TY - JOUR A1 - Janus, Kevin Alexander A1 - Achtsnicht, Stefan A1 - Drinic, Aleksander A1 - Kopp, Alexander A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Transient magnesium-based thin-film temperature sensor on a flexible, bioabsorbable substrate for future medical applications JF - Applied Research N2 - In this work, the bioabsorbable materials, namely fibroin, polylactide acid (PLA), magnesium and magnesium oxide are investigated for their application as transient, resistive temperature detectors (RTD). For this purpose, a thin-film magnesium-based meander-like electrode is deposited onto a flexible, bioabsorbable substrate (fibroin or PLA) and encapsulated (passivated) by additional magnesium oxide layers on top and below the magnesium-based electrode. The morphology of different layered RTDs is analyzed by scanning electron microscopy. The sensor performance and lifetime of the RTD is characterized both under ambient atmospheric conditions between 30°C and 43°C, and wet tissue-like conditions with a constant temperature regime of 37°C. The latter triggers the degradation process of the magnesium-based layers. The 3-layers RTDs on a PLA substrate could achieve a lifetime of 8.5 h. These sensors also show the best sensor performance under ambient atmospheric conditions with a mean sensitivity of 0.48 Ω/°C ± 0.01 Ω/°C. KW - Silk fibroin KW - Polylactide acid KW - Bioabsorbable KW - Resistive temperature detector Y1 - 2023 U6 - http://dx.doi.org/10.1002/appl.202300102 SN - 2702-4288 (Print) N1 - Corresponding author: Michael Josef Schöning IS - Accepted manuscript PB - Wiley-VCH ER - TY - JOUR A1 - Janus, Kevin Alexander A1 - Achtsnicht, Stefan A1 - Tempel, Laura A1 - Drinic, Aleksaner A1 - Kopp, Alexander A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Influence of fibroin membrane composition and curing parameters on the performance of a biodegradable enzymatic biosensor manufactured from Silicon-Free Carbon JF - Physica status solidi : pss. A, Applications and materials science N2 - Herein, fibroin, polylactide (PLA), and carbon are investigated for their suitability as biocompatible and biodegradable materials for amperometric biosensors. For this purpose, screen-printed carbon electrodes on the biodegradable substrates fibroin and PLA are modified with a glucose oxidase membrane and then encapsulated with the biocompatible material Ecoflex. The influence of different curing parameters of the carbon electrodes on the resulting biosensor characteristics is studied. The morphology of the electrodes is investigated by scanning electron microscopy, and the biosensor performance is examined by amperometric measurements of glucose (0.5–10 mM) in phosphate buffer solution, pH 7.4, at an applied potential of 1.2 V versus a Ag/AgCl reference electrode. Instead of Ecoflex, fibroin, PLA, and wound adhesive are tested as alternative encapsulation compounds: a series of swelling tests with different fibroin compositions, PLA, and Ecoflex has been performed before characterizing the most promising candidates by chronoamperometry. Therefore, the carbon electrodes are completely covered with the particular encapsulation material. Chronoamperometric measurements with H2O2 concentrations between 0.5 and 10 mM enable studying the leakage current behavior. KW - amperometric biosensors KW - biocompatible KW - biodegradabl KW - encapsulation materials KW - fibroin Y1 - 2023 U6 - http://dx.doi.org/10.1002/pssa.202300081 SN - 1862-6300 (Print) SN - 1862-6319 (Online) N1 - Corresponding author: Michael J. Schöning VL - 220 IS - 22 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Kirchner, Patrick A1 - Baltes, Klaus A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - Development of an in-line evaporation unit for the production of gas mixtures containing hydrogen peroxide – numerical modeling and experimental results JF - International Journal of Heat and Mass Transfer N2 - Hydrogen peroxide (H2O2) is a typical surface sterilization agent for packaging materials used in the pharmaceutical, food and beverage industries. We use the finite-elements method to analyze the conceptual design of an in-line thermal evaporation unit to produce a heated gas mixture of air and evaporated H2O2 solution. For the numerical model, the required phase-transition variables of pure H2O2 solution and of the aerosol mixture are acquired from vapor-liquid equilibrium (VLE) diagrams derived from vapor-pressure formulations. This work combines homogeneous single-phase turbulent flow with heat-transfer physics to describe the operation of the evaporation unit. We introduce the apparent heat-capacity concept to approximate the non-isothermal phase-transition process of the H2O2-containing aerosol. Empirical and analytical functions are defined to represent the temperature- and pressure-dependent material properties of the aqueous H2O2 solution, the aerosol and the gas mixture. To validate the numerical model, the simulation results are compared to experimental data on the heating power required to produce the gas mixture. This shows good agreement with the deviations below 10%. Experimental observations on the formation of deposits due to the evaporation of stabilized H2O2 solution fits the prediction made from simulation results. Y1 - 2019 U6 - http://dx.doi.org/10.1016/j.ijheatmasstransfer.2019.118519 SN - 0017-9310 VL - 143 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Kremers, Alexander A1 - Wagner, Torsten A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - FEM-based modeling of a calorimetric gas sensor for hydrogen peroxide monitoring JF - physica status solidi a : applications and materials sciences N2 - A physically coupled finite element