TY - JOUR A1 - Schlamann, Marc A1 - Yoon, Min-Suk A1 - Maderwald, Stefan A1 - Pietrzyk, Thomas A1 - Bitz, Andreas A1 - Gerwig, Marcus A1 - Forsting, Michael A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Kastrup, Oliver T1 - Short term effects of magnetic resonance imaging on excitability of the motor cortex at 1.5T and 7T JF - Academic Radiology N2 - Rationale and Objectives The increasing spread of high-field and ultra-high-field magnetic resonance imaging (MRI) scanners has encouraged new discussion of the safety aspects of MRI. Few studies have been published on possible cognitive effects of MRI examinations. The aim of this study was to examine whether changes are measurable after MRI examinations at 1.5 and 7 T by means of transcranial magnetic stimulation (TMS). Materials and Methods TMS was performed in 12 healthy, right-handed male volunteers. First the individual motor threshold was specified, and then the cortical silent period (SP) was measured. Subsequently, the volunteers were exposed to the 1.5-T MRI scanner for 63 minutes using standard sequences. The MRI examination was immediately followed by another TMS session. Fifteen minutes later, TMS was repeated. Four weeks later, the complete setting was repeated using a 7-T scanner. Control conditions included lying in the 1.5-T scanner for 63 minutes without scanning and lying in a separate room for 63 minutes. TMS was performed in the same way in each case. For statistical analysis, Wilcoxon's rank test was performed. Results Immediately after MRI exposure, the SP was highly significantly prolonged in all 12 subjects at 1.5 and 7 T. The motor threshold was significantly increased. Fifteen minutes after the examination, the measured value tended toward normal again. Control conditions revealed no significant differences. Conclusion MRI examinations lead to a transient and highly significant alteration in cortical excitability. This effect does not seem to depend on the strength of the static magnetic field. Y1 - 2010 U6 - https://doi.org/10.1016/j.acra.2009.10.004 SN - 1076-6332 VL - 17 IS - 3 SP - 277 EP - 281 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Ferrein, Alexander T1 - golog.lua: Towards a Non-Prolog Implementation of Golog for Embedded Systems Y1 - 2010 SP - 20 EP - 28 ER - TY - JOUR A1 - Ferrein, Alexander T1 - golog.lua: Towards a Non-Prolog Implementation of Golog for Embedded Systems JF - Cognitive Robotics / Lakemeyer, Gerhard (ed.) Y1 - 2010 N1 - Dagstuhl Seminar Proceedings ; 10081 SP - 1 EP - 15 ER - TY - JOUR A1 - Niemüller, Tim A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - A Lua-based Behavior Engine for Controlling the Humanoid Robot Nao JF - RoboCup 2009: Robot Soccer World Cup XIII Y1 - 2010 N1 - Lecture Notes in Computer Science ; 5949 SP - 240 EP - 251 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Siebel, Nils T. A1 - Steinbauer, Gerald T1 - Hybrid control for autonomous systems — Integrating learning, deliberation and reactive control JF - Robotics and Autonomous Systems Y1 - 2010 SN - 0921-8890 VL - 58 IS - 9 SP - 1037 EP - 1038 ER - TY - JOUR A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Frauenrath, Tobias A1 - Prothmann, Marcel A1 - Dieringer, Matthias A. A1 - Hezel, Fabian A1 - Renz, Wolfgang A1 - Kretschel, Kerstin A1 - Niendorf, Thoralf A1 - Schulz-Menger, Jeanette T1 - Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla—a pilot study N2 - Objectives Interest in cardiovascular magnetic resonance (CMR) at 7 T is motivated by the expected increase in spatial and temporal resolution, but the method is technically challenging. We examined the feasibility of cardiac chamber quantification at 7 T. Methods A stack of short axes covering the left ventricle was obtained in nine healthy male volunteers. At 1.5 T, steady-state free precession (SSFP) and fast gradient echo (FGRE) cine imaging with 7 mm slice thickness (STH) were used. At 7 T, FGRE with 7 mm and 4 mm STH were applied. End-diastolic volume, end-systolic volume, ejection fraction and mass were calculated. Results All 7 T examinations provided excellent blood/myocardium contrast for all slice directions. No significant difference was found regarding ejection fraction and cardiac volumes between SSFP at 1.5 T and FGRE at 7 T, while volumes obtained from FGRE at 1.5 T were underestimated. Cardiac mass derived from FGRE at 1.5 and 7 T was larger than obtained from SSFP at 1.5 T. Agreement of volumes and mass between SSFP at 1.5 T and FGRE improved for FGRE at 7 T when combined with an STH reduction to 4 mm. Conclusions This pilot study demonstrates that cardiac chamber quantification at 7 T using FGRE is feasible and agrees closely with SSFP at 1.5 T. Y1 - 2010 U6 - https://doi.org/10.1007/s00330-010-1888-2 SN - 0938-7994 VL - 20 SP - 2844 EP - 2852 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Hezel, Fabian A1 - Renz, Wolfgang A1 - de Geyer d'Orth, Thibaut A1 - Dieringer, Matthias A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Prothmann, Marcel