TY - JOUR A1 - Reisert, Steffen A1 - Henkel, Hartmut A1 - Schneider, Andreas A1 - Schäfer, Daniel A1 - Friedrich, Peter A1 - Berger, Jörg A1 - Schöning, Michael Josef T1 - Development of a handheld sensor system for the online measurement of hydrogen peroxide in aseptic filling systems JF - Physica Status Solidi (A) N2 - A handheld sensor system for the online measurement of hydrogen peroxide (H2O2) in aseptic sterilisation processes has been developed. It is based on a calorimetric-type gas sensor that consists of a differential set-up of two temperature sensors, of which one is catalytically activated and the second one is passivated and used as reference. The sensor principle relies in detecting a rise in temperature on the active sensor due to the exothermic reaction of H2O2 on the catalytic surface. To characterise the sensor system towards H2O2 sensitivity and other influencing factors, measurements have been carried out both at an experimental set-up and a manufacturer's sterilisation machine. Physical sensor characterisation was done by means of the optical microscopy. Y1 - 2010 U6 - https://doi.org/10.1002/pssa.200983304 SN - 1862-6300 N1 - Special Issue: Engineering of Functional Interfaces EnFI 2009 VL - 207 IS - 4 SP - 913 EP - 918 PB - Wiley-VCH CY - Berlin ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Sugawara, Yuri A1 - Kanoh, Shin´ichiro A1 - Yoshinobu, Tatsuo A1 - Wagner, Torsten A1 - Schöning, Michael Josef T1 - Image correction method for the chemical imaging sensor JF - Sensors and Actuators B: Chemical N2 - The chemical imaging sensor is a semiconductor-based chemical sensor that can visualize the spatial distribution of chemical species. For the practical application of this sensor, artifacts in the chemical images due to defects of the semiconductor substrate and contamination of the sensing surface etc. have been a major problem. An image correction method was developed to eliminate the influence of nonuniformity of individual sensor plate. Y1 - 2010 U6 - https://doi.org/10.1016/j.snb.2008.10.069 SN - 0925-4005 N1 - 22nd International Conference on Eurosensors: Dresden, Germany, 7-10 September 2008 VL - 144 IS - 2 SP - 344 EP - 348 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Abouzar, Maryam H. A1 - Poghossian, Arshak A1 - Siqueira, José R. Jr. A1 - Oliveira, Osvaldo N. Jr. A1 - Moritz, Werner A1 - Schöning, Michael Josef T1 - Capacitive electrolyte-insulator-semiconductor structures functionalised with a polyelectrolyte/enzyme multilayer: New strategy for enhanced field-effect biosensing JF - Physica Status Solidi (A) N2 - A novel strategy for enhanced field-effect biosensing using capacitive electrolyte–insulator–semiconductor (EIS) structures functionalised with pH-responsive weak polyelectrolyte/enzyme or dendrimer/enzyme multilayers is presented. The feasibility of the proposed approach is exemplarily demonstrated by realising a penicillin biosensor based on a capacitive p-Si–SiO2 EIS structure functionalised with a poly(allylamine hydrochloride) (PAH)/penicillinase and a poly(amidoamine) dendrimer/penicillinase multilayer. The developed sensors response to changes in both the local pH value near the gate surface and the charge of macromolecules induced via enzymatic reaction, resulting in a higher sensitivity. For comparison, an EIS penicillin biosensor with adsorptively immobilised penicillinase has been also studied. The highest penicillin sensitivity of 100 mV/dec has been observed for the EIS sensor functionalised with the PAH/penicillinase multilayer. The lower and upper detection limit was around 20 µM and 10 mM, respectively. In addition, an incorporation of enzymes in a multilayer prepared by layer-by-layer technique provides a larger amount of immobilised enzymes per sensor area, reduces enzyme leaching effects and thus, enhances the biosensor lifetime (the loss of penicillin sensitivity after 2 months was 10–12%). Y1 - 2010 U6 - https://doi.org/10.1002/pssa.200983317 SN - 1862-6300 N1 - Special Issue: Engineering of Functional Interfaces EnFI 2009 VL - 207 IS - 4 SP - 884 EP - 890 PB - Wiley-VCH CY - Berlin ER - TY - JOUR A1 - Schöning, Michael Josef A1 - Tsarouchas, D. A1 - Beckers, Leah A1 - Schubert, J. A1 - Zander, W. A1 - Kordoš, P. A1 - Lüth, Hans T1 - A highly long-term stable silicon-based pH sensor using pulsed laser deposition technique JF - Sensors and Actuators B. 35 (1996), H. 1-3 Y1 - 1996 SN - 0925-4005 SP - 228 EP - 233 ER - TY - JOUR A1 - Poghossian, Arshak A1 - Jablonski, Melanie A1 - Koch, Claudia A1 - Bronder, Thomas A1 - Rolka, David A1 - Wege, Christina A1 - Schöning, Michael Josef T1 - Field-effect biosensor using virus particles as scaffolds for enzyme immobilization JF - Biosensors and Bioelectronics N2 - A field-effect biosensor employing tobacco mosaic virus (TMV) particles as scaffolds for enzyme immobilization is presented. Nanotubular TMV scaffolds allow a dense immobilization of precisely positioned enzymes with retained activity. To demonstrate feasibility of this new strategy, a penicillin sensor has been developed by coupling a penicillinase with virus particles as a model system. The developed field-effect penicillin biosensor consists of an Al-p-Si-SiO₂-Ta₂O₅-TMV structure and has been electrochemically characterized in buffer solutions containing different concentrations of penicillin G. In addition, the morphology of the biosensor surface with virus particles was characterized by scanning electron microscopy and atomic force microscopy methods. The sensors possessed a high penicillin sensitivity of ~ 92 mV/dec in a nearly-linear range from 0.1 mM to 10 mM, and a low detection limit of about 50 µM. The long-term stability of the penicillin biosensor was periodically tested over a time period of about one year without any significant loss of sensitivity. The biosensor has also been successfully applied for penicillin detection in bovine milk samples. Y1 - 2018 U6 - https://doi.org/10.1016/j.bios.2018.03.036 SN - 0956-5663 VL - 110 SP - 168 EP - 174 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Schöning, Michael Josef A1 - Kirchner, Patrick A1 - Ng, Yue Ann A1 - Spelthahn, Heiko A1 - Schneider, Andreas A1 - Henkel, Hartmut A1 - Friedrich, Peter A1 - Kolstad, Jens A1 - Berger, Jörg A1 - Keusgen, Michael T1 - Gas sensor investigation based on a catalytically activated thin-film thermopile for H2O2 detection JF - Physica Status Solidi (A) N2 - In aseptic filling systems, hydrogen peroxide vapour is commonly used for the reduction of microbial contaminations in carton packages. In this process, the germicidal efficiency of the vapour depends especially on the H₂O₂ concentration. To monitor the H₂O₂ concentration, a calorimetric H₂O₂ gas sensor based on a catalytically activated thin-film thermopile is investigated. Two different sensor layouts, namely a circular and a linear form, as well as two various material pairs such as tungsten/nickel and gold/nickel, have been examined for the realization of a thin-film thermopile. Additionally, manganese oxide and palladium particles have been compared as responsive catalysts towards H₂O₂. The thin-film sensors have been investigated at various H₂O₂ concentrations, gas temperatures and flow rates. Y1 - 2010 U6 - https://doi.org/10.1002/pssa.200983309 SN - 1862-6300 N1 - Special Issue: Engineering of Functional Interfaces EnFI 2009 VL - 207 IS - 4 SP - 787 EP - 792 PB - Wiley-VCH CY - Berlin ER - TY - JOUR A1 - Huck, Christina A1 - Poghossian, Arshak A1 - Bäcker, Matthias A1 - Chaudhuri, S. A1 - Zander, W. A1 - Schubert, J. A1 - Begoyan, Vardges K. A1 - Buniatyan, V. V. A1 - Wagner, P. A1 - Schöning, Michael Josef T1 - Capacitively coupled electrolyte-conductivity sensor based on high-k material of barium strontium titanate JF - Sensors and actuators. B: Chemical Y1 - 2014 U6 - https://doi.org/10.1016/j.snb.2014.02.103 