method (FEM) model is developed to study the response behavior of a calorimetric gas sensor. The modeled sensor serves as a monitoring device of the concentration of gaseous hydrogen peroxide (H2 O2) in a high temperature mixture stream in aseptic sterilization processes. The principle of operation of a calorimetric H2 O2 sensor is analyzed and the results of the numerical model have been validated by using previously published sensor experiments. The deviation in the results between the FEM model and experimental data are presented and discussed. Y1 - 2017 U6 - http://dx.doi.org/10.1002/pssa.201600912 SN - 1862-6319 IS - Early View PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Vahidpour, Farnoosh A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - Development of a package-sterilization process for aseptic filling machines: A numerical approach and validation for surface treatment with hydrogen peroxide JF - Sensor and Actuators A: Physical N2 - Within the present work a sterilization process by a heated gas mixture that contains hydrogen peroxide (H₂O₂) is validated by experiments and numerical modeling techniques. The operational parameters that affect the sterilization efficacy are described alongside the two modes of sterilization: gaseous and condensed H₂O₂. Measurements with a previously developed H₂O₂ gas sensor are carried out to validate the applied H₂O₂ gas concentration during sterilization. We performed microbiological tests at different H₂O₂ gas concentrations by applying an end-point method to carrier strips, which contain different inoculation loads of Geobacillus stearothermophilus spores. The analysis of the sterilization process of a pharmaceutical glass vial is performed by numerical modeling. The numerical model combines heat- and advection-diffusion mass transfer with vapor–pressure equations to predict the location of condensate formation and the concentration of H₂O₂ at the packaging surfaces by changing the gas temperature. For a sterilization process of 0.7 s, a H₂O₂ gas concentration above 4% v/v is required to reach a log-count reduction above six. The numerical results showed the location of H₂O₂ condensate formation, which decreases with increasing sterilant-gas temperature. The model can be transferred to different gas nozzle- and packaging geometries to assure the absence of H₂O₂ residues. Y1 - 2020 U6 - http://dx.doi.org/10.1016/j.sna.2019.111691 SN - 0924-4247 VL - 303 IS - 111691 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Oberländer, Jan A1 - Kirchner, Patrick A1 - Keusgen, Michael A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - Experimental and Numerical Analyzes of a Sensor Based on Interdigitated Electrodes for Studying Microbiological Alterations JF - physica status solidi (a): applications and materials science N2 - In this work, a cell-based biosensor to evaluate the sterilization efficacy of hydrogen peroxide vapor sterilization processes is characterized. The transducer of the biosensor is based on interdigitated gold electrodes fabricated on an inert glass substrate. Impedance spectroscopy is applied to evaluate the sensor behavior and the alteration of test microorganisms due to the sterilization process. These alterations are related to changes in relative permittivity and electrical conductivity of the bacterial spores. Sensor measurements are conducted with and without bacterial spores (Bacillus atrophaeus), as well as after an industrial sterilization protocol. Equivalent two-dimensional numerical models based on finite element method of the periodic finger structures of the interdigitated gold electrodes are designed and validated using COMSOL® Multiphysics software by the application of known dielectric properties. The validated models are used to compute the electrical properties at different sensor states (blank, loaded with spores, and after sterilization). As a final result, we will derive and tabulate the frequency-dependent electrical parameters of the spore layer using a novel model that combines experimental data with numerical optimization techniques. Y1 - 2018 U6 - http://dx.doi.org/10.1002/pssa.201700920 SN - 1862-6319 VL - 215 IS - 15 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Oberländer, Jan A1 - Kirchner, Patrick A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - Thermocatalytic Behavior of Manganese (IV) Oxide as Nanoporous Material on the Dissociation of a Gas Mixture Containing Hydrogen Peroxide JF - Nanomaterials N2 - In this article, we present an overview on the thermocatalytic reaction of hydrogen peroxide (H₂O₂) gas on a manganese (IV) oxide (MnO₂) catalytic structure. The principle of operation and manufacturing techniques are introduced for a calorimetric H₂O₂ gas sensor based on porous MnO₂. Results from surface analyses by X-ray photoelectron spectroscopy (XPS) and scanning electron microscopy (SEM) of the catalytic material provide indication of the H₂O₂ dissociation reaction schemes. The correlation between theory and the experiments is documented in numerical models of the catalytic reaction. The aim of the numerical models is to provide further information on the reaction kinetics and performance enhancement of the porous MnO₂ catalyst. Y1 - 2018 U6 - http://dx.doi.org/10.3390/nano8040262 SN - 2079-4991 VL - 8 IS - 4 PB - MDPI CY - Basel ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Wagner, Patrick H. A1 - Schöning, Michael Josef T1 - Sterilization of Objects, Products, and Packaging Surfaces and Their Characterization in Different Fields of Industry: The Status in 2020 JF - physica status solidi (a) applications and materials science N2 - The treatment method to deactivate viable microorganisms from objects