TY - JOUR A1 - Kirchner, Patrick A1 - Reisert, Steffen A1 - Pütz, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Characterisation of polymeric materials as passivation layer for calorimetric H2O2 gas sensors JF - Physica Status Solidi (a) N2 - Calorimetric gas sensors for monitoring the H₂O₂ concentration at elevated temperatures in industrial sterilisation processes have been presented in previous works. These sensors are built up in form of a differential set-up of a catalytically active and passive temperature-sensitive structure. Although, various types of catalytically active dispersions have been studied, the passivation layer has to be established and therefore, chemically as well as physically characterised. In the present work, fluorinated ethylene propylene (FEP), perfluoralkoxy (PFA) and epoxy-based SU-8 photoresist as temperature-stable polymeric materials have been investigated for sensor passivation in terms of their chemical inertness against H₂O₂, their hygroscopic properties as well as their morphology. The polymeric materials were deposited via spin-coating on the temperature-sensitive structure, wherein spin-coated FEP and PFA show slight agglomerates. However, they possess a low absorption of humidity due to their hydrophobic surface, whereas the SU-8 layer has a closed surface but shows a slightly higher absorption of water. All of them were inert against gaseous H₂O₂ during the characterisation in H₂O₂ atmosphere that demonstrates their suitability as passivation layer for calorimetric H₂O₂ gas sensors. Y1 - 2012 U6 - https://doi.org/10.1002/pssa.201100773 SN - 1862-6319 VL - 209 IS - 5 SP - 859 EP - 863 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Turek, Monika A1 - Schöning, Michael Josef A1 - Kloock, Joachim P. A1 - Schubert, Jürgen A1 - Zander, Willi A1 - Krüger, Peter A1 - Keusgen, Michael T1 - Herstellung und Charakterisierung eines hybriden Sensorarrays auf Halbleiterbasis für die Umweltanalytik JF - Sensoren und Messsysteme 2008 : 14. Fachtagung Ludwigsburg, 11. und 12. März 2008 / VDI/VDE-Gesellschaft Mess- und Automatisierungstechnik Y1 - 2008 SN - 978-3-18-092011-5 N1 - VDI-Berichte ; 2011 ; Sensoren und Messsysteme 2008, 14. GMA/ITG-Fachtagung, VDI/VDE- Gesellschaft Mess- und Automatisierungstechnik, + CD-ROM, Ludwigsburg, DE, 11.-12. Mar, 2008 SP - 803 EP - 810 PB - VDI-Verl. CY - Düsseldorf ER - TY - JOUR A1 - Oberländer, Jan A1 - Bromm, Alexander A1 - Wendeler, Luisa A1 - Iken, Heiko A1 - Palomar Duran, Marlena A1 - Greeff, Anton A1 - Kirchner, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Towards a biosensor to monitor the sterilisation efficiency of aseptic filling machines JF - Physica status solidi (a) N2 - Sterilisation processes are compulsory in medicine, pharmacy, and food industries to prevent infections of consumers and microbiological contaminations of products. Monitoring the sterilisation by conventional microbiological methods is time- and lab-consuming. To overcome this problem, in this work a novel biosensor has been proposed. The sensor enables a fast method to evaluate sterilisation processes. By means of thin-film technology the sensor's transducer structures in form of IDEs (interdigitated electrodes) have been fabricated on a silicon substrate. Physical characterisation of the developed sensor was done by AFM, SEM, and profilometry. Impedance analyses were conducted for the electrical characterisation. As microbiological layer spores of B. atrophaeus have been immobilised on the sensing structure; spores of this type are a well-known sterilisation test organism. Impedance measurements at a fixed frequency over time were performed to monitor the immobilisation process. A sterilisation process according to aseptic filling machines was applied to demonstrate the sensor functionality. After both, immobilisation and sterilisation, a change in impedance could successfully be detected. Y1 - 2015 U6 - https://doi.org/10.1002/pssa.201431900 SN - 1862-6319 VL - 212 IS - 6 SP - 1299 EP - 1305 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Pilas, Johanna A1 - Iken, Heiko A1 - Selmer, Thorsten A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Development of a multi‐parameter sensor chip for the simultaneous detection of organic compounds in biogas processes JF - Physica status solidi (a) N2 - An enzyme-based multi-parameter biosensor is developed for monitoring the concentration of formate, d-lactate, and l-lactate in biological samples. The sensor is based on the specific dehydrogenation by an oxidized β-nicotinamide adenine dinucleotide (NAD+)-dependent dehydrogenase (formate dehydrogenase, d-lactic dehydrogenase, and l-lactic dehydrogenase, respectively) in combination with a diaphorase from Clostridium kluyveri (EC 1.8.1.4). The enzymes are immobilized on a platinum working electrode by cross-linking