TY - JOUR A1 - Umutlu, Lale A1 - Orzada, Stephan A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Bronte, Irina A1 - Bitz, Andreas A1 - Kraff, Oliver A1 - Ladd, Susanne C. A1 - Antoch, Gerald A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Lauenstein, Thomas C. T1 - Renal imaging at 7 Tesla: preliminary results JF - European Radiology N2 - Objective To investigate the feasibility of 7T MR imaging of the kidneys utilising a custom-built 8-channel transmit/receive radiofrequency body coil. Methods In vivo unenhanced MR was performed in 8 healthy volunteers on a 7T whole-body MR system. After B0 shimming the following sequences were obtained: 1) 2D and 3D spoiled gradient-echo sequences (FLASH, VIBE), 2) T1-weighted 2D in and opposed phase 3) True-FISP imaging and 4) a T2-weighted turbo spin echo (TSE) sequence. Visual evaluation of the overall image quality was performed by two radiologists. Results Renal MRI at 7T was feasible in all eight subjects. Best image quality was found using T1-weighted gradient echo MRI, providing high anatomical details and excellent conspicuity of the non-enhanced vasculature. With successful shimming, B1 signal voids could be effectively reduced and/or shifted out of the region of interest in most sequence types. However, T2-weighted TSE imaging remained challenging and strongly impaired because of signal heterogeneities in three volunteers. Conclusion The results demonstrate the feasibility and diagnostic potential of dedicated 7T renal imaging. Further optimisation of imaging sequences and dedicated RF coil concepts are expected to improve the acquisition quality and ultimately provide high clinical diagnostic value. Y1 - 2011 SN - 1432-1084 VL - 21 IS - 4 SP - 841 EP - 849 PB - Springer CY - Berlin ER - TY - JOUR A1 - Umutlu, Lale A1 - Kraff, Oliver A1 - Fischer, Anja A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Nassenstein, Kai A1 - Nensa, Felix A1 - Grüneisen, Johannes A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Forsting, Michael A1 - Ladd, Mark E. A1 - Lauenstein, Thomas C. T1 - Seven-Tesla MRI of the female pelvis JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-013-2868-0 SN - 1432-1084 VL - 23 IS - 9 SP - 2364 EP - 2373 PB - Springer CY - Berlin ER - TY - JOUR A1 - Umutlu, Lale A1 - Bitz, Andreas A1 - Maderwald, Stefan A1 - Orzada, Stephan A1 - Kinner, Sonja A1 - Kraff, Oliver A1 - Brote, Irina A1 - Ladd, Susanne C. A1 - Schroeder, Tobias A1 - Forsting, Michael T1 - Contrast-enhanced ultra-high-field liver MRI: a feasibility trial JF - European Journal of Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1016/j.ejrad.2011.07.004 SN - 0720-048X VL - 82 IS - 5 SP - 760 EP - 767 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Schlamann, Marc A1 - Yoon, Min-Suk A1 - Maderwald, Stefan A1 - Pietrzyk, Thomas A1 - Bitz, Andreas A1 - Gerwig, Marcus A1 - Forsting, Michael A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Kastrup, Oliver T1 - Short term effects of magnetic resonance imaging on excitability of the motor cortex at 1.5T and 7T JF - Academic Radiology N2 - Rationale and Objectives The increasing spread of high-field and ultra-high-field magnetic resonance imaging (MRI) scanners has encouraged new discussion of the safety aspects of MRI. Few studies have been published on possible cognitive effects of MRI examinations. The aim of this study was to examine whether changes are measurable after MRI examinations at 1.5 and 7 T by means of transcranial magnetic stimulation (TMS). Materials and Methods TMS was performed in 12 healthy, right-handed male volunteers. First the individual motor threshold was specified, and then the cortical silent period (SP) was measured. Subsequently, the volunteers were exposed to the 1.5-T MRI scanner for 63 minutes using standard sequences. The MRI examination was immediately followed by another TMS session. Fifteen minutes later, TMS was repeated. Four weeks later, the complete setting was repeated using a 7-T scanner. Control conditions included lying in the 1.5-T scanner for 63 minutes without scanning and lying in a separate room for 63 minutes. TMS was performed in the same way in each case. For statistical analysis, Wilcoxon's rank test was performed. Results Immediately after MRI exposure, the SP was highly significantly prolonged in all 12 subjects at 1.5 and 7 T. The motor threshold was significantly increased. Fifteen minutes after the examination, the measured value tended toward normal again. Control conditions revealed no significant differences. Conclusion MRI examinations lead to a transient and highly significant alteration in cortical excitability. This effect does not seem to depend on the strength of the static magnetic field. Y1 - 2010 U6 - http://dx.doi.org/10.1016/j.acra.2009.10.004 SN - 1076-6332 VL - 17 IS - 3 SP - 277 EP - 281 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Schlamann, Marc A1 - Voigt, Melanie A. A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Kraff, Oliver A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Forsting, Michael A1 - Wilhelm, Hans T1 - Exposure to high-field MRI does not affect cognitive function JF - Journal of Magnetic Resonance Imaging N2 - Purpose To assess potential cognitive deficits under the influence of static magnetic fields at various field strengths some studies already exist. These studies were not focused on attention as the most vulnerable cognitive function. Additionally, mostly no magnetic resonance imaging (MRI) sequences were performed. Materials and Methods In all, 25 right-handed men were enrolled in