TY - JOUR A1 - Bhattarai, Aroj A1 - Jabbari, Medisa A1 - Anding, Ralf A1 - Staat, Manfred T1 - Surgical treatment of vaginal vault prolapse using different prosthetic mesh implants: a finite element analysis JF - tm - Technisches Messen N2 - Particularly multiparous elderly women may suffer from vaginal vault prolapse after hysterectomy due to weak support from lax apical ligaments. A decreased amount of estrogen and progesterone in older age is assumed to remodel the collagen thereby reducing tissue stiffness. Sacrocolpopexy is either performed as open or laparoscopic surgery using prosthetic mesh implants to substitute lax ligaments. Y-shaped mesh models (DynaMesh, Gynemesh, and Ultrapro) are implanted in a 3D female pelvic floor finite element model in the extraperitoneal space from the vaginal cuff to the first sacral (S1) bone below promontory. Numerical simulations are conducted during Valsalva maneuver with weakened tissues modeled by reduced tissue stiffness. Tissues are modeled as incompressible, isotropic hyperelastic materials whereas the meshes are modeled either as orthotropic linear elastic or as isotropic hyperlastic materials. The positions of the vaginal cuff and the bladder base are calculated from the pubococcygeal line for female pelvic floor at rest, for prolapse and after repair using the three meshes. Due to mesh mechanics and mesh pore deformation along the loaded direction, the DynaMesh with regular rectangular mesh pores is found to provide better mechanical support to the organs than the Gynemesh and the Ultrapro with irregular hexagonal mesh pores. Insbesondere ältere, mehrgebährende Frauen leiden häufiger an einem Scheidenvorfall nach einer Hysterektomie aufgrund der schwachen Unterstützung durch laxe apikale Bänder. Es wird angenommen, dass eine verringerte Menge an Östrogen und Progesteron im höheren Alter das Kollagen umformt, wodurch die Gewebesteifigkeit reduziert wird. Die Sakrokolpopexie ist eine offene oder laparoskopische Operation, die mit prothetischen Netzimplantaten durchgeführt wird, um laxe Bänder zu ersetzen. Y-förmige Netzmodelle (DynaMesh, Gynemesh und Ultrapro) werden in einem 3D-Modell des weiblichen Beckenbodens im extraperitonealen Raum vom Vaginalstumpf bis zum Promontorium implantiert. Numerische Simulationen werden während des Valsalva-Manövers mit geschwächtem Gewebe durchgeführt, das durch eine reduzierte Gewebesteifigkeit modelliert wird. Die Gewebe werden als inkompressible, isotrop hyperelastische Materialien modelliert, während die Netze entweder als orthotrope linear elastische oder als isotrope hyperlastische Materialien modelliert werden. Die Positionen des Vaginalstumpfs, der Blase und der Harnröhrenachse werden anhand der Pubococcygeallinie aus der Ruhelage, für den Prolaps und nach der Reparatur unter Verwendung der drei Netze berechnet. Aufgrund der Netzmechanik und der Netzporenverformung bietet das DynaMesh mit regelmäßigen rechteckigen Netzporen eine bessere mechanische Unterstützung und eine Neupositionierung des Scheidengewölbes, der Blase und der Urethraachse als Gynemesh und Ultrapro mit unregelmäßigen hexagonalen Netzporen. Y1 - 2018 U6 - http://dx.doi.org/10.1515/teme-2017-0115 SN - 2196-7113 VL - 85 IS - 5 SP - 331 EP - 342 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Horbach, Andreas A1 - Staat, Manfred T1 - Optical strain measurement for the modeling of surgical meshes and their porosity JF - Current Directions in Biomedical Engineering N2 - The porosity of surgical meshes makes them flexible for large elastic deformation and establishes the healing conditions of good tissue in growth. The biomechanic modeling of orthotropic and compressible materials requires new materials models and simulstaneoaus fit of deformation in the load direction as well as trannsversely to to load. This nonlinear modeling can be achieved by an optical deformation measurement. At the same time the full field deformation measurement allows the dermination of the change of porosity with deformation. Also the socalled effective porosity, which has been defined to asses the tisssue interatcion with the mesh implants, can be determined from the global deformation of the surgical meshes. Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0045 SN - 2364-5504 VL - Band 4 IS - 1 SP - 181 EP - 184 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Laura, C.O. A1 - Drechsler, Klaus A1 - Erdt, M. A1 - Wesarg, S. A1 - Bale, R. T1 - Intervention assessment tool for primary tumors in the liver JF - Current Directions in Biomedical Engineering N2 - After a liver tumor intervention the medical doctor has to compare both pre and postoperative CT acquisitions to ensure that all carcinogenic cells are destroyed. A correct assessment of the intervention is of vital importance, since it will reduce the probability of tumor recurrence. Some methods have been proposed to support the medical doctors during the assessment process, however, all of them focus on secondary tumors. In this paper a tool is presented that enables the outcome validation for both primary and secondary tumors. Therefore, a multiphase registration (preoperative arterial and portal phases) followed by a registration between the pre and postoperative CT images is carried out. The first registration is in charge of the primary tumors that are only visible in the arterial phase. The secondary tumors will be incorporated in the second registration step. Finally, the part of the tumor that was not covered by the necrosis is quantified and visualized. The method has been tested in 9 patients, with an average registration error of 1.41 mm. Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0081 SN - 2364-5504 VL - 4 IS - 1 SP - 337 EP - 340 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Lapitan, Denis G. A1 - Rogatkin, Dmitrii A. A1 - Persheyev, Sydulla K. A1 - Kotliar, Konstantin T1 - False spectra formation in the differential two-channel scheme of the laser Doppler flowmeter JF - Biomedizinische Technik N2 - Noise in the differential two-channel scheme of a classic laser Doppler flowmetry (LDF) instrument was studied. Formation of false spectral components in the output signal due to beating of electrical signals in the differential amplifier was found out. The improved block-diagram of the flowmeter was developed allowing to reduce the noise. Y1 - 2018 U6 - http://dx.doi.org/10.1515/bmt-2017-0060 SN - 0013-5585 VL - 63 IS - 4 SP - 439 EP - 444 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Bhattarai, Aroj A1 - Staat, Manfred T1 - Computational comparison of different textile implants to correct apical prolapse in females JF - Current Directions in Biomedical Engineering N2 - Prosthetic textile implants of different shapes, sizes and polymers are used to correct the apical prolapse after hysterectomy (removal of the uterus). The selection of the implant before or during minimally invasive surgery depends on the patient’s anatomical defect, intended function after reconstruction and most importantly the surgeon’s preference. Weakness or damage of the supporting tissues during childbirth, menopause or previous pelvic surgeries may put females in higher risk of prolapse. Numerical simulations of reconstructed pelvic floor with weakened tissues and organ supported by textile product models: DynaMesh®-PRS soft, DynaMesh®-PRP soft and DynaMesh®-CESA from FEG Textiletechnik mbH, Germany are compared. Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0159 VL - 4 IS - 1 SP - 661 EP - 664 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Ketelhut, Maike A1 - Göll, Fabian A1 - Braunstein, Björn A1 - Albracht, Kirsten A1 - Abel, Dirk T1 - Comparison of different training algorithms for the leg extension training with an industrial robot JF - Current Directions in Biomedical Engineering N2 - In the past, different training scenarios have been developed and implemented on robotic research platforms, but no systematic analysis and comparison have been done so far. This paper deals with the comparison of an isokinematic (motion with constant velocity) and an isotonic (motion against constant weight) training algorithm. Both algorithms