or products is termed sterilization. There are multiple forms of sterilization, each intended to be applied for a specific target, which depends on—but not limited to—the thermal, physical, and chemical stability of that target. Herein, an overview on the currently used sterilization processes in the global market is provided. Different sterilization techniques are grouped under a category that describes the method of treatment: radiation (gamma, electron beam, X-ray, and ultraviolet), thermal (dry and moist heat), and chemical (ethylene oxide, ozone, chlorine dioxide, and hydrogen peroxide). For each sterilization process, the typical process parameters as defined by regulations and the mode of antimicrobial activity are summarized. Finally, the recommended microorganisms that are used as biological indicators to validate sterilization processes in accordance with the rules that are established by various regulatory agencies are summarized. KW - bioburdens KW - sterility tests KW - sterilization efficacy KW - sterilization methods KW - validation methods Y1 - 2021 U6 - http://dx.doi.org/10.1002/pssa.202000732 SN - 1862-6319 N1 - Corresponding author: Michael J. Schöning VL - 218 IS - 13 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Jildeh, Zaid B. A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Pieper, Martin T1 - Simulating the electromagnetic‐thermal treatment of thin aluminium layers for adhesion improvement JF - Physica status solidi (a) N2 - A composite layer material used in packaging industry is made from joining layers of different materials using an adhesive. An important processing step in the production of aluminium-containing composites is the surface treatment and consequent coating of adhesive material on the aluminium surface. To increase adhesion strength between aluminium layer and the adhesive material, the foil is heat treated. For efficient heating, induction heating was considered as state-of-the-art treatment process. Due to the complexity of the heating process and the unpredictable nature of the heating source, the control of the process is not yet optimised. In this work, a finite element analysis of the process was established and various process parameters were studied. The process was simplified and modelled in 3D. The numerical model contains an air domain, an aluminium layer and a copper coil fitted with a magnetic field concentrating material. The effect of changing the speed of the aluminium foil (or rolling speed) was studied with the change of the coil current. Statistical analysis was used for generating a general control equation of coil current with changing rolling speed. Y1 - 2015 U6 - http://dx.doi.org/10.1002/pssa.201431893 SN - 1862-6319 VL - Vol. 212 IS - 6 SP - 1234 EP - 1241 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Karschuck, T. L. A1 - Filipov, Y. A1 - Bollella, P. A1 - Schöning, Michael Josef A1 - Katz, E. T1 - Not-XOR (NXOR) logic gate based on an enzyme-catalyzed reaction JF - International Journal of Unconventional Computing N2 - Enzyme-catalyzed reactions have been designed to mimic various Boolean logic gates in the general framework of unconventional biomolecular computing. While some of the logic gates, particularly OR, AND, are easy to realize with biocatalytic reactions and have been reported in numerous publications, some other, like NXOR, are very challenging and have not been realized yet with enzyme reactions. The paper reports on a novel approach to mimicking the NXOR logic gate using the bell-shaped enzyme activity dependent on pH values. Shifting pH from the optimum value to the acidic or basic values by using acid or base inputs (meaning 1,0 and 0,1 inputs) inhibits the enzyme reaction, while keeping the optimum pH (assuming 0,0 and 1,1 input combinations) preserves a high enzyme activity. The challenging part of the present approach is the selection of an enzyme with a well-demonstrated bell-shape activity dependence on the pH value. While many enzymes can satisfy this condition, we selected pyrroloquinoline quinone (PQQ)-dependent glucose dehydrogenase as this enzyme has the optimum pH center-located on the pH scale allowing the enzyme activity change by the acidic and basic pH shift from the optimum value corresponding to the highest activity. The present NXOR gate is added to the biomolecular “toolbox” as a new example of Boolean logic gates based on enzyme reactions. Y1 - 2019 SN - 1548-7199 VL - 14 IS - 3-4 SP - 235 EP - 242 PB - Old City Publishing CY - Philadelphia ER - TY - JOUR A1 - Karschuck, Tobias A1 - Poghossian, Arshak A1 - Ser, Joey A1 - Tsokolakyan, Astghik A1 - Achtsnicht, Stefan A1 - Wagner, Patrick A1 - Schöning, Michael Josef T1 - Capacitive model of enzyme-modified field-effect biosensors: Impact of enzyme coverage JF - Sensors and Actuators B: Chemical N2 - Electrolyte-insulator-semiconductor capacitors (EISCAP) belong to field-effect sensors having an attractive transducer architecture for constructing various biochemical sensors. In this study, a capacitive model of enzyme-modified EISCAPs has been developed and the impact of the surface coverage of immobilized enzymes on its capacitance-voltage and constant-capacitance characteristics was studied theoretically and experimentally. The used multicell arrangement enables a multiplexed electrochemical characterization of up to sixteen EISCAPs. Different enzyme coverages have been achieved by means of parallel electrical connection of bare and enzyme-covered single EISCAPs in diverse combinations. As predicted by the model, with increasing the enzyme coverage, both the shift of capacitance-voltage curves and the amplitude of the constant-capacitance signal increase, resulting