this study. All subjects underwent one MRI examination of 63 minutes at 1.5 T and one at 7 T within an interval of 10 to 30 days. The order of the examinations was randomized. Subjects were referred to six standardized neuropsychological tests strictly focused on attention immediately before and after each MRI examination. Differences in neuropsychological variables between the timepoints before and after each MRI examination were assessed and P-values were calculated Results Only six subtests revealed significant differences between pre- and post-MRI. In these tests the subjects achieved better results in post-MRI testing than in pre-MRI testing (P = 0.013–0.032). The other tests revealed no significant results. Conclusion The improvement in post-MRI testing is only explicable as a result of learning effects. MRI examinations, even in ultrahigh-field scanners, do not seem to have any persisting influence on the attention networks of human cognition immediately after exposure. Y1 - 2010 U6 - http://dx.doi.org/10.1002/jmri.22065 SN - 1522-2586 VL - 31 IS - 5 SP - 1061 EP - 1066 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Rietsch, Stefan H. G. A1 - Brunheim, Sascha A1 - Orzada, Stephan A1 - Voelker, Maximilian N. A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Gratz, Marcel A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - Development and evaluation of a 16-channel receive-only RF coil to improve 7T ultra-high field body MRI with focus on the spine JF - Magnetic Resonance in Medicine Y1 - 2019 U6 - http://dx.doi.org/10.1002/mrm.27731 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Orzada, Stephan A1 - Solbach, Klaus A1 - Gratz, Marcel A1 - Brunheim, Sascha A1 - Fiedler, Thomas M. A1 - Johst, Sören A1 - Bitz, Andreas A1 - Shooshtary, Samaneh A1 - Abuelhaija, Asjraf A1 - Voelker, Maximilian N. A1 - Rietsch, Stefan H. G. A1 - Kraff, Oliver A1 - Maderwald, Stefan A1 - Flöser, Martina A1 - Oehmingen, Mark A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - A 32-channel parallel transmit system add-on for 7T MRI JF - Plos one Y1 - 2019 U6 - http://dx.doi.org/10.1371/journal.pone.0222452 ER - TY - JOUR A1 - Orzada, Stephan A1 - Maderwald, Stefan A1 - Poser, Benedikt Andreas A1 - Bitz, Andreas A1 - Quick, Harald H. A1 - Ladd, Mark E. T1 - RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high-field MRI JF - Magnetic Resonance in Medicine N2 - As the field strength and, therefore, the operational frequency in MRI is increased, the wavelength approaches the size of the human head/body, resulting in wave effects, which cause signal decreases and dropouts. Several multichannel approaches have been proposed to try to tackle these problems, including RF shimming, where each element in an array is driven by its own amplifier and modulated with a certain (constant) amplitude and phase relative to the other elements, and Transmit SENSE, where spatially tailored RF pulses are used. In this article, a relatively inexpensive and easy to use imaging scheme for 7 Tesla imaging is proposed to mitigate signal voids due to B1 field inhomogeneity. Two time-interleaved images are acquired using a different excitation mode for each. By forming virtual receive elements, both images are reconstructed together using GRAPPA to achieve a more homogeneous image, with only small SNR and SAR penalty in head and body imaging at 7 Tesla. Y1 - 2010 U6 - http://dx.doi.org/10.1002/mrm.22527 SN - 1522-2594 VL - 64 IS - 2 SP - 327 EP - 333 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Orzada, Stephan A1 - Johst, Sören A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Solbach, Klaus A1 - Ladd, Mark E. T1 - Mitigation of B1(+) inhomogeneity on single-channel transmit systems with TIAMO JF - Magnetic Resonance in Medicine Y1 - 2013 U6 - http://dx.doi.org/10.1002/mrm.24453 SN - 1522-2594 VL - 70 IS - 1 SP - 290 EP - 294 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Schäfer, Lena C. A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Maderwald, Stefan T1 - Open design eight-channel transmit/receive coil for high-resolution and real-time ankle imaging at 7 T JF - Medical Physics N2 - Purpose: At 1.5 T, real-time MRI of joint movement has been shown to be feasible. However, 7 T, provides higher SNR and thus an improved potential for parallel imaging acceleration. The purpose of this work was to build an open, U-shaped eight-channel transmit/receive microstrip coil for 7 T MRI to enable high-resolution and real-time imaging of the moving ankle joint. Methods: A U-shaped eight-channel transmit/receive array for the human ankle was built.urn:x-wiley:00942405:mp3399:equation:mp3399-math-0001-parameters and urn:x-wiley:00942405:mp3399:equation:mp3399-math-0002-factor were measured. SAR calculations of different ankle postures were performed to ensure patient safety. Inhomogeneities in the transmit field consequent to the open design were compensated for by the use of static RF shimming. High-resolution and real-time imaging was performed in human volunteers. Results: The presented array showed good performance with regard to patient comfort and image quality. High acceleration factors of up to 4 are feasible without visible acceleration artifacts. Reasonable image homogeneity was achieved with RF shimming. Conclusions: Open, noncylindrical designs for transmit/receive coils are practical at 7 T and real-time imaging of the moving joint is feasible with the presented coil design. Y1 - 2011 U6 - http://dx.doi.org/10.1118/1.3553399 SN - 2473-4209 VL - 38 IS - 3 SP - 1162 EP - 1167 PB - Wiley CY - Hoboken ER -