TY - JOUR A1 - Werkhausen, Amelie A1 - Cronin, Neil J. A1 - Albracht, Kirsten A1 - Paulsen, Gøran A1 - Larsen, Askild V. A1 - Bojsen-Møller, Jens A1 - Seynnes, Olivier R. T1 - Training-induced increase in Achilles tendon stiffness affects tendon strain pattern during running JF - PeerJ N2 - Background During the stance phase of running, the elasticity of the Achilles tendon enables the utilisation of elastic energy and allows beneficial contractile conditions for the triceps surae muscles. However, the effect of changes in tendon mechanical properties induced by chronic loading is still poorly understood. We tested the hypothesis that a training-induced increase in Achilles tendon stiffness would result in reduced tendon strain during the stance phase of running, which would reduce fascicle strains in the triceps surae muscles, particularly in the mono-articular soleus. Methods Eleven subjects were assigned to a training group performing isometric singleleg plantarflexion contractions three times per week for ten weeks, and another ten subjects formed a control group. Before and after the training period, Achilles tendon stiffness was estimated, and muscle-tendon mechanics were assessed during running at preferred speed using ultrasonography, kinematics and kinetics. Results Achilles tendon stiffness increased by 18% (P <0:01) in the training group, but the associated reduction in strain seen during isometric contractions was not statistically significant. Tendon elongation during the stance phase of running was similar after training, but tendon recoil was reduced by 30% (P <0:01), while estimated tendon force remained unchanged. Neither gastrocnemius medialis nor soleus fascicle shortening during stance was affected by training. Discussion These results show that a training-induced increase in Achilles tendon stiffness altered tendon behaviour during running. Despite training-induced changes in tendon mechanical properties and recoil behaviour, the data suggest that fascicle shortening patterns were preserved for the running speed that we examined. The asymmetrical changes in tendon strain patterns supports the notion that simple inseries models do not fully explain the mechanical output of the muscle-tendon unit during a complex task like running. KW - Achilles tendon KW - Stiffness KW - Running KW - Tendon properties KW - Architectural gear ratio Y1 - 2019 U6 - http://dx.doi.org/10.7717/peerj.6764 SN - 21678359 PB - Peer CY - London ER - TY - GEN A1 - Blottner, Dieter A1 - Hastermann, Maria A1 - Muckelt, Paul A1 - Albracht, Kirsten A1 - Schoenrock, Britt A1 - Salanova, Michele A1 - Warner, Martin A1 - Gunga, Hans-Christian A1 - Stokes, Maria T1 - MYOTONES - Inflight muscle health status monitoring during long-duration space missions onboard the International Space Station: a single case study T2 - IAC Papers Archive N2 - The MYOTONES experiment is the first to monitor changes in the basic biomechanical properties (tone, elasticity and stiffness) of the resting human myofascial system due to microgravity with a oninvasive, portable device on board the ISS. The MyotonPRO device applies several brief mechanical stimuli to the surface of the skin, and the natural oscillation signals of the tissue beneath are detected and computed by the MyotonPRO. Thus, an objective, quick and easy determination of the state of the underlying tissue is possible. Two preflight, four inflight and four post flight measurements were performed on a male astronaut using the same 10 measurement points (MP) for each session. MPs were located on the plantar fascia, Achilles tendon, M. soleus, M. gastrocnemius, M. multifidus, M. splenius capitis, M. deltoideus anterior, M. rectus femoris, infrapatellar tendon, M. tibialis anterior. Subcutaneous tissues thickness above the MPs was measured using ultrasound imaging. Magnetic resonance images (MRI) of lower limb muscles and functional tests were also performed pre- and postflight. Our first measurements on board the ISS confirmed increased tone and stiffness of the lumbar multifidus muscle, an important trunk stabilizer, dysfunction of which is known to be associated with back pain. Furthermore, reduced tone and stiffness of Achilles tendon and plantar fascia were observed inflight vs. preflight, confirming previous findings from terrestrial analog studies and parabolic flights. Unexpectedly, the deltoid showed