TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Schmidt, Simon A1 - Stelter, Jonathan K. A1 - Wittrich, Marco A1 - Quick, Harald H. A1 - Bitz, Andreas A1 - Ladd, Mark E. T1 - Performance and safety assessment of an integrated transmitarray for body imaging at 7 T under consideration of specificabsorption rate, tissue temperature, and thermal dose JF - NMR in Biomedicine N2 - In this study, the performance of an integrated body-imaging array for 7 T with 32 radiofrequency (RF) channels under consideration of local specific absorption rate (SAR), tissue temperature, and thermal dose limits was evaluated and the imaging performance was compared with a clinical 3 T body coil. Thirty-two transmit elements were placed in three rings between the bore liner and RF shield of the gradient coil. Slice-selective RF pulse optimizations for B1 shimming and spokes were performed for differently oriented slices in the body under consideration of realistic constraints for power and local SAR. To improve the B1+ homogeneity, safety assessments based on temperature and thermal dose were performed to possibly allow for higher input power for the pulse optimization than permissible with SAR limits. The results showed that using two spokes, the 7 T array outperformed the 3 T birdcage in all the considered regions of interest. However, a significantly higher SAR or lower duty cycle at 7 T is necessary in some cases to achieve similar B1+ homogeneity as at 3 T. The homogeneity in up to 50 cm-long coronal slices can particularly benefit from the high RF shim performance provided by the 32 RF channels. The thermal dose approach increases the allowable input power and the corresponding local SAR, in one example up to 100 W/kg, without limiting the exposure time necessary for an MR examination. In conclusion, the integrated antenna array at 7 T enables a clinical workflow for body imaging and comparable imaging performance to a conventional 3 T clinical body coil. KW - body imaging at 7 T MRI KW - thermal dose KW - tissue temperature KW - transmit antenna arrays Y1 - 2022 U6 - http://dx.doi.org/10.1002/nbm.4656 SN - 0952-3480 (Print) SN - 1099-1492 (Online) VL - 35 IS - 5 SP - 1 EP - 17 PB - Wiley ER - TY - JOUR A1 - Fagan, Andrew J. A1 - Bitz, Andreas A1 - Björkman-Burtscher, Isabella M. A1 - Collins, Christopher M. A1 - Kimbrell, Vera A1 - Raaijmakers, Alexander J. E. T1 - 7T MR Safety JF - Journal of Magnetic Resonance Imaging (JMRI) Y1 - 2021 U6 - http://dx.doi.org/10.1002/jmri.27319 SN - 1522-2586 VL - 53 IS - 2 SP - 333 EP - 346 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Clemens, Markus A1 - Bitz, Andreas T1 - Safety of subjects during radiofrequency exposure in ultra-high-field magnetic resonance imaging JF - IEEE Letters on Electromagnetic Compatibility Practice and Applications N2 - Magnetic resonance imaging (MRI) is one of the most important medical imaging techniques. Since the introduction of MRI in the mid-1980s, there has been a continuous trend toward higher static magnetic fields to obtain i.a. a higher signal-to-noise ratio. The step toward ultra-high-field (UHF) MRI at 7 Tesla and higher, however, creates several challenges regarding the homogeneity of the spin excitation RF transmit field and the RF exposure of the subject. In UHF MRI systems, the wavelength of the RF field is in the range of the diameter of the human body, which can result in inhomogeneous spin excitation and local SAR hotspots. To optimize the homogeneity in a region of interest, UHF MRI systems use parallel transmit systems with multiple transmit antennas and time-dependent modulation of the RF signal in the individual transmit channels. Furthermore, SAR increases with increasing field strength, while the SAR limits remain unchanged. Two different approaches to generate the RF transmit field in UHF systems using antenna arrays close and remote to the body are investigated in this letter. Achievable imaging performance is evaluated compared to typical clinical RF transmit systems at lower field strength. The evaluation has been performed under consideration of RF exposure based on local SAR and tissue temperature. Furthermore, results for thermal dose as an alternative RF exposure