with glutaraldehyde (GA). The principle of the determination scheme in case of l-lactate is as follows: l-lactic dehydrogenase (l-LDH) converts l-lactate into pyruvate by reaction with NAD+. In the presence of hexacyanoferrate(III), the resulting reduced β-nicotinamide adenine dinucleotide (NADH) is then regenerated enzymatically by diaphorase. The electrochemical detection is based on the current generated by oxidation of hexacyanoferrate(II) at an applied potential of +0.3 V vs. an Ag/AgCl reference electrode. The biosensor will be electrochemically characterized in terms of linear working range and sensitivity. Additionally, the successful practical application of the sensor is demonstrated in an extract from maize silage. Y1 - 2015 U6 - https://doi.org/10.1002/pssa.201431894 SN - 1862-6319 VL - 212 IS - 6 SP - 1306 EP - 1312 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Oberländer, Jan A1 - Jildeh, Zaid B. A1 - Kirchner, Patrick A1 - Wendeler, Luisa A1 - Bromm, Alexander A1 - Iken, Heiko A1 - Wagner, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Study of Interdigitated Electrode Arrays Using Experiments and Finite Element Models for the Evaluation of Sterilization Processes JF - Sensors N2 - In this work, a sensor to evaluate sterilization processes with hydrogen peroxide vapor has been characterized. Experimental, analytical and numerical methods were applied to evaluate and study the sensor behavior. The sensor set-up is based on planar interdigitated electrodes. The interdigitated electrode structure consists of 614 electrode fingers spanning over a total sensing area of 20 mm2. Sensor measurements were conducted with and without microbiological spores as well as after an industrial sterilization protocol. The measurements were verified using an analytical expression based on a first-order elliptical integral. A model based on the finite element method with periodic boundary conditions in two dimensions was developed and utilized to validate the experimental findings. Y1 - 2015 U6 - https://doi.org/10.3390/s151026115 SN - 1424-8220 N1 - This article belongs to the Special Issue "Gas Sensors—Designs and Applications" VL - 15 IS - 10 SP - 26115 EP - 26127 PB - MDPI CY - Basel ER - TY - JOUR A1 - Arreola, Julio A1 - Mätzkow, Malte A1 - Durán, Marlena Palomar A1 - Greeff, Anton A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Optimization of the immobilization of bacterial spores on glass substrates with organosilanes JF - Physica status solidi (A) : Applications and materials science N2 - Spores can be immobilized on biosensors to function as sensitive recognition elements. However, the immobilization can affect the sensitivity and reproducibility of the sensor signal. In this work, three different immobilization strategies with organosilanes were optimized and characterized to immobilize Bacillus atrophaeus spores on glass substrates. Five different silanization parameters were investigated: nature of the solvent, concentration of the silane, silanization time, curing process, and silanization temperature. The resulting silane layers were resistant to a buffer solution (e.g., Ringer solution) with a polysorbate (e.g., Tween®80) and sonication. KW - silanization KW - organosilanes KW - immobilization KW - endospores KW - biosensors KW - Bacillus atrophaeus Y1 - 2016 U6 - https://doi.org/10.1002/pssa.201532914 SN - 1862-6319 VL - 213 IS - 6 SP - 1463 EP - 1470 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Oberländer, Jan A1 - Kirchner, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Strategies in developing thin-film sensors for monitoring aseptic food processes : Theoretical considerations and investigations of passivation materials JF - Electrochimica Acta N2 - The sterilization of packages in aseptic food processes is highly significant to maintain a consumer-safe product with extended shelf-life. Today, the sterilization of food packages is predominantly accomplished by gaseous hydrogen peroxide (H2O2) in combination with heat. In order to monitor this sterilization process, calorimetric gas sensors as differential set-up of two platinum temperature sensors representing a catalytically active (additionally deposition of MnO2) and a passive segment have been recently developed. The temperature rise of the exothermic decomposition serves as an indicator of the present H2O2 concentration. In the present work, a theoretical approach considering the sensor’s thermochemistry and physical transport phenomena was formulated to evaluate the temperature rise based on the energy content of gaseous H2O2. In a further part of this work, three polymers have been analyzed with respect to their application as passivation materials. The examined polymers are photoresist SU-8, perfluoroalkoxy (PFA) and fluorinated ethylene propylene (FEP). Thermal analyses by