negative inflight changes in tone and stiffness, and increased elasticity, suggesting a potential risk of muscle atrophy in longer spaceflight that should be addressed by adequate inflight countermeasure protocols. Most values from limb and back MPS showed deflected patterns (in either directions) from inflight shortly after the re-entry phase on the landing day and one week later. Most parameter values then normalized to baseline after 3 weeks likely due to 1G re-adaptation and possible outcome of the reconditioning protocol. No major changes in subcutaneous tissues thickness above the MPs were found inflight vs preflight, suggesting no bias (i.e., fluid shift, extreme tissue thickening or loss). Pre- and postflight MRI and functional tests showed negligible changes in calf muscle size, power and force, which is likely due to training effects from current inflight exercise protocols. The MYOTONES experiment is currently ongoing to collect data from further crew members. The potential impact of this research is to better understand the effects of microgravity and countermeasures over the time course of an ISS mission cycle. This will enable exercise countermeasures to be tailored Y1 - 2019 SN - 00741795 N1 - International Astronautical Congress: space: the power of the past, the promise of the future - Washington DC, USA/Vereinigte Staaten von Amerika Dauer: 21.10.2019 → 25.10.2019 PB - Pergamon CY - Oxford ER - TY - CHAP A1 - Ketelhut, Maike A1 - Göll, Fabian A1 - Braunstein, Bjoern A1 - Albracht, Kirsten A1 - Abel, Dirk T1 - Iterative learning control of an industrial robot for neuromuscular training T2 - 2019 IEEE Conference on Control Technology and Applications N2 - Effective training requires high muscle forces potentially leading to training-induced injuries. Thus, continuous monitoring and controlling of the loadings applied to the musculoskeletal system along the motion trajectory is required. In this paper, a norm-optimal iterative learning control algorithm for the robot-assisted training is developed. The algorithm aims at minimizing the external knee joint moment, which is commonly used to quantify the loading of the medial compartment. To estimate the external knee joint moment, a musculoskeletal lower extremity model is implemented in OpenSim and coupled with a model of an industrial robot and a force plate mounted at its end-effector. The algorithm is tested in simulation for patients with varus, normal and valgus alignment of the knee. The results show that the algorithm is able to minimize the external knee joint moment in all three cases and converges after less than seven iterations. KW - Knee KW - Training KW - Load modeling KW - Force KW - Iterative learning control Y1 - 2019 SN - 978-1-7281-2767-5 (ePub) SN - 978-1-7281-2766-8 (USB) SN - 978-1-7281-2768-2 (PoD) U6 - http://dx.doi.org/10.1109/CCTA.2019.8920659 N1 - 2019 IEEE Conference on Control Technology and Applications (CCTA) Hong Kong, China, August 19-21, 2019 PB - IEEE CY - New York ER - TY - JOUR A1 - Ketelhut, Maike A1 - Kolditz, Melanie A1 - Göll, Fabian A1 - Braunstein, Bjoern A1 - Albracht, Kirsten A1 - Abel, Dirk T1 - Admittance control of an industrial robot during resistance training JF - IFAC-PapersOnLine N2 - Neuromuscular strength training of the leg extensor muscles plays an important role in the rehabilitation and prevention of age and wealth related diseases. In this paper, we focus on the design and implementation of a Cartesian admittance control scheme for isotonic training, i.e. leg extension and flexion against a predefined weight. For preliminary testing and validation of the designed algorithm an experimental research and development platform consisting of an industrial robot and a force plate mounted at its end-effector has been used. Linear, diagonal and arbitrary two-dimensional motion trajectories with different weights for the leg extension and flexion part are applied. The proposed algorithm is easily adaptable to trajectories consisting of arbitrary six-dimensional poses and allows the implementation of individualized trajectories. KW - Assistive technology KW - Rehabilitation engineering KW - Human-Computer interaction KW - Automatic control Y1 - 2019 U6 - http://dx.doi.org/10.1016/j.ifacol.2019.12.102 SN - 2405-8963 N1 - 14th IFAC Symposium on Analysis, Design, and Evaluation of Human Machine Systems HMS 2019 Tallinn, Estonia, 16–91 