TY - JOUR A1 - Kraff, Oliver A1 - Bitz, Andreas A1 - Dammann, Philipp A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - An eight-channel transmit/receive multipurpose coil for musculoskeletal MR imaging at 7 T JF - Medical Physics N2 - Purpose: MRI plays a leading diagnostic role in assessing the musculoskeletal (MSK) system and is well established for most questions at clinically used field strengths (up to 3 T). However, there are still limitations in imaging early stages of cartilage degeneration, very fine tendons and ligaments, or in locating nerve lesions, for example. 7 T MRI of the knee has already received increasing attention in the current published literature, but there is a strong need to develop new radiofrequency (RF) coils to assess more regions of the MSK system. In this work, an eight-channel transmit/receive RF array was built as a multipurpose coil for imaging some of the thus far neglected regions. An extensive coil characterization protocol and first in vivo results of the human wrist, shoulder, elbow, knee, and ankle imaged at 7 T will be presented. Methods: Eight surface loop coils with a dimension ofurn:x-wiley:00942405:media:mp7176:mp7176-math-0001 were machined from FR4 circuit board material. To facilitate easy positioning, two coil clusters, each with four loop elements, were combined to one RF transmit/receive array. An overlapped and shifted arrangement of the coil elements was chosen to reduce the mutual inductance between neighboring coils. A phantom made of body-simulating liquid was used for tuning and matching on the bench. Afterward, the S-parameters were verified on a human wrist, elbow, and shoulder. For safety validation, a detailed compliance test was performed including full wave simulations of the RF field distribution and the corresponding specific absorption rate (SAR) for all joints. In vivo images of four volunteers were assessed with gradient echo and spin echo sequences modified to obtain optimal image contrast, full anatomic coverage, and the highest spatial resolution within a reasonable acquisition time. The performance of the RF coil was additionally evaluated by in vivo B1 mapping. Results: A comparison of B1 per unit power, flip angle distribution, and anatomic images showed a fairly homogeneous excitation for the smaller joints (elbow, wrist, and ankle), while for the larger joints, the shoulder and especially the knee, B1 inhomogeneities and limited penetration depth were more pronounced. However, the greater part of the shoulder joint could be imaged.In vivo images rendered very fine anatomic details such as fascicles of the median nerve and the branching of the nerve bundles. High-resolution images of cartilage, labrum, and tendons could be acquired. Additionally, turbo spin echo (TSE) and inversion recovery sequences performed very well. Conclusions: This study demonstrates that the concept of two four-channel transmit/receive RF arrays can be used as a multipurpose coil for high-resolutionin vivo MR imaging of the musculoskeletal system at 7 T. Not only gradient echo but also typical clinical and SAR-intensive sequences such as STIR and TSE performed well. Imaging of small structures and peripheral nerves could in particular benefit from this technique. Y1 - 2010 U6 - https://doi.org/10.1118/1.3517176 SN - 2473-4209 VL - 37 IS - 12 SP - 6368 EP - 6376 PB - Wiley CY - Hoboken, NJ ER - TY - JOUR A1 - Orzada, Stephan A1 - Maderwald, Stefan A1 - Poser, Benedikt Andreas A1 - Bitz, Andreas A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high-field MRI JF - Magnetic Resonance in Medicine N2 - As the field strength and, therefore, the operational frequency in MRI is increased, the wavelength approaches the size of the human head/body, resulting in wave effects, which cause signal decreases and dropouts. Several multichannel approaches have been proposed to try to tackle these problems, including RF shimming, where each element in an array is driven by its own amplifier and modulated with a certain (constant) amplitude and phase relative to the other elements, and Transmit SENSE, where spatially tailored RF pulses are used. In this article, a relatively inexpensive and easy to use imaging scheme for 7 Tesla imaging is proposed to mitigate signal voids due to B1 field inhomogeneity. Two time-interleaved images are acquired using a different excitation mode for each. By forming virtual receive elements, both images are reconstructed together using GRAPPA to achieve a more homogeneous image, with only small SNR and SAR penalty in head and body imaging at 7 Tesla. Y1 - 2010 U6 - https://doi.org/10.1002/mrm.22527 SN - 1522-2594 VL - 64 IS - 2 SP - 327 EP - 333 PB - Wiley-Liss CY - New York ER - TY - CHAP A1 - Abel, Thomas A1 - Bonin, Dominik A1 - Albracht, Kirsten A1 - Zeller, Sebastian A1 - Brüggemann, Gert-Peter A1 - Burkett, Brendan A1 - Strüder, Heiko K. T1 - Kinematic profile of the elite handcyclist T2 - 28th International Conference on Biomechanics in Sports, Marquette, Michigan, USA, July 19 – 23, 2010 Y1 - 2017 SN - 1999-4168 SP - 140 EP - 141 ER - TY - THES A1 - Albracht, Kirsten T1 - Influence of mechanical properties of the leg extensor muscletendon units on running economy Y1 - 2010 N1 - Cologne, German Sport Univ., Diss., 2010 PB - Deutsche Sporthochschule Köln CY - Köln ER - TY - JOUR A1 - Aggeloussis, Nickos A1 - Giannakou, Erasmia A1 - Albracht, Kirsten A1 - Arampatzis, Adamantios T1 - Reproducibility of fascicle length and pennation angle of gastrocnemius medialis in human gait in vivo JF - Gait and Posture N2 - The purpose of the current study was to examine the reproducibility of fascicle length and pennation angle of gastrocnemius medialis while human walking. To the best of our knowledge, this