TY - JOUR A1 - Orzada, Stephan A1 - Maderwald, Stefan A1 - Poser, Benedikt Andreas A1 - Bitz, Andreas A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high-field MRI JF - Magnetic Resonance in Medicine N2 - As the field strength and, therefore, the operational frequency in MRI is increased, the wavelength approaches the size of the human head/body, resulting in wave effects, which cause signal decreases and dropouts. Several multichannel approaches have been proposed to try to tackle these problems, including RF shimming, where each element in an array is driven by its own amplifier and modulated with a certain (constant) amplitude and phase relative to the other elements, and Transmit SENSE, where spatially tailored RF pulses are used. In this article, a relatively inexpensive and easy to use imaging scheme for 7 Tesla imaging is proposed to mitigate signal voids due to B1 field inhomogeneity. Two time-interleaved images are acquired using a different excitation mode for each. By forming virtual receive elements, both images are reconstructed together using GRAPPA to achieve a more homogeneous image, with only small SNR and SAR penalty in head and body imaging at 7 Tesla. Y1 - 2010 U6 - http://dx.doi.org/10.1002/mrm.22527 SN - 1522-2594 VL - 64 IS - 2 SP - 327 EP - 333 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Kobus, Thiele A1 - Bitz, Andreas A1 - Uden, Mark J. van A1 - Lagemaat, Miram W. A1 - Rothgang, Eva A1 - Orzada, Stephan A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - In vivo 31P MR spectroscopic imaging of the human prostate at 7 T: safety and feasibility JF - Magnetic Resonance in Medicine N2 - 31P MR spectroscopic imaging of the human prostate provides information about phosphorylated metabolites that could be used for prostate cancer characterization. The sensitivity of a magnetic field strength of 7 T might enable 3D 31P MR spectroscopic imaging with relevant spatial resolution in a clinically acceptable measurement time. To this end, a 31P endorectal coil was developed and combined with an eight-channel 1H body-array coil to relate metabolic information to anatomical location. An extensive safety validation was performed to evaluate the specific absorption rate, the radiofrequency field distribution, and the temperature distribution of both coils. This validation consisted of detailed Finite Integration Technique simulations, confirmed by MR thermometry and Burn:x-wiley:07403194:media:MRM24175:tex2gif-stack-1 measurements in a phantom and in vivo temperature measurements. The safety studies demonstrated that the presence of the 31P endorectal coil had no influence on the specific absorption rate levels and temperature distribution of the external eight-channel 1H array coil. To stay within a 10 g averaged local specific absorption rate of 10 W/kg, a maximum time-averaged input power of 33 W for the 1H array coil was allowed. For transmitting with the 31P endorectal coil, our safety limit of less than 1°C temperature increase in vivo during a 15-min MR spectroscopic imaging experiment was reached at a time-averaged input power of 1.9 W. With this power setting, a second in vivo measurement was performed on a healthy volunteer. Using adiabatic excitation, 3D 31P MR spectroscopic imaging produced spectra from the entire prostate in 18 min with a spatial resolution of 4 cm3. The spectral resolution enabled the separate detection of phosphocholine, phosphoethanolamine, inorganic phosphate, and other metabolites that could play an important role in the characterization of prostate cancer. Y1 - 2012 U6 - http://dx.doi.org/10.1002/mrm.24175 SN - 1522-2594 VL - 68 IS - 6 SP - 1683 EP - 1695 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Orzada, S. A1 - Maderwald, S. A1 - Poser, B. A. A1 - Johst, S. A1 - Kannengiesser, S. A1 - Ladd, M. E. A1 - Bitz, Andreas T1 - Time-interleaved acquisition of modes: an analysis of SAR and image contrast implications JF - Magnetic Resonance in Medicine N2 - s the magnetic field strength and therefore the operational frequency in MRI are increased, the radiofrequency wavelength approaches the size of the human head/body, resulting in wave effects which cause signal decreases and dropouts. Especially, whole-body imaging at 7 T and higher is therefore challenging. Recently, an acquisition scheme called time-interleaved acquisition of modes has been proposed to tackle the inhomogeneity problems in high-field MRI. The basic premise is to excite two (or more) different Burn:x-wiley:07403194:media:MRM23081:tex2gif-stack-1 modes using static radiofrequency shimming in an interleaved acquisition, where the complementary radiofrequency patterns of the two modes can be exploited to improve overall signal homogeneity. In this work, the impact of time-interleaved acquisition of mode on image contrast as well as on time-averaged specific absorption rate is addressed in detail. Time-interleaved acquisition of mode is superior in Burn:x-wiley:07403194:media:MRM23081:tex2gif-stack-2 homogeneity compared with conventional radiofrequency shimming while being highly specific absorption rate efficient. Time-interleaved acquisition of modes can enable almost homogeneous high-field imaging throughout the entire field of view in PD, T2, and T2*-weighted imaging and, if a specified homogeneity criterion is met, in T1-weighted imaging as well. Y1 - 2012 U6 - http://dx.doi.org/10.1002/mrm.23081 SN - 1522-2594 VL - 67 IS - 4 SP - 1033 EP - 1041 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Orzada, Stephan A1 - Johst, Sören A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Solbach, Klaus A1 - Ladd, Mark E. T1 - Mitigation of B1(+) inhomogeneity on single-channel transmit systems with TIAMO JF - Magnetic Resonance in Medicine Y1 - 2013 U6 - http://dx.doi.org/10.1002/mrm.24453 