TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Schmidt, Simon A1 - Stelter, Jonathan K. A1 - Wittrich, Marco A1 - Quick, Harald H. A1 - Bitz, Andreas A1 - Ladd, Mark E. T1 - Performance and safety assessment of an integrated transmit array for body imaging at 7 T under consideration of specificabsorption rate, tissue temperature, and thermal dose JF - NMR in Biomedicine N2 - In this study, the performance of an integrated body-imaging array for 7 T with 32 radiofrequency (RF) channels under consideration of local specific absorption rate (SAR), tissue temperature, and thermal dose limits was evaluated and the imaging performance was compared with a clinical 3 T body coil. Thirty-two transmit elements were placed in three rings between the bore liner and RF shield of the gradient coil. Slice-selective RF pulse optimizations for B1 shimming and spokes were performed for differently oriented slices in the body under consideration of realistic constraints for power and local SAR. To improve the B1+ homogeneity, safety assessments based on temperature and thermal dose were performed to possibly allow for higher input power for the pulse optimization than permissible with SAR limits. The results showed that using two spokes, the 7 T array outperformed the 3 T birdcage in all the considered regions of interest. However, a significantly higher SAR or lower duty cycle at 7 T is necessary in some cases to achieve similar B1+ homogeneity as at 3 T. The homogeneity in up to 50 cm-long coronal slices can particularly benefit from the high RF shim performance provided by the 32 RF channels. The thermal dose approach increases the allowable input power and the corresponding local SAR, in one example up to 100 W/kg, without limiting the exposure time necessary for an MR examination. In conclusion, the integrated antenna array at 7 T enables a clinical workflow for body imaging and comparable imaging performance to a conventional 3 T clinical body coil. KW - body imaging at 7 T MRI KW - thermal dose KW - tissue temperature KW - transmit antenna arrays Y1 - 2022 U6 - https://doi.org/10.1002/nbm.4656 SN - 0952-3480 (Print) SN - 1099-1492 (Online) VL - 35 IS - 5 SP - 1 EP - 17 PB - Wiley ER - TY - JOUR A1 - Schulte-Tigges, Joschua A1 - Förster, Marco A1 - Nikolovski, Gjorgji A1 - Reke, Michael A1 - Ferrein, Alexander A1 - Kaszner, Daniel A1 - Matheis, Dominik A1 - Walter, Thomas T1 - Benchmarking of various LiDAR sensors for use in self-driving vehicles in real-world environments JF - Sensors N2 - Abstract In this paper, we report on our benchmark results of the LiDAR sensors Livox Horizon, Robosense M1, Blickfeld Cube, Blickfeld Cube Range, Velodyne Velarray H800, and Innoviz Pro. The idea was to test the sensors in different typical scenarios that were defined with real-world use cases in mind, in order to find a sensor that meet the requirements of self-driving vehicles. For this, we defined static and dynamic benchmark scenarios. In the static scenarios, both LiDAR and the detection target do not move during the measurement. In dynamic scenarios, the LiDAR sensor was mounted on the vehicle which was driving toward the detection target. We tested all mentioned LiDAR sensors in both scenarios, show the results regarding the detection accuracy of the targets, and discuss their usefulness for deployment in self-driving cars. KW - Lidar KW - Benchmark KW - Self-driving Y1 - 2022 U6 - https://doi.org/10.3390/s22197146 SN - 1424-8220 N1 - This article belongs to the Special Issue "Sensor Fusion for Vehicles Navigation and Robotic Systems" VL - 22 IS - 19 PB - MDPI CY - Basel ER - TY - JOUR A1 - Hoffmann, Andreas A1 - Rohrbach, Felix A1 - Uhl, Matthias A1 - Ceblin, Maximilian A1 - Bauer, Thomas A1 - Mallah, Marcel A1 - Jacob, Timo A1 - Heuermann, Holger A1 - Kuehne, Alexander J. C. T1 - Atmospheric pressure plasma-jet treatment of polyacrylonitrile-nonwovens—Stabilization and roll-to-roll processing JF - Journal of Applied Polymer Science N2 - Carbon nanofiber nonwovens represent a powerful class of materials with prospective application in filtration technology or as electrodes with high surface area in batteries, fuel cells, and supercapacitors. While new precursor-to-carbon conversion processes have been explored to overcome productivity restrictions for carbon fiber tows, alternatives for the two-step thermal conversion of polyacrylonitrile precursors into carbon fiber nonwovens are absent. In this work, we develop a continuous roll-to-roll stabilization process using an atmospheric pressure microwave plasma jet. We explore the influence