TY - JOUR A1 - Khodaverdi, M. A1 - Chatziioannou, A. F. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. A1 - Pietrzyk, U. T1 - Investigation of different MicroCT scanner configurations by GEANT4 simulations JF - IEEE Transactions on Nuclear Science N2 - This study has been performed to design the combination of the new ClearPET (ClearPET is a trademark of the Crystal Clear Collaboration), a small animal positron emission tomography (PET) system, with a micro-computed tomography (microCT) scanner. The properties of different microCT systems have been determined by simulations based on GEANT4. We will demonstrate the influence of the detector material and the X-ray spectrum on the obtained contrast. Four different detector materials (selenium, cadmium zinc telluride, cesium iodide and gadolinium oxysulfide) and two X-ray spectra (a molybdenum and a tungsten source) have been considered. The spectra have also been modified by aluminum filters of varying thickness. The contrast between different tissue types (water, air, brain, bone and fat) has been simulated by using a suitable phantom. The results indicate the possibility to improve the image contrast in microCT by an optimized combination of the X-ray source and detector material. Y1 - 2005 SN - 0018-9499 VL - 52 IS - 1 SP - 188 EP - 192 ER - TY - JOUR A1 - Khodaverdi, M. A1 - Chaziioannou, A. F. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. A1 - Pietrzyk, U. T1 - Investigation of different microCT scanner configurations by GEANT4 simulations JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4 N2 - This study has been performed to design the combination of the new ClearPET TM (ClearPET is a trademark of the Crystal Clear Collaboration), a small animal Positron Emission Tomography (PET) system, with a microComputed Tomography (microCT) scanner. The properties of different microCT systems have been determined by simulations based on GEANT4. We demonstrate the influence of the detector material and the X-ray spectrum on the obtained contrast. Four different detector materials (selenium, cadmium zinc telluride, cesium iodide and gadolinium oxysulfide) and two X-ray spectra (a molybdenum and a tungsten source) have been considered. The spectra have also been modified by aluminum filters of varying thickness. The contrast between different tissue types (water, air, brain, bone and fat) has been simulated by using a suitable phantom. The results indicate the possibility to improve the image contrast in microCT by an optimized combination of the X-ray source and detector material. Y1 - 2004 SN - 1082-3654 SP - 2989 EP - 2993 ER - TY - JOUR A1 - Christ, D. A1 - Hollendung, A. A1 - Larue, H. A1 - Parl, C. A1 - Streun, M. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. T1 - Homogenization of the MultiChannel PM gain by inserting light attenuating masks JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4 N2 - MultiChannel Photomultipliers (PM), like the R7600-00-M64 or R5900-00-M64 from Hamamatsu, are often chosen as photodetectors in high-resolution positron emission tomography (PET). A major problem of this PM is the nonuniform channel gain. In order to solve this problem, light attenuating masks were created. The aim of the masks is a homogenization of the output of all 64 channels using different hole sizes at the channel positions. The hole area, which is individually defined for the different channels, is inversely proportional to the channel gain. The measurements by inserting light attenuating masks improved a homogenization to a ratio of 1:1.2. Y1 - 2004 SN - 1082-3654 SP - 2382 EP - 2385 ER - TY - CHAP A1 - Terstegge, Andreas A1 - Weber, S. A1 - Herzog, H. A1 - Müller-Gärtner, H. W. A1 - Halling, H. T1 - Design and implementation aspects of a 3D reconstruction algorithm for the Jülich TierPET system T2 - 1997 International Meeting on Fully ThreeDimensional Image Reconstruction in Radiology and Nuclear Medicine : 3D97 Y1 - 1997 SP - 170 EP - 173 ER - TY - CHAP A1 - Weber, S. A1 - Terstegge, Andreas A1 - Engels, R. A1 - Herzog, H. A1 - Reinartz, R. A1 - Reinhart, P. A1 - Rongen, F. A1 - Müller-Gärtner, H. W. A1 - Halling, H. T1 - The KFA TierPET: performance characteristics and measurements T2 - Conference record / 1996 IEEE Nuclear Science Symposium [and Medical Imaging], November 2 - 9, 1996, Anaheim, California ; vol. 2 Y1 - 1996 SN - 0-7803-3534-1 SN - 1082-3654 SP - 1117 EP - 1119 PB - IEEE CY - Piscataway, NJ ER - TY - CHAP A1 - Weber, S. A1 - Terstegge, Andreas A1 - Halling, H. A1 - Herzog, H. A1 - Reinartz, R. A1 - Reinhart, P. A1 - Rongen, F. A1 - Müller-Gärtner, H.-W. T1 - The design of an animal PET: flexible geometry for achieving optimal spatial resolution or high sensitivity T2 - Conference record / 1995 IEEE Nuclear Science Symposium and Medical Imaging, October 21 - 28, 1995, San Francisco ; vol. 2 Y1 - 1995 SN - 078033180X ; 0780331818 ; 0780331826 SP - 1002 EP - 1005 PB - IEEE CY - Piscataway, NJ ER -