TY - CHAP A1 - Mandekar, Swati A1 - Jentsch, Lina A1 - Lutz, Kai A1 - Behbahani, Mehdi A1 - Melnykowycz, Mark T1 - Earable design analysis for sleep EEG measurements T2 - UbiComp '21 N2 - Conventional EEG devices cannot be used in everyday life and hence, past decade research has been focused on Ear-EEG for mobile, at-home monitoring for various applications ranging from emotion detection to sleep monitoring. As the area available for electrode contact in the ear is limited, the electrode size and location play a vital role for an Ear-EEG system. In this investigation, we present a quantitative study of ear-electrodes with two electrode sizes at different locations in a wet and dry configuration. Electrode impedance scales inversely with size and ranges from 450 kΩ to 1.29 MΩ for dry and from 22 kΩ to 42 kΩ for wet contact at 10 Hz. For any size, the location in the ear canal with the lowest impedance is ELE (Left Ear Superior), presumably due to increased contact pressure caused by the outer-ear anatomy. The results can be used to optimize signal pickup and SNR for specific applications. We demonstrate this by recording sleep spindles during sleep onset with high quality (5.27 μVrms). KW - EEG KW - sensors KW - Impedance Spectroscopy KW - Sleep EEG KW - biopotential electrodes Y1 - 2021 U6 - http://dx.doi.org/10.1145/3460418.3479328 N1 - UbiComp '21: Adjunct Proceedings of the 2021 ACM International Joint Conference on Pervasive and Ubiquitous Computing and Proceedings of the 2021 ACM International Symposium on Wearable Computers, September 21–26, 2021, Virtual, USA SP - 171 EP - 175 ER - TY - CHAP A1 - Schmitz, Annika A1 - Apandi, Shah Eiman Amzar Shah A1 - Spillner, Jan A1 - Hima, Flutura A1 - Behbahani, Mehdi ED - Digel, Ilya ED - Staat, Manfred ED - Trzewik, Jürgen ED - Sielemann, Stefanie ED - Erni, Daniel ED - Zylka, Waldemar T1 - Effect of different cannula positions in the pulmonary artery on blood flow and gas exchange using computational fluid dynamics analysis T2 - 4th YRA MedTech Symposium 2024 : February 1 / 2024 / FH Aachen N2 - Pulmonary arterial cannulation is a common and effective method for percutaneous mechanical circulatory support for concurrent right heart and respiratory failure [1]. However, limited data exists to what effect the positioning of the cannula has on the oxygen perfusion throughout the pulmonary artery (PA). This study aims to evaluate, using computational fluid dynamics (CFD), the effect of different cannula positions in the PA with respect to the oxygenation of the different branching vessels in order for an optimal cannula position to be determined. The four chosen different positions (see Fig. 1) of the cannulas are, in the lower part of the main pulmonary artery (MPA), in the MPA at the junction between the right pulmonary artery (RPA) and the left pulmonary artery (LPA), in the RPA at the first branch of the RPA and in the LPA at the first branch of the LPA. Y1 - 2024 SN - 978-3-940402-65-3 U6 - http://dx.doi.org/10.17185/duepublico/81475 SP - 29 EP - 30 PB - Universität Duisburg-Essen CY - Duisburg ER - TY - JOUR A1 - Thiebes, Anja Lena A1 - Klein, Sarah A1 - Zingsheim, Jonas A1 - Möller, Georg H. A1 - Gürzing, Stefanie A1 - Reddemann, Manuel A. A1 - Behbahani, Mehdi A1 - Cornelissen, Christian G. T1 - Effervescent atomizer as novel cell spray technology to decrease the gas-to-liquid ratio JF - pharmaceutics N2 - Cell spraying has become a feasible application method for cell therapy and tissue engineering approaches. Different devices have been used with varying success. Often, twin-fluid atomizers are used, which require a high gas velocity for optimal aerosolization characteristics. To decrease the amount and velocity of required air, a custom-made atomizer was designed based on the effervescent principle. Different designs were evaluated regarding spray characteristics and their influence on human adipose-derived mesenchymal stromal cells. The arithmetic mean diameters of the droplets were 15.4–33.5 µm with decreasing diameters for increasing gas-to-liquid ratios. The survival rate was >90% of the control for the lowest gas-to-liquid ratio. For higher ratios, cell survival decreased to approximately 50%. Further experiments were performed with the design, which had shown the highest survival rates. After seven days, no significant differences in metabolic activity were observed. The apoptosis rates were not influenced by aerosolization, while high gas-to-liquid ratios caused increased necrosis levels. Tri-lineage differentiation potential