SN - 1522-2594 VL - 70 IS - 1 SP - 290 EP - 294 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Umutlu, Lale A1 - Orzada, Stephan A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Bronte, Irina A1 - Bitz, Andreas A1 - Kraff, Oliver A1 - Ladd, Susanne C. A1 - Antoch, Gerald A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Lauenstein, Thomas C. T1 - Renal imaging at 7 Tesla: preliminary results JF - European Radiology N2 - Objective To investigate the feasibility of 7T MR imaging of the kidneys utilising a custom-built 8-channel transmit/receive radiofrequency body coil. Methods In vivo unenhanced MR was performed in 8 healthy volunteers on a 7T whole-body MR system. After B0 shimming the following sequences were obtained: 1) 2D and 3D spoiled gradient-echo sequences (FLASH, VIBE), 2) T1-weighted 2D in and opposed phase 3) True-FISP imaging and 4) a T2-weighted turbo spin echo (TSE) sequence. Visual evaluation of the overall image quality was performed by two radiologists. Results Renal MRI at 7T was feasible in all eight subjects. Best image quality was found using T1-weighted gradient echo MRI, providing high anatomical details and excellent conspicuity of the non-enhanced vasculature. With successful shimming, B1 signal voids could be effectively reduced and/or shifted out of the region of interest in most sequence types. However, T2-weighted TSE imaging remained challenging and strongly impaired because of signal heterogeneities in three volunteers. Conclusion The results demonstrate the feasibility and diagnostic potential of dedicated 7T renal imaging. Further optimisation of imaging sequences and dedicated RF coil concepts are expected to improve the acquisition quality and ultimately provide high clinical diagnostic value. Y1 - 2011 SN - 1432-1084 VL - 21 IS - 4 SP - 841 EP - 849 PB - Springer CY - Berlin ER - TY - JOUR A1 - Umutlu, L. A1 - Maderwald, S. A1 - Kinner, S. A1 - Kraff, O. A1 - Bitz, Andreas A1 - Orzada, S. A1 - Johst, S. A1 - Wrede, K. A1 - Forsting, M. A1 - Ladd, M. E. A1 - Lauenstein, T. C. A1 - Quick, H. H. T1 - First-pass contrast-enhanced renal MRA at 7 Tesla: initial results JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-012-2666-0 SN - 1432-1084 VL - 23 IS - 4 SP - 1059 EP - 1066 PB - Springer CY - Berlin ER - TY - JOUR A1 - Kraff, Oliver A1 - Wrede, Karsten H. A1 - Schoemberg, Tobias A1 - Dammann, Philipp A1 - Noureddine, Yacine A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - MR safety assessment of potential RF heating from cranial fixation plates at 7 T JF - Medical Physics Y1 - 2013 U6 - http://dx.doi.org/10.1118/1.4795347 SN - 2473-4209 VL - 40 IS - 4 SP - 042302-1 EP - 042302-10 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Bitz, Andreas A1 - Felder, Jorg A1 - Wittig, Tilmann T1 - Designing MRI Coils with Aid of Simulation JF - Microwaves & RF Y1 - 2013 SN - 0745-2993 VL - 52 IS - 7 SP - 56 PB - Penton CY - Cleveland, Ohio ER - TY - JOUR A1 - Umutlu, Lale A1 - Kraff, Oliver A1 - Fischer, Anja A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Nassenstein, Kai A1 - Nensa, Felix A1 - Grüneisen, Johannes A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Forsting, Michael A1 - Ladd, Mark E. A1 - Lauenstein, Thomas C. T1 - Seven-Tesla MRI of the female pelvis JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-013-2868-0 SN - 1432-1084 VL - 23 IS - 9 SP - 2364 EP - 2373 PB - Springer CY - Berlin ER - TY - JOUR A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Lagemaat, Miriam W. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Kobus, Thiele A1 - Kraff, Oliver A1 - Maderwald, Stefan A1 - Ladd, Mark E. A1 - Scheenen, Tom W. J. T1 - Feasibility of T₂-weighted turbo spin echo imaging of the human prostate at 7 tesla JF - Magnetic Resonance in Medicine N2 - Purpose To demonstrate that high quality T₂-weighted (T2w) turbo spin-echo (TSE) imaging of the complete prostate can be achieved routinely and within safety limits at 7 T, using an external transceive body array coil only. Methods Nine healthy volunteers and 12 prostate cancer patients were scanned on a 7 T whole-body system. Preparation consisted of B₀ and radiofrequency shimming and localized flip angle calibration. T₁ and T₂ relaxation times were measured and used to define the T2w-TSE protocol. T2w imaging was performed using a TSE sequence (pulse repetition time/echo time 3000–3640/71 ms) with prolonged excitation and refocusing pulses to reduce specific absorption rate. Results High quality T2w TSE imaging was performed in less than 2 min in all subjects. Tumors of patients with gold-standard tumor localization (MR-guided biopsy or prostatectomy) were well visualized on 7 T imaging (n = 3). The number of consecutive slices achievable within a 10-g averaged specific absorption rate limit of 10 W/kg was ≥28 in all subjects, sufficient for full prostate coverage with 3-mm slices in at least one direction. Conclusion High quality T2w TSE prostate imaging can be performed routinely and within specific absorption rate limits at 7 T with an external transceive body array. Y1 - 2014 U6 - http://dx.doi.org/10.1002/mrm.24818 SN - 1522-2594 VL - 71 IS - 5 SP - 1711 EP - 1719 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Vos, Eline K. A1 - Maas, Marnix C. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Uden, Mark J. van A1 - Kobus, Thiele A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - Phosphorus magnetic resonance spectroscopic imaging at 7 T in patients with prostate cancer JF - Investigative Radiology N2 - Objectives The aim of this study was to identify characteristics of phosphorus (³¹P) spectra of the human prostate and to investigate changes of individual phospholipid metabolites in prostate cancer through in vivo ³¹P magnetic resonance spectroscopic imaging (MRSI) at 7 T. Materials and Methods In this institutional review board–approved study, 15 patients with biopsy-proven prostate cancer underwent T₂-weighted magnetic resonance imaging and 3-dimensional ³¹P MRSI at 7 T. Voxels were selected at the tumor location, in normal-appearing peripheral zone tissue, normal-appearing transition zone tissue, and in the base of the prostate close to the seminal vesicles. Phosphorus metabolite ratios were determined and compared between tissue types. Results Signals of phosphoethanolamine (PE) and phosphocholine (PC) were present and well resolved in most ³¹P spectra in the prostate. Glycerophosphocholine signals were observable in 43% of the voxels in malignant tissue, but in only 10% of the voxels in normal-appearing tissue away from the seminal vesicles. In many spectra, independent of tissue type, 2 peaks resonated in the chemical shift range of inorganic phosphate, possibly representing 2 separate pH compartments. The PC/PE ratio in the seminal vesicles was highly elevated compared with the prostate in 5 patients. A considerable overlap of ³¹P metabolite ratios was found between prostate cancer and normal-appearing prostate tissue, preventing direct discrimination of these tissues. The only 2 patients with high Gleason scores tumors (≥4+5) presented with high PC and glycerophosphocholine levels in their cancer lesions. Conclusions Phosphorus MRSI at 7 T shows distinct features of phospholipid metabolites in the prostate gland and its surrounding structures. In this exploratory study, no differences in ³¹P metabolite ratios were observed between prostate cancer and normal-appearing prostate tissue possibly because of the partial volume effects of small tumor foci in large MRSI voxels. Y1 - 2014 U6 - http://dx.doi.org/10.1097/RLI.0000000000000012 SN - 1536-0210 VL - 49 IS - 5 SP - 363 EP - 372 PB - Lippincott Williams & Wilkins CY - Philadelphia, Pa. ER - TY - JOUR A1 - Theysohn, Jens M. A1 - Kraff, Oliver A1 - Eilers, Kristina A1 - Andrade, Dorian A1 - Gerwig, Marcus A1 - Timmann, Dagmar A1 - Schmitt, Franz A1 - Ladd, Mark E. A1 - Ladd, Susanne C. A1 - Bitz, Andreas T1 - Vestibular effects of a 7 Tesla MRI examination compared to 1.5 T and 0 T in healthy volunteers JF - PLoS one N2 - Ultra-high-field MRI (7 Tesla (T) and above) elicits more temporary side-effects compared to 1.5 T and 3 T, e.g. dizziness or “postural instability” even after exiting the scanner. The current study aims to assess quantitatively vestibular performance before and after exposure to different MRI scenarios at 7 T, 1.5 T and 0 T. Sway path and body axis rotation (Unterberger's stepping test) were quantitatively recorded in a total of 46 volunteers before, 2 minutes after, and 15 minutes after different exposure scenarios: 7 T head MRI (n = 27), 7 T no RF (n = 22), 7 T only B₀ (n = 20), 7 T in & out B₀ (n = 20), 1.5 T no RF (n = 20), 0 T (n = 15). All exposure scenarios lasted 30 minutes except for brief one minute exposure in 7 T in & out B₀. Both measures were documented utilizing a 3D ultrasound system. During sway path evaluation, the experiment was repeated with eyes both open and closed. Sway paths for all long-lasting 7 T scenarios (normal, no RF, only B₀) with eyes closed were significantly prolonged 2 minutes after exiting the scanner, normalizing after 15 minutes. Brief exposure to 7 T B₀ or 30 minutes exposure to 1.5 T or 0 T did not show significant changes. End positions after Unterberger's stepping test were significantly changed counter-clockwise after all 7 T scenarios, including the brief in & out B₀ exposure. Shorter exposure resulted in a smaller alteration angle. In contrast to sway path, reversal of changes in body axis rotation was incomplete after 15 minutes. 1.5 T caused no rotational changes. The results show that exposure to the 7 Tesla static magnetic field causes only a temporary dysfunction or “over-compensation” of the vestibular system not measurable at 1.5 or 0 Tesla. Radiofrequency fields, gradient switching, and orthostatic dysregulation do not seem to play a role. Y1 - 2014 U6 - http://dx.doi.org/10.1371/journal.pone.0092104 SN - 1932-6203 VL - 9 IS - 3 PB - PLOS CY - San Francisco ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Weiland, Elisabeth A1 - Uden, Mark J. van A1 - Kobus, Thiele A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - (31) P MR spectroscopic imaging of the human prostate at 7 T: T1 relaxation times, Nuclear Overhauser Effect, and spectral characterization JF - Magnetic Resonance in Medicine Y1 - 2015 U6 - http://dx.doi.org/10.1002/mrm.25209 SN - 1522-2594 VL - 73 IS - 3 SP - 909 EP - 920 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Vos, E. K. A1 - Lagemaat, M. W. A1 - Barentsz, J. O. A1 - Fütterer, J. J. A1 - Zamecnik, P. A1 - Roozen, H. A1 - Orzada, S. A1 - Bitz, Andreas A1 - Maas, M. C. A1 - Scheenen, T. W. J. T1 - Image quality and cancer visibility of T2-weighted Magnetic Resonance Imaging of the prostate at 7 Tesla JF - European Radiology N2 - Objectives To assess the image quality of T2-weighted (T2w) magnetic resonance imaging of the prostate and the visibility of prostate cancer at 7 Tesla (T). Materials & methods Seventeen prostate cancer patients underwent T2w imaging at 7T with only an external transmit/receive array coil. Three radiologists independently scored images for image quality, visibility of anatomical structures, and presence of artefacts. Krippendorff’s alpha and weighted kappa statistics were used to assess inter-observer agreement. Visibility of prostate cancer lesions was assessed by directly linking the T2w images to the confirmed location of prostate cancer on histopathology. Results T2w imaging at 7T was achievable with ‘satisfactory’ (3/5) to ‘good’ (4/5) quality. Visibility of anatomical structures was