means of differential scanning calorimetry (DSC) and thermogravimetric analysis (TGA) have been conducted to determine the operation limits of the polymers. The overall chemical resistance and stability of the polymers against the harsh environmental conditions during the sterilization process have been examined by attenuated total reflection Fourier transform infrared spectroscopy (ATR-FTIR). Y1 - 2015 U6 - https://doi.org/10.1016/j.electacta.2015.06.126 SN - 0013-4686 VL - 183 SP - 130 EP - 136 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Oberländer, Jan A1 - Jildeh, Zaid B. A1 - Kirchner, Patrick A1 - Wendeler, Luisa A1 - Bromm, Alexander A1 - Iken, Heiko A1 - Wagner, Patrick A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Experimental and numerical evaluation of interdigitated electrode array for monitoring gaseous sterilization processes T2 - 12. Dresdner Sensor-Symposium 2015 Y1 - 2015 U6 - https://doi.org/10.5162/12dss2015/P3.11 SP - 163 EP - 168 ER - TY - JOUR A1 - Jablonski, Melanie A1 - Münstermann, Felix A1 - Nork, Jasmina A1 - Molinnus, Denise A1 - Muschallik, Lukas A1 - Bongaerts, Johannes A1 - Wagner, Torsten A1 - Keusgen, Michael A1 - Siegert, Petra A1 - Schöning, Michael Josef T1 - Capacitive field‐effect biosensor applied for the detection of acetoin in alcoholic beverages and fermentation broths JF - physica status solidi (a) applications and materials science N2 - An acetoin biosensor based on a capacitive electrolyte–insulator–semiconductor (EIS) structure modified with the enzyme acetoin reductase, also known as butane-2,3-diol dehydrogenase (Bacillus clausii DSM 8716ᵀ), is applied for acetoin detection in beer, red wine, and fermentation broth samples for the first time. The EIS sensor consists of an Al/p-Si/SiO₂/Ta₂O₅ layer structure with immobilized acetoin reductase on top of the Ta₂O₅ transducer layer by means of crosslinking via glutaraldehyde. The unmodified and enzyme-modified sensors are electrochemically characterized by means of leakage current, capacitance–voltage, and constant capacitance methods, respectively. KW - acetoin KW - acetoin reductase KW - alcoholic beverages KW - biosensors KW - capacitive field-effect sensors Y1 - 2021 U6 - https://doi.org/10.1002/pssa.202000765 SN - 1862-6319 N1 - Corresponding author: Melanie Jablonski VL - 218 IS - 13 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Wagner, Torsten A1 - Vornholt, Wolfgang A1 - Werner, Frederik A1 - Yoshinobu, Tatsuo A1 - Miyamoto, Ko-Ichiro A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Light-addressable potentiometric sensor (LAPS) combined with magnetic beads for pharmaceutical screening JF - Physics in medicine N2 - The light-addressable potentiometric sensor (LAPS) has the unique feature to address different regions of a sensor surface without the need of complex structures. Measurements at different locations on the sensor surface can be performed in a common analyte solution, which distinctly simplifies the fluidic set-up. However, the measurement in a single analyte chamber prevents the application of different drugs or different concentrations of a drug to each measurement spot at the same time as in the case of multi-reservoir-based set-ups. In this work, the authors designed a LAPS-based set-up for cell culture screening that utilises magnetic beads loaded with the endotoxin (lipopolysaccharides, LPS), to generate a spatially distributed gradient of analyte concentration. Different external magnetic fields can be adjusted to move the magnetic beads loaded with a specific drug within the measurement cell. By recording the metabolic activities of a cell layer cultured on top of the LAPS surface, this work shows the possibility to apply different concentrations of a sample along the LAPS measurement spots within a common analyte solution. Y1 - 2016 U6 - https://doi.org/10.1016/j.phmed.2016.03.001 SN - 2352-4510 VL - 2016 IS - 1 SP - 2 EP - 7 ER - TY - JOUR A1 - Arreola, Julio A1 - Oberländer, Jan A1 - Mätzkow, M. A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Surface functionalization for spore-based biosensors with organosilanes JF - Electrochimica Acta N2 - In the present work, surface functionalization of different sensor materials was studied. Organosilanes are well known to serve as coupling agent for biomolecules or cells on inorganic materials. 