TY - JOUR A1 - Frotscher, Ralf A1 - Staat, Manfred T1 - Stresses produced by different textile mesh implants in a tissue equivalent JF - BioNanoMaterials N2 - Two single-incision mini-slings used for treating urinary incontinence in women are compared with respect to the stresses they produce in their surrounding tissue. In an earlier paper we experimentally observed that these implants produce considerably different stress distributions in a muscle tissue equivalent. Here we perform 2D finite element analyses to compare the shear stresses and normal stresses in the tissue equivalent for the two meshes and to investigate their failure behavior. The results clearly show that the Gynecare TVT fails for increasing loads in a zipper-like manner because it gradually debonds from the surrounding tissue. Contrary to that, the tissue at the ends of the DynaMesh-SIS direct may rupture but only at higher loads. The simulation results are in good agreement with the experimental observations thus the computational model helps to interpret the experimental results and provides a tool for qualitative evaluation of mesh implants. Y1 - 2014 U6 - http://dx.doi.org/10.1515/bnm-2014-0003 SN - 2191-4672 (E-Journal); 2193-066X (E-Journal); 0011-8656 (Print); 1616-0177 (Print); 2193-0651 (Print) VL - 15 IS - 1-2 SP - 25 EP - 30 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Bhattarai, Aroj A1 - Jabbari, Medisa A1 - Anding, Ralf A1 - Staat, Manfred T1 - Surgical treatment of vaginal vault prolapse using different prosthetic mesh implants: a finite element analysis JF - tm - Technisches Messen N2 - Particularly multiparous elderly women may suffer from vaginal vault prolapse after hysterectomy due to weak support from lax apical ligaments. A decreased amount of estrogen and progesterone in older age is assumed to remodel the collagen thereby reducing tissue stiffness. Sacrocolpopexy is either performed as open or laparoscopic surgery using prosthetic mesh implants to substitute lax ligaments. Y-shaped mesh models (DynaMesh, Gynemesh, and Ultrapro) are implanted in a 3D female pelvic floor finite element model in the extraperitoneal space from the vaginal cuff to the first sacral (S1) bone below promontory. Numerical simulations are conducted during Valsalva maneuver with weakened tissues modeled by reduced tissue stiffness. Tissues are modeled as incompressible, isotropic hyperelastic materials whereas the meshes are modeled either as orthotropic linear elastic or as isotropic hyperlastic materials. The positions of the vaginal cuff and the bladder base are calculated from the pubococcygeal line for female pelvic floor at rest, for prolapse and after repair using the three meshes. Due to mesh mechanics and mesh pore deformation along the loaded direction, the DynaMesh with regular rectangular mesh pores is found to provide better mechanical support to the organs than the Gynemesh and the Ultrapro with irregular hexagonal mesh pores. Insbesondere ältere, mehrgebährende Frauen leiden häufiger an einem Scheidenvorfall nach einer Hysterektomie aufgrund der schwachen Unterstützung durch laxe apikale Bänder. Es wird angenommen, dass eine verringerte Menge an Östrogen und Progesteron im höheren Alter das Kollagen umformt, wodurch die Gewebesteifigkeit reduziert wird. Die Sakrokolpopexie ist eine offene oder laparoskopische Operation, die mit prothetischen Netzimplantaten durchgeführt wird, um laxe Bänder zu ersetzen. Y-förmige Netzmodelle (DynaMesh, Gynemesh und Ultrapro) werden in einem 3D-Modell des weiblichen Beckenbodens im extraperitonealen Raum vom Vaginalstumpf bis zum Promontorium implantiert. Numerische Simulationen werden während des Valsalva-Manövers mit geschwächtem Gewebe durchgeführt, das durch eine reduzierte Gewebesteifigkeit modelliert wird. Die Gewebe werden als inkompressible, isotrop hyperelastische Materialien modelliert, während die Netze entweder als orthotrope linear elastische oder als isotrope hyperlastische Materialien modelliert werden. Die Positionen des Vaginalstumpfs, der Blase und der Harnröhrenachse werden anhand der Pubococcygeallinie aus der Ruhelage, für den Prolaps und nach der Reparatur unter Verwendung der drei Netze berechnet. Aufgrund der Netzmechanik und der Netzporenverformung bietet das DynaMesh mit regelmäßigen rechteckigen Netzporen eine bessere mechanische Unterstützung und eine