SN - 1873-3077 (E-Journal); 0925-4005 (Print) IS - 198 SP - 102 EP - 109 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Buniatyan, Vahe V. A1 - Abouzar, Maryam H. A1 - Martirosyan, Norayr W. A1 - Schubert, Jürgen A1 - Gevorgian, Spartak A1 - Schöning, Michael Josef A1 - Poghossian, Arshak T1 - pH-sensitive properties of barium strontium titanate (BST) thin films prepared by pulsed laser deposition technique JF - Physica Status Solidi (A) N2 - pH-sensitive properties of barium strontium titanate (BST) high-k thin films as alternative gate material for field-effect capacitive (bio-)chemical sensors based on an electrolyte-insulator-semiconductor system have been investigated. The BST films of different compositions (Ba0.31Sr0.69TiO3, Ba0.25Sr0.75TiO3 and Mg-doped Ba0.8Sr0.2Mg0.1Ti0.9O3) were deposited by pulsed laser deposition technique from targets fabricated by self-propagating high-temperature synthesis. The realised sensors have been electrochemically characterised by means of impedance-spectroscopy, capacitance–voltage and constant-capacitance method. The sensors possess a Nernstian-like pH sensitivity in the concentration range between pH 3 and 11 with a response time of 5–10 s. An equivalent circuit model for the BST-based capacitive field-effect sensor is discussed. Y1 - 2010 U6 - https://doi.org/10.1002/pssa.200983310 SN - 1862-6300 N1 - Special Issue: Engineering of Functional Interfaces EnFI 2009 VL - 207 IS - 4 SP - 824 EP - 830 PB - Wiley-VCH CY - Berlin ER - TY - JOUR A1 - Turek, Monik A1 - Heiden, Wolfgang A1 - Guo, Sharon A1 - Riesen, Alfred A1 - Schubert, Jürgen A1 - Zander, Willi A1 - Krüger, Peter A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Simultaneous detection of cyanide and heavy metals for environmental analysis by means of µISEs JF - Physica Status Solidi (A) N2 - In environmental analysis, cyanide and heavy metals play an important role, because these substances are highly toxic for biological systems. They can lead to chronic and acute diseases. Due to the chemical properties of cyanide it is frequently used for industrial processes such as extraction of silver and gold. Heavy metals can be found as trace elements in nature and are often applied in industries e.g., galvanization processes. Up to now, cyanide and heavy metals can be detected by several sensors separately and their detection is often limited to laboratory investigations. In this publication, with regard to an in situ analysis, a new miniaturized silicon-based sensor system for the simultaneous detection of cyanide and heavy metals in aqueous solutions is presented that is based on chalcogenide glass-based micro ion-selective electrodes (µISEs). The µISEs are incorporated into a specially designed measuring system for the simultaneous detection of heavy metals and cyanide in solutions and validated by simultaneous measurements of Cu2+- and CN−-ions, Cd2+- and CN−- ions and Pb2+- and CN−-ions. The particular sensor system has shown good sensor properties in the µ-molar ion-concentration range. For simultaneous measurements in complex heavy metal and cyanide solutions an intelligent software using fuzzy logic is discussed. Y1 - 2010 U6 - https://doi.org/10.1002/pssa.200983303 SN - 1862-6300 N1 - Special Issue: Engineering of Functional Interfaces EnFI 2009 VL - 207 IS - 4 SP - 817 EP - 823 PB - Wiley-VCH CY - Berlin ER - TY - JOUR A1 - Jansen, S. V. A1 - Behbahani, Mehdi A1 - Laumen, M. A1 - Kaufmann, T. A1 - Hormes, M. A1 - Behr, Marek A1 - Schmitz-Rode, T. A1 - Steinseifer, U. T1 - Investigation of Steady Flow Through a Realistic Model of the Thoracic Human Aorta Using 3D Stereo PIV and CFD-Simulation JF - ASAIO Journal Y1 - 2010 U6 - https://doi.org/10.1097/01.mat.0000369377.65122.a3 N1 - American Society of Artificial Organs (ASAIO), Baltimore, USA, May 27-29, 2010 VL - 56 IS - 2 SP - 98 PB - Lippincott Williams & Wilkins CY - Philadelphia ER -