into adipocytes, chondrocytes, and osteoblasts was not negatively influenced by aerosolization. Thus, the effervescent aerosolization principle was proven suitable for cell applications requiring reduced amounts of supplied air. This is the first time an effervescent atomizer was used for cell processing. KW - tri-lineage differentiation KW - survival KW - twin-fluid atomizer KW - adipose-derived stromal cells (ASCs) KW - cell atomization KW - cell aerosolization Y1 - 2022 U6 - http://dx.doi.org/10.3390/pharmaceutics14112421 N1 - This article belongs to the Special Issue "Stromal, Stem, Signaling Cells: The Multiple Roles and Applications of Mesenchymal Cells" VL - 14 IS - 11 PB - MDPI CY - Basel ER - TY - JOUR A1 - Probst, M. A1 - Behbahani, Mehdi A1 - Borrmann, E. A1 - Elgeti, S. A1 - Nicolai, M. A1 - Behr, M. T1 - Hemodynamic Modeling for Numerical Analysis and Design of Medical Devices Y1 - 2010 N1 - Posterpresentation ; NIC Symposium 2010 ; 24 - 25 February 2010 Jülich, Germany ER - TY - JOUR A1 - Doorschodt, B. M. A1 - Schreinemachers, M. C. J. M. A1 - Behbahani, Mehdi A1 - Florquin, S. A1 - Weis, J. A1 - Staat, Manfred A1 - Tolba, R. H. T1 - Hypothermic machine perfusion of kidney grafts: which pressure is preferred JF - Annals of Biomedical Engineering. 39 (2011), H. 3 Y1 - 2011 SN - 1573-9686 SP - 1051 EP - 1059 PB - Springer CY - Berlin ER - TY - JOUR A1 - Brockhaus, Moritz K. A1 - Behbahani, Mehdi A1 - Muris, Farina A1 - Jansen, Sebastian V. A1 - Schmitz- Rode, Thomas A1 - Steinseifer, Ulrich A1 - Clauser, Johanna C. T1 - In vitro thrombogenicity testing of pulsatile mechanical circulatory support systems: Design and proof-of-concept JF - Artificial Organs N2 - Thrombogenic complications are a main issue in mechanical circulatory support (MCS). There is no validated in vitro method available to quantitatively assess the thrombogenic performance of pulsatile MCS devices under realistic hemodynamic conditions. The aim of this study is to propose a method to evaluate the thrombogenic potential of new designs without the use of complex in-vivo trials. This study presents a novel in vitro method for reproducible thrombogenicity testing of pulsatile MCS systems using low molecular weight heparinized porcine blood. Blood parameters are continuously measured with full blood thromboelastometry (ROTEM; EXTEM, FIBTEM and a custom-made analysis HEPNATEM). Thrombus formation is optically observed after four hours of testing. The results of three experiments are presented each with two parallel loops. The area of thrombus formation inside the MCS device was reproducible. The implantation of a filter inside the loop catches embolizing thrombi without a measurable increase of platelet activation, allowing conclusions of the place of origin of thrombi inside the device. EXTEM and FIBTEM parameters such as clotting velocity (α) and maximum clot firmness (MCF) show a total decrease by around 6% with a characteristic kink after 180 minutes. HEPNATEM α and MCF rise within the first 180 minutes indicate a continuously increasing activation level of coagulation. After 180 minutes, the consumption of clotting factors prevails, resulting in a decrease of α and MCF. With the designed mock loop and the presented protocol we are able to identify thrombogenic hot spots inside a pulsatile pump and characterize their thrombogenic potential. Y1 - 2021 U6 - http://dx.doi.org/10.1111/aor.14046 SN - 1525-1594 VL - 45 IS - 12 SP - 1513 EP - 1521 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Lenz, Maximilian A1 - Kahmann, Stephanie Lucina A1 - Behbahani, Mehdi A1 - Pennig, Lenhard A1 - Hackl, Michael A1 - Leschinger, Tim A1 - Müller, Lars Peter A1 - Wegmann, Kilian T1 - Influence of rotator cuff preload on fracture configuration in proximal humerus fractures: a proof of concept for fracture simulation JF - Archives of Orthopaedic and Trauma Surgery N2 - Introduction In regard of surgical training, the reproducible simulation of life-like proximal humerus fractures in human cadaveric specimens is desirable. The aim of the present study was to develop a technique that allows simulation of realistic proximal humerus fractures and to analyse the influence of rotator cuff preload on the generated lesions in regards of fracture configuration. Materials and methods Ten cadaveric specimens (6 left, 4 right) were fractured using a custom-made drop-test bench, in two groups. Five specimens were fractured without rotator cuff preload, while the other five were fractured with the tendons of the rotator cuff