are designed for a robotic research platform consisting of a 3D force plate and a high payload industrial robot, which allows leg extension training with arbitrary six-dimensional motion trajectories. In the isokinematic as well as the isotonic training algorithm, individual paths are defined i n C artesian s pace by sufficient s upport p oses. I n t he i sotonic t raining s cenario, the trajectory is adapted to the measured force as the robot should only move along the trajectory as long as the force applied by the user exceeds a minimum threshold. In the isotonic training scenario however, the robot’s acceleration is a function of the force applied by the user. To validate these findings, a simulative experiment with a simple linear trajectory is performed. For this purpose, the same force path is applied in both training scenarios. The results illustrate that the algorithms differ in the force dependent trajectory adaption. KW - Rehabilitation Technology and Prosthetics KW - Surgical Navigation and Robotics Y1 - 2018 U6 - http://dx.doi.org/10.1515/cdbme-2018-0005 SN - 2364-5504 VL - 4 IS - 1 SP - 17 EP - 20 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Birgel, Stefan A1 - Leschinger, Tim A1 - Wegmann, Kilian A1 - Staat, Manfred T1 - Calculation of muscle forces and joint reaction loads in the shoulder area via an OpenSim based computer model JF - tm - Technisches Messen N2 - Using the OpenSim software and verified anatomical data, a computer model for the calculation of biomechanical parameters is developed and used to determine the effect of a reattachment of the Supraspinatus muscle with a medial displacement of the muscle attachment point, which may be necessary for a rupture of the supraspinatus tendon. The results include the influence of the operation on basic biomechanical parameters such as the lever arm, as well as the calculated the muscle activations for the supraspinatus and deltoid. In addition, the influence on joint stability is examined by an analysis of the joint reaction force. The study provides a detailed description of the used model, as well as medical findings to a reattachment of the supraspinatus. Mit der Software OpenSim und überprüften anatomischen Daten wird ein Computermodell zur Berechnung von biomechanischen Parametern entwickelt und genutzt, um den Effekt einer Refixierung des Supraspinatusmuskels mit einer medialen Verschiebung des Muskelansatzpunktes zu ermitteln, wie sie unter anderem nach einem Riss der Supraspinatussehne notwendig sein kann. Die Ergebnisse umfassen hierbei den Einfluss der Operation auf grundlegende biomechanische Parameter wie den Hebelarm sowie die berechneten Muskelaktivierungen für den Supraspinatus und Deltoideus. Zusätzlich wird der Einfluss auf die Gelenkstabilität betrachtet und durch eine Analyse der Gelenkreaktionskraft untersucht. Die Studie bietet eine detaillierte Beschreibung des genutzten Modells, sowie medizinische Erkenntnisse zu einer Refixierung des Supraspinatus. Y1 - 2018 U6 - http://dx.doi.org/10.1515/teme-2017-0114 SN - 2196-7113 VL - 85 IS - 5 SP - 321 EP - 330 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Kampker, Achim A1 - Schleser, Markus A1 - Triebs, Johannes A1 - Berg, Alexander-Karl A1 - Hansen, Jan Ole A1 - Fiedler, Falko T1 - Bauteilintegrierte Vorrichtungsfunktionen im Karosseriebau : Aktuelle Ergebnisse im Forschungsprojekt ProeK JF - ZWF Zeitschrift für wirtschaftlichen Fabrikbetrieb N2 - Das Forschungsprojekt Produktionseffizienz in der Kleinserie (ProeK) erarbeitet kostengünstige und effiziente Lösungsansätze für Prozessketten im Zukunftsfeld der Elektromobilität. Das Teilprojekt Karosserie setzt diese Zielsetzung durch innovative und praxisorientierte Produkt- und Prozesskonzepte mit neuartigen bauteilintegrierten Vorrichtungsfunktionen (BiV) um. Im Teilprojekt Außenhaut sollen Toleranzen adaptiv durch Anpassungen der Prozessparameter sowie Bauteilmanipulation kompensiert werden. Y1 - 2018 U6 - http://dx.doi.org/10.3139/104.112012 SN - 0947-0085 VL - 113 IS - 11 SP - 767 EP - 770 PB - De Gruyter CY - Berlin ER -