compatibility Practice and Applications N2 - Magnetic resonance imaging (MRI) is one of the most important medical imaging techniques. Since the introduction of MRI in the mid-1980s, there has been a continuous trend toward higher static magnetic fields to obtain i.a. a higher signal-to-noise ratio. The step toward ultra-high-field (UHF) MRI at 7 Tesla and higher, however, creates several challenges regarding the homogeneity of the spin excitation RF transmit field and the RF exposure of the subject. In UHF MRI systems, the wavelength of the RF field is in the range of the diameter of the human body, which can result in inhomogeneous spin excitation and local SAR hotspots. To optimize the homogeneity in a region of interest, UHF MRI systems use parallel transmit systems with multiple transmit antennas and time-dependent modulation of the RF signal in the individual transmit channels. Furthermore, SAR increases with increasing field strength, while the SAR limits remain unchanged. Two different approaches to generate the RF transmit field in UHF systems using antenna arrays close and remote to the body are investigated in this letter. Achievable imaging performance is evaluated compared to typical clinical RF transmit systems at lower field strength. The evaluation has been performed under consideration of RF exposure based on local SAR and tissue temperature. Furthermore, results for thermal dose as an alternative RF exposure metric are presented. Y1 - 2020 SN - 2637-6423 U6 - http://dx.doi.org/10.1109/LEMCPA.2020.3029747 VL - 2 IS - 3 SP - 1 EP - 8 PB - IEEE CY - New York, NY ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - SAR Simulations & Safety JF - NeuroImage Y1 - 2017 U6 - http://dx.doi.org/10.1016/j.neuroimage.2017.03.035 SN - 1053-8119 IS - Epub ahead of print PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus temperature limits JF - Medical Physics Y1 - 2017 U6 - http://dx.doi.org/10.1002/mp.12034 N1 - This article is corrected by: Errata: Erratum: “RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus tissue temperature limits” [Med. Phys. 44(1), 143–157 (2017)] Volume 44, Issue 2, 772 VL - 44 IS - 1 SP - 143 EP - 157 ER - TY - JOUR A1 - Fagan, Andrew J. A1 - Bitz, Andreas A1 - Björkman-Burtscher, Isabella M. A1 - Collins, Christopher M. A1 - Kimbrell, Vera A1 - Raaijmakers, Alexander J. E. T1 - 7T MR Safety JF - Journal of Magnetic Resonance Imaging (JMRI) Y1 - 2021 U6 - http://dx.doi.org/10.1002/jmri.27319 SN - 1522-2586 VL - 53 IS - 2 SP - 333 EP - 346 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - El Quardi, A. A1 - Streckert, J. A1 - Bitz, Andreas A1 - Münkner, S. A1 - Engel, J. A1 - Hansen, V. T1 - New fin-line devices for radiofrequency exposure of small biological samples in vitro allowing whole-cell patch clamp recordings JF - Bioelectromagnetics N2 - The development and analysis of three waveguides for the exposure of small biological in vitro samples to mobile communication signals at 900 MHz (GSM, Global System for Mobile Communications), 1.8 GHz (GSM), and 2 GHz (UMTS, Universal Mobile Telecommunications System) is presented. The waveguides were based on a fin-line concept and the chamber containing the samples bathed in extracellular solution was placed onto two fins with a slot in between, where the exposure field concentrates. Measures were taken to allow for patch clamp recordings during radiofrequency (RF) exposure. The necessary power for the achievement of the maximum desired specific absorption rate (SAR) of 20 W/kg (average over the mass of the solution) was approximately Pin = 50 mW, Pin = 19 mW, and Pin = 18 mW for the 900 MHz, 1800 MHz, and 2 GHz devices, respectively. At 20 W/kg, a slight RF-induced temperature elevation in the solution of no more than 0.3 °C was detected, while no thermal offsets due to the electromagnetic exposure could be detected at the lower SAR settings (2, 0.2, and 0.02 W/kg). A deviation of 10% from the intended solution volume yielded a calculated SAR deviation of 8% from the desired value. A maximum ±10% variation in the local SAR could occur when the position of the patch clamp electrode was altered within the area where the cells to be investigated were located. Y1 - 2011 U6 - http://dx.doi.org/10.1002/bem.20621 SN - 1521-186X VL - 32 IS - 2 SP - 102 EP - 112 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Chen, Bixia A1 - Schoemberg, Tobias A1 - Kraff, Oliver A1 - Dammann, Philipp A1 - Bitz, Andreas A1 - Schlamann, Marc A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Sure, Ulrich A1 - Wrede, Karsten H. T1 - Cranial fixation plates in cerebral magnetic resonance imaging: a 3 and 7 Tesla in vivo image quality study JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective This study assesses and quantifies impairment of postoperative magnetic resonance imaging (MRI) at 7 Tesla (T) after implantation of titanium cranial fixation plates (CFPs) for neurosurgical bone flap fixation. Materials and methods The study group comprised five patients who were intra-individually examined with 3 and 7 T MRI preoperatively and postoperatively (within 72 h/3 months) after implantation of CFPs. Acquired sequences included T₁-weighted magnetization-prepared rapid-acquisition gradient-echo (MPRAGE), T₂-weighted turbo-spin-echo (TSE) imaging, and susceptibility-weighted imaging (SWI). Two experienced neurosurgeons and a neuroradiologist rated image quality and the presence of artifacts in consensus reading. Results Minor artifacts occurred around the CFPs in MPRAGE and T2 TSE at both field strengths, with no significant differences between 3 and 7 T. In SWI, artifacts were accentuated in the early postoperative scans at both field strengths due to intracranial air and hemorrhagic remnants. After resorption, the brain tissue directly adjacent to skull bone could still be assessed. Image quality after 3 months was equal to the preoperative examinations at 3 and 7 T. Conclusion Image quality after CFP implantation was not significantly impaired in 7 T MRI, and artifacts were comparable to those in 3 T MRI. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s10334-016-0548-1 SN - 1352-8661 VL - 29 IS - 3 SP - 389 EP - 398 PB - Springer CY - Berlin ER - TY - JOUR A1 - Bitz, Andreas A1 - Zhou, Yi A1 - El Quardi, Abdessamad A1 - Streckert, Joachim T1 - Occupational Exposure at Mobile Communication Base Station Antenna Sites JF - Frequenz Y1 - 2009 U6 - http://dx.doi.org/10.1515/FREQ.2009.63.7-8.123 SN - 2191-6349 VL - 63 IS - 7-8 SP - 123 EP - 128 ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J.R. A1 - Hansen, V.W. A1 - Lerchl, A. ED - Danesy, Dorothea T1 - Freely moving or restrained animals in bioelec-tromagnetic experiments - pros and cons T2 - AP 2000 : Millennium Conference on Antennas & Propagation, Davos, Switzerland, 9 - 14 April 2000 Y1 - 2000 IS - Band 1 SP - 489 PB - ESA Publications Division, ESTEC CY - Noordwijk ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Stogbauer, F. ED - Hietanen, Maila ED - Jokela, Kari ED - Juutilainen, Jukka ED - Työterveyslaitos, T1 - Shielded RF-exposure system with an integrated LF-measuring system for in vitro experiments with cell-layers T2 - Proceedings of the Ebea 2001 : 5th International Congress of the European Bioelectromagnetics Association (Ebea) ; 6-8 September 2001, Marina Congress Center, Helsinki, Finland Y1 - 2001 SN - 9789518024401 SP - 67 EP - 68 PB - Finnish Institute of Occupational Health CY - Helsinki ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Buschmann, J. T1 - RF exposure of non restrained animals in an overmoded radial waveguide T2 - 22nd BEMS Annual Meeting, Munich, Germany, 2000 Y1 - 2000 SP - 63 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Kraff, O. A1 - Orzada, S. A1 - Maderwald, S. A1 - Brote, I. A1 - Ladd, M. T1 - Experimental and Numerical Assessment of RF Safety of Transmit Coils at 7 Tesla T2 - ISMRM workshop on MR safety 2010 : RF heating of the human in MRI : workshop series. The Washington County Historic Courthouse, Stillwater, Minnesota, USA, 15 - 17 October 2010 Y1 - 2010 SN - 978-1-62276-088-6 SP - 195 ER - TY - CHAP A1 - Bitz, Andreas A1 - Kraff, O. A1 - Orzada, S. A1 - Maderwald, S. A1 - Brote, I. A1 - Johst, S. A1 - Ladd, E. T1 - Assessment of RF Safety of Transmit Coils at 7 Tesla by Experimental and Numerical Procedures (490.) T2 - 19th annual ISMRM scientific meeting and exhibition 2011 : Montreal, Quebec, Canada, 7 - 13 May 2011 Y1 - 2012 SN - 978-1-61839-284-8 IS - Volume 1 SP - 475 PB - Curran CY - Red Hook, NY ER - TY - CHAP A1 - Bitz, Andreas A1 - Kobus, Thiele A1 - Scheenen, Tom W. J. A1 - Ladd, Mark E. T1 - RF Safety of the Combination of a 31P Tx/Rx Endorectal Coil & a 1H Tx/Rx Body Array for 31P MRSI of the Prostate at 7T (311.) T2 - 20th Annual ISMRM scientific meeting and exhibition 2012 : Melbourne, Australia, 5 - 11 May 2012 Y1 - 2013 SN - 978-1-62276-943-8 SN - 1545-4428 IS - Volume 1 SP - 311 PB - Curran CY - Red Hook, NY ER - TY - CHAP A1 - Bitz, Andreas A1 - Klomp, D.W. A1 - Ladd, M.E. T1 - Experimental and numerical determination of SAR and temperature distribution of a human endorectal coil for MR imaging of the prostate at 7T (903.) T2 - 16th annual ISMRM scientific meeting and exhibition 2008 : Toronto, Ontario, Canada, 3 - 9 May 2008 Y1 - 2009 SN - 978-1-61567-196-0 PB - Curran CY - Red Hook, NY ER -