in an enhancement of analyte sensitivity of the EISCAP biosensor. In addition, the capability of the multicell arrangement with multi-enzyme covered EISCAPs for sequentially detecting multianalytes (penicillin and urea) utilizing the enzymes penicillinase and urease has been experimentally demonstrated and discussed. KW - Field-effect biosensor KW - Capacitive model KW - Enzyme coverage KW - Multianalyte detection KW - Penicillin Y1 - 2024 U6 - http://dx.doi.org/10.1016/j.snb.2024.135530 SN - 0925-4005 (Print) SN - 1873-3077 (Online) N1 - Corresponding Author: Michael J. Schöning VL - 408 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Karschuck, Tobias A1 - Schmidt, Stefan A1 - Achtsnicht, Stefan A1 - Poghossian, Arshak A1 - Wagner, Patrick A1 - Schöning, Michael Josef T1 - Multiplexing system for automated characterization of a capacitive field-effect sensor array JF - Physica Status Solidi A N2 - In comparison to single-analyte devices, multiplexed systems for a multianalyte detection offer a reduced assay time and sample volume, low cost, and high throughput. Herein, a multiplexing platform for an automated quasi-simultaneous characterization of multiple (up to 16) capacitive field-effect sensors by the capacitive–voltage (C–V) and the constant-capacitance (ConCap) mode is presented. The sensors are mounted in a newly designed multicell arrangement with one common reference electrode and are electrically connected to the impedance analyzer via the base station. A Python script for the automated characterization of the sensors executes the user-defined measurement protocol. The developed multiplexing system is tested for pH measurements and the label-free detection of ligand-stabilized, charged gold nanoparticles. KW - Capacitive field-effect sensor KW - Gold nanoparticles KW - Label-free detection KW - Multicell KW - Multiplexing Y1 - 2023 U6 - http://dx.doi.org/10.1002/pssa.202300265 SN - 1862-6300 (Print) SN - 1862-6319 (Online) N1 - Corresponding author: Michael Josef Schöning VL - 220 IS - 22 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Kassab, T. A1 - Han, Y. A1 - Poghossian, Arshak A1 - Ingebrandt, S. A1 - Offenhäusser, A. A1 - Schöning, Michael Josef T1 - Detection of layerby-layer adsorbed polyelectrolytes by means of field-effect based capacitive EIS structures JF - Biomedizinische Technik. 49 (2004), H. 2 Y1 - 2004 SN - 0932-4666 SP - 1034 EP - 1035 ER - TY - JOUR A1 - Katz, Evgeny A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Enzyme-based logic gates and circuits - analytical applications and interfacing with electronics JF - Analytical and Bioanalytical Chemistry N2 - The paper is an overview of enzyme-based logic gates and their short circuits, with specific examples of Boolean AND and OR gates, and concatenated logic gates composed of multi-step enzyme-biocatalyzed reactions. Noise formation in the biocatalytic reactions and its decrease by adding a “filter” system, converting convex to sigmoid response function, are discussed. Despite the fact that the enzyme-based logic gates are primarily considered as components of future biomolecular computing systems, their biosensing applications are promising for immediate practical use. Analytical use of the enzyme logic systems in biomedical and forensic applications is discussed and exemplified with the logic analysis of biomarkers of various injuries, e.g., liver injury, and with analysis of biomarkers characteristic of different ethnicity found in blood samples on a crime scene. Interfacing of enzyme logic systems with modified electrodes and semiconductor devices is discussed, giving particular attention to the interfaces functionalized with signal-responsive materials. Future perspectives in the design of the biomolecular logic systems and their applications are discussed in the conclusion. Y1 - 2017 U6 - http://dx.doi.org/10.1007/s00216-016-0079-7 SN - 1618-2650 VL - 409 SP - 81 EP - 94 PB - Springer CY - Berlin ER - TY - JOUR A1 - Keusgen, M. A1 - Jünger, M. A1 - Knobbe, D.-T. A1 - Kloock, Joachim P. A1 - Schöning, Michael Josef T1 - Entwicklung eines potentiometrischen Cyanid-Biosensors auf der Basis von EIS-Strukturen JF - Sensoren für zukünftige Hochtechnologien und Neuentwicklungen für die Verfahrenstechnik / 6. Dresdner Sensor-Symposium, 8. - 10. Dezember 2003, Dresden. Jörg Peter Baselt; Gerald Gerlach (Hg.) Y1 - 2003 SN - 3-935712-92-8 SP - 235 EP - 238 PB - w.e.b.-Univ.-Verl. CY - Dresden ER - TY - JOUR A1 - Keusgen, M. A1 - Jünger, M. A1 - Krest, I. A1 - Schöning, Michael Josef T1 - Development of a biosensor specific for cysteine sulfoxides JF - Biosensors & Bioelectronics. 18 (2003), H. 5-6 Y1 - 2003 SN - 0956-5663 SP - 805 EP - 812 ER - TY - JOUR A1 - Keusgen, M. A1 - Jünger, M. A1 - Schöning, Michael Josef T1 - Biosensoric detection of the cysteine sulphoxide alliin JF - Book of abstracts / ed. by J. Saneistr. Y1 - 2002 SN - 80-01-02576-4 N1 - Eurosensors ; (16, 2002, Praha) SP - 1175 EP - 1178 PB - Czech Technical University, Faculty of Electrical Engineering, Department of Measurement CY - Prague ER - TY - JOUR A1 - Keusgen, M. A1 - Kloock, Joachim P. A1 - Knobbe, D.-T. A1 - Jünger, M. A1 - Krest, I. A1 - Goldbach, M. A1 - Klein, W. A1 - Schöning, Michael Josef T1 - Direct determination of cyanides by potentiometric biosensors JF - Sensors and Actuators B. 103 (2004), H. 1-2 Y1 - 2004 SN - 0925-4005 SP - 380 EP - 385 ER - TY - JOUR A1 - Keusgen, M. A1 - Schöning, Michael Josef T1 - Strategies for biosensoric detection of potential drugs in nature JF - Biomedizinische Technik. 