predominantly scored as ‘satisfactory’ (3/5) and ‘good’ (4/5). If artefacts were present, they were mostly motion artefacts and, to a lesser extent, aliasing artefacts and noise. Krippendorff’s analysis revealed an α = 0.44 between three readers for the overall image quality scores. Clinically significant cancer lesions in both peripheral zone and transition zone were visible at 7T. Conclusion T2w imaging with satisfactory to good quality can be routinely acquired, and cancer lesions were visible in patients with prostate cancer at 7T using only an external transmit/receive body array coil. Y1 - 2014 U6 - http://dx.doi.org/10.1007/s00330-014-3234-6 SN - 1432-1084 VL - 24 IS - 8 SP - 1950 EP - 1958 PB - Springer CY - Cham ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Breukels, Vincent A1 - Vos, Eline K. A1 - B., Adam A1 - Uden, Mark J. van A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Maas, Marnix C. A1 - Scheenen, Tom W. J. T1 - ¹H MR spectroscopic imaging of the prostate at 7T using spectral-spatial pulses JF - Magnetic Resonance in Medicine N2 - Purpose To assess the feasibility of prostate ¹H MR spectroscopic imaging (MRSI) using low-power spectral-spatial (SPSP) pulses at 7T, exploiting accurate spectral selection and spatial selectivity simultaneously. Methods A double spin-echo sequence was equipped with SPSP refocusing pulses with a spectral selectivity of 1 ppm. Three-dimensional prostate ¹H-MRSI at 7T was performed with the SPSP-MRSI sequence using an 8-channel transmit array coil and an endorectal receive coil in three patients with prostate cancer and in one healthy subject. No additional water or lipid suppression pulses were used. Results Prostate ¹H-MRSI could be obtained well within specific absorption rate (SAR) limits in a clinically feasible time (10 min). Next to the common citrate signals, the prostate spectra exhibited high spermine signals concealing creatine and sometimes also choline. Residual lipid signals were observed at the edges of the prostate because of limitations in spectral and spatial selectivity. Conclusion It is possible to perform prostate ¹H-MRSI at 7T with a SPSP-MRSI sequence while using separate transmit and receive coils. This low-SAR MRSI concept provides the opportunity to increase spatial resolution of MRSI within reasonable scan times. Y1 - 2016 U6 - http://dx.doi.org/10.1002/mrm.25569 SN - 1522-2594 VL - 75 IS - 3 SP - 933 EP - 945 PB - International Society for Magnetic Resonance in Medicine ER - TY - JOUR A1 - Noureddine, Yacine A1 - Bitz, Andreas A1 - Ladd, Mark E. A1 - Thürling, Markus A1 - Ladd, Susanne C. A1 - Schaefers, Gregor A1 - Kraff, Oliver T1 - Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study JF - Magnetic Resonance Materials in Physics, Biology and Medicine Y1 - 2015 U6 - http://dx.doi.org/10.1007/s10334-015-0499-y SN - 1352-8661 VL - 28 IS - 6 SP - 577 EP - 590 PB - Springer CY - Berlin ER - TY - JOUR A1 - Bank, Bart L. van de A1 - Orzada, Stephan A1 - Smits, Frits A1 - Lagemaat, Miriam W. A1 - Rodgers, Christopher T. A1 - Bitz, Andreas A1 - Scheenen, Tom W. J. T1 - Optimized (31) P MRS in the human brain at 7 T with a dedicated RF coil setup JF - NMR in Biomedicine Y1 - 2015 U6 - http://dx.doi.org/10.1002/nbm.3422 SN - 1099-1492 VL - 28 IS - 11 SP - 1570 EP - 1578 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Schmidt, K. A1 - Forkmann, K. A1 - Sinke, C. A1 - Gratz, M. A1 - Bitz, Andreas A1 - Bingel, U. T1 - The differential effect of trigeminal vs. peripheral pain stimulation on visual processing and memory encoding is influenced by pain-related fear JF - NeuroImage N2 - Compared to peripheral pain, trigeminal pain elicits higher levels of fear, which is assumed to enhance the interruptive effects of pain on concomitant cognitive processes. In this fMRI study we examined the behavioral and neural effects of trigeminal (forehead) and peripheral (hand) pain on visual processing and memory encoding. Cerebral activity was measured in 23 healthy subjects performing a visual categorization task that was immediately followed by a surprise recognition task. During the categorization task subjects received concomitant noxious electrical stimulation on the forehead or hand. Our data show that fear ratings were significantly higher for trigeminal pain. Categorization and recognition performance did not differ between pictures that were presented with trigeminal and peripheral pain. However, object categorization in the presence of trigeminal pain was associated with stronger activity in task-relevant visual areas (lateral occipital complex, LOC), memory encoding areas (hippocampus and parahippocampus) and areas implicated in emotional processing (amygdala) compared to peripheral pain. Further, individual differences in neural activation between the trigeminal and the peripheral condition were positively related to differences in fear ratings between both conditions. Functional connectivity between amygdala and LOC was increased during trigeminal compared to peripheral painful stimulation. Fear-driven compensatory resource activation seems to be enhanced for trigeminal stimuli, presumably due to their exceptional biological relevance. Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.neuroimage.2016.03.026 SN - 1053-8119 VL - 134 SP - 386 EP - 395 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Chen, Bixia A1 - Schoemberg, Tobias A1 - Kraff, Oliver A1 - Dammann, Philipp A1 - Bitz, Andreas A1 - Schlamann, Marc A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Sure, Ulrich A1 - Wrede, Karsten H. T1 - Cranial fixation plates in cerebral magnetic resonance imaging: a 3 and 7 Tesla in vivo image quality study JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective This study assesses and quantifies impairment of postoperative magnetic resonance imaging (MRI) at 7 Tesla (T) after implantation of titanium cranial fixation plates (CFPs) for neurosurgical bone flap fixation. Materials and methods The study group comprised five patients who were intra-individually examined with 3 and 7 T MRI preoperatively and postoperatively (within 72 h/3 months) after implantation of CFPs. Acquired sequences included T₁-weighted magnetization-prepared rapid-acquisition gradient-echo (MPRAGE), T₂-weighted turbo-spin-echo (TSE) imaging, and susceptibility-weighted imaging (SWI). Two experienced neurosurgeons and a neuroradiologist rated image quality and the presence of artifacts in consensus reading. Results Minor artifacts occurred around the CFPs in MPRAGE and T2 TSE at both field strengths, with no significant differences between 3 and 7 T. In SWI, artifacts were accentuated in the early postoperative scans at both field strengths due to intracranial air and hemorrhagic remnants. After resorption, the brain tissue directly adjacent to skull bone could still be assessed. Image quality after 3 months was equal to the preoperative examinations at 3 and 7 T. Conclusion Image quality after CFP implantation was not significantly impaired in 7 T MRI, and artifacts were comparable to those in 3 T MRI. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s10334-016-0548-1 SN - 1352-8661 VL - 29 IS - 3 SP - 389 EP - 398 PB - Springer CY - Berlin ER - TY - JOUR A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - A method to approximate maximum local SAR in multichannel transmit MR systems without transmit phase information JF - Magnetic Resonance in Medicine N2 - Purpose To calculate local specific absorption rate (SAR) correctly, both the amplitude and phase of the signal in each transmit channel have to be known. In this work, we propose a method to derive a conservative upper bound for the local SAR, with a reasonable safety margin without knowledge of the transmit phases of the channels. Methods The proposed method uses virtual observation points (VOPs). Correction factors are calculated for each set of VOPs that prevent underestimation of local SAR when the VOPs are applied with the correct amplitudes but fixed phases. Results The proposed method proved to be superior to the worst-case calculation based on the maximum eigenvalue of the VOPs. The mean overestimation for six coil setups could be reduced, whereas no underestimation of the maximum local SAR occurred. In the best investigated case, the overestimation could be reduced from a factor of 3.3 to a factor of 1.7. Conclusion The upper bound for the local SAR calculated with the proposed method allows a fast estimation of the local SAR based on power measurements in the transmit channels and facilitates SAR monitoring in systems that do not have the capability to monitor transmit phases Y1 - 2016 U6 - http://dx.doi.org/10.1002/mrm.26398 SN - 1522-2594 VL - 78 IS - 2 SP - 805 EP - 811 PB - International Society for Magnetic Resonance in Medicine ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus temperature limits JF - Medical Physics Y1 - 2017 U6 - http://dx.doi.org/10.1002/mp.12034 N1 - This article is corrected by: Errata: Erratum: “RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus tissue temperature limits” [Med. Phys. 44(1), 143–157 (2017)] Volume 44, Issue 2, 772 VL - 44 IS - 1 SP - 143 EP - 157 ER - TY - JOUR A1 - Noureddine, Yacine A1 - Kraff, Oliver A1 - Ladd, Mark E. A1 - Wrede, Karsten H. A1 - Chen, Bixia A1 - Quick, Harald H. A1 - Schaefers, Gregor A1 - Bitz, Andreas T1 - In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla JF - Magnetic Resonance in Medicine Y1 - 2017 U6 - http://dx.doi.org/10.1002/mrm.26650 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - SAR Simulations & Safety JF - NeuroImage Y1 - 2017 U6 - http://dx.doi.org/10.1016/j.neuroimage.2017.03.035 SN - 1053-8119 IS - Epub ahead of print PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - 20 Years of RoboCup - A Subjective Retrospection JF - KI - Künstliche Intelligenz N2 - This summer, RoboCup competitions were held for the 20th time in Leipzig, Germany. It was the second time that RoboCup took place in Germany, 10 years after the 2006 RoboCup in Bremen. In this article, we give an overview on the latest developments of RoboCup and what happened in the different leagues over the last decade. With its 20th edition, RoboCup clearly is a success story and a role model for robotics competitions. From our personal view point, we acknowledge this by giving a retrospection about what makes RoboCup such a success. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s13218-016-0449-5 SN - 1610-1987 VL - 30 IS - 3 SP - 225 EP - 232 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ferrein, Alexander A1 - Schiffer, Stefan A1 - Booysen, T. A1 - Stopforth, R. T1 - Why it is harder to run RoboCup in South Africa: Experiences from German South African collaborations JF - International Journal of Advanced Robotic Systems N2 - Robots are widely used as a vehicle to spark interest in science and technology in learners. A number of initiatives focus on this issue, for instance, the Roberta Initiative, the FIRST Lego League, the World Robot Olympiad and RoboCup Junior. Robotic competitions are valuable not only for school learners but also for university students, as the RoboCup initiative shows. Besides technical skills, the students get some project exposure and experience what it means to finish their tasks on time. But qualifying students for future high-tech areas should not only be for students from developed countries. In this article, we present our experiences with research and education in robotics within the RoboCup initiative, in Germany and South Africa; we report on our experiences with trying to get the