of various plasma-jet parameters on the morphology of the nonwoven and compare the stabilized nonwoven to thermally stabilized samples using scanning electron microscopy, differential scanning calorimetry, and infrared spectroscopy. We show that stabilization with a non-equilibrium plasma-jet can be twice as productive as the conventional thermal stabilization in a convection furnace, while producing electrodes of comparable electrochemical performance. KW - batteries and fuel cells KW - electrospinning KW - fibers KW - irradiation KW - porous materials Y1 - 2022 U6 - https://doi.org/10.1002/app.52887 SN - 0021-8995 (Print) SN - 1097-4628 (Online) N1 - Weitere Informationen: Bundesministerium für Bildung und Forschung, Fördernummer: 13XP5036E. Deutsche Forschungsgemeinschaft, Fördernummern: 390874152, 441209207, 327886311 VL - 139 IS - 37 SP - 1 EP - 9 PB - Wiley ER - TY - JOUR A1 - Elsen, Ingo A1 - Kraiss, Karl-Friedrich T1 - System concept and realization of a scalable neurocomputing architecture JF - Systems Analysis Modelling Simulation N2 - This paper describes the realization of a novel neurocomputer which is based on the concepts of a coprocessor. In contrast to existing neurocomputers the main interest was the realization of a scalable, flexible system, which is capable of computing neural networks of arbitrary topology and scale, with full independence of special hardware from the software's point of view. On the other hand, computational power should be added, whenever needed and flexibly adapted to the requirements of the application. Hardware independence is achieved by a run time system which is capable of using all available computing power, including multiple host CPUs and an arbitrary number of neural coprocessors autonomously. The realization of arbitrary neural topologies is provided through the implementation of the elementary operations which can be found in most neural topologies. Y1 - 1999 SN - 0232-9298 SN - 1029-4902 VL - 35 IS - 4 SP - 399 EP - 419 PB - Gordon and Breach Science Publishers CY - Amsterdam ER - TY - JOUR A1 - Elsen, Ingo A1 - Kraiss, Karl-Friedrich A1 - Krumbiegel, Dirk A1 - Walter, Peter A1 - Wickel, Jochen T1 - Visual information retrieval for 3D product identification: a midterm report JF - KI - Künstliche Intelligenz Y1 - 1999 SN - 1610-1987 SN - 0933-1875 VL - 13 IS - 1 SP - 64 EP - 67 PB - Springer CY - Berlin ER - TY - JOUR A1 - Elsen, Ingo A1 - Hartung, Frank A1 - Horn, Uwe A1 - Kampmann, Markus A1 - Peters, Liliane ED - Voas, Jeffrey T1 - Streaming technology in 3G mobile communication systems JF - Computer : innovative technology for computer professionals N2 - Third-generation mobile communication systems will combine standardized streaming with a range of unique services to provide high-quality Internet content that meets the specific needs of the rapidly growing mobile market. Y1 - 2001 SN - 0018-9162 SN - 1558-0814 VL - 34 IS - 9 Seiten SP - 46 EP - 52 PB - IEEE CY - New York ER - TY - JOUR A1 - Kowalewski, Paul A1 - Bragard, Michael A1 - Hüning, Felix A1 - De Doncker, Rik W. T1 - An inexpensive Wiegand-sensor-based rotary encoder without rotating magnets for use in electrical drives JF - IEEE Transactions on Instrumentation and Measurement N2 - This paper introduces an inexpensive Wiegand-sensor-based rotary encoder that avoids rotating magnets and is suitable for electrical-drive applications. So far, Wiegand-sensor-based encoders usually include a magnetic pole wheel with rotating permanent magnets. These encoders combine the disadvantages of an increased magnet demand and a limited maximal speed due to the centripetal force acting on the rotating magnets. The proposed approach reduces the total demand of permanent magnets drastically. Moreover, the rotating part is manufacturable from a single piece of steel, which makes it very robust and cheap. This work presents the theoretical operating principle of the proposed approach and validates its benefits on a hardware prototype. The presented proof-of-concept prototype achieves a mechanical resolution of 4.5 ° by using only 4 permanent magnets, 2Wiegand sensors and a rotating steel gear wheel with 20 teeth. KW - Rotary encoder KW - Wiegand sensor Y1 - 2023 U6 - https://doi.org/10.1109/TIM.2023.3326166 SN - 0018-9456 (Print) SN - 1557-9662 (Online) N1 - Early Access SP - 10 Seiten PB - IEEE ER - TY - JOUR A1 - Coll-Perales, Baldomero A1 - Schulte-Tigges, Joschua A1 - Rondinone, Michele A1 - Gozalvez, Javier A1 - Reke, Michael A1 - Matheis, Dominik A1 - Walter, Thomas T1 - Prototyping and evaluation of infrastructure-assisted