is the first study of the reproducibility of fascicle length and pennation angle of gastrocnemius medialis in vivo during human gait. Twelve males performed 10 gait trials on a treadmill, in 2 separate days. B-mode ultrasonography, with the ultrasound probe firmly adjusted in the transverse and frontal planes using a special cast, was used to measure the fascicle length and the pennation angle of the gastrocnemius medialis (GM). A Vicon 624 system with three cameras operating at 120 Hz was also used to record the ankle and knee joint angles. The results showed that measurements of fascicle length and pennation angle showed high reproducibility during the gait cycle, both within the same day and between different days. Moreover, the root mean square differences between the repeated waveforms of both variables were very small, compared with their ranges (fascicle length: RMS = ∼3 mm, range: 38–63 mm; pennation angle: RMS = ∼1.5°, range: 22–32°). However, their reproducibility was lower compared to the joint angles. It was found that representative data have to be derived by a wide number of gait trials (fascicle length ∼six trials, pennation angle more than 10 trials), to assure the reliability of the fascicle length and pennation angle in human gait. Y1 - 2010 U6 - https://doi.org/10.1016/j.gaitpost.2009.08.249 SN - 0966-6362 VL - 31 IS - 1 SP - 73 EP - 77 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Arampatzis, Adamantios A1 - Peper, Andreas A1 - Bierbaum, Stefanie A1 - Albracht, Kirsten T1 - Plasticity of human Achilles tendon mechanical and morphological properties in response to cyclic strain JF - Journal of Biomechanics N2 - The purpose of the current study in combination with our previous published data (Arampatzis et al., 2007) was to examine the effects of a controlled modulation of strain magnitude and strain frequency applied to the Achilles tendon on the plasticity of tendon mechanical and morphological properties. Eleven male adults (23.9±2.2 yr) participated in the study. The participants exercised one leg at low magnitude tendon strain (2.97±0.47%), and the other leg at high tendon strain magnitude (4.72±1.08%) of similar frequency (0.5 Hz, 1 s loading, 1 s relaxation) and exercise volume (integral of the plantar flexion moment over time) for 14 weeks, 4 days per week, 5 sets per session. The exercise volume was similar to the intervention of our earlier study (0.17 Hz frequency; 3 s loading, 3 s relaxation) allowing a direct comparison of the results. Before and after the intervention ankle joint moment has been measured by a dynamometer, tendon–aponeurosis elongation by ultrasound and cross-sectional area of the Achilles tendon by magnet resonance images (MRI). We found a decrease in strain at a given tendon force, an increase in tendon–aponeurosis stiffness and tendon elastic modulus of the Achilles tendon only in the leg exercised at high strain magnitude. The cross-sectional area (CSA) of the Achilles tendon did not show any statistically significant (P>0.05) differences to the pre-exercise values in both legs. The results indicate a superior improvement in tendon properties (stiffness, elastic modulus and CSA) at the low frequency (0.17 Hz) compared to the high strain frequency (0.5 Hz) protocol. These findings provide evidence that the strain magnitude applied to the Achilles tendon should exceed the value, which occurs during habitual activities to trigger adaptational effects and that higher tendon strain duration per contraction leads to superior tendon adaptational responses. Y1 - 2010 U6 - https://doi.org/10.1016/j.jbiomech.2010.08.014 SN - 0021-9290 VL - 43 IS - 16 SP - 3073 EP - 3079 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Evers, Jörg A1 - Fleischhaker, Robert A1 - Pálffy, A. A1 - Keitel, C. T1 - Light propagation : From atomic to nuclear quantum optics T2 - Modern optics and photonics: atoms and structured media Y1 - 2010 SN - 978-981431327-8 U6 - https://doi.org/10.1142/9789814313278_0001 SP - 3 EP - 15 PB - World Scientific Publishing Co. ER - TY - JOUR A1 - Fleischhaker, Robert A1 - Evers, Jörg A1 - Dey, Tarak N. T1 - Phase modulation induced by cooperative effects in electromagnetically induced transparency JF - Physical Review A - Atomic, molecular, and optical physics N2 - We analyze the influence of dipole-dipole interactions in an electromagnetically induced transparency set up for a density at the onset of cooperative effects. To this end, we include mean-field models for the influence of local-field corrections and radiation trapping into our calculation. We show both analytically and numerically that the polarization contribution to the local field strongly modulates the phase of a weak pulse. We give an intuitive explanation for this local-field-induced phase modulation and demonstrate that it distinctively differs from the nonlinear self-phase-modulation that a strong pulse experiences in a Kerr medium. Y1 - 2010 U6 - https://doi.org/10.1103/PhysRevA.82.013815 SN - 1050-2947 VL - 82 IS - 1 PB - AIP Publishing CY - Melville, NY ER - TY - JOUR A1 - Butenweg, Christoph T1 - Bewertung der Erdbebensicherheit von bestehenden Bauwerken JF - Österreichische Ingenieur- und Architekten-Zeitschrift : ÖIAZ ; Zeitschr. d. ÖIAV / Österreichischer Ingenieur- und Architekten-Verein Y1 - 2010 SN - 0721-9415 VL - 155 IS - 1-3 und 4-6 SP - 11 EP - 17 PB - ÖAIV CY - Wien ER - TY - CHAP A1 - Meskouris, Konstantin A1 - Holtschoppen, Britta A1 - Butenweg, Christoph A1 - Park, Jin T1 - Seismische Auslegung von Silo- und Tankbauwerken T2 - Festschrift zum 60. Geburtstag von Univ.-Prof. Dr.-Ing. Ingbert Mangerig. (Berichte aus dem konstruktiven Ingenieurbau ; 2010,6) Y1 - 2010 SP - 215 EP - 224 PB - Univ. der Bundeswehr CY - Neubiberg ER -