RoboCup initiative in South Africa going. RoboCup has a huge support base of academic institutions in Germany; this is not the case in South Africa. We present our ‘north–south’ collaboration initiatives in RoboCup between Germany and South Africa and discuss some of the reasons why we think it is harder to run RoboCup in South Africa. Y1 - 2016 U6 - http://dx.doi.org/10.1177/1729881416662789 SN - 1729-8806 VL - 13 IS - 5 SP - 1 EP - 13 ER - TY - JOUR A1 - Haag, S. A1 - Zontar, D. A1 - Schleupen, Josef A1 - Müller, T. A1 - Brecher, C. T1 - Chain of refined perception in self-optimizing assembly of micro-optical systems JF - Journal of sensors and sensor systems N2 - Today, the assembly of laser systems requires a large share of manual operations due to its complexity regarding the optimal alignment of optics. Although the feasibility of automated alignment of laser optics has been shown in research labs, the development effort for the automation of assembly does not meet economic requirements – especially for low-volume laser production. This paper presents a model-based and sensor-integrated assembly execution approach for flexible assembly cells consisting of a macro-positioner covering a large workspace and a compact micromanipulator with camera attached to the positioner. In order to make full use of available models from computer-aided design (CAD) and optical simulation, sensor systems at different levels of accuracy are used for matching perceived information with model data. This approach is named "chain of refined perception", and it allows for automated planning of complex assembly tasks along all major phases of assembly such as collision-free path planning, part feeding, and active and passive alignment. The focus of the paper is put on the in-process image-based metrology and information extraction used for identifying and calibrating local coordinate systems as well as the exploitation of that information for a part feeding process for micro-optics. Results will be presented regarding the processes of automated calibration of the robot camera as well as the local coordinate systems of part feeding area and robot base. Y1 - 2014 U6 - http://dx.doi.org/10.5194/jsss-3-87-2014 SN - 2194-878X VL - 3 IS - 1 SP - 87 EP - 95 PB - Copernicus Publ. CY - Göttingen ER - TY - JOUR A1 - Leingartner, Max A1 - Maurer, Johannes A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - Evaluation of Sensors and Mapping Approaches for Disasters in Tunnels JF - Journal of Field Robotics N2 - Ground or aerial robots equipped with advanced sensing technologies, such as three-dimensional laser scanners and advanced mapping algorithms, are deemed useful as a supporting technology for first responders. A great deal of excellent research in the field exists, but practical applications at real disaster sites are scarce. Many projects concentrate on equipping robots with advanced capabilities, such as autonomous exploration or object manipulation. In spite of this, realistic application areas for such robots are limited to teleoperated reconnaissance or search. In this paper, we investigate how well state-of-the-art and off-the-shelf components and algorithms are suited for reconnaissance in current disaster-relief scenarios. The basic idea is to make use of some of the most common sensors and deploy some widely used algorithms in a disaster situation, and to evaluate how well the components work for these scenarios. We acquired the sensor data from two field experiments, one from a disaster-relief operation in a motorway tunnel, and one from a mapping experiment in a partly closed down motorway tunnel. Based on these data, which we make publicly available, we evaluate state-of-the-art and off-the-shelf mapping approaches. In our analysis, we integrate opinions and replies from first responders as well as from some algorithm developers on the usefulness of the data and the limitations of the deployed approaches, respectively. We discuss the lessons we learned during the two missions. These lessons are interesting for the community working in similar areas of urban search and rescue, particularly reconnaissance and search. Y1 - 2016 U6 - http://dx.doi.org/10.1002/rob.21611 SN - 1556-4967 VL - 33 IS - 8 SP - 1037 EP - 1057 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Schiffer, Stefan A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - Abstracting Away Low-Level Details in Service Robotics with Fuzzy Fluents JF - Model-Driven Knowledge Engineering for Improved Software Modularity in Robotics and Automation. Workshop at European Robotics Forum 2015 Vienna, Austria, March 11-13, 2015. Y1 - 2015 SP - 1 EP - 4 ER - TY - JOUR A1 - Klinge, Uwe A1 - Otto, Jens A1 - Mühl, Thomas T1 - High Structural Stability of Textile Implants Prevents Pore Collapse and Preserves Effective Porosity at Strain JF - BioMed Research International Y1 - 2015 U6 - http://dx.doi.org/10.1155/2015/953209 SN - 2314-6133 (Print) SN - 2314-6141 (Online) N1 - Article ID: 953209 VL - 2015 ER - TY - JOUR A1 - Bragard, Michael A1 - van Hoek, H. A1 - De Doncker, R. W. T1 - A major design step in IETO concept realization that allows overcurrent protection and pushes limits of switching performance JF - IEEE transactions on power electronics N2 - This paper presents the latest prototype of the integrated emitter turn-off thyristor concept, which potentially ranks among thyristor high-power devices like the gate turn-off thyristor and the integrated gate-commutated thyristor (IGCT). Due to modifications of the external driver stage and mechanical press-pack design optimization, this prototype allows for full device characterization. The turn-off capability was increased to 1600 A with an active silicon area of 823mm2 . This leads to a transient peak power of 