TY - JOUR A1 - Wu, Chunsheng A1 - Bronder, Thomas A1 - Poghossian, Arshak A1 - Werner, Frederik A1 - Bäcker, Matthias A1 - Schöning, Michael Josef T1 - Label-free electrical detection of DNA with a multi-spot LAPS: First step towards light-addressable DNA chips JF - Physica status solidi A : Applications and materials science N2 - A multi-spot (4 × 4 spots) light-addressable potentiometric sensor (MLAPS) consisting of an Al–p-Si–SiO2 structure has been applied for the label-free electrical detection of DNA (deoxyribonucleic acid) immobilization and hybridization by the intrinsic molecular charge for the first time. Single-stranded probe ssDNA molecules (20 bases) were covalently immobilized onto the silanized SiO2 gate surface. The unspecific adsorption of mismatch ssDNA on the MLAPS gate surface was blocked by bovine serum albumin molecules. To reduce the screening effect and to achieve a high sensor signal, the measurements were performed in a low ionic-strength solution. The photocurrent–voltage (I–V) curves were simultaneously recorded on all 16 spots after each surface functionalization step. Large shifts of I–V curves of 25 mV were registered after the DNA immobilization and hybridization event. In contrast, a small potential shift (∼5 mV) was observed in case of mismatch ssDNA, revealing good specificity of the sensor. The obtained results demonstrate the potential of the MLAPS as promising transducer platform for the multi-spot label-free electrical detection of DNA molecules by their intrinsic molecular charge. Y1 - 2014 U6 - http://dx.doi.org/10.1002/pssa.201330442 SN - 1521-396X (E-Journal); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1423 EP - 1428 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Wagner, Patrick A1 - Doll, Theodor A1 - Schöning, Michael Josef T1 - Engineering of functional interfaces / Patrick Wagner ; Theodor Doll ; Michael J. Schöning (eds.) JF - Physica status solidi (A) : Applications and materials science Y1 - 2014 U6 - http://dx.doi.org/10.1002/pssa.201470241 SN - 1521-396X (E-Book); 1862-6319 (E-Book); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1339 EP - 1339 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Takenaga, Shoko A1 - Biselli, Manfred A1 - Schnitzler, Thomas A1 - Öhlschläger, Peter A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Toward multi-analyte bioarray sensors: LAPS-based on-chip determination of a Michaelis–Menten-like kinetics for cell culturing JF - Physica status solidi A : Applications and materials science N2 - The metabolic activity of Chinese hamster ovary (CHO) cells was observed using a light-addressable potentiometric sensor (LAPS). The dependency toward different glucose concentrations (17–200 mM) follows a Michaelis–Menten kinetics trajectory with Kₘ = 32.8 mM, and the obtained Kₘ value in this experiment was compared with that found in literature. In addition, the pH shift induced by glucose metabolism of tumor cells transfected with the HPV-16 genome (C3 cells) was successfully observed. These results indicate the possibility to determine the tumor cells metabolism with a LAPS-based measurement device. Y1 - 2014 U6 - http://dx.doi.org/10.1002/pssa.201330464 SN - 1521-396X (E); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1410 EP - 1415 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Siqueira, Jose R. A1 - Molinnus, Denise A1 - Beging, Stefan A1 - Schöning, Michael Josef T1 - Incorporating a hybrid urease-carbon nanotubes sensitive nanofilm on capacitive field-effect sensors for urea detection JF - Analytical chemistry N2 - The ideal combination among biomolecules and nanomaterials is the key for reaching biosensing units with high sensitivity. The challenge, however, is to find out a stable and sensitive film architecture that can be incorporated on the sensor’s surface. In this paper, we report on the benefits of incorporating a layer-by-layer (LbL) nanofilm of polyamidoamine (PAMAM) dendrimer and carbon nanotubes (CNTs) on capacitive electrolyte-insulator-semiconductor (EIS) field-effect sensors for detecting urea. Three sensor arrangements were studied in order to investigate the adequate film architecture, involving the LbL film with the enzyme urease: (i) urease immobilized directly onto a bare EIS [EIS-urease] sensor; (ii) urease atop the LbL film over the EIS [EIS-(PAMAM/CNT)-urease] sensor; and (iii) urease sandwiched between the LbL film and another CNT layer [EIS-(PAMAM/CNT)-urease-CNT]. The surface morphology of all three urea-based EIS biosensors was investigated by atomic force microscopy (AFM), while the biosensing abilities were studied by means of capacitance–voltage (C/V) and dynamic constant-capacitance (ConCap) measureaments at urea concentrations