49 (2004), H. 2 Y1 - 2004 SN - 0932-4666 SP - 1004 EP - 1005 ER - TY - JOUR A1 - Keusgen, Michael A1 - Jünger, Martina A1 - Krest, Ingo A1 - Schöning, Michael Josef T1 - Biosensoric detection of the cysteine sulphoxide alliin JF - Sensors and Actuators B. 95 (2003), H. 1-3 Y1 - 2003 SN - 0925-4005 SP - 297 EP - 302 ER - TY - JOUR A1 - Khaydukova, M. M. A1 - Zadorozhnaya, O. A. A1 - Kirsanov, D. O. A1 - Iken, Heiko A1 - Rolka, David A1 - Schöning, Michael Josef A1 - Babain, V. A. A1 - Vlasov, Yu. G. A1 - Legin, A. V. T1 - Multivariate processing of atomic-force microscopy images for detection of the response of plasticized polymeric membranes JF - Russian journal of applied chemistry N2 - The possibility of using the atomic-force microscopy as a method for detection of the analytical signal from plasticized polymeric sensor membranes was analyzed. The surfaces of cadmium-selective membranes based on two polymeric matrices were examined. The digital images were processed with multivariate image analysis techniques. A correlation was found between the surface profile of an ion-selective membrane and the concentration of the ion in solution. Y1 - 2014 U6 - http://dx.doi.org/10.1134/S1070427214030112 SN - 1608-3296 (E-Journal); 1070-4272 (Print) VL - 87 IS - 3 SP - 307 EP - 314 PB - Springer CY - Dordrecht ER - TY - CHAP A1 - Kirchner, Patrick A1 - Henkel, H. A1 - Näther, Niko A1 - Spelthahn, H. A1 - Schneider, A. A1 - Berger, J. A1 - Kolstad, J. A1 - Friedrich, P. A1 - Schöning, Michael Josef T1 - RFID-basiertes Sensorsystem zur Realisierung intelligenter Verpackungen für die Nahrungsmittelindustrie T2 - KMU - innovativ: IKT 2008. CD-ROM : BMBF-Statustagung KMU - innovativ: IKT, Darmstadt, 17. - 18. Nov. 2008 Y1 - 2008 IS - CD-ROM-Ausg. PB - BMBF CY - Berlin ER - TY - JOUR A1 - Kirchner, Patrick A1 - Li, B. A1 - Spelthahn, H. A1 - Henkel, H. A1 - Friedrich, P. A1 - Kolstad, J. A1 - Keusgen, M. A1 - Schöning, Michael Josef T1 - Thin-film calorimetric H2O2 gas sensor for the validation of germicidal effectivity in aseptic filling processes JF - Procedia Chemistry. 1 (2009), H. 1 Y1 - 2009 SN - 1876-6196 N1 - Proceedings of the Eurosensors XXIII conference ; Eurosensors 23 SP - 983 EP - 986 ER - TY - JOUR A1 - Kirchner, Patrick A1 - Li, B. A1 - Spelthahn, H. A1 - Henkel, H. A1 - Schneider, A. A1 - Kolstad, J. A1 - Friedrich, P. A1 - Schöning, Michael Josef T1 - Realisierung eines miniaturisierten Gassensors in Chiptechnologie basierend auf einer Dünnschicht-Thermosäule zur H2O2-Detektion in aseptischen Abfüllanlagen JF - 9. Dresdner Sensor-Symposium : Dresden, 07.-09. Dezember 2009 / Gerlach, Gerald ; Hauptmann, Peter [Hrsg.] Y1 - 2009 SN - 978-3-941298-44-6 N1 - Dresdner Sensor-Symposium ; (9, 2009, Dresden) SP - 293 EP - 296 PB - TUDpress CY - Dresden ER - TY - JOUR A1 - Kirchner, Patrick A1 - Li, Bin A1 - Spelthahn, Heiko A1 - Henkel, Hartmut A1 - Schneider, Andreas A1 - Friedrich, Peter A1 - Kolstad, Jens A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Thin-film calorimetric H2O2 gas sensor for the validation of germicidal effectivity in aseptic filling processes JF - Sensors and Actuators B: Chemical. 154 (2011), H. 2 Y1 - 2011 SN - 1873-3077 N1 - EUROSENSORS XXIII SP - 257 EP - 263 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Friedrich, Peter A1 - Berger, Jörg A1 - Rysstad, Gunnar A1 - Schöning, Michael Josef A1 - Keusgen, Michael T1 - Realisation of a calorimetric gas sensor on polyimide foil for applications in aseptic food industry JF - Sensors and Actuators B: Chemical N2 - A calorimetric gas sensor is presented for the monitoring of vapour-phase H2O2 at elevated temperature during sterilisation processes in aseptic food industry. The sensor was built up on a flexible polyimide foil (thickness: 25 μm) that has been chosen due to its thermal stability and low thermal conductivity. The sensor set-up consists of two temperature-sensitive platinum thin-film resistances passivated by a layer of SU-8 photo resist and catalytically activated by manganese(IV) oxide. Instead of an active heating structure, the calorimetric sensor utilises the elevated temperature of the evaporated H2O2 aerosol. In an experimental test rig, the sensor has shown a sensitivity of 4.78 °C/(%, v/v) in a H2O2 concentration range of 0%, v/v to 8%, v/v. Furthermore, the sensor possesses the same, unchanged sensor signal even at varied medium temperatures between 210 °C and 270 °C of the gas stream. At flow rates of the gas stream from 8 m3/h to 12 m3/h, the sensor has shown only a slightly reduced sensitivity at a low flow rate of 8 m3/h. The sensor characterisation demonstrates the suitability of the calorimetric gas sensor for monitoring the efficiency of industrial sterilisation processes. KW - Sterilisation process KW - Hydrogen peroxide KW - Polyimide KW - Calorimetric gas sensor Y1 - 2012 U6 - http://dx.doi.org/10.1016/j.snb.2011.01.032 SN - 0925-4005 N1 - Part of special issue "Eurosensors XXIV, 2010" VL - 170 SP - 60 EP - 66 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Friedrich, Peter A1 - Berger, Jörg A1 - Suso, Henri-Pierre A1 - Kupyna, Andriy A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Optimisation and fabrication of a calorimetric gas sensor built up on a polyimide substrate for H2O2 monitoring JF - Physica status solidi (a) : applications and material science. 208 (2011), H. 6 Y1 - 2011 SN - 1862-6319 SP - 1235 EP - 1240 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Friedrich, Peter A1 - Rysstad, G. A1 - Berger, J. A1 - Keusgen, M. A1 - Schöning, Michael Josef T1 - Realization of a calorimetric gas sensor on polyimide foil for applications in aseptic food industry JF - Procedia Engineering. 