TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Brökel, M. A1 - Fuss, L. A1 - Larue, H. A1 - Parl, C. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - The ClearPET data acquisition JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 5 N2 - Within the Crystal Clear Collaboration a modular system for a small animal PET scanner (ClearPET™) has been developed. The modularity allows the assembly of scanners of different sizes and characteristics in order to fit the specific needs of the individual member institutions. Now a first demonstrator is being completed in Julich. The system performs depth of interaction detection by using a phoswich arrangement combining LSO and LuYAP scintillators which are coupled to multi-channel photomultipliers (PMTs). A free-running ADC digitizes the signal from the PMT and the complete scintillation pulses are sampled by an FPGA and sent with 20 MB/S to a PC for preprocessing. The pulse provides information about the gamma energy and the scintillator material which identifies the interaction layer. Furthermore, the exact pulse starting time is obtained from the sampled data. This is important as no hardware coincidence detection is implemented. All single events are recorded and coincidences are identified by software. An advantage of that is that the coincidence window and the dimensions of the field of view can be adjusted easily. The ClearPET™ demonstrator is equipped with 10240 crystals on 80 PMTs. This paper presents an overview of the data acquisition system. Y1 - 2004 SN - 1082-3654 SP - 3097 EP - 3100 ER - TY - JOUR A1 - Ziemons, Karl A1 - Auffray, E. A1 - Barbier, R. A1 - Brandenburg, G. T1 - The ClearPET TM LSO/LuYAP phoswich scanner: a high performance small animal PET system JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 3 N2 - A 2nd generation high performance small animal PET scanner, called ClearPET™, has been designed and a first prototype is built by working groups of the Crystal Clear Collaboration (CCC). In order to achieve high sensitivity and maintain good uniform spatial resolution over the field of view in high resolution PET systems, it is necessary to extract the depth of interaction (DOI) information and correct for spatial degradation. The design of the first ClearPET™ Demonstrator based on the use of the multi-anode photomultiplier tube (Hamamatsu R7600-M64) and a LSO/LuYAP phoswich matrix. The two crystal layers of 8*8 crystals (2*2*10 mm3) are stacked on each other and mounted without light guide as one to one on the PMT. A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the crystal ring is 137 mm, the axial detector length is 110 mm. The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the pulse arrival time. Single gamma interactions are recorded and coincidences are found by software. The gantry allows rotation of the detector modules around the field of view. The measurements have been done using the first LSO/LuYAP detector cassettes. Y1 - 2004 SN - 1082-3654 SP - 1728 EP - 1732 ER -