TY - JOUR A1 - Wert, Stefan A1 - Iken, Heiko A1 - Schöning, Michael Josef A1 - Matysik, Frank-Michael T1 - Development of a temperature‐pulse enhanced electrochemical glucose biosensor and characterization of its stability via scanning electrochemical microscopy JF - Electroanalysis N2 - Glucose oxidase (GOx) is an enzyme frequently used in glucose biosensors. As increased temperatures can enhance the performance of electrochemical sensors, we investigated the impact of temperature pulses on GOx that was drop-coated on flattened Pt microwires. The wires were heated by an alternating current. The sensitivity towards glucose and the temperature stability of GOx was investigated by amperometry. An up to 22-fold increase of sensitivity was observed. Spatially resolved enzyme activity changes were investigated via scanning electrochemical microscopy. The application of short (<100 ms) heat pulses was associated with less thermal inactivation of the immobilized GOx than long-term heating. Y1 - 2021 U6 - http://dx.doi.org/10.1002/elan.202100089 SN - 1521-4109 IS - Early View PB - Wiley-VCH CY - Weinheim ER - TY - CHAP A1 - Werner, Frederik A1 - Yoshinobu, T. A1 - Miyamoto, K. A1 - Schöning, Michael Josef A1 - Wagner, Torsten T1 - Semiconductor-based sensors for imaging of chemical processes T2 - Sensoren und Messsysteme 2014 ; Beiträge der 17. GMA/ITG-Fachtagung vom 3. bis 4. Juni 2014 in Nürnberg. (ITG-Fachbericht ; 250) Y1 - 2014 SN - 978-3-8007-3622-5 SP - 1 EP - 5 PB - VDE-Verl. CY - Düsseldorf ER - TY - JOUR A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Frequency behaviour of light-addressable potentiometric sensors JF - Physica Status Solidi (A) N2 - Light-addressable potentiometric sensors (LAPS) are semiconductor-based potentiometric sensors, with the advantage to detect the concentration of a chemical species in a liquid solution above the sensor surface in a spatially resolved manner. The addressing is achieved by a modulated and focused light source illuminating the semiconductor and generating a concentration-depending photocurrent. This work introduces a LAPS set-up that is able to monitor the electrical impedance in addition to the photocurrent. The impedance spectra of a LAPS structure, with and without illumination, as well as the frequency behaviour of the LAPS measurement are investigated. The measurements are supported by electrical equivalent circuits to explain the impedance and the LAPS-frequency behaviour. The work investigates the influence of different parameters on the frequency behaviour of the LAPS. Furthermore, the phase shift of the photocurrent, the influence of the surface potential as well as the changes of the sensor impedance will be discussed. Y1 - 2013 U6 - http://dx.doi.org/10.1002/pssa.201200929 SN - 1521-396X ; 0031-8965 VL - 210 IS - 5 SP - 884 EP - 891 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Miyamoto, Ko-ichiro A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - High speed and high resolution chemical imaging based on a new type of OLED-LAPS set-up JF - Sensors and Actuators B: Chemical N2 - Light-addressable potentiometric sensors (LAPS) are field-effect-based sensors. A modulated light source is used to define the particular measurement spot to perform spatially resolved measurements of chemical species and to generate chemical images. In this work, an organic-LED (OLED) display has been chosen as a light source. This allows high measurement resolution and miniaturisation of the system. A new developed driving method for the OLED display optimised for LAPS-based measurements is demonstrated. The new method enables to define modulation frequencies between 1 kHz and 16 kHz and hence, reduces the measurement time of a chemical image by a factor of 40 compared to the traditional addressing of an OLED display. Y1 - 2012 U6 - http://dx.doi.org/10.1016/j.snb.2011.12.102 SN - 0925-4005 N1 - Part of special issue "Selected Papers presented at Eurosensors XXV" VL - 175 SP - 118 EP - 122 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Werner, Frederik A1 - Wagner, Torsten A1 - Miyamoto, Ko-ichiro A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - High speed and high resolution chemical imaging based on a new type of OLED-LAPS set-up JF - Procedia Engineering. 