TY - JOUR A1 - Hugenroth, Kristin A1 - Borchardt, Ralf A1 - Ritter, Philine A1 - Groß‑Hardt, Sascha A1 - Meyns, Bart A1 - Verbelen, Tom A1 - Steinseifer, Ulrich A1 - Kaufmann, Tim A. S. A1 - Engelmann, Ulrich M. T1 - Optimizing cerebral perfusion and hemodynamics during cardiopulmonary bypass through cannula design combining in silico, in vitro and in vivo input JF - Scientific Reports N2 - Cardiopulmonary bypass (CPB) is a standard technique for cardiac surgery, but comes with the risk of severe neurological complications (e.g. stroke) caused by embolisms and/or reduced cerebral perfusion. We report on an aortic cannula prototype design (optiCAN) with helical outflow and jet-splitting dispersion tip that could reduce the risk of embolic events and restores cerebral perfusion to 97.5% of physiological flow during CPB in vivo, whereas a commercial curved-tip cannula yields 74.6%. In further in vitro comparison, pressure loss and hemolysis parameters of optiCAN remain unaffected. Results are reproducibly confirmed in silico for an exemplary human aortic anatomy via computational fluid dynamics (CFD) simulations. Based on CFD simulations, we firstly show that optiCAN design improves aortic root washout, which reduces the risk of thromboembolism. Secondly, we identify regions of the aortic intima with increased risk of plaque release by correlating areas of enhanced plaque growth and high wall shear stresses (WSS). From this we propose another easy-to-manufacture cannula design (opti2CAN) that decreases areas burdened by high WSS, while preserving physiological cerebral flow and favorable hemodynamics. With this novel cannula design, we propose a cannulation option to reduce neurological complications and the prevalence of stroke in high-risk patients after CPB. Y1 - 2021 U6 - https://doi.org/10.1038/s41598-021-96397-2 SN - 2045-2322 VL - 11 IS - Art. No. 16800 SP - 1 EP - 12 PB - Springer CY - Berlin ER - TY - JOUR A1 - Neumaier, Felix A1 - Weiss, Miriam A1 - Veldeman, Michael A1 - Kotliar, Konstantin A1 - Wiesmann, Martin A1 - Schulze-Steinen, Henna A1 - Höllig, Anke A1 - Clusmann, Hans A1 - Schubert, Gerrit Alexander A1 - Albanna, Walid T1 - Changes in endogenous daytime melatonin levels after aneurysmal subarachnoid hemorrhage – preliminary findings from an observational cohort study JF - Clinical Neurology and Neurosurgery N2 - Aneurysmal subarachnoid hemorrhage (aSAH) is associated with early and delayed brain injury due to several underlying and interrelated processes, which include inflammation, oxidative stress, endothelial, and neuronal apoptosis. Treatment with melatonin, a cytoprotective neurohormone with anti-inflammatory, anti-oxidant and anti-apoptotic effects, has been shown to attenuate early brain injury (EBI) and to prevent delayed cerebral vasospasm in experimental aSAH models. Less is known about the role of endogenous melatonin for aSAH outcome and how its production is altered by the pathophysiological cascades initiated during EBI. In the present observational study, we analyzed changes in melatonin levels during the first three weeks after aSAH. KW - constructive alignment KW - examination KW - long-term retention KW - multimodal KW - practical learning Y1 - 2021 U6 - https://doi.org/10.1016/j.clineuro.2021.106870 SN - 0303-8467 VL - 208 IS - Article No.: 106870 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Brockhaus, Moritz K. A1 - Behbahani, Mehdi A1 - Muris, Farina A1 - Jansen, Sebastian V. A1 - Schmitz- Rode, Thomas A1 - Steinseifer, Ulrich A1 - Clauser, Johanna C. T1 - In vitro thrombogenicity testing of pulsatile mechanical circulatory support systems: Design and proof-of-concept JF - Artificial Organs N2 - Thrombogenic complications are a main issue in mechanical circulatory support (MCS). There is no validated in vitro method available to quantitatively assess the thrombogenic performance of pulsatile MCS devices under realistic hemodynamic conditions. The aim of this study is to propose a method to evaluate the thrombogenic potential of new designs without the use of complex in-vivo trials. This study presents a novel