TY - JOUR A1 - Ciritsis, Alexander A1 - Horbach, Andreas A1 - Staat, Manfred A1 - Kuhl, Christiane K. A1 - Kraemer, Nils Andreas T1 - Porosity and tissue integration of elastic mesh implants evaluated in vitro and in vivo JF - Journal of Biomedical Materials Research: Part B: Applied Biomaterials N2 - Purpose In vivo, a loss of mesh porosity triggers scar tissue formation and restricts functionality. The purpose of this study was to evaluate the properties and configuration changes as mesh deformation and mesh shrinkage of a soft mesh implant compared with a conventional stiff mesh implant in vitro and in a porcine model. Material and Methods Tensile tests and digital image correlation were used to determine the textile porosity for both mesh types in vitro. A group of three pigs each were treated with magnetic resonance imaging (MRI) visible conventional stiff polyvinylidene fluoride meshes (PVDF) or with soft thermoplastic polyurethane meshes (TPU) (FEG Textiltechnik mbH, Aachen, Germany), respectively. MRI was performed with a pneumoperitoneum at a pressure of 0 and 15 mmHg, which resulted in bulging of the abdomen. The mesh-induced signal voids were semiautomatically segmented and the mesh areas were determined. With the deformations assessed in both mesh types at both pressure conditions, the porosity change of the meshes after 8 weeks of ingrowth was calculated as an indicator of preserved elastic properties. The explanted specimens were examined histologically for the maturity of the scar (collagen I/III ratio). Results In TPU, the in vitro porosity increased constantly, in PVDF, a loss of porosity was observed under mild stresses. In vivo, the mean mesh areas of TPU were 206.8 cm2 (± 5.7 cm2) at 0 mmHg pneumoperitoneum and 274.6 cm2 (± 5.2 cm2) at 15 mmHg; for PVDF the mean areas were 205.5 cm2 (± 8.8 cm2) and 221.5 cm2 (± 11.8 cm2), respectively. The pneumoperitoneum-induced pressure increase resulted in a calculated porosity increase of 8.4% for TPU and of 1.2% for PVDF. The mean collagen I/III ratio was 8.7 (± 0.5) for TPU and 4.7 (± 0.7) for PVDF. Conclusion The elastic properties of TPU mesh implants result in improved tissue integration compared to conventional PVDF meshes, and they adapt more efficiently to the abdominal wall. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 827–833, 2018. Y1 - 2018 U6 - http://dx.doi.org/10.1002/jbm.b.33877 SN - 1552-4981 VL - 106 IS - 2 SP - 827 EP - 833 PB - Wiley CY - New York, NY ER - TY - JOUR A1 - Digel, Ilya A1 - Wehlitz, V. A1 - Kayser, Peter A1 - Figiel-Lange, A. A1 - Bassam, R. A1 - Rundstedt, F. von T1 - Suspension depletion approach for exemption of infected Solanum jasminoides cells from pospiviroids JF - Plant Pathology N2 - Despite numerous studies, viroid elimination from infected plants remains a very challenging task. This study introduces for the first time a novel ‘suspension depletion’ approach for exemption of Solanum jasminoides plants from viroids. The proposed method implies initial establishment of suspension cultures of the infected plant cells. The suspended cells were then physically treated (mild thermotherapy, 33 °C), which presumably delayed the replication of the viroid. The viroid concentration in the treated biomass was monitored weekly using pospiviroid-specific PCR. After 10–12 weeks of continuous treatment, a sufficient decrease in viroid concentration was observed such that the infection became undetectable by PCR. The treated single cells then gave rise to microcolonies on a solid culture medium and the obtained viroid-negative clones were further promoted to regenerate into viroid-free plants. Three years of accumulated experimental data suggests feasibility, broad applicability, and good efficacy of the proposed approach. Y1 - 2018 U6 - http://dx.doi.org/10.1111/ppa.12750 SN - 1365-3059 VL - 67 IS - 2 SP - 358 EP - 365 PB - Wiley CY - Oxford ER -