TY - JOUR A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Frequency behaviour of light-addressable potentiometric sensors JF - Physica Status Solidi (A) N2 - Light-addressable potentiometric sensors (LAPS) are semiconductor-based potentiometric sensors, with the advantage to detect the concentration of a chemical species in a liquid solution above the sensor surface in a spatially resolved manner. The addressing is achieved by a modulated and focused light source illuminating the semiconductor and generating a concentration-depending photocurrent. This work introduces a LAPS set-up that is able to monitor the electrical impedance in addition to the photocurrent. The impedance spectra of a LAPS structure, with and without illumination, as well as the frequency behaviour of the LAPS measurement are investigated. The measurements are supported by electrical equivalent circuits to explain the impedance and the LAPS-frequency behaviour. The work investigates the influence of different parameters on the frequency behaviour of the LAPS. Furthermore, the phase shift of the photocurrent, the influence of the surface potential as well as the changes of the sensor impedance will be discussed. Y1 - 2013 U6 - https://doi.org/10.1002/pssa.201200929 SN - 1521-396X ; 0031-8965 VL - 210 IS - 5 SP - 884 EP - 891 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Huck, Christina A1 - Schiffels, Johannes A1 - Herrera, Cony N. A1 - Schelden, Maximilian A1 - Selmer, Thorsten A1 - Poghossian, Arshak A1 - Baumann, Marcus A1 - Wagner, Patrick A1 - Schöning, Michael Josef T1 - Metabolic responses of Escherichia coli upon glucose pulses captured by a capacitive field-effect sensor JF - Physica Status Solidi (A) N2 - Living cells are complex biological systems transforming metabolites taken up from the surrounding medium. Monitoring the responses of such cells to certain substrate concentrations is a challenging task and offers possibilities to gain insight into the vitality of a community influenced by the growth environment. Cell-based sensors represent a promising platform for monitoring the metabolic activity and thus, the “welfare” of relevant organisms. In the present study, metabolic responses of the model bacterium Escherichia coli in suspension, layered onto a capacitive field-effect structure, were examined to pulses of glucose in the concentration range between 0.05 and 2 mM. It was found that acidification of the surrounding medium takes place immediately after glucose addition and follows Michaelis–Menten kinetic behavior as a function of the glucose concentration. In future, the presented setup can, therefore, be used to study substrate specificities on the enzymatic level and may as well be used to perform investigations of more complex metabolic responses. Conclusions and perspectives highlighting this system are discussed. Y1 - 2013 U6 - https://doi.org/10.1002/pssa.201200900 SN - 0031-8965 VL - 210 IS - 5 SP - 926 EP - 931 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Bäcker, Matthias A1 - Schusser, Sebastian A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Degradationsverhalten bioabbaubarer Polymere : siliziumbasierte Sensorik zur Überwachung JF - GIT Labor-Fachzeitschrift Y1 - 2013 SN - 0016-3538 SP - 32 EP - 33 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Gräßl, Andreas A1 - Renz, Wolfgang A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Winter, Lukas A1 - Özerdem, Celal A1 - Rieger, Jan A1 - Kellmann, Peter A1 - Santoro, Davide A1 - Lindel, Tomasz D. A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Niendorf, Thoralf T1 - Modular 32-channel transceiver coil array for cardiac MRI at 7.0T JF - Magnetic Resonance in Medicine N2 - Purpose To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. Methods The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Results Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. Conclusion The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T. Y1 - 2013 U6 - https://doi.org/10.1002/mrm.24903 SN - 1522-2594 VL - 72 IS - 1 SP - 276 EP - 290 PB - Wiley-Liss CY - New York ER -