TY - JOUR A1 - Mosset, J.-B. A1 - Devroede, O. A1 - Krieguer, M. A1 - Rey, M. A1 - Vieira, J.-M. A1 - Jung, J. H. A1 - Kuntner, C. A1 - Streun, M. A1 - Ziemons, Karl A1 - Auffray, E. A1 - Sempere-Roldan, P. A1 - Lecoq, P. A1 - Bruyndonckx, P. A1 - Loude, J.-F. A1 - Tavernier, S. A1 - Morcel, C. T1 - Development of an optimized LSO/LuYAP phoswich detector head for the Lausanne ClearPET demonstrator JF - IEEE Transactions on Nuclear Science N2 - This paper describes the LSO/LuYAP phoswich detector head developed for the ClearPET small animal PET scanner demonstrator that is under construction in Lausanne within the Crystal Clear Collaboration. The detector head consists of a dual layer of 8×8 LSO and LuYAP crystal arrays coupled to a multi-anode photomultiplier tube (Hamamatsu R7600-M64). Equalistion of the LSO/LuYAP light collection is obtained through partial attenuation of the LSO scintillation light using a thin aluminum deposit of 20-35 nm on LSO and appropriate temperature regulation of the phoswich head between 30°C to 60°C. At 511keV, typical FWHM energy resolutions of the pixels of a phoswich head amounts to (28±2)% for LSO and (25±2)% for LuYAP. The LSO versus LuYAP crystal identification efficiency is better than 98%. Six detector modules have been mounted on a rotating gantry. Axial and tangential spatial resolutions were measured up to 4 cm from the scanner axis and compared to Monte Carlo simulations using GATE. FWHM spatial resolution ranges from 1.3 mm on axis to 2.6 mm at 4 cm from the axis. Y1 - 2006 SN - 0018-9499 VL - 53 IS - 1 SP - 25 EP - 29 ER -