ranging from 0.1 mM to 100 mM. The EIS-urease and EIS-(PAMAM/CNT)-urease sensors showed similar sensitivity (∼18 mV/decade) and a nonregular signal behavior as the urea concentration increased. On the other hand, the EIS-(PAMAM/CNT)-urease-CNT sensor exhibited a superior output signal performance and higher sensitivity of about 33 mV/decade. The presence of the additional CNT layer was decisive to achieve a urea based EIS sensor with enhanced properties. Such sensitive architecture demonstrates that the incorporation of an adequate hybrid enzyme-nanofilm as sensing unit opens new prospects for biosensing applications using the field-effect sensor platform. Y1 - 2014 U6 - http://dx.doi.org/10.1021/ac500458s SN - 1520-6882 (E-Journal); 0003-2700 (Print); 0096-4484 (Print) VL - 86 IS - 11 SP - 5370 EP - 5375 PB - ACS Publications CY - Columbus ER - TY - JOUR A1 - Schusser, Sebastian A1 - Bäcker, Matthias A1 - Krischer, M. A1 - Wenzel, L. A1 - Leinhos, Marcel A1 - Poghossian, Arshak A1 - Biselli, Manfred A1 - Wagner, P. A1 - Schöning, Michael Josef T1 - Enzymatically catalyzed degradation of biodegradable polymers investigated by means of a semiconductor-based field-effect sensor JF - Procedia Engineering N2 - A semiconductor field-effect device has been used for an enzymatically catalyzed degradation of biopolymers for the first time. This novel technique is capable to monitor the degradation process of multiple samples in situ and in real-time. As model system, the degradation of the biopolymer poly(D, L-lactic acid) has been monitored in the degradation medium containing the enzyme lipase from Rhizomucor miehei. The obtained results demonstrate the potential of capacitive field-effect sensors for degradation studies of biodegradable polymers. KW - Field-effect sensor KW - enzymatic (bio)degradation KW - poly(d, l-lactic acid) KW - in-situ monitoring KW - impedance spectroscopy Y1 - 2014 U6 - http://dx.doi.org/10.1016/j.proeng.2014.11.689 SN - 1877-7058 N1 - EUROSENSORS 2014 ; European Conference on Solid-State Transducers <28, 2014> VL - 87 SP - 1314 EP - 1317 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Riedel, Marc A1 - Kartchemnik, Julia A1 - Schöning, Michael Josef A1 - Lisdat, Fred T1 - Impedimetric DNA detection – steps forward to sensorial application JF - Analytical chemistry N2 - This study describes a label-free impedimetric sensor based on short ssDNA recognition elements for the detection of hybridization events. We concentrate on the elucidation of the influence of target length and recognition sequence position on the sensorial performance. The impedimetric measurements are performed in the presence of the redox system ferri-/ferrocyanide and show an increase in charge transfer resistance upon hybridization of ssDNA to the sensor surface. Investigations on the impedimetric signal stability demonstrate a clear influence of the buffers used during the sensor preparation and the choice of the passivating mercaptoalcanol compound. A stable sensor system has been developed, enabling a reproducible detection of 25mer target DNA in the low nanomolar range. After hybridization, a sensor regeneration can be reached with deionized water by adjustment of effective convection conditions, ensuring a sensor reusability. By investigations of longer targets with overhangs exposed to the solution, we can demonstrate applicability of the impedimetric detection for longer ssDNA. However, a decreasing charge transfer resistance change (ΔRct) is found by extending the overhang. As a strategy to increase the impedance change for longer target strands, the position of the recognition sequence can be designed in a way that a small overhang is exposed to the electrode surface. This is found to result in an increase in the relative Rct change. These results suggest that DNA and consequently negative charge near the electrode possess a larger impact on the impedimetric signal than DNA further away. Y1 - 2014 U6 - http://dx.doi.org/10.1021/ac501800q SN - 1520-6882 (E-Journal); 0003-2700 (Print); 0096-4484 (Print) VL - 86 (2014) IS - 15 SP - 7867 EP - 7874 PB - ACS Publications CY - Columbus ER - TY - JOUR A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Label-free sensing of biomolecules with field-effect devices for clinical applications JF - Electroanalysis N2 - Among the variety of transducer concepts proposed for label-free detection of biomolecules, the semiconductor field-effect device (FED) is one of the most attractive platforms. As medical techniques continue to progress towards diagnostic and therapies based on biomarkers, the ability of FEDs for a label-free, fast and real-time detection of multiple pathogenic and physiologically relevant molecules with