TY - JOUR A1 - Lindner, Simon A1 - Burger, René A1 - Rutledge, Douglas N. A1 - Do, Xuan Tung A1 - Rumpf, Jessica A1 - Diehl, Bernd W. K. A1 - Schulze, Margit A1 - Monakhova, Yulia T1 - Is the calibration transfer of multivariate calibration models between high- and low-field NMR instruments possible? A case study of lignin molecular weight JF - Analytical chemistry N2 - Although several successful applications of benchtop nuclear magnetic resonance (NMR) spectroscopy in quantitative mixture analysis exist, the possibility of calibration transfer remains mostly unexplored, especially between high- and low-field NMR. This study investigates for the first time the calibration transfer of partial least squares regressions [weight average molecular weight (Mw) of lignin] between high-field (600 MHz) NMR and benchtop NMR devices (43 and 60 MHz). For the transfer, piecewise direct standardization, calibration transfer based on canonical correlation analysis, and transfer via the extreme learning machine auto-encoder method are employed. Despite the immense resolution difference between high-field and low-field NMR instruments, the results demonstrate that the calibration transfer from high- to low-field is feasible in the case of a physical property, namely, the molecular weight, achieving validation errors close to the original calibration (down to only 1.2 times higher root mean square errors). These results introduce new perspectives for applications of benchtop NMR, in which existing calibrations from expensive high-field instruments can be transferred to cheaper benchtop instruments to economize. Y1 - 2022 SN - 1520-6882 U6 - https://doi.org/10.1021/acs.analchem.1c05125 VL - 94 IS - 9 SP - 3997 EP - 4004 PB - ACS Publications CY - Washington, DC ER - TY - CHAP A1 - Gaigall, Daniel T1 - On Consistent Hypothesis Testing In General Hilbert Spaces T2 - Proceedings of the 4th International Conference on Statistics: Theory and Applications (ICSTA’22) N2 - Inference on the basis of high-dimensional and functional data are two topics which are discussed frequently in the current statistical literature. A possibility to include both topics in a single approach is working on a very general space for the underlying observations, such as a separable Hilbert space. We propose a general method for consistently hypothesis testing on the basis of random variables with values in separable Hilbert spaces. We avoid concerns with the curse of dimensionality due to a projection idea. We apply well-known test statistics from nonparametric inference to the projected data and integrate over all projections from a specific set and with respect to suitable probability measures. In contrast to classical methods, which are applicable for real-valued random variables or random vectors of dimensions lower than the sample size, the tests can be applied to random vectors of dimensions larger than the sample size or even to functional and high-dimensional data. In general, resampling procedures such as bootstrap or permutation are suitable to determine critical values. The idea can be extended to the case of incomplete observations. Moreover, we develop an efficient algorithm for implementing the method. Examples are given for testing goodness-of-fit in a one-sample situation in [1] or for testing marginal homogeneity on the basis of a paired sample in [2]. Here, the test statistics in use can be seen as generalizations of the well-known Cramérvon-Mises test statistics in the one-sample and two-samples case. The treatment of other testing problems is possible as well. By using the theory of U-statistics, for instance, asymptotic null distributions of the test statistics are obtained as the sample size tends to infinity. Standard continuity assumptions ensure the asymptotic exactness of the tests under the null hypothesis and that the tests detect any alternative in the limit. Simulation studies demonstrate size and power of the tests in the finite sample case, confirm the theoretical findings, and are used for the comparison with concurring procedures. A possible application of the general approach is inference for stock market returns, also in high data frequencies. In the field of empirical finance, statistical inference of stock market prices usually takes place on the basis of related log-returns as data. In the classical models for stock prices, i.e., the exponential Lévy model, Black-Scholes model, and Merton model, properties such as independence and stationarity of the increments ensure an independent and identically structure of the data. Specific trends during certain periods of the stock price processes can cause complications in this regard. In fact, our approach can compensate those effects by the treatment of the log-returns as random vectors or even as functional data. Y1 - 2022 U6 - https://doi.org/10.11159/icsta22.157 N1 - 4th International Conference on Statistics: Theory and Applications (ICSTA’22), Prague, Czech Republic – July 28- 30 SP - Paper No. 157 PB - Avestia Publishing CY - Orléans, Kanada ER - TY - JOUR A1 - Vahidpour, Farnoosh A1 - Alghazali, Yousef H. M. A1 - Akca, Sevilay A1 - Hommes, Gregor A1 - Schöning, Michael Josef T1 - An Enzyme-Based Interdigitated Electrode-Type Biosensor for Detecting Low Concentrations of H₂O₂ Vapor/Aerosol JF - Chemosensors N2 - This work introduces a novel method for the detection of H₂O₂ vapor/aerosol of low concentrations, which is mainly applied in the sterilization of equipment in medical industry. Interdigitated electrode (IDE) structures have been fabricated by means of microfabrication techniques. A differential setup of IDEs was prepared, containing an active sensor element (active IDE) and a passive sensor element (passive IDE), where the former was immobilized with an enzymatic membrane of horseradish peroxidase that is selective towards H₂O₂. Changes in the IDEs’ capacitance values (active sensor element versus passive sensor element) under H₂O₂ vapor/aerosol atmosphere proved the detection in the concentration range up to 630 ppm with a fast response time (<60 s). The influence of relative humidity was also tested with regard to the sensor signal, showing no cross-sensitivity. The repeatability assessment of the IDE biosensors confirmed their stable capacitive signal in eight subsequent cycles of exposure to H₂O₂ vapor/aerosol. Room-temperature detection of H₂O₂ vapor/aerosol with such miniaturized biosensors will allow a future three-dimensional, flexible mapping of aseptic chambers and help to evaluate sterilization assurance in medical industry. Y1 - 2022 U6 - https://doi.org/10.3390/chemosensors10060202 SN - 2227-9040 N1 - This article belongs to the Special Issue "Bioinspired Chemical Sensors and Micro-Nano Devices" VL - 10 IS - 6 PB - MDPI CY - Basel ER - TY - JOUR A1 - Pourshahidi, Ali Mohammad A1 - Engelmann, Ulrich M. A1 - Offenhäusser, Andreas A1 - Krause, Hans-Joachim T1 - Resolving ambiguities in core size determination of magnetic nanoparticles from magnetic frequency mixing data JF - Journal of Magnetism and Magnetic Materials N2 - Frequency mixing magnetic detection (FMMD) has been widely utilized as a measurement technique in magnetic immunoassays. It can also be used for the characterization and distinction (also known as “colourization”) of different types of magnetic nanoparticles (MNPs) based on their core sizes. In a previous work, it was shown that the large particles contribute most of the FMMD signal. This leads to ambiguities in core size determination from fitting since the contribution of the small-sized particles is almost undetectable among the strong responses from the large ones. In this work, we report on how this ambiguity can be overcome by modelling the signal intensity using the Langevin model in thermodynamic equilibrium including a lognormal core size distribution fL(dc,d0,σ) fitted to experimentally measured FMMD data of immobilized MNPs. For each given median diameter d0, an ambiguous amount of best-fitting pairs of parameters distribution width σ and number of particles Np with R2 > 0.99 are extracted. By determining the samples’ total iron mass, mFe, with inductively coupled plasma optical emission spectrometry (ICP-OES), we are then able to identify the one specific best-fitting pair (σ, Np) one uniquely. With this additional externally measured parameter, we resolved the ambiguity in core size distribution and determined the parameters (d0, σ, Np) directly from FMMD measurements, allowing precise MNPs sample characterization. Y1 - 2022 U6 - https://doi.org/10.1016/j.jmmm.2022.169969 SN - 0304-8853 VL - 563 IS - In progress, Art. No. 169969 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Monakhova, Yulia A1 - Diehl, Bernd W.K. T1 - Nuclear magnetic resonance spectroscopy as an elegant tool for a complete quality control of crude heparin material JF - Journal of Pharmaceutical and Biomedical Analysis N2 - Nuclear magnetic resonance (NMR) spectrometric methods for the quantitative analysis of pure heparin in crude heparin is proposed. For quantification, a two-step routine was developed using a USP heparin reference sample for calibration and benzoic acid as an internal standard. The method was successfully validated for its accuracy, reproducibility, and precision. The methodology was used to analyze 20 authentic porcine heparinoid samples having heparin content between 4.25 w/w % and 64.4 w/w %. The characterization of crude heparin products was further extended to a simultaneous analysis of these common ions: sodium, calcium, acetate and chloride. A significant, linear dependence was found between anticoagulant activity and assayed heparin content for thirteen heparinoids samples, for which reference data were available. A Diffused-ordered NMR experiment (DOSY) can be used for qualitative analysis of specific glycosaminoglycans (GAGs) in heparinoid matrices and, potentially, for quantitative prediction of molecular weight of GAGs. NMR spectrometry therefore represents a unique analytical method suitable for the simultaneous quantitative