TY - JOUR A1 - Miyamoto, Koichiro A1 - Seki, Kosuke A1 - Suto, Takeyuki A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Improved spatial resolution of the chemical imaging sensor with a hybrid illumination that suppresses lateral diffusion of photocarriers JF - Sensor and Actuators B: Chemical N2 - The chemical imaging sensor is a semiconductor-based chemical sensor capable of visualizing pH and ion distributions. The spatial resolution depends on the lateral diffusion of photocarriers generated by illumination of the semiconductor substrate. In this study, two types of optical setups, one based on a bundle of optical fibers and the other based on a binocular tube head, were developed to project a hybrid illumination of a modulated light beam and a ring-shaped constant illumination onto the sensor plate. An improved spatial resolution was realized by the ring-shaped constant illumination, which suppressed lateral diffusion of photocarriers by enhanced recombination due to the increased carrier concentration. Y1 - 2018 U6 - http://dx.doi.org/10.1016/j.snb.2018.07.016 SN - 0925-4005 VL - 273 SP - 1328 EP - 1333 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Yu, Bing A1 - Isoda, Hiroko A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Visualization of the recovery process of defects in a cultured cell layer by chemical imaging sensor JF - Sensors and Actuators B: Chemical N2 - The chemical imaging sensor is a field-effect sensor which is able to visualize both the distribution of ions (in LAPS mode) and the distribution of impedance (in SPIM mode) in the sample. In this study, a novel cell assay is proposed, in which the chemical imaging sensor operated in SPIM mode is applied to monitor the recovery of defects in a cell layer brought into proximity of the sensing surface. A reduced impedance at a defect formed artificially in a cell layer was successfully visualized in a photocurrent image. The cell layer was cultured over two weeks, during which the temporal change of the photocurrent distribution corresponding to the recovery of the defect was observed. Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.snb.2016.04.018 SN - 0925-4005 VL - 236 SP - 965 EP - 969 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Yoshida, Midori A1 - Sakai, Taito A1 - Matsuzaka, Atsushi A1 - Wagner, Torsten A1 - Kanoh, Sanoh A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - Differential setup of light-addressable potentiometric sensor with an enzyme reactor in a flow channel JF - Japanese Journal of Applied Physics. 50 (2011) Y1 - 2011 SN - 0021-4922 SP - 04DL08-1 EP - 04DL08-5 PB - Japan Society of Applied Physics CY - Bristol ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Kanoh, Shin`ichiro A1 - Schöning, Michael Josef T1 - Phase-mode LAPS and its application to chemical imaging JF - Sensors and Actuators B: Chemical. 154 (2011), H. 1 Y1 - 2011 SN - 1873-3077 SP - 28 EP - 32 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Kanoh, Shin`ichiro A1 - Schöning, Michael Josef T1 - Phase-mode operation of FDM-LAPS JF - Sensor letters Y1 - 2011 SN - 1546-1971 VL - 9 IS - 2 SP - 691 EP - 694 PB - American Scientific Publishers CY - Stevenson Ranch, Calif. ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Wagner, Torsten A1 - Mimura, Shuhei A1 - Kanoh, Shiníchiro A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - Constant-phase-mode operation of the light-addressable potentiometric sensor JF - Sensors and Actuators B: Chemical. 154 (2011), H. 2 Y1 - 2011 SN - 1873-3077 N1 - EUROSENSORS XXIII SP - 119 EP - 123 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Wagner, Torsten A1 - Mimura, Shuhei A1 - Kanoh, Shin`ichiro A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - Constant-phase-mode operation of the light-addressable potentiometric sensor JF - Procedia Chemistry. 1 (2009), H. 1 Y1 - 2009 SN - 1876-6196 N1 - Proceedings of the Eurosensors XXIII conference ; Eurosensors 23 SP - 1487 EP - 1490 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Miyamoto, Ko-ichiro A1 - Suto, Takeyuki A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Restraining the Diffusion of Photocarriers to Improve the Spatial Resolution of the Chemical Imaging Sensor T2 - MDPI Proceedings Y1 - 2017 U6 - http://dx.doi.org/10.3390/proceedings1040477 N1 - Eurosensors 2017 Conference, Paris, France, 3–6 September 2017 VL - 1 IS - 4 ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Sugawara, Yuri A1 - Kanoh, Shin´ichiro A1 - Yoshinobu, Tatsuo A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Image correction method for the chemical imaging sensor JF - Sensors and Actuators B: Chemical. 144 (2010), H. 2 Y1 - 2010 N1 - 22nd International Conference on Eurosensors - Dresden, Germany, 7-10 September 2008 ; Eurosensors ; (22, 2008, Dresden) SP - 344 EP - 348 ER - TY - JOUR A1 - Miyamoto, Ko-Ichiro A1 - Sato, Takuya A1 - Abe, Minami A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Light-addressable potentiometric sensor as a sensing element in plug-based microfluidic devices JF - Micromachines N2 - A plug-based microfluidic system based on the principle of the light-addressable potentiometric sensor (LAPS) is proposed. The LAPS is a semiconductor-based chemical sensor, which has a free addressability of the measurement point on the sensing surface. By combining a microfluidic device and LAPS, ion sensing can be performed anywhere inside the microfluidic channel. In this study, the sample solution to be measured was introduced into the channel in a form of a plug with a volume in the range of microliters. Taking advantage of the light-addressability, the position of the plug could be monitored and pneumatically controlled. With the developed system, the pH value of a plug with a volume down to 400 nL could be measured. As an example of plug-based operation, two plugs were merged in the channel, and the pH change was detected by differential measurement. KW - light-addressable potentiometric sensor KW - plug-based microfluidic device KW - chemical sensor Y1 - 2016 U6 - http://dx.doi.org/10.3390/mi7070111 SN - 2072-666X N1 - This article belongs to the Special Issue "Micro/Nano Devices for Chemical Analysis" VL - 7 IS - 7 SP - 111 PB - MDPI CY - Basel ER -