in vitro method for reproducible thrombogenicity testing of pulsatile MCS systems using low molecular weight heparinized porcine blood. Blood parameters are continuously measured with full blood thromboelastometry (ROTEM; EXTEM, FIBTEM and a custom-made analysis HEPNATEM). Thrombus formation is optically observed after four hours of testing. The results of three experiments are presented each with two parallel loops. The area of thrombus formation inside the MCS device was reproducible. The implantation of a filter inside the loop catches embolizing thrombi without a measurable increase of platelet activation, allowing conclusions of the place of origin of thrombi inside the device. EXTEM and FIBTEM parameters such as clotting velocity (α) and maximum clot firmness (MCF) show a total decrease by around 6% with a characteristic kink after 180 minutes. HEPNATEM α and MCF rise within the first 180 minutes indicate a continuously increasing activation level of coagulation. After 180 minutes, the consumption of clotting factors prevails, resulting in a decrease of α and MCF. With the designed mock loop and the presented protocol we are able to identify thrombogenic hot spots inside a pulsatile pump and characterize their thrombogenic potential. Y1 - 2021 U6 - https://doi.org/10.1111/aor.14046 SN - 1525-1594 VL - 45 IS - 12 SP - 1513 EP - 1521 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Hackl, Michael A1 - Nacov, Julia A1 - Kammerlohr, Sandra A1 - Staat, Manfred A1 - Buess, Eduard A1 - Leschinger, Tim A1 - Müller, Lars P. A1 - Wegmann, Kilian T1 - Intratendinous Strain Variations of the Supraspinatus Tendon Depending on Repair Technique: A Biomechanical Analysis Regarding the Cause of Medial Cuff Failure JF - The American Journal of Sports Medicine Y1 - 2021 U6 - https://doi.org/10.1177/03635465211006138 SN - 1552-3365 SN - 0363-5465 VL - 49 IS - 7 SP - 1847 EP - 1853 PB - Sage CY - London ER - TY - JOUR A1 - Neumaier, Felix A1 - Kotliar, Konstantin A1 - Haeren, Roel Hubert Louis A1 - Temel, Yasin A1 - Lüke, Jan Niklas A1 - Seyam, Osama A1 - Lindauer, Ute A1 - Clusmann, Hans A1 - Hescheler, Jürgen A1 - Schubert, Gerrit Alexander A1 - Schneider, Toni A1 - Albanna, Walid T1 - Retinal Vessel Responses to Flicker Stimulation Are Impaired in Ca v 2.3-Deficient Mice—An in- vivo Evaluation Using Retinal Vessel Analysis (RVA) JF - Frontiers in Neurology Y1 - 2021 U6 - https://doi.org/10.3389/fneur.2021.659890 VL - 12 SP - 1 EP - 11 PB - Frontiers ER - TY - JOUR A1 - Arampatzis, Adamantios A1 - Karamanidis, Kiros A1 - Albracht, Kirsten T1 - Adaptational responses of the human Achilles tendon by modulation of the applied cyclic strain magnitude JF - Journal of Experimental Biology Y1 - 2007 U6 - https://doi.org/10.1242/jeb.003814 SN - 0022-0949 VL - 210 IS - 15 SP - 2743 EP - 2753 ER - TY - JOUR A1 - Albracht, Kirsten A1 - Arampatzis, Adamantios T1 - Influence of the mechanical properties of the muscle-tendon unit on force generation in runners with different running economy JF - Biological Cybernetics Y1 - 2006 U6 - https://doi.org/10.1007/s00422-006-0070-z SN - 1432-0770 VL - 95 IS - 1 SP - 87 EP - 96 ER - TY - JOUR A1 - Gamella, Maria A1 - Zakharchenko, Andrey A1 - Guz, Nataliia A1 - Masi, Madeline A1 - Minko, Sergiy A1 - Kolpashchikov, Dmitry M. A1 - Iken, Heiko A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Katz, Evgeny T1 - DNA computing system activated by electrochemically triggered DNA realease from a polymer-brush-modified electrode array JF - Electroanalysis N2 - An array of four independently wired indium tin oxide (ITO) electrodes was used for electrochemically stimulated DNA release and activation of DNA-based Identity, AND and XOR logic gates. Single-stranded DNA molecules were loaded on the mixed poly(N,N-dimethylaminoethyl methacrylate) (PDMAEMA)/poly(methacrylic acid) (PMAA) brush covalently attached to the ITO electrodes. The DNA deposition was performed at pH 5.0 when the polymer brush is positively charged due to protonation of tertiary amino groups in PDMAEMA, thus resulting in electrostatic attraction of the negatively charged