control of organic and inorganic composition of crude heparin samples (especially heparin content) as well as an estimation of other physical and quality parameters (molecular weight, animal origin and activity). KW - NMR spectroscopy KW - Heparin KW - Crude heparin KW - USP KW - Ions Y1 - 2022 U6 - https://doi.org/10.1016/j.jpba.2022.114915 SN - 0731-7085 VL - 219 IS - Article number: 114915 PB - Elsevier CY - New York, NY ER - TY - JOUR A1 - Burger, René A1 - Lindner, Simon A1 - Rumpf, Jessica A1 - Do, Xuan Tung A1 - Diehl, Bernd W.K. A1 - Rehahn, Matthias A1 - Monakhova, Yulia A1 - Schulze, Margit T1 - Benchtop versus high field NMR: Comparable performance found for the molecular weight determination of lignin JF - Journal of Pharmaceutical and Biomedical Analysis N2 - Lignin is a promising renewable biopolymer being investigated worldwide as an environmentally benign substitute of fossil-based aromatic compounds, e.g. for the use as an excipient with antioxidant and antimicrobial properties in drug delivery or even as active compound. For its successful implementation into process streams, a quick, easy, and reliable method is needed for its molecular weight determination. Here we present a method using 1H spectra of benchtop as well as conventional NMR systems in combination with multivariate data analysis, to determine lignin’s molecular weight (Mw and Mn) and polydispersity index (PDI). A set of 36 organosolv lignin samples (from Miscanthus x giganteus, Paulownia tomentosa and Silphium perfoliatum) was used for the calibration and cross validation, and 17 samples were used as external validation set. Validation errors between 5.6% and 12.9% were achieved for all parameters on all NMR devices (43, 60, 500 and 600 MHz). Surprisingly, no significant difference in the performance of the benchtop and high-field devices was found. This facilitates the application of this method for determining lignin’s molecular weight in an industrial environment because of the low maintenance expenditure, small footprint, ruggedness, and low cost of permanent magnet benchtop NMR systems. KW - NMR KW - PLS-regression KW - Molecular weight determination KW - Chemometrics KW - Biomass Y1 - 2022 SN - 0731-7085 U6 - https://doi.org/10.1016/j.jpba.2022.114649 VL - 212 IS - Article number: 114649 PB - Elsevier CY - New York, NY ER - TY - JOUR A1 - Monakhova, Yulia A1 - Soboleva, Polina M. A1 - Fedotova, Elena S. A1 - Musina, Kristina T. A1 - Burmistrova, Natalia A. T1 - Quantum chemical calculations of IR spectra of heparin disaccharide subunits JF - Computational and Theoretical Chemistry N2 - Heparin is a natural polysaccharide, which plays essential role in many biological processes. Alterations in building blocks can modify biological roles of commercial heparin products, due to significant changes in the conformation of the polymer chain. The variability structure of heparin leads to difficulty in quality control using different analytical methods, including infrared (IR) spectroscopy. In this paper molecular modelling of heparin disaccharide subunits was performed using quantum chemistry. The structural and spectral parameters of these disaccharides have been calculated using RHF/6-311G. In addition, over-sulphated chondroitin sulphate disaccharide was studied as one of the most widespread contaminants of heparin. Calculated IR spectra were analyzed with respect to specific structure parameters. IR spectroscopic fingerprint was found to be sensitive to substitution pattern of disaccharide subunits. Vibrational assignments of calculated spectra were correlated with experimental IR spectral bands of native heparin. Chemometrics was used to perform multivariate analysis of simulated spectral data. KW - IR spectroscopy KW - Chemometrics KW - Quantum chemistry KW - Molecular modelling KW - Quality control Y1 - 2022 SN - 2210-271X U6 - https://doi.org/10.1016/j.comptc.2022.113891 VL - 1217 IS - Article number: 113891 PB - Elsevier CY - New York, NY ER - TY - JOUR A1 - Haeger, Gerrit A1 - Bongaerts, Johannes A1 - Siegert, Petra T1 - A convenient ninhydrin assay in 96-well format for amino acid-releasing enzymes using an air-stable reagent JF - Analytical Biochemistry N2 - An improved and convenient ninhydrin assay for aminoacylase activity measurements was developed using the commercial EZ Nin™ reagent. Alternative reagents from literature were also evaluated and compared. The addition of DMSO to the reagent enhanced the solubility of Ruhemann's purple (RP). Furthermore, we found that the use of a basic, aqueous buffer enhances stability of RP. An acidic protocol for the quantification of lysine was developed by addition of glacial acetic acid. The assay allows for parallel processing in a 96-well format with measurements microtiter plates. Y1 - 2022 U6 - https://doi.org/10.1016/j.ab.2022.114819 SN - 1096-0309 IS - 624 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Langohr, Philipp A1 - Bung, Daniel Bernhard A1 - Crookston, Brian M. ED - Ortega-Sánchez, Miguel T1 - Hybrid investigation