TY - JOUR A1 - Achtsnicht, Stefan A1 - Tödter, Julia A1 - Niehues, Julia A1 - Telöken, Matthias A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim A1 - Schröper, Florian T1 - 3D printed modular immunofiltration columns for frequency mixing-based multiplex magnetic immunodetection JF - Sensors N2 - For performing point-of-care molecular diagnostics, magnetic immunoassays constitute a promising alternative to established enzyme-linked immunosorbent assays (ELISA) because they are fast, robust and sensitive. Simultaneous detection of multiple biomolecular targets from one body fluid sample is desired. The aim of this work is to show that multiplex magnetic immunodetection based on magnetic frequency mixing by means of modular immunofiltration columns prepared for different targets is feasible. By calculations of the magnetic response signal, the required spacing between the modules was determined. Immunofiltration columns were manufactured by 3D printing and antibody immobilization was performed in a batch approach. It was shown experimentally that two different target molecules in a sample solution could be individually detected in a single assaying step with magnetic measurements of the corresponding immobilization filters. The arrangement order of the filters and of a negative control did not influence the results. Thus, a simple and reliable approach to multi-target magnetic immunodetection was demonstrated. Y1 - 2019 U6 - http://dx.doi.org/10.3390/s19010148 SN - 1424-8220 VL - 19 IS - 1 PB - MDPI CY - Basel ER - TY - JOUR A1 - Achtsnicht, Stefan A1 - Neuendorf, Christian A1 - Faßbender, Tobias A1 - Nölke, Greta A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim A1 - Schröper, Florian T1 - Sensitive and rapid detection of cholera toxin subunit B using magnetic frequency mixing detection JF - Plos One N2 - Cholera is a life-threatening disease caused by the cholera toxin (CT) as produced by some Vibrio cholerae serogroups. In this research we present a method which directly detects the toxin’s B subunit (CTB) in drinking water. For this purpose we performed a magnetic sandwich immunoassay inside a 3D immunofiltration column. We used two different commercially available antibodies to capture CTB and for binding to superparamagnetic beads. ELISA experiments were performed to select the antibody combination. The beads act as labels for the magnetic frequency mixing detection technique. We show that the limit of detection depends on the type of magnetic beads. A nonlinear Hill curve was fitted to the calibration measurements by means of a custom-written python software. We achieved a sensitive and rapid detection of CTB within a broad concentration range from 0.2 ng/ml to more than 700 ng/ml. Y1 - 2019 U6 - http://dx.doi.org/10.1371/journal.pone.0219356 SN - 1932-6203 VL - 14 IS - 7 PB - Plos CY - San Francisco ER -