5 (2010) Y1 - 2010 SN - 1877-7058 N1 - Eurosensor XXIV Conference SP - 264 EP - 267 ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Suco, Henri-Pierre A1 - Rysstad, Gunnar A1 - Schöning, Michael Josef T1 - Monitoring the microbicidal effectiveness of gaseous hydrogen peroxide in sterilisation processes by means of a calorimetric gas sensor JF - Food control N2 - In the present work, a novel method for monitoring sterilisation processes with gaseous H2O2 in combination with heat activation by means of a specially designed calorimetric gas sensor was evaluated. Therefore, the sterilisation process was extensively studied by using test specimens inoculated with Bacillus atrophaeus spores in order to identify the most influencing process factors on its microbicidal effectiveness. Besides the contact time of the test specimens with gaseous H2O2 varied between 0.2 and 0.5 s, the present H2O2 concentration in a range from 0 to 8% v/v (volume percent) had a strong influence on the microbicidal effectiveness, whereas the change of the vaporiser temperature, gas flow and humidity were almost negligible. Furthermore, a calorimetric H2O2 gas sensor was characterised in the sterilisation process with gaseous H2O2 in a wide range of parameter settings, wherein the measurement signal has shown a linear response against the H2O2 concentration with a sensitivity of 4.75 °C/(% v/v). In a final step, a correlation model by matching the measurement signal of the gas sensor with the microbial inactivation kinetics was established that demonstrates its suitability as an efficient method for validating the microbicidal effectiveness of sterilisation processes with gaseous H2O2. KW - hydrogen peroxide KW - sterilisation KW - Bacillus atrophaeus KW - calorimetric gas sensor Y1 - 2012 U6 - http://dx.doi.org/10.1016/j.foodcont.2012.11.048 SN - 0956-7135 VL - 31 IS - 2 SP - 530 EP - 538 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Suso, Henri-Pierre A1 - Rysstad, Gunnar A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Towards a wireless sensor system for real-time H2O2 monitoring in aseptic food processes JF - Physica status solidi (a) N2 - A wireless sensor system based on the industrial ZigBee standard for low-rate wireless networking was developed that enables real-time monitoring of gaseous H2O2 during the package sterilization in aseptic food processes. The sensor system consists of a remote unit connected to a calorimetric gas sensor, which was already established in former works, and an external base unit connected to a laptop computer. The remote unit was built up by an XBee radio frequency (RF) module for data communication and a programmable system-on-chip controller to read out the sensor signal and process the sensor data, whereas the base unit is a second XBee RF module. For the rapid H2O2 detection on various locations inside the package that has to be sterilized, a novel read-out strategy of the calorimetric gas sensor was established, wherein the sensor response is measured within the short sterilization time and correlated with the present H2O2 concentration. In an exemplary measurement application in an aseptic filling machinery, the suitability of the new, wireless sensor system was demonstrated, wherein the influence of the gas velocity on the H2O2 distribution inside a package was determined and verified with microbiological tests. KW - calorimetric gas sensor;hydrogen peroxide;wireless sensor system Y1 - 2013 U6 - http://dx.doi.org/10.1002/pssa.201200920 SN - 1862-6319 VL - 210 IS - 5 SP - 877 EP - 883 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Kirchner, Patrick A1 - Reisert, Steffen A1 - Pütz, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Characterisation of polymeric materials as passivation layer for calorimetric H2O2 gas sensors JF - Physica Status Solidi (a) N2 - Calorimetric gas sensors for monitoring the H₂O₂ concentration at elevated temperatures in industrial sterilisation processes have been presented in previous works. These sensors are built up in form of a differential set-up of a catalytically active and passive temperature-sensitive structure. Although, various types of catalytically active dispersions have been studied, the passivation layer has to be established and therefore, chemically as well as physically characterised. In the present work, fluorinated ethylene propylene (FEP), perfluoralkoxy (PFA) and epoxy-based SU-8 photoresist as temperature-stable polymeric materials have been investigated for sensor passivation in terms of their chemical inertness against H₂O₂, their hygroscopic properties as well as their morphology. The polymeric materials were deposited via spin-coating on the temperature-sensitive structure, wherein spin-coated FEP and PFA show slight agglomerates. However, they possess a low absorption of humidity due to their hydrophobic surface, whereas the SU-8 layer has a closed surface but shows a slightly higher absorption of water. All of them were inert against gaseous H₂O₂ during the characterisation in H₂O₂ atmosphere that demonstrates their suitability as passivation layer for calorimetric H₂O₂ gas sensors. Y1 - 2012 U6 - http://dx.doi.org/10.1002/pssa.201100773 SN - 1862-6319 VL - 209 IS - 5 SP - 859 EP - 863 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Kirchner, Patrick A1 - Spelthahn, H. A1 - Schöning, Michael Josef A1 - Henkel, H. A1 - Schneider, A. A1 - Friedrich, P. A1 - Kolstad, J. A1 - Berger, J. T1 - Realisierung eines Polyimid-basierten kalorimetrischen Gassensors zur Inline-Überwachung der H2O2-Konzentration in aseptischen Abfüllanlagen JF - Sensoren und Messsysteme 2010 [Elektronische Ressource] : Vorträge der 15. ITG/GMA-Fachtagung vom 18. bis 19. Mai 2010 in Nürnberg / Informationstechnische Gesellschaft im VDE (ITG); VDI/VDE-Gesellschaft Mess- und Automatisierungstechnik (GMA) Y1 - 2010 SN - 978-3-8007-3260-9 N1 - Fachtagung Sensoren und Messsysteme 15, 2010, Nürnberg ; Gesellschaft Mess- und Automatisierungstechnik SP - 607 EP - 612 PB - VDE Verlag CY - Berlin ER - TY - JOUR A1 - Kloock, Joachim P. A1 - Moreno, Lia A1 - Bratov, A. A1 - Huachupoma, S. A1 - Xu, J. A1 - Wagner, Torsten A1 - Yoshinobu, T. A1 - Ermolenko, Y. A1 - Vlasov, Y. G. A1 - Schöning, Michael Josef T1 - PLD-prepared cadmium sensors based on chalcogenide glasses —ISFET, LAPS and μISE semiconductor structures JF - Sensors and Actuators B: Chemical. 