TY - JOUR A1 - Gräßl, Andreas A1 - Renz, Wolfgang A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Winter, Lukas A1 - Özerdem, Celal A1 - Rieger, Jan A1 - Kellmann, Peter A1 - Santoro, Davide A1 - Lindel, Tomasz Dawid A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Niendorf, Thoralf T1 - Modular 32-channel transceiver coil array for cardiac MRI at 7.0T JF - Magnetic Resonance in Medicine N2 - Purpose To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. Methods The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Results Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. Conclusion The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T. Y1 - 2013 U6 - https://doi.org/10.1002/mrm.24903 SN - 1522-2594 VL - 72 IS - 1 SP - 276 EP - 290 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Orzada, Stephan A1 - Fiedler, Thomas M. A1 - Bitz, Andreas A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - Local SAR compression with overestimation control to reduce maximum relative SAR overestimation and improve multi-channel RF array performance JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective In local SAR compression algorithms, the overestimation is generally not linearly dependent on actual local SAR. This can lead to large relative overestimation at low actual SAR values, unnecessarily constraining transmit array performance. Method Two strategies are proposed to reduce maximum relative overestimation for a given number of VOPs. The first strategy uses an overestimation matrix that roughly approximates actual local SAR; the second strategy uses a small set of pre-calculated VOPs as the overestimation term for the compression. Result Comparison with a previous method shows that for a given maximum relative overestimation the number of VOPs can be reduced by around 20% at the cost of a higher absolute overestimation at high actual local SAR values. Conclusion The proposed strategies outperform a previously published strategy and can improve the SAR compression where maximum relative overestimation constrains the performance of parallel transmission. Y1 - 2020 SN - 1352-8661 U6 - https://doi.org/10.1007/s10334-020-00890-0 IS - 34 (2021) SP - 153 EP - 164 PB - Springer CY - Heidelberg ER - TY - JOUR A1 - Scholl, Ingrid A1 - Palm, Christoph A1 - Lehmann, Thomas Martin A1 - Spitzer, Klaus T1 - Nutzung eines Farbkonstanz-Algorithmus zur Entfernung von Glanzlichtern in laryngoskopischen Bildern. Palm, Christoph; Scholl, Ingrid, Lehmann, Thomas; Spitzer, Klaus. JF - Methoden der Medizinischen Informatik, Biometrie und Epidemiologie in der modernen Informationsgesellschaft Y1 - 1998 SN - 3-8208-1335-7 N1 - Jahrestagung der GMDS ; 43 SP - 300 EP - 303 PB - MVV, Medien und Medizin CY - München ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - SAR Simulations & Safety JF - NeuroImage Y1 - 2017 U6 - https://doi.org/10.1016/j.neuroimage.2017.03.035 SN - 1053-8119 IS - Epub ahead of print PB - Elsevier CY - Amsterdam ER - TY - GEN A1 - Frauenrath, Tobias A1 - Fuchs, Katharina A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Rieger, Jan A1 - Niendorf, Thoralf T1 - Improved cardiac triggering by combining multiple physiological signals: a cardiac MR feasibility study at 7.0 T T2 - 2012 ISMRM Annual Meeting Proceedings N2 - In current clinical cardiovascular MR (CMR) practice cardiac motion is commonly dealt with using ECG based synchronization. However, ECG is corrupted by magneto-hydrodynamic (MHD) effects in magnetic fields. This leads to artifacts in the ECG trace and evokes severe T-wave elevations, which might be misinterpreted as R-waves resulting in erroneous triggering. At (ultra)high field strengths, the propensity of ECG recordings to MHD effects is further pronounced. Pulse oximetry (POX) being inherently sensitive to blood oxygenation provides an alternative approach for cardiac gating. However, due to the travel time of the blood the peak of maximum oxygenation and hence the trigger is delayed by approx. 