high specificity and sensitivity offers very promising prospects for their application in point-of-care and personalized medicine for an early diagnosis and treatment of diseases. The presented paper reviews recent advances and current trends in research and development of different FEDs for label-free, direct electrical detection of charged biomolecules by their intrinsic molecular charge. The authors are mainly focusing on the detection of the DNA hybridization event, antibody-antigen affinity reaction as well as clinically relevant biomolecules such as cardiac and cancer biomarkers. Y1 - 2014 U6 - http://dx.doi.org/10.1002/elan.201400073 SN - 1521-4109 (E-Journal); 1040-0397 (Print) VL - 26 IS - 6 SP - 1197 EP - 1213 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Oberländer, Jan A1 - Kirchner, Patrick A1 - Boyen, Hans-Gerd A1 - Schöning, Michael Josef T1 - Detection of hydrogen peroxide vapor by use of manganese(IV) oxide as catalyst for calorimetric gas sensors JF - Physica status solidi A: Applications and materials science N2 - In this work, the catalyst manganese(IV) oxide (MnO2), of calorimetric gas sensors (to monitor the sterilization agent vaporized hydrogen peroxide) has been investigated in more detail. Chemical analyses by means of X-ray-induced photoelectron spectroscopy have been performed to unravel the surface chemistry prior and after exposure to hydrogen peroxide vapor at elevated temperature, as applied in the sterilization processes of beverage cartons. The surface characterization reveals a change in oxidation states of the metal oxide catalyst after exposure to hydrogen peroxide. Additionally, a cleaning effect of the catalyst, which itself is attached to the sensor surface by means of a polymer interlayer, could be observed. Y1 - 2014 U6 - http://dx.doi.org/10.1002/pssa.201330359 SN - 1521-396X (E-Journal); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1372 EP - 1376 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Murib, Mohammed Sharif A1 - Yeap, Weng-Siang A1 - Martens, Daan A1 - Bienstman, Peter A1 - Ceuninck, Ward de A1 - Grinsven, Bart van A1 - Schöning, Michael Josef A1 - Michiels, Luc A1 - Haenen, Ken A1 - Ameloot, Marcel A1 - Serpengüzel, Ali A1 - Wagner, Patrick T1 - Photonic detection and characterization of DNA using sapphire microspheres JF - Journal of biomedical optics N2 - A microcavity-based deoxyribonucleic acid (DNA) optical biosensor is demonstrated for the first time using synthetic sapphire for the optical cavity. Transmitted and elastic scattering intensity at 1510 nm are analyzed from a sapphire microsphere (radius 500  μm, refractive index 1.77) on an optical fiber half coupler. The 0.43 nm angular mode spacing of the resonances correlates well with the optical size of the sapphire sphere. Probe DNA consisting of a 36-mer fragment was covalently immobilized on a sapphire microsphere and hybridized with a 29-mer target DNA. Whispering gallery modes (WGMs) were monitored before the sapphire was functionalized with DNA and after it was functionalized with single-stranded DNA (ssDNA) and double-stranded DNA (dsDNA). The shift in WGMs from the surface modification with DNA was measured and correlated well with the estimated thickness of the add-on DNA layer. It is shown that ssDNA is more uniformly oriented on the sapphire surface than dsDNA. In addition, it is shown that functionalization of the sapphire spherical surface with DNA does not affect the quality factor (Q≈104) of the sapphire microspheres. The use of sapphire is especially interesting because this material is chemically resilient, biocompatible, and widely used for medical implants. Y1 - 2014 U6 - http://dx.doi.org/10.1117/1.JBO.19.9.097006 SN - 1560-2281 (E-Journal); 1083-3668 (Print) VL - 19 IS - 9 SP - 097006 PB - SPIE CY - Bellingham ER - TY - JOUR A1 - Moseley, Fiona A1 - Halamek, Jan A1 - Kramer, Friederike A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Katz, Evgeny T1 - An enzyme-based reversible CNOT logic gate realized in a flow system JF - Analyst N2 - An enzyme system organized in a flow device was used to mimic a reversible Controlled NOT (CNOT) gate with two input and two output signals. Reversible conversion of NAD⁺ and NADH cofactors was used to perform a XOR logic operation, while biocatalytic hydrolysis of p-nitrophenyl phosphate resulted in an Identity operation working in parallel. The first biomolecular realization of a CNOT gate is promising for integration into complex biomolecular networks and future biosensor/biomedical applications. Y1 - 2014 U6 - http://dx.doi.org/10.1039/C4AN00133H SN - 1364-5528 (E-Journal) ; 0003-2654 (Print) VL - 139 IS - 8 SP - 1839 EP - 1842 PB - Royal Society of Chemistry CY - Cambridge ER -