3-aminopropyltriethoxysilane (APTES) was used to attach microbiological spores time to an interdigitated sensor surface. The functionality and physical properties of APTES were studied on isolated sensor materials, namely silicon dioxide (SiO2) and platinum (Pt) as well as the combined material on sensor level. A predominant immobilization of spores could be demonstrated on SiO2 surfaces. Additionally, the impedance signal of APTES-functionalized biosensor chips has been investigated. Y1 - 2017 U6 - https://doi.org/10.1016/j.electacta.2017.04.157 SN - 0013-4686 VL - 241 SP - 237 EP - 243 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Oberländer, Jan A1 - Mayer, Marlena A1 - Greeff, Anton A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Spore-based biosensor to monitor the microbicidal efficacy of gaseous hydrogen peroxide sterilization processes JF - Biosensors and Bioelectronics N2 - In this work, a spore-based biosensor is evaluated to monitor the microbicidal efficacy of sterilization processes applying gaseous hydrogen peroxide (H2O2). The sensor is based on interdigitated electrode structures (IDEs) that have been fabricated by means of thin-film technologies. Impedimetric measurements are applied to study the effect of sterilization process on spores of Bacillus atrophaeus. This resilient microorganism is commonly used in industry to proof the sterilization efficiency. The sensor measurements are accompanied by conventional microbiological challenge tests, as well as morphological characterizations with scanning electron microscopy (SEM) and transmission electron microscopy (TEM). The sensor measurements are correlated with the microbiological test routines. In both methods, namely the sensor-based and microbiological one, a tailing effect has been observed. The results are evaluated and discussed in a three-dimensional calibration plot demonstrating the sensor's suitability to enable a rapid process decision in terms of a successfully performed sterilization. Y1 - 2018 U6 - https://doi.org/10.1016/j.bios.2017.12.045 SN - 0956-5663 VL - 104 SP - 87 EP - 94 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Arreola, Julio A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Toward an immobilization method for spore-based biosensors in oxidative environment JF - Electrochimica Acta Y1 - 2019 U6 - https://doi.org/10.1016/j.electacta.2019.01.148 VL - 302 SP - 394 EP - 401 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Arreola, Julio A1 - Keusgen, Michael A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Combined calorimetric gas- and spore-based biosensor array for online monitoring and sterility assurance of gaseous hydrogen peroxide in aseptic filling machines JF - Biosensors and Bioelectronics Y1 - 2019 U6 - https://doi.org/10.1016/j.bios.2019.111628 SN - 0956-5663 VL - 143 IS - 111628 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Pilas, Johanna A1 - Selmer, Thorsten A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Screen-printed carbon electrodes modified with graphene oxide for the design of a reagent-free NAD+-dependent biosensor array JF - Analytical Chemistry Y1 - 2019 U6 - https://doi.org/10.1021/acs.analchem.9b04481 VL - 91 IS - 23 SP - 15293 EP - 15299 PB - ACS Publications CY - Washington ER - TY - JOUR A1 - Arreola, Julio A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Effect of O2 plasma on properties of electrolyte-insulator-semiconductor structures JF - physica status solidi a : applications and materials sciences N2 - Prior to immobilization of biomolecules or cells onto biosensor surfaces, the surface must be physically or chemically activated for further functionalization. Organosilanes are a versatile option as they facilitate the immobilization through their terminal groups and also display self-assembly. Incorporating hydroxyl groups is one of the important methods for primary immobilization. This can be done, for example, with oxygen plasma treatment. However, this treatment can affect the performance of the biosensors and this effect is not quite well understood for surface functionalization. In this work, the effect of O2 plasma treatment on EIS sensors was investigated by means of electrochemical characterizations: capacitance–voltage (C–V) and constant capacitance (ConCap) measurements. After O2 plasma treatment, the potential of the EIS sensor dramatically shifts to a more negative value. This was successfully reset by using an annealing process. KW - surface functionalization KW - O2 plasma KW - hydroxylation KW - electrolyte-insulator semiconductor sensor (EIS) KW - annealing Y1 - 2017 U6 - https://doi.org/10.1002/pssa.201700025 SN - 1862-6319 VL - 214 PB - Wiley-VCH CY - Weinheim ER - TY - CHAP A1 - Oberländer, Jan A1 - Arreola, Julio A1 - Hansen, Christina A1 - Greeff, Anton A1 - Mayer, Marlena A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Impedimetric Biosensor to Enable Fast Evaluation of Gaseous Sterilization Processes T2 - MDPI Proceedings Y1 - 2017 U6 - https://doi.org/10.3390/proceedings1040435 N1 - Eurosensors 2017 Conference, Paris, France, 3–6 September 2017 VL - 1 IS - 4 ER - TY - JOUR A1 - Pilas, Johanna A1 - Yazici, Yasemen A1 - Selmer, Thorsten A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Optimization of an amperometric biosensor array for simultaneous measurement