preloaded with 2 kg each. The humeral shaft and the shortened scapula were potted. The humerus was positioned at 90° of abduction and 10° of internal rotation to simulate a fall on the elevated arm. In two specimens of each group, the emergence of the fractures was documented with high-speed video imaging. Pre-fracture radiographs were taken to evaluate the deltoid-tuberosity index as a measure of bone density. Post-fracture X-rays and CT scans were performed to define the exact fracture configurations. Neer’s classification was used to analyse the fractures. Results In all ten cadaveric specimens life-like proximal humerus fractures were achieved. Two III-part and three IV-part fractures resulted in each group. The preloading of the rotator cuff muscles had no further influence on the fracture configuration. High-speed videos of the fracture simulation revealed identical fracture mechanisms for both groups. We observed a two-step fracture mechanism, with initial impaction of the head segment against the glenoid followed by fracturing of the head and the tuberosities and then with further impaction of the shaft against the acromion, which lead to separation of the tuberosities. Conclusion A high energetic axial impulse can reliably induce realistic proximal humerus fractures in cadaveric specimens. The preload of the rotator cuff muscles had no influence on initial fracture configuration. Therefore, fracture simulation in the proximal humerus is less elaborate. Using the presented technique, pre-fractured specimens are available for real-life surgical education. KW - Proximal humerus fracture KW - Biomechanical simulation KW - Fracture configuration KW - Fracture simulation KW - Rotator cuff Y1 - 2022 U6 - http://dx.doi.org/10.1007/s00402-022-04471-9 SN - 1434-3916 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Kopp, Alexander A1 - Schunck, Laura A1 - Gosau, Martin A1 - Smeets, Ralf A1 - Burg, Simon A1 - Fuest, Sandra A1 - Kröger, Nadja A1 - Zinser, Max A1 - Krohn, Sebastian A1 - Behbahani, Mehdi A1 - Köpf, Marius A1 - Lauts, Lisa A1 - Rutkowski, Rico T1 - Influence of the casting concentration on the mechanical and optical properties of Fa/CaCl2-derived silk fibroin membranes JF - International Journal of Molecular Sciences N2 - In this study, we describe the manufacturing and characterization of silk fibroin membranes derived from the silkworm Bombyx mori. To date, the dissolution process used in this study has only been researched to a limited extent, although it entails various potential advantages, such as reduced expenses and the absence of toxic chemicals in comparison to other conventional techniques. Therefore, the aim of this study was to determine the influence of different fibroin concentrations on the process output and resulting membrane properties. Casted membranes were thus characterized with regard to their mechanical, structural and optical assets via tensile testing, SEM, light microscopy and spectrophotometry. Cytotoxicity was evaluated using BrdU, XTT, and LDH assays, followed by live–dead staining. The formic acid (FA) dissolution method was proven to be suitable for the manufacturing of transparent and mechanically stable membranes. The fibroin concentration affects both thickness and transparency of the membranes. The membranes did not exhibit any signs of cytotoxicity. When compared to other current scientific and technical benchmarks, the manufactured membranes displayed promising potential for various biomedical applications. Further research is nevertheless necessary to improve reproducible manufacturing, including a more uniform thickness, less impurity and physiological pH within the membranes. Y1 - 2020 U6 - http://dx.doi.org/10.3390/ijms21186704 SN - 1422-0067 N1 - Special issue: Optimization of Biomaterials for Reconstructive and Regenerative Medicine VL - 21 IS - 18 art. no. 6704 PB - MDPI CY - Basel ER - TY - JOUR A1 - Behbahani, Mehdi A1 - Nicolai, M. T1 - Interaktive Exploration von Blutströmungsprozessen in Herzpumpen JF - VRCA-Jahresbericht 2005/2006 Y1 - 2006 N1 - Virtual Reality Center SP - 88 EP - 90 PB - Virtual Reality Center CY - Aachen ER - TY - JOUR A1 - Jansen, S. V. A1 - Behbahani, Mehdi A1 - Laumen, M. A1 - Kaufmann, T. A1 - Hormes, M. A1 - Behr, M. A1 - Schmitz-Rode, T. A1 - Steinseifer, U. T1 - Investigation of Steady Flow Through a Realistic Model of the Thoracic Human Aorta Using 3D Stereo PIV and CFD-Simulation Y1 - 2010 N1 - Posterpresentation ; American Society of Artificial Organs (ASAIO), Baltimore, USA, May 27-29, 2010 ER -