672.1kW/cm² . Within this paper, measurements and concept assessment are presented and a comparison to state-of-the-art IGCT devices is provided. Y1 - 2012 U6 - http://dx.doi.org/10.1109/TPEL.2012.2189136 SN - 0885-8993 VL - 27 IS - 9 SP - 4163 EP - 4171 PB - IEEE CY - New York ER - TY - JOUR A1 - Bragard, Michael A1 - Conrad, M. A1 - van Hoek, H. A1 - De Doncker, R. W. T1 - The integrated emitter turn-off thyristor (IETO) : an innovative thyristor-based high power semiconductor device using MOS assisted turn-off JF - IEEE transactions on industry applications Y1 - 2011 U6 - http://dx.doi.org/10.1109/TIA.2011.2161432 SN - 0093-9994 VL - 47 IS - 5 SP - 2175 EP - 2182 PB - IEEE CY - New York ER - TY - JOUR A1 - Bragard, Michael A1 - Soltau, N. A1 - Thomas, S. A1 - De Doncker, R. W. T1 - The balance of renewable sources and user demands in grids : power electronics for modular battery energy storage systems JF - IEEE transactions on power electronics N2 - The continuously growing amount of renewable sources starts compromising the stability of electrical grids. Contradictory to fossil fuel power plants, energy production of wind and photovoltaic (PV) energy is fluctuating. Although predictions have significantly improved, an outage of multi-MW offshore wind farms poses a challenging problem. One solution could be the integration of storage systems in the grid. After a short overview, this paper focuses on two exemplary battery storage systems, including the required power electronics. The grid integration, as well as the optimal usage of volatile energy reserves, is presented for a 5- kW PV system for home application, as well as for a 100- MW medium-voltage system, intended for wind farm usage. The efficiency and cost of topologies are investigated as a key parameter for large-scale integration of renewable power at medium- and low-voltage. Y1 - 2010 U6 - http://dx.doi.org/10.1109/TPEL.2010.2085455 SN - 0885-8993 VL - 25 IS - 12 SP - 3049 EP - 3056 PB - IEEE CY - New York ER - TY - JOUR A1 - Köllensperger, P. A1 - Bragard, Michael A1 - Plum, T. A1 - De Doncker, R. W. T1 - The dual GCT : new high-power device using optimized GCT technology JF - IEEE transactions on industry applications Y1 - 2009 U6 - http://dx.doi.org/10.1109/TIA.2009.2027364 SN - 0093-9994 VL - 45 IS - 5 SP - 1754 EP - 1762 ER - TY - JOUR A1 - Holtrup, S. A1 - Sadeghfam, Arash A1 - Heuermann, Holger A1 - Awakowicz, P. T1 - Characterization and optimization technique for microwave-driven high-intensity discharge lamps using hot S-parameters JF - IEEE transactions on microwave theories and techniques N2 - High-intensity discharge lamps can be driven by radio-frequency signals in the ISM frequency band at 2.45 GHz, using a matching network to transform the impedance of the plasma to the source impedance. To achieve an optimal operating condition, a good characterization of the lamp in terms of radio frequency equivalent circuits under operating conditions is necessary, enabling the design of an efficient matching network. This paper presents the characterization technique for such lamps and presents the design of the required matching network. For the characterization, a high-intensity discharge lamp was driven by a monofrequent large signal at 2.45 GHz, whereas a frequency sweep over 300 MHz was performed across this signal to measure so-called small-signal hot S-parameters using a vector network analyzer. These parameters are then used as an equivalent load in a circuit simulator to design an appropriate matching network. Using the measured data as a black-box model in the simulation results in a quick and efficient method to simulate and design efficient matching networks in spite of the complex plasma behavior. Furthermore, photometric analysis of high-intensity discharge lamps are carried out, comparing microwave operation to conventional operation. Y1 - 2014 U6 - http://dx.doi.org/10.1109/TMTT.2014.2342652 SN - 0018-9480 VL - 62 IS - 10 SP - 2471 EP - 2480 PB - IEEE CY - New York ER - TY - JOUR A1 - Alhwarin, Faraj A1 - Ferrein, Alexander A1 - Scholl, Ingrid T1 - IR stereo kinect: improving depth images by combining structured light with IR stereo Y1 - 2014 SP - 1 EP - 9 ER - TY - JOUR A1 - Ferrein, Alexander T1 - Towards applying soccer moves in the RoboCup Standard Platform League JF - ÖGAI-Journal Y1 - 2008 SN - 0254-4326 VL - Bd. 27 IS - H. 3 SP - 4 EP - 9 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Niemüller, Tim A1 - Potgieter, Anet T1 - Team Zadeat 2009 : team report Y1 - 2009 VL - 6 SP - 1 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Niemüller, Tim A1 - Steinbauer, Gerald T1 - Team Zadeat 2010 : application for participation Y1 - 2010 ER - TY - JOUR A1 - Beck, Daniel A1 - Buchleitner, Martin A1 - Ferrein, Alexander A1 - Niemüller, Tim A1 - Steinbauer, Gerald T1 - Mostly Harmless & AllemaniACs - mixed innovations Y1 - 2014 SP - 1 EP - 8 ER - TY - JOUR A1 - Vorst, Phillip A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - AllemaniACs3D team description Y1 - 2006 SP - 1 EP - 6 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Potgieter, Anet T1 - Establishing the RoboCup Standard League in Africa - applying for the RoboCup Standard League with a German-Austrian-South African Research Project Y1 - 2007 SP - 1 EP - 5 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Potgieter, Anet T1 - RoboCup Standard Platform League - Team Zadeat : an intercontinental research effort Y1 - 2008 SP - 1 EP - 7 ER - TY - JOUR A1 - Gligorevic, Snjezana T1 - Airport surface propagation channel in the C-Band: measurements and modeling JF - IEEE transactions on antennas and propagation Y1 - 2013 SN - 0018-926x VL - Vol. 61 IS - Iss. 9 SP - 4792 EP - 4802 PB - IEEE CY - New York ER - TY - JOUR A1 - Gligorevic, Snjezana T1 - Joint channel estimation