September 2019 VL - 52 IS - 19 SP - 223 EP - 228 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Achtsnicht, Stefan A1 - Schönenborn, Kristina A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim T1 - Measurement of the magnetophoretic velocity of different superparamagnetic beads JF - Journal of Magnetism and Magnetic Materials N2 - The movement of magnetic beads due to a magnetic field gradient is of great interest in different application fields. In this report we present a technique based on a magnetic tweezers setup to measure the velocity factor of magnetically actuated individual superparamagnetic beads in a fluidic environment. Several beads can be tracked simultaneously in order to gain and improve statistics. Furthermore we show our results for different beads with hydrodynamic diameters between 200 and 1000 nm from diverse manufacturers. These measurement data can, for example, be used to determine design parameters for a magnetic separation system, like maximum flow rate and minimum separation time, or to select suitable beads for fixed experimental requirements. KW - magnetophoretic velocity KW - superparamagnetic bead KW - magnetic tweezers KW - magnetic separation KW - magnetic actuation Y1 - 2019 U6 - http://dx.doi.org/10.1016/j.jmmm.2018.10.066 SN - 0304-8853 VL - 477 IS - 1 SP - 244 EP - 248 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Achtsnicht, Stefan A1 - Tödter, Julia A1 - Niehues, Julia A1 - Telöken, Matthias A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim A1 - Schröper, Florian T1 - 3D printed modular immunofiltration columns for frequency mixing-based multiplex magnetic immunodetection JF - Sensors N2 - For performing point-of-care molecular diagnostics, magnetic immunoassays constitute a promising alternative to established enzyme-linked immunosorbent assays (ELISA) because they are fast, robust and sensitive. Simultaneous detection of multiple biomolecular targets from one body fluid sample is desired. The aim of this work is to show that multiplex magnetic immunodetection based on magnetic frequency mixing by means of modular immunofiltration columns prepared for different targets is feasible. By calculations of the magnetic response signal, the required spacing between the modules was determined. Immunofiltration columns were manufactured by 3D printing and antibody immobilization was performed in a batch approach. It was shown experimentally that two different target molecules in a sample solution could be individually detected in a single assaying step with magnetic measurements of the corresponding immobilization filters. The arrangement order of the filters and of a negative control did not influence the results. Thus, a simple and reliable approach to multi-target magnetic immunodetection was demonstrated. Y1 - 2019 U6 - http://dx.doi.org/10.3390/s19010148 SN - 1424-8220 VL - 19 IS - 1 PB - MDPI CY - Basel ER - TY - JOUR A1 - Achtsnicht, Stefan A1 - Pourshahidi, Ali Mohammad A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim T1 - Multiplex detection of different magnetic beads using frequency scanning in magnetic frequency mixing technique JF - Sensors N2 - In modern bioanalytical methods, it is often desired to detect several targets in one sample within one measurement. Immunological methods including those that use superparamagnetic beads are an important group of techniques for these applications. The goal of this work is to investigate the feasibility of simultaneously detecting different superparamagnetic beads acting as markers using the magnetic frequency mixing technique. The frequency of the magnetic excitation field is scanned while the lower driving frequency is kept constant. Due to the particles’ nonlinear magnetization, mixing frequencies are generated. To record their amplitude and phase information, a direct digitization of the pickup-coil’s signal with subsequent Fast Fourier Transformation is performed. By synchronizing both magnetic beads using frequency scanning in magnetic frequency mixing technique magnetic fields, a stable phase information is gained. In this research, it is shown that the amplitude of the dominant mixing component is proportional to the amount of superparamagnetic beads inside a sample. Additionally, it is shown that the phase does not show this behaviour. Excitation frequency scans of different bead types were performed, showing