DNA. By applying electrolysis at −1.0 V(vs. Ag/AgCl reference) electrochemical oxygen reduction resulted in the consumption of hydrogen ions and local pH increase near the electrode surface. The process resulted in recharging the polymer brush to the negative state due to dissociation of carboxylic groups of PMAA, thus repulsing the negatively charged DNA and releasing it from the electrode surface. The DNA release was performed in various combinations from different electrodes in the array assembly. The released DNA operated as input signals for activation of the Boolean logic gates. The developed system represents a step forward in DNA computing, combining for the first time DNA chemical processes with electronic input signals. Y1 - 2017 U6 - https://doi.org/10.1002/elan.201600389 SN - 1521-4109 VL - 29 IS - 2 SP - 398 EP - 408 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Sousa, Marcos A. M. A1 - Siqueira, Jose R. Jr. A1 - Vercik, Andres A1 - Schöning, Michael Josef A1 - Oliveira, Osvaldo N. Jr. T1 - Determining the optimized layer-by-layer film architecture with dendrimer/carbon nanotubes for field-effect sensors JF - IEEE Sensors Journal N2 - The capacitive electrolyte–insulator–semiconductor (EIS) structure is a typical device based on a field-effect sensor platform. With a simple silicon-based structure, EIS have been useful for several sensing applications, especially with incorporation of nanostructured films to modulate the ionic transport and the flat-band potential. In this paper, we report on ion transport and changes in flat-band potential in EIS sensors made with layer-by-layer films containing poly(amidoamine) (PAMAM) dendrimer and single-walled carbon nanotubes (SWNTs) adsorbed on p-Si/SiO 2 /Ta 2 O 5 chips with an Al ohmic contact. The impedance spectra were fitted using an equivalent circuit model, from which we could determine parameters such as the double-layer capacitance. This capacitance decreased with the number of bilayers owing to space charge accumulated at the electrolyte–insulator interface, up to three PAMAM/SWNTs bilayers, after which it stabilized. The charge-transfer resistance was also minimum for three bilayers, thus indicating that this is the ideal architecture for an optimized EIS performance. The understanding of the influence of nanostructures and the fine control of operation parameters pave the way for optimizing the design and performance of new EIS sensors. Y1 - 2017 U6 - https://doi.org/10.1109/JSEN.2017.2653238 SN - 1558-1748 VL - 17 IS - 6 SP - 1735 EP - 1740 PB - IEEE CY - New York ER - TY - JOUR A1 - Yoshinobu, Tatsuo A1 - Miyamoto, Ko-ichiro A1 - Werner, Frederik A1 - Poghossian, Arshak A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Light-addressable potentiometric sensors for quantitative spatial imaging of chemical species JF - Annual Review of Analytical Chemistry N2 - A light-addressable potentiometric sensor (LAPS) is a semiconductor-based chemical sensor, in which a measurement site on the sensing surface is defined by illumination. This light addressability can be applied to visualize the spatial distribution of pH or the concentration of a specific chemical species, with potential applications in the fields of chemistry, materials science, biology, and medicine. In this review, the features of this chemical imaging sensor technology are compared with those of other technologies. Instrumentation, principles of operation, and various measurement modes of chemical imaging sensor systems are described. The review discusses and summarizes state-of-the-art technologies, especially with regard to the spatial resolution and measurement speed; for example, a high spatial resolution in a submicron range and a readout speed in the range of several tens of thousands of pixels per second have been achieved with the LAPS. The possibility of combining this technology with microfluidic devices and other potential future developments are discussed. Y1 - 2017 U6 - https://doi.org/10.1146/annurev-anchem-061516-045158 SN - 1936-1327 VL - 10 SP - 225 EP - 246 PB - Annual Reviews CY - Palo Alto, Calif. ER -