and equalisation of fast time-varying frequency-selective channels JF - European transactions on telecommunications Y1 - 2008 SN - 1541-8251; 2161-3915; 1120-3862; 1124-318X VL - Vol. 19 IS - Iss. 1 SP - 1 EP - 13 ER - TY - JOUR A1 - Brück, Stefan A1 - Sorger, Ulrich A1 - Gligorevic, Snjezana A1 - Stolte, Norbert T1 - Interleaving for outer convolutional codes in DS-CDMA systems JF - IEEE transactions on communications Y1 - 2000 SN - 0090-6778 VL - Vol. 48 IS - Iss. 7 SP - 1100 EP - 1107 ER - TY - JOUR A1 - Niemüller, Tim A1 - Lakemeyer, Gerhard A1 - Ferrein, Alexander A1 - Reuter, S. A1 - Ewert, D. A1 - Jeschke, S. A1 - Pensky, D. A1 - Karras, Ulrich T1 - Proposal for advancements to the LLSF in 2014 and beyond Y1 - 2013 SP - Publ. online ER - TY - JOUR A1 - Jens, Otto A1 - Kaldenhoff, E. A1 - Kirschner-Hermanns, R. A1 - Mühl, Thomas A1 - Klinge, Uwe T1 - Elongation of textile pelvic floor implants under load is related to complete loss of effective porosity, thereby favoring incorporation in scar plates JF - Journal of biomedical materials research. Part A N2 - Use of textile structures for reinforcement of pelvic floor structures has to consider mechanical forces to the implant, which are quite different to the tension free conditions of the abdominal wall. Thus, biomechanical analysis of textile devices has to include the impact of strain on stretchability and effective porosity. Prolift® and Prolift + M®, developed for tension free conditions, were tested by measuring stretchability and effective porosity applying mechanical strain. For comparison, we used Dynamesh-PR4®, which was designed for pelvic floor repair to withstand mechanical strain. Prolift® at rest showed moderate porosity with little stretchability but complete loss of effective porosity at strain of 4.9 N/cm. Prolift + M® revealed an increased porosity at rest, but at strain showed high stretchability, with subsequent loss of effective porosity at strain of 2.5 N/cm. Dynamesh PR4® preserved its high porosity even under strain, but as consequence of limited stretchability. Though in tension free conditions Prolift® and Prolift + M® can be considered as large pore class I meshes, application of mechanical strain rapidly lead to collapse of pores. The loss of porosity at mechanical stress can be prevented by constructions with high structural stability. Assessment of porosity under strain was found helpful to define requirements for pelvic floor devices. Clinical studies have to prove whether devices with high porosity as well as high structural stability can improve the patients' outcome. Y1 - 2014 U6 - http://dx.doi.org/10.1002/jbm.a.34767 SN - 1552-4965 VL - 102 IS - 4 SP - 1079 EP - 1084 PB - Wiley CY - New York ER - TY - JOUR A1 - Ferrein, Alexander A1 - Meyer, Thomas T1 - A Brief Overview of Artificial Intelligence in South Africa JF - AI Magazine N2 - South Africa in recent years is the establishment of a number of research hubs involved in AI activities ranging from mobile robotics and computational intelligence, to knowledge representation and reasoning, and human language technologies. In this survey we take the reader through a quick tour of the research being conducted at these hubs, and touch on an initiative to maintain and extend the current level of interest in AI research in the country. Y1 - 2012 U6 - http://dx.doi.org/10.1609/aimag.v33i1.2357 SN - 0738-4602 VL - 33 IS - 1 SP - 99 EP - 101 PB - AAAI CY - Menlo Park ER - TY - JOUR A1 - Rens, Gavin A1 - Ferrein, Alexander T1 - Belief-node condensation for online POMDP algorithms N2 - Slightly extended version of the paper accepted at the Robotics and Artificial Intelligence Workshop, a special track of IEEE AFRICON-2013, held in Mauritius, 9-12 September 2013 Y1 - 2013 SP - 1 EP - 7 PB - IEEE CY - New York ER - TY - JOUR A1 - Schiffer, Stefan A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - Caesar: an intelligent domestic service robot JF - Intelligent service robotics N2 - In this paper we present CAESAR, an intelligent domestic service robot. In domestic settings for service robots complex tasks have to be accomplished. Those tasks benefit from deliberation, from robust action execution and from flexible methods for human–robot interaction that account for qualitative notions used in natural language as well as human fallibility. Our robot CAESAR deploys AI techniques on several levels of its system architecture. On the low-level side, system modules for localization or navigation make, for instance, use of path-planning methods, heuristic search, and Bayesian filters. For face recognition and human–machine interaction, random trees and well-known methods from natural language processing are deployed. For deliberation, we use the robot programming and plan language READYLOG, which was developed for the high-level control of agents and robots; it allows combining programming the behaviour using planning to find a course of action. READYLOG is a variant of the robot programming language Golog. We extended READYLOG to be able to cope with qualitative notions of space frequently used by humans, such as “near” and “far”. This facilitates human–robot interaction by bridging the gap between human natural language and the numerical values needed by the robot. Further, we use READYLOG to increase the flexible interpretation of human commands with decision-theoretic planning. We give an overview of the different methods deployed in CAESAR and show the applicability of a system equipped with these AI techniques in domestic service robotics Y1 - 2012 U6 - http://dx.doi.org/10.1007/s11370-012-0118-y SN - 1861-2776 N1 - Special Issue on Artificial Intelligence Techniques for Robotics: Sensing, Representation and Action, Part I VL - 5 IS - 4 SP - 259 EP - 276 PB - Springer CY - Berlin ER -