A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf T1 - Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla JF - Journal of Cardiovascular Magnetic Resonance N2 - Background To demonstrate the applicability of acoustic cardiac triggering (ACT) for imaging of the heart at ultrahigh magnetic fields (7.0 T) by comparing phonocardiogram, conventional vector electrocardiogram (ECG) and traditional pulse oximetry (POX) triggered 2D CINE acquisitions together with (i) a qualitative image quality analysis, (ii) an assessment of the left ventricular function parameter and (iii) an examination of trigger reliability and trigger detection variance derived from the signal waveforms. Results ECG was susceptible to severe distortions at 7.0 T. POX and ACT provided waveforms free of interferences from electromagnetic fields or from magneto-hydrodynamic effects. Frequent R-wave mis-registration occurred in ECG-triggered acquisitions with a failure rate of up to 30% resulting in cardiac motion induced artifacts. ACT and POX triggering produced images free of cardiac motion artefacts. ECG showed a severe jitter in the R-wave detection. POX also showed a trigger jitter of approximately Δt = 72 ms which is equivalent to two cardiac phases. ACT showed a jitter of approximately Δt = 5 ms only. ECG waveforms revealed a standard deviation for the cardiac trigger offset larger than that observed for ACT or POX waveforms. Image quality assessment showed that ACT substantially improved image quality as compared to ECG (image quality score at end-diastole: ECG = 1.7 ± 0.5, ACT = 2.4 ± 0.5, p = 0.04) while the comparison between ECG vs. POX gated acquisitions showed no significant differences in image quality (image quality score: ECG = 1.7 ± 0.5, POX = 2.0 ± 0.5, p = 0.34). Conclusions The applicability of acoustic triggering for cardiac CINE imaging at 7.0 T was demonstrated. ACT's trigger reliability and fidelity are superior to that of ECG and POX. ACT promises to be beneficial for cardiovascular magnetic resonance at ultra-high field strengths including 7.0 T. KW - Interval Time Series KW - Image Quality Score KW - Image Quality Assessment KW - Sound Pressure Level KW - Cardiovascular Magnetic Resonance Y1 - 2010 U6 - https://doi.org/10.1186/1532-429X-12-67 SN - 1532-429X VL - 12 IS - 1 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Niemüller, Tim A1 - Ferrein, Alexander A1 - Beck, Daniel A1 - Lakemeyer, Gerhard T1 - Design Principles of the Component-Based Robot Software Framework Fawkes JF - Simulation, Modeling, and Programming for Autonomous Robots Y1 - 2010 N1 - Lecture Notes in Computer Science ; 6472 ; Second International Conference, SIMPAR 2010, Darmstadt, Germany, November 15-18, 2010. Proceedings SP - 300 EP - 311 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - On the Way to High-Level Programming for Resource-Limited Embedded Systems with Golog JF - Simulation, Modeling, and Programming for Autonomous Robots Y1 - 2010 N1 - Lecture Notes in Computer Science ; 6472 ; Second International Conference, SIMPAR 2010, Darmstadt, Germany, November 15-18, 2010. Proceedings SP - 229 EP - 240 ER - TY - JOUR A1 - Ferrein, Alexander T1 - Robot controllers for highly dynamic environments with real-time constraints JF - Künstliche Intelligenz : KI. 24 (2010), H. 2 Y1 - 2010 SN - 1610-1987 SP - 175 EP - 178 ER - TY - JOUR A1 - Schiffer, Stefan A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - Proceedings of the Fourth International Conference on Intelligent Robotics and Applications (ICIRA 2011) JF - Proceedings of the Fourth International Conference on Intelligent Robotics and Applications (ICIRA 2011) Y1 - 2010 SP - 1 EP - 10 ER - TY - JOUR A1 - Bragard, Michael A1 - Soltau, N. A1 - Thomas, S. A1 - De Doncker, R. W. T1 - The balance of renewable sources and user demands in grids : power electronics for modular battery energy storage systems JF - IEEE transactions on power electronics N2 - The continuously growing amount of renewable sources starts compromising the stability of electrical grids. Contradictory to fossil fuel power plants, energy production of wind and photovoltaic (PV) energy is fluctuating. Although predictions have significantly improved, an outage of multi-MW offshore wind farms poses a challenging problem. One solution could be the integration of storage systems in the grid. After a short overview, this paper focuses on two exemplary battery storage systems, including the required power electronics. The grid integration, as well as the optimal usage of volatile energy reserves, is presented for a 5- kW PV system for home application, as well as for a 100- MW medium-voltage system, intended for wind farm usage. The efficiency and cost of topologies are investigated as a key parameter for large-scale integration of renewable power at medium- and low-voltage. Y1 - 2010 U6 - https://doi.org/10.1109/TPEL.2010.2085455 SN - 0885-8993 VL - 25 IS - 12 SP - 3049 EP - 3056 PB - IEEE CY - New York ER -