118 (2006), H. 1-2 Y1 - 2006 SN - 0925-4005 N1 - Eurosensors XIX - Eurosensors XIX - The 19th European Conference on Solid-State Transducers SP - 149 EP - 155 ER - TY - CHAP A1 - Kloock, Joachim P. A1 - Moreno, Lia A1 - Huachupoma, S. A1 - Xu, J. A1 - Wagner, Torsten A1 - Bratov, A. A1 - Doll, T. A1 - Vlasov, Y. A1 - Schöning, Michael Josef ED - Gerlach, Gerald T1 - Halbleiterbasierte Schwermetallsensorik auf der Basis von Chalkogenidgläsern für zukünftige „Lab on Chip“-Anwendungen T2 - 7. Dresdner Sensor-Symposium - Neue Herausforderungen und Anwendungen in der Sensortechnik Y1 - 2005 SN - 3-938863-29-3 SP - 221 EP - 224 PB - TUDpress, Verl. der Wissenschaften CY - Dresden ER - TY - JOUR A1 - Kloock, Joachim P. A1 - Mourzina, Y. A1 - Schubert, J. A1 - Ermelenko, Y. A1 - Schöning, Michael Josef T1 - Pulsed laser deposition: A tool for fabricating thin-film microsensors JF - Biomedizinische Technik. 49 (2004), H. 2 Y1 - 2004 SN - 0932-4666 SP - 1032 EP - 1033 ER - TY - JOUR A1 - Kloock, Joachim P. A1 - Mourzina, Y.G. A1 - Ermolenko, Y. A1 - Doll, T. A1 - Schubert, J. A1 - Schöning, Michael Josef T1 - Inorganic thin-film sensor membranes with PLD-prepared chalcogenide glasses: Challenges and implementation JF - Sensors. 4 (2004), H. 10 Y1 - 2004 SN - 1424-8220 SP - 156 EP - 162 ER - TY - JOUR A1 - Kloock, Joachim P. A1 - Mourzina, Y.G. A1 - Yoshinobu, T. A1 - Schubert, J. A1 - Schöning, Michael Josef T1 - Chalkogenidglassensoren für die Schwermetallanalytik JF - Sensoren und Mess-Systeme : Vorträge der 11. ITG/GMA-Fachtagung vom 11. bis 12. März 2002 in Ludwigsburg / Veranst.: Informationstechnische Gesellschaft im VDE (ITG) ... Wiss. Tagungsleitung: C. D. Kohl .... Y1 - 2002 SN - 3-8007-2675-0 N1 - Sensoren und Mess-Systeme 2002 SP - 395 EP - 398 PB - VDE-Verl. CY - Berlin [u.a.] ER - TY - JOUR A1 - Kloock, Joachim P. A1 - Poghossian, Arshak A1 - Schumacher, K. A1 - Rosenkranz, C. A1 - Schultze, J. W. A1 - Müller-Veggian, Mattea A1 - Schöning, Michael Josef T1 - Funktionsprüfung und Charakterisierung von ionensensitiven Feldeffekttransistoren (ISFETs) auf Waferebene mittels Mikrotropfenzelle für den zukünftigen Einsatz in der Sensorproduktion JF - Sensoren und Messsysteme 2006 : Vorträge der 13. ITG/GMA-Fachtagung vom 13. bis 14. März 2006 in Freiburg/Breisgau / Veranst.: Informationstechnische Gesellschaft im VDE (ITG) ; VDE/VDI-Gesellschaft Mess- und Automatisierungstechnik (GMA). Wiss. Tagungsleitung: L. M. Reindl Y1 - 2006 SN - 3-8007-2939-3 N1 - Fachtagung Sensoren und Messsysteme <13, 2006, Freiburg, Breisgau> SP - 257 EP - 260 PB - VDE-Verl. CY - Berlin ER - TY - CHAP A1 - Kloock, Joachim P. A1 - Schubert, J. A1 - Ermelenko, Y. A1 - Vlasov, Y. G. A1 - Bratov, A. A1 - Schöning, Michael Josef T1 - Thin-film sensors with chalcogenide glass materials – a general survey T2 - Biochemical sensing utilisation of micro- and nanotechnologies : Warsaw, [23rd - 26th] November 2005 / ed. by M. Mascini ... Y1 - 2006 SP - 92 EP - 97 CY - Warsaw ER - TY - CHAP A1 - Kloock, Joachim P. A1 - Schöning, Michael Josef T1 - Heavy metal detection with semiconductor devices based on PLD-prepared chalcogenide glass thin films T2 - Armenian Journal of Physics Y1 - 2007 SN - 1829-1171 SP - 95 EP - 98 ER - TY - JOUR A1 - Knobbe, D.-T. A1 - Schöning, Michael Josef A1 - Poghossian, Arshak A1 - Mourzina, Y. T1 - Charakterisierung von kapazitiven EMISSensoren mittels Impedanzspektroskopie, Kapazitäts-Spannungs- und Konstant-Kapazitäts-Messung JF - Sensoren und Messsysteme 2004 : Tagung Ludwigsburg, 15. und 16. März 2004 / VDI-VDE-Gesellschaft Mess- und Automatisierungstechnik Y1 - 2004 SN - 3-18-091829-2 N1 - Auch erschienen als: VDI-Berichte ; 1829 SP - 851 EP - 854 PB - VDI CY - Düsseldorf ER - TY - JOUR A1 - Koch, Claudia A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Wege, Christian T1 - Penicillin Detection by Tobacco Mosaic Virus-Assisted Colorimetric Biosensors JF - Nanotheranostics N2 - The presentation of enzymes on viral scaffolds has beneficial effects such as an increased enzyme loading and a prolonged reusability in comparison to conventional immobilization platforms. Here, we used modified tobacco mosaic virus (TMV) nanorods as enzyme carriers in penicillin G detection for the first time. Penicillinase enzymes were conjugated with streptavidin and coupled to TMV rods by use of a bifunctional biotin-linker. Penicillinase-decorated TMV particles were characterized extensively in halochromic dye-based biosensing. Acidometric analyte detection was performed with bromcresol purple as pH indicator and spectrophotometry. The TMV-assisted sensors exhibited increased enzyme loading and strongly improved reusability, and higher analysis rates compared to layouts without viral adapters. They extended the half-life of the sensors from 4 - 6 days to 5 weeks and thus allowed an at least 8-fold longer use of the sensors. Using a commercial budget-priced penicillinase preparation, a detection limit of 100 µM penicillin was obtained. Initial experiments also indicate that the system may be transferred to label-free detection layouts. Y1 - 2018 U6 - http://dx.doi.org/10.7150/ntno.22114 SN - 2206-7418 VL - 2 IS - 2 SP - 184 EP - 196 PB - Ivyspring CY - Sydney