300 ms with respect to the ECG's R-wave. Also the peak of maximum oxygenation shows a jitter of up to 65 ms. Alternative triggering approaches include acoustic cardiac triggering (ACT). In current clinical practice cardiac gating / triggering commonly relies on using single physiological signals only. Realizing this limitation this study proposes a combined triggering approach which exploits multiple physiological signals including ECG, POX or ACT to track cardiac activity. The feasibility of the coupled approach is examined for LV function assessment at 7.0 T. For this purpose, breath-held 2D-CINE imaging in conjunction with cardiac synchronization was performed paralleled by real time logging of physiological waveforms to track (mis)synchronization between the cardiac cycle and data acquisition. Combinations of the ECG, POX and ACT signals were evaluated and processed in real time to facilitate reliable trigger information. Y1 - 2012 SN - 1545-4428 N1 - ISMRM 20th Annual Meeting & Exhibition, 5-11 May 2012, Melbourne, Australia ER - TY - GEN A1 - Gräßl, Andreas A1 - Renz, Wolfgang A1 - Hezel, Fabian A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Hoffmann, Werner A1 - Kellmann, Peter A1 - Martin, Conrad A1 - Niendorf, Thoralf T1 - Design, evaluation and application of a modular 32 channel transmit/receive surface coil array for cardiac MRI at 7T T2 - 2012 ISMRM Annual Meeting Proceedings N2 - Cardiac MR (CMR) at ultrahigh (≥7.0 T) fields is regarded as one of the most challenging MRI applications. At 7.0 T image quality is not always exclusively defined by signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Detrimental effects bear the potential to spoil the signal-to-noise (SNR) and contrast-to-noise (CNR) benefits of cardiac MR (CMR) at 7.0 T. B₁⁺-inhomogeneities and signal voids represent the main challenges. Various pioneering coil concepts have been proposed to tackle these issues, enabling cardiac MRI at 7.0 T. This includes a trend towards an ever larger number of transmit and receive channels. This approach affords multi-dimensional B₁⁺ modulations to improve B₁⁺ shimming performance and to enhance RF efficiency. Also, parallel imaging benefits from a high number of receive channels enabling two-dimensional acceleration. Realizing the limitations of existing coil designs tailored for UHF CMR and recognizing the opportunities of a many element TX/RX channel architecture this work proposes a modular, two dimensional 32-channel transmit and receive array using loop elements and examines its efficacy for enhanced B¹+ homogeneity and improved parallel imaging performance. Y1 - 2012 SN - 1545-4428 N1 - ISMRM 20th Annual Meeting & Exhibition, 5-11 May 2012, Melbourne, Australia ER - TY - GEN A1 - Tkachenko, Valeriy A1 - von Knobelsdorff-Brenkenhoff, Florian A1 - Kleindienst, Denise A1 - Winter, Lukas A1 - Rieger, Jan A1 - Frauenrath, Tobias A1 - Dieringer, Matthias A. A1 - Santoro, Davide A1 - Niendorf, Thoralf A1 - Schulz-Menger, Jeanette T1 - Cardiovasular MR at 7Tesla: assessment of the right ventricle T2 - 2012 ISMRM Annual Meeting Proceedings N2 - The assessment of the right ventricle (RV) is a challenge in today's cardiology, but of growing clinical impact regarding patient prognosis in different cardiac diseases. The detection and differentiation of small wall motion abnormalities may help to enhance the differentiation of cardiomyopathies including Arrhythmogenic Rightventricular Cardiomyopathy. Cardiovascular magnetic resonance (CMR) at 1.5T is the accepted gold standard for RV quantification. The higher spatial resolution achievable at ultrahigh field strength (UHF) offers the potential to gain new insights into the structure and function of the RV. To approach this goal accurate RV chamber quantification at 7T has to be proven. Consequently this study examines the feasibility of assessment of RV dimensions and function at 7T using improved spatial resolution enabled by the intrinsic sensitivity gain of UHF CMR. For this purpose, a dedicated 16 channel TX/RX RF coil array is used together with 2D CINE fast gradient echo (FGRE) imaging. For comparison RV chamber quantification is conducted at 1.5T using a SSFP based state of the art clinical protocol. Y1 - 2012 SN - 1545-4428 N1 - ISMRM 20th Annual Meeting & Exhibition, 5-11 May 2012, Melbourne, Australia ER - TY - GEN A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Winter, Lukas A1 - Gräßl, Andreas A1 - Santoro, Davide A1 - Özerdem, Celal A1 - Renz, Wolfgang A1 - Greiser, Andreas A1 - Niendorf, Thoralf T1 - Lessons learned from cardiac MRI at 7.0 T: LV function assessment at 3.0 T using local multi-channel transceiver coil arrays T2 - 2012 ISMRM Annual Meeting Proceedings N2 - Cardiac MR (CMR) is of proven clinical value but also an area of vigorous ongoing research since image quality is not always exclusively defined by signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Recent developments of CMR at 7.0 T have been driven by pioneering explorations into novel multichannel transmit and receive coil array technology to tackle the challenges B1+-field inhomogeneities, to offset