TY - JOUR A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Schäfer, Lena C. A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Maderwald, Stefan T1 - Open design eight-channel transmit/receive coil for high-resolution and real-time ankle imaging at 7 T JF - Medical Physics N2 - Purpose: At 1.5 T, real-time MRI of joint movement has been shown to be feasible. However, 7 T, provides higher SNR and thus an improved potential for parallel imaging acceleration. The purpose of this work was to build an open, U-shaped eight-channel transmit/receive microstrip coil for 7 T MRI to enable high-resolution and real-time imaging of the moving ankle joint. Methods: A U-shaped eight-channel transmit/receive array for the human ankle was built.urn:x-wiley:00942405:mp3399:equation:mp3399-math-0001-parameters and urn:x-wiley:00942405:mp3399:equation:mp3399-math-0002-factor were measured. SAR calculations of different ankle postures were performed to ensure patient safety. Inhomogeneities in the transmit field consequent to the open design were compensated for by the use of static RF shimming. High-resolution and real-time imaging was performed in human volunteers. Results: The presented array showed good performance with regard to patient comfort and image quality. High acceleration factors of up to 4 are feasible without visible acceleration artifacts. Reasonable image homogeneity was achieved with RF shimming. Conclusions: Open, noncylindrical designs for transmit/receive coils are practical at 7 T and real-time imaging of the moving joint is feasible with the presented coil design. Y1 - 2011 U6 - https://doi.org/10.1118/1.3553399 SN - 2473-4209 VL - 38 IS - 3 SP - 1162 EP - 1167 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Kraff, Oliver A1 - Bitz, Andreas A1 - Dammann, Philipp A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - An eight-channel transmit/receive multipurpose coil for musculoskeletal MR imaging at 7 T JF - Medical Physics N2 - Purpose: MRI plays a leading diagnostic role in assessing the musculoskeletal (MSK) system and is well established for most questions at clinically used field strengths (up to 3 T). However, there are still limitations in imaging early stages of cartilage degeneration, very fine tendons and ligaments, or in locating nerve lesions, for example. 7 T MRI of the knee has already received increasing attention in the current published literature, but there is a strong need to develop new radiofrequency (RF) coils to assess more regions of the MSK system. In this work, an eight-channel transmit/receive RF array was built as a multipurpose coil for imaging some of the thus far neglected regions. An extensive coil characterization protocol and first in vivo results of the human wrist, shoulder, elbow, knee, and ankle imaged at 7 T will be presented. Methods: Eight surface loop coils with a dimension ofurn:x-wiley:00942405:media:mp7176:mp7176-math-0001 were machined from FR4 circuit board material. To facilitate easy positioning, two coil clusters, each with four loop elements, were combined to one RF transmit/receive array. An overlapped and shifted arrangement of the coil elements was chosen to reduce the mutual inductance between neighboring coils. A phantom made of body-simulating liquid was used for tuning and matching on the bench. Afterward, the S-parameters were verified on a human wrist, elbow, and shoulder. For safety validation, a detailed compliance test was performed including full wave simulations of the RF field distribution and the corresponding specific absorption rate (SAR) for all joints. In vivo images of four volunteers were assessed with gradient echo and spin echo sequences modified to obtain optimal image contrast, full anatomic coverage, and the highest spatial resolution within a reasonable acquisition time. The performance of the RF coil was additionally evaluated by in vivo B1 mapping. Results: A comparison of B1 per unit power, flip angle distribution, and anatomic images showed a fairly homogeneous excitation for the smaller joints (elbow, wrist, and ankle), while for the larger joints, the shoulder and especially the knee, B1 inhomogeneities and limited penetration depth were