of ethanol, formate, d- and l-lactate JF - Electrochimica Acta N2 - The immobilization of NAD+-dependent dehydrogenases, in combination with a diaphorase, enables the facile development of multiparametric sensing devices. In this work, an amperometric biosensor array for simultaneous determination of ethanol, formate, d- and l-lactate is presented. Enzyme immobilization on platinum thin-film electrodes was realized by chemical cross-linking with glutaraldehyde. The optimization of the sensor performance was investigated with regard to enzyme loading, glutaraldehyde concentration, pH, cofactor concentration and temperature. Under optimal working conditions (potassium phosphate buffer with pH 7.5, 2.5 mmol L-1 NAD+, 2.0 mmol L-1 ferricyanide, 25 °C and 0.4% glutaraldehyde) the linear working range and sensitivity of the four sensor elements was improved. Simultaneous and cross-talk free measurements of four different metabolic parameters were performed successfully. The reliable analytical performance of the biosensor array was demonstrated by application in a clarified sample of inoculum sludge. Thereby, a promising approach for on-site monitoring of fermentation processes is provided. KW - Simultaneous determination KW - Enzymatic biosensor KW - Diaphorase KW - Dehydrogenase Y1 - 2017 U6 - https://doi.org/10.1016/j.electacta.2017.07.119 SN - 0013-4686 VL - 251 SP - 256 EP - 262 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Röhlen, Desiree A1 - Pilas, Johanna A1 - Dahmen, Markus A1 - Keusgen, Michael A1 - Selmer, Thorsten A1 - Schöning, Michael Josef T1 - Toward a Hybrid Biosensor System for Analysis of Organic and Volatile Fatty Acids in Fermentation Processes JF - Frontiers in Chemistry N2 - Monitoring of organic acids (OA) and volatile fatty acids (VFA) is crucial for the control of anaerobic digestion. In case of unstable process conditions, an accumulation of these intermediates occurs. In the present work, two different enzyme-based biosensor arrays are combined and presented for facile electrochemical determination of several process-relevant analytes. Each biosensor utilizes a platinum sensor chip (14 × 14 mm²) with five individual working electrodes. The OA biosensor enables simultaneous measurement of ethanol, formate, d- and l-lactate, based on a bi-enzymatic detection principle. The second VFA biosensor provides an amperometric platform for quantification of acetate and propionate, mediated by oxidation of hydrogen peroxide. The cross-sensitivity of both biosensors toward potential interferents, typically present in fermentation samples, was investigated. The potential for practical application in complex media was successfully demonstrated in spiked sludge samples collected from three different biogas plants. Thereby, the results obtained by both of the biosensors were in good agreement to the applied reference measurements by photometry and gas chromatography, respectively. The proposed hybrid biosensor system was also used for long-term monitoring of a lab-scale biogas reactor (0.01 m³) for a period of 2 months. In combination with typically monitored parameters, such as gas quality, pH and FOS/TAC (volatile organic acids/total anorganic carbonate), the amperometric measurements of OA and VFA concentration could enhance the understanding of ongoing fermentation processes. Y1 - 2018 U6 - https://doi.org/10.3389/fchem.2018.00284 IS - 6 PB - Frontiers CY - Lausanne ER - TY - JOUR A1 - Molinnus, Denise A1 - Muschallik, Lukas A1 - Gonzalez, Laura Osorio A1 - Bongaerts, Johannes A1 - Wagner, Torsten A1 - Selmer, Thorsten A1 - Siegert, Petra A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Development and characterization of a field-effect biosensor for the detection of acetoin JF - Biosensors and Bioelectronics N2 - A capacitive electrolyte-insulator-semiconductor (EIS) field-effect biosensor for acetoin detection has been presented for the first time. The EIS sensor consists of a layer structure of Al/p-Si/SiO₂/Ta₂O₅/enzyme acetoin reductase. The enzyme, also referred to as butane-2,3-diol dehydrogenase from B. clausii DSM 8716T, has been recently characterized. The enzyme catalyzes the (R)-specific reduction of racemic acetoin to (R,R)- and meso-butane-2,3-diol, respectively. Two different enzyme immobilization strategies (cross-linking by using glutaraldehyde and adsorption) have been studied. Typical biosensor parameters such as optimal pH working range, sensitivity, hysteresis, linear concentration range and long-term stability have been examined by means of constant-capacitance (ConCap) mode measurements. Furthermore, preliminary experiments have been successfully carried out for the detection of acetoin in diluted white wine samples. Y1 - 2018 U6 - https://doi.org/10.1016/j.bios.2018.05.023 VL - 115 SP - 1 EP - 6 PB - Elsevier CY - Amsterdam ER -