transition of control for cooperative automated vehicles JF - IEEE Transactions on Intelligent Transportation Systems N2 - Automated driving is now possible in diverse road and traffic conditions. However, there are still situations that automated vehicles cannot handle safely and efficiently. In this case, a Transition of Control (ToC) is necessary so that the driver takes control of the driving. Executing a ToC requires the driver to get full situation awareness of the driving environment. If the driver fails to get back the control in a limited time, a Minimum Risk Maneuver (MRM) is executed to bring the vehicle into a safe state (e.g., decelerating to full stop). The execution of ToCs requires some time and can cause traffic disruption and safety risks that increase if several vehicles execute ToCs/MRMs at similar times and in the same area. This study proposes to use novel C-ITS traffic management measures where the infrastructure exploits V2X communications to assist Connected and Automated Vehicles (CAVs) in the execution of ToCs. The infrastructure can suggest a spatial distribution of ToCs, and inform vehicles of the locations where they could execute a safe stop in case of MRM. This paper reports the first field operational tests that validate the feasibility and quantify the benefits of the proposed infrastructure-assisted ToC and MRM management. The paper also presents the CAV and roadside infrastructure prototypes implemented and used in the trials. The conducted field trials demonstrate that infrastructure-assisted traffic management solutions can reduce safety risks and traffic disruptions. KW - Automated driving KW - automated vehicles KW - connected automated vehicles KW - CAV KW - experimental evaluation Y1 - 2021 U6 - https://doi.org/10.1109/TITS.2021.3061085 SN - 1524-9050 (Print) SN - 1558-0016 (Online) VL - 23 IS - 7 SP - 6720 EP - 6736 PB - IEEE ER - TY - JOUR A1 - Gräßl, Andreas A1 - Renz, Wolfgang A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Winter, Lukas A1 - Özerdem, Celal A1 - Rieger, Jan A1 - Kellmann, Peter A1 - Santoro, Davide A1 - Lindel, Tomasz D. A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Niendorf, Thoralf T1 - Modular 32-channel transceiver coil array for cardiac MRI at 7.0T JF - Magnetic Resonance in Medicine N2 - Purpose To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. Methods The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Results Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. Conclusion The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T. Y1 - 2013 U6 - https://doi.org/10.1002/mrm.24903 SN - 1522-2594 VL - 72 IS - 1 SP - 276 EP - 290 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Fuchs, Katharina A1 - Dieringer, Matthias A. A1 - Özerdem, Celal A1 - Patel, Nishan A1 - Renz, Wolfgang A1 - Greiser, Andreas A1 - Elgeti, Thomas A1 - Niendorf, Thoralf T1 - Detailing the use of magnetohydrodynamic effects for synchronization of MRI with the cardiac cycle: A feasibility study JF - Journal of Magnetic Resonance Imaging N2 - Purpose: To investigate the feasibility of using magnetohydrodynamic (MHD) effects for synchronization of magnetic resonance imaging (MRI) with the cardiac cycle. Materials and Methods: The MHD effect was scrutinized using a pulsatile flow phantom at B0 = 7.0 T. MHD effects were examined in vivo in healthy volunteers (n = 10) for B0 ranging from 0.05–7.0 T. Noncontrast-enhanced MR angiography (MRA) of the carotids was performed using a gated steady-state free-precession (SSFP) imaging technique in conjunction with electrocardiogram (ECG) and MHD synchronization. Results: The MHD potential correlates with flow velocities derived from phase contrast MRI. MHD voltages depend on the orientation between B0 and the flow of a conductive fluid. An increase in the interelectrode spacing along the flow increases the MHD potential. In vivo measurement of the MHD effect provides peak voltages of 1.5 mV for surface areas close to the common carotid artery at B0 = 7.0 T. Synchronization of MRI with the cardiac cycle using MHD triggering is feasible. MHD triggered MRA of the carotids at 3.0 T showed an overall image quality and richness of anatomic detail, which is comparable to ECG-triggered MRAs. Conclusion: This feasibility study demonstrates the use of MHD effects for synchronization of MR acquisitions with the cardiac cycle. J. Magn. Reson. Imaging 2012;36:364–372. © 2012 Wiley Periodicals, Inc. Y1 - 2012 U6 - https://doi.org/10.1002/jmri.23634 SN - 1522-2586 VL - 36 IS - 2 SP - 364 EP - 372 PB - Wiley-Liss CY - New York ER -