more pronounced. However, the greater part of the shoulder joint could be imaged.In vivo images rendered very fine anatomic details such as fascicles of the median nerve and the branching of the nerve bundles. High-resolution images of cartilage, labrum, and tendons could be acquired. Additionally, turbo spin echo (TSE) and inversion recovery sequences performed very well. Conclusions: This study demonstrates that the concept of two four-channel transmit/receive RF arrays can be used as a multipurpose coil for high-resolutionin vivo MR imaging of the musculoskeletal system at 7 T. Not only gradient echo but also typical clinical and SAR-intensive sequences such as STIR and TSE performed well. Imaging of small structures and peripheral nerves could in particular benefit from this technique. Y1 - 2010 U6 - https://doi.org/10.1118/1.3517176 SN - 2473-4209 VL - 37 IS - 12 SP - 6368 EP - 6376 PB - Wiley CY - Hoboken, NJ ER - TY - JOUR A1 - Bank, Bart L. van de A1 - Orzada, Stephan A1 - Smits, Frits A1 - Lagemaat, Miriam W. A1 - Rodgers, Christopher T. A1 - Bitz, Andreas A1 - Scheenen, Tom W. J. T1 - Optimized (31) P MRS in the human brain at 7 T with a dedicated RF coil setup JF - NMR in Biomedicine Y1 - 2015 U6 - https://doi.org/10.1002/nbm.3422 SN - 1099-1492 VL - 28 IS - 11 SP - 1570 EP - 1578 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Lagemaat, Miriam W. A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Weiland, Elisabeth A1 - Uden, Mark J. van A1 - Kobus, Thiele A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - (31) P MR spectroscopic imaging of the human prostate at 7 T: T1 relaxation times, Nuclear Overhauser Effect, and spectral characterization JF - Magnetic Resonance in Medicine Y1 - 2015 U6 - https://doi.org/10.1002/mrm.25209 SN - 1522-2594 VL - 73 IS - 3 SP - 909 EP - 920 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Ciritsis, Alexander A1 - Horbach, Andreas A1 - Staat, Manfred A1 - Kuhl, Christiane K. A1 - Kraemer, Nils Andreas T1 - Porosity and tissue integration of elastic mesh implants evaluated in vitro and in vivo JF - Journal of Biomedical Materials Research: Part B: Applied Biomaterials N2 - Purpose In vivo, a loss of mesh porosity triggers scar tissue formation and restricts functionality. The purpose of this study was to evaluate the properties and configuration changes as mesh deformation and mesh shrinkage of a soft mesh implant compared with a conventional stiff mesh implant in vitro and in a porcine model. Material and Methods Tensile tests and digital image correlation were used to determine the textile porosity for both mesh types in vitro. A group of three pigs each were treated with magnetic resonance imaging (MRI) visible conventional stiff polyvinylidene fluoride meshes (PVDF) or with soft thermoplastic polyurethane meshes (TPU) (FEG Textiltechnik mbH, Aachen, Germany), respectively. MRI was performed with a pneumoperitoneum at a pressure of 0 and 15 mmHg, which resulted in bulging of the abdomen. The mesh-induced signal voids were semiautomatically segmented and the mesh areas were determined. With the deformations assessed in both mesh types at both pressure conditions, the porosity change of the meshes after 8 weeks of ingrowth was calculated as an indicator of preserved elastic properties. The explanted specimens were examined histologically for the maturity of the scar (collagen I/III ratio). Results In TPU, the in vitro porosity increased constantly, in PVDF, a loss of porosity was observed under mild stresses. In vivo, the mean mesh areas of TPU were 206.8 cm2 (± 5.7 cm2) at 0 mmHg pneumoperitoneum and 274.6 cm2 (± 5.2 cm2) at 15 mmHg; for PVDF the mean areas were 205.5 cm2 (± 8.8 cm2) and 221.5 cm2 (± 11.8 cm2), respectively. The pneumoperitoneum-induced pressure increase resulted in a calculated porosity increase of 8.4% for TPU and of 1.2% for PVDF. The mean collagen I/III ratio was 8.7 (± 0.5) for TPU and 4.7 (± 0.7) for PVDF. Conclusion The elastic properties of TPU mesh implants result in improved tissue integration compared to conventional PVDF meshes, and they adapt more efficiently to the abdominal wall. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 827–833, 2018. Y1 - 2018 U6 - https://doi.org/10.1002/jbm.b.33877 SN - 1552-4981 VL - 106 IS - 2 SP - 827 EP - 833 PB - Wiley CY - New York, NY ER - TY - JOUR A1 - El Quardi, A. A1 - Streckert, J. A1 - Bitz, Andreas A1 - Münkner, S. A1 - Engel, J. A1 - Hansen, V. T1 - New fin-line devices for radiofrequency exposure of small biological samples in vitro allowing whole-cell patch clamp recordings JF - Bioelectromagnetics N2 - The development and analysis of three waveguides for the exposure of small biological in vitro samples to mobile communication signals at 900 MHz (GSM, Global System for Mobile Communications), 1.8 GHz (GSM), and 2 GHz (UMTS, Universal Mobile Telecommunications System) is presented. The waveguides were based on a fin-line concept and the chamber containing the samples bathed in extracellular solution was placed onto two fins with a slot in between, where the exposure field concentrates. Measures were taken to allow for patch clamp recordings during radiofrequency (RF) exposure. The necessary power for the achievement of the maximum desired specific absorption rate (SAR) of 20 W/kg (average over the mass of the solution) was approximately Pin = 50 mW, Pin = 19 mW, and Pin = 18 mW for the 900 MHz, 1800 MHz, and 2 GHz devices, respectively. At 20 W/kg, a slight RF-induced temperature elevation in the solution of no more than 0.3 °C was detected, while no thermal offsets due to the electromagnetic exposure could be detected at the lower SAR settings (2, 0.2, and 0.02 W/kg). A deviation of 10% from the intended solution volume yielded a calculated SAR deviation of 8% from the desired value. A maximum ±10% variation in the local SAR could occur when the position of the patch clamp electrode was altered within the area where the cells to be investigated were located. Y1 - 2011 U6 - https://doi.org/10.1002/bem.20621 SN - 1521-186X VL - 32 IS - 2 SP - 102 EP - 112 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Noureddine, Yacine A1 - Kraff, Oliver A1 - Ladd, Mark E. A1 - Wrede, Karsten H. A1 - Chen, Bixia A1 - Quick, Harald H. A1 - Schaefers, Gregor A1 - Bitz, Andreas T1 - In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla JF - Magnetic Resonance in Medicine Y1 - 2017 U6 - https://doi.org/10.1002/mrm.26650 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Kraff, Oliver A1 - Wrede, Karsten H. A1 - Schoemberg, Tobias A1 - Dammann, Philipp A1 - Noureddine, Yacine A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - MR safety assessment of potential RF heating from cranial fixation plates at 7 T JF - Medical Physics Y1 - 2013 U6 - https://doi.org/10.1118/1.4795347 SN - 2473-4209 VL - 40 IS - 4 SP - 042302-1 EP - 042302-10 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Alexopoulos, Spiros A1 - Hoffschmidt, Bernhard T1 - Advances in solar tower technology JF - Wiley interdisciplinary reviews : Energy and Environment : WIREs Y1 - 2017 U6 - https://doi.org/10.1002/wene.217 SN - 2041-840X VL - 6 IS - 1 SP - 1 EP - 19 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Nguyen-Xuan, H. A1 - Rabczuk, T. A1 - Nguyen-Thoi, T. A1 - Tran, Thanh Ngoc A1 - Nguyen-Thanh, N. T1 - Computation of limit and shakedown loads using a node-based smoothed finite element method JF - International Journal for Numerical Methods in Engineering N2 - This paper presents a novel numerical procedure for computing limit and shakedown loads of structures using a node-based smoothed FEM in combination with a primal–dual algorithm. An associated primal–dual form based on the von Mises yield criterion is adopted. The primal-dual algorithm together with a Newton-like iteration are then used to solve this associated primal–dual form to determine simultaneously both approximate upper and quasi-lower bounds of the plastic collapse limit and the shakedown limit. The present formulation uses only linear approximations and its implementation into finite element programs is quite simple. Several numerical examples are given to show the reliability, accuracy, and generality of the present formulation compared with other available methods. Y1 - 2011 U6 - https://doi.org/10.1002/nme.3317 SN - 1097-0207 VL - 90 IS - 3 SP - 287 EP - 310 PB - Wiley CY - Weinheim ER -