different phases, without correlation to their diverse amplitudes. Two commercially available beads were selected and a determination of their amount in a mixture is performed as a demonstration for multiplex measurements. KW - frequency mixing magnetic detection KW - magnetic sandwich immunoassay KW - multiparametric immunoassays Y1 - 2019 U6 - http://dx.doi.org/10.3390/s19112599 SN - 1424-8220 VL - 19 IS - 11 PB - MDPI CY - Basel ER - TY - JOUR A1 - Achtsnicht, Stefan A1 - Neuendorf, Christian A1 - Faßbender, Tobias A1 - Nölke, Greta A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim A1 - Schröper, Florian T1 - Sensitive and rapid detection of cholera toxin subunit B using magnetic frequency mixing detection JF - Plos One N2 - Cholera is a life-threatening disease caused by the cholera toxin (CT) as produced by some Vibrio cholerae serogroups. In this research we present a method which directly detects the toxin’s B subunit (CTB) in drinking water. For this purpose we performed a magnetic sandwich immunoassay inside a 3D immunofiltration column. We used two different commercially available antibodies to capture CTB and for binding to superparamagnetic beads. ELISA experiments were performed to select the antibody combination. The beads act as labels for the magnetic frequency mixing detection technique. We show that the limit of detection depends on the type of magnetic beads. A nonlinear Hill curve was fitted to the calibration measurements by means of a custom-written python software. We achieved a sensitive and rapid detection of CTB within a broad concentration range from 0.2 ng/ml to more than 700 ng/ml. Y1 - 2019 U6 - http://dx.doi.org/10.1371/journal.pone.0219356 SN - 1932-6203 VL - 14 IS - 7 PB - Plos CY - San Francisco ER - TY - JOUR A1 - Baringhaus, Ludwig A1 - Gaigall, Daniel T1 - On an asymptotic relative efficiency concept based on expected volumes of confidence regions JF - Statistics - A Journal of Theoretical and Applied Statistic N2 - The paper deals with an asymptotic relative efficiency concept for confidence regions of multidimensional parameters that is based on the expected volumes of the confidence regions. Under standard conditions the asymptotic relative efficiencies of confidence regions are seen to be certain powers of the ratio of the limits of the expected volumes. These limits are explicitly derived for confidence regions associated with certain plugin estimators, likelihood ratio tests and Wald tests. Under regularity conditions, the asymptotic relative efficiency of each of these procedures with respect to each one of its competitors is equal to 1. The results are applied to multivariate normal distributions and multinomial distributions in a fairly general setting. KW - Volume of confidence regions KW - asymptotic relative efficiency KW - likelihood ratio test KW - multivariate normal distribution KW - multinomial distribution Y1 - 2019 U6 - http://dx.doi.org/10.1080/02331888.2019.1683560 SN - 1029-4910 VL - 53 IS - 6 SP - 1396 EP - 1436 PB - Taylor & Francis CY - London ER - TY - JOUR A1 - Gaigall, Daniel T1 - On a new approach to the multi-sample goodness-of-fit problem JF - Communications in Statistics - Simulation and Computation N2 - Suppose we have k samples X₁,₁,…,X₁,ₙ₁,…,Xₖ,₁,…,Xₖ,ₙₖ with different sample sizes ₙ₁,…,ₙₖ and unknown underlying distribution functions F₁,…,Fₖ as observations plus k families of distribution functions {G₁(⋅,ϑ);ϑ∈Θ},…,{Gₖ(⋅,ϑ);ϑ∈Θ}, each indexed by elements ϑ from the same parameter set Θ, we consider the new goodness-of-fit problem whether or not (F₁,…,Fₖ) belongs to the parametric family {(G₁(⋅,ϑ),…,Gₖ(⋅,ϑ));ϑ∈Θ}. New test statistics are presented and a parametric bootstrap procedure for the approximation of the unknown null distributions is discussed. Under regularity assumptions, it is proved that the approximation works asymptotically, and the limiting distributions of the test statistics in the null hypothesis case are determined. Simulation studies investigate the quality of the new approach for small and moderate sample sizes. Applications to real-data sets illustrate how the idea can be used for verifying model assumptions. KW - Goodness-of-fit test KW - Multi-sample problem KW - Parametric bootstrap Y1 - 2019 U6 - http://dx.doi.org/10.1080/03610918.2019.1618472 SN - 1532-4141 VL - 53 IS - 10 SP - 2971 EP - 2989 PB - Taylor & Francis CY - London ER -