of labyrinth weirs: Discharge capacity and energy dissipation T2 - Proceedings of the 39th IAHR World Congress N2 - The replacement of existing spillway crests or gates with labyrinth weirs is a proven techno-economical means to increase the discharge capacity when rehabilitating existing structures. However, additional information is needed regarding energy dissipation of such weirs, since due to the folded weir crest, a three-dimensional flow field is generated, yielding more complex overflow and energy dissipation processes. In this study, CFD simulations of labyrinth weirs were conducted 1) to analyze the discharge coefficients for different discharges to compare the Cd values to literature data and 2) to analyze and improve energy dissipation downstream of the structure. All tests were performed for a structure at laboratory scale with a height of approx. P = 30.5 cm, a ratio of the total crest length to the total width of 4.7, a sidewall angle of 10° and a quarter-round weir crest shape. Tested headwater ratios were 0.089 ≤ HT/P ≤ 0.817. For numerical simulations, FLOW-3D Hydro was employed, solving the RANS equations with use of finite-volume method and RNG k-ε turbulence closure. In terms of discharge capacity, results were compared to data from physical model tests performed at the Utah Water Research Laboratory (Utah State University), emphasizing higher discharge coefficients from CFD than from the physical model. For upstream heads, some discrepancy in the range of ± 1 cm between literature, CFD and physical model tests was identified with a discussion regarding differences included in the manuscript. For downstream energy dissipation, variable tailwater depths were considered to analyze the formation and sweep-out of a hydraulic jump. It was found that even for high discharges, relatively low downstream Froude numbers were obtained due to high energy dissipation involved by the three-dimensional flow between the sidewalls. The effects of some additional energy dissipation devices, e.g. baffle blocks or end sills, were also analyzed. End sills were found to be non-effective. However, baffle blocks with different locations may improve energy dissipation downstream of labyrinth weirs. Y1 - 2022 SN - 978-90-832612-1-8 U6 - https://doi.org/10.3850/IAHR-39WC252171192022738 SN - 2521-7119 (print) SN - 2521-716X (online) N1 - 39th IAHR World Congress, 19. - 24. Juni 2022, Granada SP - 2313 EP - 2318 PB - International Association for Hydro-Environment Engineering and Research (IAHR) CY - Madrid ER - TY - CHAP A1 - Crookston, Brian M. A1 - Bung, Daniel Bernhard ED - Ortega-Sánchez, Miguel T1 - Application of RGB-D cameras in hydraulic laboratory studies T2 - Proceedings of the 39th IAHR World Congress N2 - Non-intrusive measuring techniques have attained a lot of interest in relation to both hydraulic modeling and prototype applications. Complimenting acoustic techniques, significant progress has been made for the development of new optical methods. Computer vision techniques can help to extract new information, e. g. high-resolution velocity and depth data, from videos captured with relatively inexpensive, consumer-grade cameras. Depth cameras are sensors providing information on the distance between the camera and observed features. Currently, sensors with different working principles are available. Stereoscopic systems reference physical image features (passive system) from two perspectives; in order to enhance the number of features and improve the results, a sensor may also estimate the disparity from a detected light to its original projection (active stereo system). In the current study, the RGB-D camera Intel RealSense D435, working on such stereo vision principle, is used in different, typical hydraulic modeling applications. All tests have been conducted at the Utah Water Research Laboratory. This paper will demonstrate the performance and limitations of the RGB-D sensor, installed as a single camera and as camera arrays, applied to 1) detect the free surface for highly turbulent, aerated hydraulic jumps, for free-falling jets and for an energy dissipation basin downstream of a labyrinth weir and 2) to monitor local scours upstream and downstream of a Piano Key Weir. It is intended to share the authors’ experiences with respect to camera settings, calibration, lightning conditions and other requirements in order to promote this useful, easily accessible device. Results will be compared to data from classical instrumentation and the literature. It will be shown that even in difficult application, e. g. the detection of a highly turbulent, fluctuating free-surface, the RGB-D sensor may yield similar accuracy as classical, intrusive probes. Y1 - 2022 SN - 978-90-832612-1-8 U6 - https://doi.org/10.3850/IAHR-39WC252171192022964 SN - 2521-7119 (print) SN - 2521-716X (online) N1 - 39th IAHR World Congress, 19. - 24. Juni 2022, Granada SP - 5127 EP - 5133 PB - International Association for Hydro-Environment Engineering and Research (IAHR) CY - Madrid ER -