Neupositionierung des Scheidengewölbes, der Blase und der Urethraachse als Gynemesh und Ultrapro mit unregelmäßigen hexagonalen Netzporen. Y1 - 2018 U6 - http://dx.doi.org/10.1515/teme-2017-0115 SN - 2196-7113 VL - 85 IS - 5 SP - 331 EP - 342 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Birgel, Stefan A1 - Leschinger, Tim A1 - Wegmann, Kilian A1 - Staat, Manfred T1 - Calculation of muscle forces and joint reaction loads in the shoulder area via an OpenSim based computer model JF - tm - Technisches Messen N2 - Using the OpenSim software and verified anatomical data, a computer model for the calculation of biomechanical parameters is developed and used to determine the effect of a reattachment of the Supraspinatus muscle with a medial displacement of the muscle attachment point, which may be necessary for a rupture of the supraspinatus tendon. The results include the influence of the operation on basic biomechanical parameters such as the lever arm, as well as the calculated the muscle activations for the supraspinatus and deltoid. In addition, the influence on joint stability is examined by an analysis of the joint reaction force. The study provides a detailed description of the used model, as well as medical findings to a reattachment of the supraspinatus. Mit der Software OpenSim und überprüften anatomischen Daten wird ein Computermodell zur Berechnung von biomechanischen Parametern entwickelt und genutzt, um den Effekt einer Refixierung des Supraspinatusmuskels mit einer medialen Verschiebung des Muskelansatzpunktes zu ermitteln, wie sie unter anderem nach einem Riss der Supraspinatussehne notwendig sein kann. Die Ergebnisse umfassen hierbei den Einfluss der Operation auf grundlegende biomechanische Parameter wie den Hebelarm sowie die berechneten Muskelaktivierungen für den Supraspinatus und Deltoideus. Zusätzlich wird der Einfluss auf die Gelenkstabilität betrachtet und durch eine Analyse der Gelenkreaktionskraft untersucht. Die Studie bietet eine detaillierte Beschreibung des genutzten Modells, sowie medizinische Erkenntnisse zu einer Refixierung des Supraspinatus. Y1 - 2018 U6 - http://dx.doi.org/10.1515/teme-2017-0114 SN - 2196-7113 VL - 85 IS - 5 SP - 321 EP - 330 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Horbach, Andreas A1 - Staat, Manfred T1 - Optical strain measurement for the modeling of surgical meshes and their porosity JF - Current Directions in Biomedical Engineering N2 - The porosity of surgical meshes makes them flexible for large elastic deformation and establishes the healing conditions of good tissue in growth. The biomechanic modeling of orthotropic and compressible materials requires new materials models and simulstaneoaus fit of deformation in the load direction as well as trannsversely to to load. This nonlinear modeling can be achieved by an optical deformation measurement. At the same time the full field deformation measurement allows the dermination of the change of porosity with deformation. Also the socalled effective porosity, which has been defined to asses the tisssue interatcion with the mesh implants, can be determined from the global deformation of the surgical meshes. Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0045 SN - 2364-5504 VL - Band 4 IS - 1 SP - 181 EP - 184 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Bhattarai, Aroj A1 - Staat, Manfred T1 - Computational comparison of different textile implants to correct apical prolapse in females JF - Current Directions in Biomedical Engineering N2 - Prosthetic textile implants of different shapes, sizes and polymers are used to correct the apical prolapse after hysterectomy (removal of the uterus). The selection of the implant before or during minimally invasive surgery depends on the patient’s anatomical defect, intended function after reconstruction and most importantly the surgeon’s preference. Weakness or damage of the supporting tissues during childbirth, menopause or previous pelvic surgeries may put females in higher risk of prolapse. Numerical simulations of reconstructed pelvic floor with weakened tissues and organ supported by textile product models: DynaMesh®-PRS soft, DynaMesh®-PRP soft and DynaMesh®-CESA from FEG Textiletechnik mbH, Germany are compared. Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0159 VL - 4 IS - 1 SP - 661 EP - 664 PB - De Gruyter CY - Berlin ER -