specific-absorption rate (SAR) constraints and to reduce banding artifacts in SSFP imaging. For this study, recognition of the benefits and performance of local surface Tx/Rx-array structures recently established at 7.0 T inspired migration to 3.0 T, where RF inhomogeneities and SAR limitations encountered in routine clinical CMR, though somewhat reduced versus the 7.0 T situation, remain significant. For all these reasons, this study was designed to build and examine the feasibility of a local four channel Tx/Rx cardiac coil array for anatomical and functional cardiac imaging at 3.0 T. For comparison, a homebuilt 4 channel Rx cardiac coil array exhibiting the same geometry as the Tx/Rx coil and a Rx surface coil array were used. Y1 - 2012 SN - 1545-4428 N1 - ISMRM 20th Annual Meeting & Exhibition, 5-11 May 2012, Melbourne, Australia ER - TY - GEN A1 - Lindel, Tomasz Dawid A1 - Greiser, Andreas A1 - Waxman, Patrick A1 - Dietterle, Martin A1 - Seifert, Frank A1 - Fontius, Ulrich A1 - Renz, Wolfgang A1 - Dieringer, Matthias A. A1 - Frauenrath, Tobias A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf A1 - Ittermann, Bernd T1 - Cardiac CINE MRI at 7 T using a transmit array T2 - 2012 ISMRM Annual Meeting Proceedings N2 - With its need for high SNR and short acquisition times, Cardiac MRI (CMR) is an intriguing target application for ultrahigh field MRI. Due to the sheer size of the upper torso, however, the known RF issues of 7T MRI are also most prominent in CMR. Recent years brought substantial progress but the full potential of the ultrahigh field for CMR is yet to be exploited. Parallel transmission (pTx) is a promising approach in this context and several groups have already reported B1 shimming for 7T CMR. In such a static pTx application amplitudes and phases of all Tx channels are adjusted individually but otherwise imaging techniques established in current clinical practice 1.5 T and 3 T are applied. More advanced forms of pTx as spatially selective excitation (SSE) using Transmit SENSE promise additional benefits like faster imaging with reduced fields of view or improved SAR control. SSE requires the full dynamic capabilities of pTx, however, and for the majority of today's implemented pTx hardware the internal synchronization of the Tx array does not easily permit external triggering as needed for CMR. Here we report a software solution to this problem and demonstrate the feasibility of CINE CMR at 7 T using a Tx array. Y1 - 2012 SN - 1545-4428 N1 - ISMRM 20th Annual Meeting & Exhibition, 5-11 May 2012, Melbourne, Australia ER - TY - JOUR A1 - Schiffer, Stefan A1 - Ferrein, Alexander A1 - Lakemeyer, Gerhard T1 - Caesar: an intelligent domestic service robot JF - Intelligent service robotics N2 - In this paper we present CAESAR, an intelligent domestic service robot. In domestic settings for service robots complex tasks have to be accomplished. Those tasks benefit from deliberation, from robust action execution and from flexible methods for human–robot interaction that account for qualitative notions used in natural language as well as human fallibility. Our robot CAESAR deploys AI techniques on several levels of its system architecture. On the low-level side, system modules for localization or navigation make, for instance, use of path-planning methods, heuristic search, and Bayesian filters. For face recognition and human–machine interaction, random trees and well-known methods from natural language processing are deployed. For deliberation, we use the robot programming and plan language READYLOG, which was developed for the high-level control of agents and robots; it allows combining programming the behaviour using planning to find a course of action. READYLOG is a variant of the robot programming language Golog. We extended READYLOG to be able to cope with qualitative notions of space frequently used by humans, such as “near” and “far”. This facilitates human–robot interaction by bridging the gap between human natural language and the numerical values needed by the robot. Further, we use READYLOG to increase the flexible interpretation of human commands with decision-theoretic planning. We give an overview of the different methods deployed in CAESAR and show the applicability of a system equipped with these AI techniques in domestic service robotics Y1 - 2012 U6 - https://doi.org/10.1007/s11370-012-0118-y SN - 1861-2776 N1 - Special Issue on Artificial Intelligence Techniques for Robotics: Sensing, Representation and Action, Part I VL - 5 IS - 4 SP - 259 EP - 276 PB - Springer CY - Berlin ER -