ER - TY - CHAP A1 - Koch, Claudia A1 - Poghossian, Arshak A1 - Wege, Christina A1 - Schöning, Michael Josef ED - Wege, Christina T1 - TMV-Based Adapter Templates for Enhanced Enzyme Loading in Biosensor Applications T2 - Virus-Derived Nanoparticles for Advanced Technologies N2 - Nanotubular tobacco mosaic virus (TMV) particles and RNA-free lower-order coat protein (CP) aggregates have been employed as enzyme carriers in different diagnostic layouts and compared for their influence on biosensor performance. In the following, we describe a label-free electrochemical biosensor for improved glucose detection by use of TMV adapters and the enzyme glucose oxidase (GOD). A specific and efficient immobilization of streptavidin-conjugated GOD ([SA]-GOD) complexes on biotinylated TMV nanotubes or CP aggregates was achieved via bioaffinity binding. Glucose sensors with adsorptively immobilized [SA]-GOD, and with [SA]-GOD cross-linked with glutardialdehyde, respectively, were tested in parallel on the same sensor chip. Comparison of these sensors revealed that TMV adapters enhanced the amperometric glucose detection remarkably, conveying highest sensitivity, an extended linear detection range and fastest response times. These results underline a great potential of an integration of virus/biomolecule hybrids with electronic transducers for applications in biosensorics and biochips. Here, we describe the fabrication and use of amperometric sensor chips combining an array of circular Pt electrodes, their loading with GOD-modified TMV nanotubes (and other GOD immobilization methods), and the subsequent investigations of the sensor performance. KW - Tobacco mosaic virus (TMV) KW - Coat protein KW - Enzyme nanocarrier KW - Glucose biosensor KW - Glucose oxidase Y1 - 2018 SN - 978-1-4939-7808-3 U6 - http://dx.doi.org/10.1007/978-1-4939-7808-3 N1 - Methods in Molecular Biology, vol 1776 SP - 553 EP - 568 PB - Humana Press CY - New York, NY ER - TY - JOUR A1 - Kramer, Friederike A1 - Halamkova, Lenka A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Katz, Evgeny A1 - Halamek, Jan T1 - Biocatalytic analysis of biomarkers for forensic identification of ethnicity between Caucasian and African American JF - The analyst. August 2013 Y1 - 2013 SN - 1364-5528 (E-Journal); 0003-2654 (Print) VL - Vol. 138 SP - 6251 EP - 6257 PB - Royal Society of Chemistry CY - Cambridge ER - TY - CHAP A1 - Kreyer, Jörg A1 - Stollenwerk, Dominik A1 - Schöning, Michael Josef T1 - Tagung des Forschernachwuchses der FH Aachen 20. November 2013 / ed.: J. Kreyer, D. Stollenwerk, M. J. Schöning Y1 - 2013 PB - FH Aachen CY - Aachen ER - TY - JOUR A1 - Krämer, Melina A1 - Pita, Marcos A1 - Zhou, Jian A1 - Ornatska, Maryna A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Katz, Evgeny T1 - Coupling of Biocomputing Systems with Electronic Chips: Electronic Interface for Transduction of Biochemical Information JF - Journal of Physical Chemistry C: Nanomaterials and Interfaces. 113 (2009), H. 6 Y1 - 2009 SN - 1932-7455 SP - 2573 EP - 2579 PB - American Cemical Society CY - Washington, DC ER - TY - JOUR A1 - Kurowski, A. A1 - Schultze, J. A1 - Lüth, H. A1 - Schöning, Michael Josef T1 - Micro- and nanopatterning of sensor chips by means of macroporous silicon. JF - Sensors and Actuators B. 83 (2002), H. 1-3 Y1 - 2002 SN - 0925-4005 N1 - Selected Papers from TRANSDUCERS '01 EUROSENSORS XV SP - 123 EP - 128 ER - TY - JOUR A1 - Kurowski, A. A1 - Schultze, J.W. A1 - Lüth, H. A1 - Schöning, Michael Josef T1 - Micro- and nanopatterning of sensor chips by means of macroporous silicon JF - Transducers '01 Eurosensors XV : digest of technical papers / the 11th International Conference on Solid-State Sensors and Actuators, June 10-14, 2001, Munich, Germany. Ernst Obermeier (Ed.) Y1 - 2001 SN - 3-540-42150-5 N1 - Eurosensors (15, 2001, München) ; International Conference on Solid State Sensors and Actuators (11, 2001, München) SP - 640 EP - 643 PB - Springer CY - Berlin [u.a.] ER - TY - JOUR A1 - Leinhos, Marcel A1 - Schusser, Sebastian A1 - Bäcker, Matthias A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Micromachined multi-parameter sensor chip for the control of polymer-degradation medium JF - Physica Status Solidi (A) : special issue on engineering and functional interfaces N2 - It is well known that the degradation environment can strongly influence the biodegradability and kinetics of biodegradation processes of polymers. Therefore, besides the monitoring of the degradation process, it is also necessary to control the medium in which the degradation takes place. In this work, a micromachined multi-parameter sensor chip for the control of the polymer-degradation medium has been developed. The chip combines a capacitive field-effect pH sensor, a four-electrode electrolyte-conductivity sensor and a thin-film Pt-temperature sensor. The results of characterization of individual sensors are presented. In addition, the multi-parameter sensor chip together with an impedimetric polymer-degradation sensor was simultaneously characterized in degradation solutions with different pH and electrolyte conductivity. The obtained results demonstrate the feasibility of the multi-parameter sensor chip for the control of the polymer-degradation medium. Y1 - 2014 U6 - http://dx.doi.org/10.1002/pssa.201330364 SN - 1521-396X (E-Journal); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1346 EP - 1351 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Lüth, H. A1 - Thust, M. A1 - Steffen, A. A1 - Kordos, P. A1 - Schöning, Michael Josef T1 - Biochemical sensors with structured and porous silicon capacitors JF - Materials Science and Engineering B. 69-70 (2000) Y1 - 2000 SN - 0921-5107 SP - 104 EP - 108 ER -