metric are presented. Y1 - 2020 SN - 2637-6423 U6 - http://dx.doi.org/10.1109/LEMCPA.2020.3029747 VL - 2 IS - 3 SP - 1 EP - 8 PB - IEEE CY - New York, NY ER - TY - JOUR A1 - Orzada, Stephan A1 - Fiedler, Thomas M. A1 - Bitz, Andreas A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - Local SAR compression with overestimation control to reduce maximum relative SAR overestimation and improve multi-channel RF array performance JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective In local SAR compression algorithms, the overestimation is generally not linearly dependent on actual local SAR. This can lead to large relative overestimation at low actual SAR values, unnecessarily constraining transmit array performance. Method Two strategies are proposed to reduce maximum relative overestimation for a given number of VOPs. The first strategy uses an overestimation matrix that roughly approximates actual local SAR; the second strategy uses a small set of pre-calculated VOPs as the overestimation term for the compression. Result Comparison with a previous method shows that for a given maximum relative overestimation the number of VOPs can be reduced by around 20% at the cost of a higher absolute overestimation at high actual local SAR values. Conclusion The proposed strategies outperform a previously published strategy and can improve the SAR compression where maximum relative overestimation constrains the performance of parallel transmission. Y1 - 2020 SN - 1352-8661 U6 - http://dx.doi.org/10.1007/s10334-020-00890-0 IS - 34 (2021) SP - 153 EP - 164 PB - Springer CY - Heidelberg ER - TY - JOUR A1 - Orzada, Stephan A1 - Solbach, Klaus A1 - Gratz, Marcel A1 - Brunheim, Sascha A1 - Fiedler, Thomas M. A1 - Johst, Sören A1 - Bitz, Andreas A1 - Shooshtary, Samaneh A1 - Abuelhaija, Asjraf A1 - Voelker, Maximilian N. A1 - Rietsch, Stefan H. G. A1 - Kraff, Oliver A1 - Maderwald, Stefan A1 - Flöser, Martina A1 - Oehmingen, Mark A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - A 32-channel parallel transmit system add-on for 7T MRI JF - Plos one Y1 - 2019 U6 - http://dx.doi.org/10.1371/journal.pone.0222452 ER - TY - JOUR A1 - Schmidt, Katharina A1 - Forkmann, Katarina A1 - Schultz, Heidrun A1 - Gratz, Marcel A1 - Bitz, Andreas A1 - Wiech, Katja A1 - Bingel, Ulrike T1 - Enhanced Neural Reinstatement for Evoked Facial Pain Compared With Evoked Hand Pain JF - The Journal of Pain Y1 - 2019 U6 - http://dx.doi.org/10.1016/j.jpain.2019.03.003 SN - 1526-5900 IS - In Press, Corrected Proof PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Noureddine, Yacine A1 - Kraff, Oliver A1 - Ladd, Mark E. A1 - Wrede, Karsten A1 - Chen, Bixia A1 - Quick, Harald H. A1 - Schaefers, Georg A1 - Bitz, Andreas T1 - Radiofrequency induced heating around aneurysm clips using a generic birdcage head coil at 7 Tesla under consideration of the minimum distance to decouple multiple aneurysm clips JF - Magnetic Resonance in Medicine Y1 - 2019 U6 - http://dx.doi.org/10.1002/mrm.27835 SN - 1522-2594 IS - Early view SP - 1 EP - 17 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Rietsch, Stefan H. G. A1 - Brunheim, Sascha A1 - Orzada, Stephan A1 - Voelker, Maximilian N. A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Gratz, Marcel A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - Development and evaluation of a 16-channel receive-only RF coil to improve 7T ultra-high field body MRI with focus on the spine JF - Magnetic Resonance in Medicine Y1 - 2019 U6 - http://dx.doi.org/10.1002/mrm.27731 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Rietsch, Stefan H. G. A1 - Pfaffenrot, Viktor A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Brunheim, Sascha A1 - Lazik-Palm, Andrea A1 - Theysohn, Jens M. A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Kraff, Oliver T1 - An 8-channel transceiver 7-channel receive RF coil setup for high SNR ultrahigh-field MRI of the shoulder at 7T JF - Medical Physics Y1 - 2017 U6 - http://dx.doi.org/10.1002/mp.12612 SN - 0094-2405 IS - Article in press PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Johst, Sören A1 - Gratz, Marcel A1 - Völker, Maximilian N. A1 - Kraff, Oliver A1 - Abuelhaija, Ashraf A1 - Fiedler, Thomas M. A1 - Solbach, Klaus A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - Analysis of an integrated 8-Channel Tx/Rx body array for use as a body coil in 7-Tesla MRI JF - Frontiers in Physics Y1 - 2017 U6 - http://dx.doi.org/10.3389/fphy.2017.00017 SN - 2296-424X N1 - Article number 17 VL - 5 IS - Jun ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Buschmann, J. T1 - RF exposure of non restrained animals in an overmoded radial waveguide T2 - 22nd BEMS Annual Meeting, Munich, Germany, 2000 Y1 - 2000 SP - 63 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J.R. A1 - Hansen, V.W. A1 - Lerchl, A. ED - Danesy, Dorothea T1 - Freely moving or restrained animals in bioelec-tromagnetic experiments - pros and cons T2 - AP 2000 : Millennium Conference on Antennas & Propagation, Davos, Switzerland, 9 - 14 April 2000 Y1 - 2000 IS - Band 1 SP - 489 PB - ESA Publications Division, ESTEC CY - Noordwijk ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Stogbauer, F. ED - Hietanen, Maila ED - Jokela, Kari ED - Juutilainen, Jukka ED - Työterveyslaitos, T1 - Shielded RF-exposure system with an integrated LF-measuring system for in vitro experiments with cell-layers T2 - Proceedings of the Ebea 2001 : 5th International Congress of the European Bioelectromagnetics Association (Ebea) ; 6-8 September 2001, Marina Congress Center, Helsinki, Finland Y1 - 2001 SN - 9789518024401 SP - 67 EP - 68 PB - Finnish Institute of Occupational Health CY - Helsinki ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V.W. T1 - Base station antennas embedded in complex environments: rf exposure in controlled areas T2 - 23rd BEMS Annual Meeting, St. Paul, Minnesota, USA, 49-50, June 2001 Y1 - 2001 SP - 49 EP - 50 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V.W. T1 - Absorption rates inside human body due to radiated electro-magnetic fields of multi-band base station antennas. Boundary value problems with electrically large and high-resolution bodies T2 - 24th BEMS Annual Meeting, Quebec City, Quebec, Canada, 23 - 27 June 2002 Y1 - 2002 SP - 40 EP - 41 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V. T1 - Combination of the hybrid/sup [2]/-method and the FDTD-method for solution of boundary value problems with electrically large and high-resolution bodies T2 - IEEE Antennas and Propagation Society International Symposium (IEEE Cat. No.02CH37313) 2002 : San Antonio, TX, USA, 16-21 June Y1 - 2002 SN - 0-7803-7330-8 U6 - http://dx.doi.org/10.1109/APS.2002.1018209 SP - 278 EP - 281 PB - IEEE CY - Piscataway, NJ ER - TY - CHAP A1 - Bitz, Andreas A1 - El Ouardi, A. A1 - Streckert, J. A1 - Hansen, V. T1 - Efficient calculation of human exposure in front of base station antennas by a combination of the FDTD and Hybrid(2)-method T2 - Proceedings of the 16th International Zurich Symposium on Electromagnetic Compatibility, Topical Meeting on Biomedical EMC, Zurich, Switzerland, February 2005 Y1 - 2005 SN - 3-9521199-9-7 SP - 115 EP - 118 PB - ETH Zentrum CY - Zürich ER - TY - CHAP A1 - Bitz, Andreas A1 - Klomp, D.W. A1 - Ladd, M.E. T1 - Experimental and numerical determination of SAR and temperature distribution of a human endorectal coil for MR imaging of the prostate at 7T (903.) T2 - 16th annual ISMRM scientific meeting and exhibition 2008 : Toronto, Ontario, Canada, 3 - 9 May 2008 Y1 - 2009 SN - 978-1-61567-196-0 PB - Curran CY - Red Hook, NY ER - TY - CHAP A1 - Bitz, Andreas A1 - Kraff, O. A1 - Orzada, S. A1 - Maderwald, S. A1 - Brote, I. A1 - Ladd, M. T1 - Experimental and Numerical Assessment of RF Safety of Transmit Coils at 7 Tesla T2 - ISMRM workshop on MR safety 2010 : RF heating of the human in MRI : workshop series. The Washington County Historic Courthouse, Stillwater, Minnesota, USA, 15 - 17 October 2010 Y1 - 2010 SN - 978-1-62276-088-6 SP - 195 ER - TY - CHAP A1 - Bitz, Andreas A1 - Kraff, O. A1 - Orzada, S. A1 - Maderwald, S. A1 - Brote, I. A1 - Johst, S. A1 - Ladd, E. T1 - Assessment of RF Safety of Transmit Coils at 7 Tesla by Experimental and Numerical Procedures (490.) T2 - 19th annual ISMRM scientific meeting and exhibition 2011 : Montreal, Quebec, Canada, 7 - 13 May 2011 Y1 - 2012 SN - 978-1-61839-284-8 IS - Volume 1 SP - 475 PB - Curran CY - Red Hook, NY ER -