25 (2011) Y1 - 2011 SN - 1877-7058 N1 - EurosensorsXXV ; Proc. Eurosensors XXV, September 4-7, 2011, Athens, Greece SP - 346 EP - 349 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Werner, Frederik A1 - Takenaga, Shoko A1 - Taki, Hidenori A1 - Sawada, Kazuaki A1 - Schöning, Michael Josef T1 - Comparison of label-free ACh-imaging sensors based on CCD and LAPS JF - Sensors and Actuators B: Chemical (2012) N2 - Semiconductor-based chemical imaging sensors, like the light-addressable potentiometric sensor (LAPS) or the pH-imaging sensor based on a charge-coupled device (CCD), are becoming a powerful tool for label-free imaging of biological phenomena. We have proposed a polyion-based enzymatic membrane to develop an acetylcholine (ACh) imaging sensor for neural cell-activity observations. In this study, a CCD-type ACh-imaging sensor and a LAPS-type ACh-imaging sensor were fabricated and the prospect of both sensors was clarified by making a comparison of their basic characteristics. Y1 - 2013 SN - 0925-4005 VL - 177 SP - 745 EP - 752 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Werner, Frederik A1 - Schusser, Sebastian A1 - Spalthahn, Heiko A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Schöning, Michael Josef T1 - Field-programmable gate array based controller for multi spot light-addressable potentiometric sensors with integrated signal correction mode JF - Electrochimica Acta N2 - A light-addressable potentiometric sensor (LAPS) can measure the concentration of one or several analytes at the sensor surface simultaneously in a spatially resolved manner. A modulated light pointer stimulates the semiconductor structure at the area of interest and a responding photocurrent can be read out. By simultaneous stimulation of several areas with light pointers of different modulation frequencies, the read out can be performed at the same time. With the new proposed controller electronic based on a field-programmable gate array (FPGA), it is possible to control the modulation frequencies, phase shifts, and light brightness of multiple light pointers independently and simultaneously. Thus, it is possible to investigate the frequency response of the sensor, and to examine the analyte concentration by the determination of the surface potential with the help of current/voltage curves and phase/voltage curves. Additionally, the ability to individually change the light intensities of each light pointer is used to perform signal correction. Y1 - 2011 U6 - http://dx.doi.org/10.1016/j.electacta.2011.03.012 SN - 0013-4686 VL - 56 IS - 26 SP - 9656 EP - 9660 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Werner, Frederik A1 - Miyamoto, Ko-ichiro A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Lateral resolution enhancement of pulse-driven light-addressable potentiometric sensor JF - Sensor and Actuators B: Chemical N2 - To study chemical and biological processes, spatially resolved determination of the concentrations of one or more analyte species is of distinct interest. With a light-addressable potentiometric sensor (LAPS), chemical images can be created, which visualize the concentration distribution above the sensor plate. One important challenge is to achieve a good lateral resolution in order to detect events that take place in a small and limited region. LAPS utilizes a focused light spot to address the measurement region. By moving this light spot along the semiconductor sensor plate, the concentration distribution can be observed. In this study, we show that utilizing a pulse as light excitation instead of a traditionally used continuously modulated light excitation, the lateral resolution can be improved by a factor of 6 or more. KW - Chemical images KW - LAPS KW - Light-addressable potentiometric sensor Y1 - 2017 U6 - http://dx.doi.org/10.1016/j.snb.2017.02.057 SN - 0925-4005 VL - 248 SP - 961 EP - 965 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Werner, Frederik A1 - Krumbe, Christoph A1 - Schumacher, Katharina A1 - Groebel, Simone A1 - Spelthahn, Heiko A1 - Stellberg, Michael A1 - Wagner, Torsten A1 - Yoshinobu, Tatsuo A1 - Selmer, Thorsten A1 - Keusgen, Michael A1 - Baumann, Marcus A1 - Schöning, Michael Josef T1 - Determination of the extracellular acidification of Escherichia coli by a light-addressable potentiometric sensor JF - Physica status solidi (a) : applications and material science. 208 (2011), H. 6 Y1 - 2011 SN - 1862-6319 SP - 1340 EP - 1344 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Werkhausen, Amelie A1 - Willwacher, Steffen A1 - Albracht, Kirsten T1 - Medial gastrocnemius muscle fascicles shorten throughout stance during sprint acceleration JF - Scandinavian Journal of Medicine & Science in Sports N2 - The compliant nature of distal limb muscle-tendon units is traditionally considered suboptimal in explosive movements when positive joint work is required. However, during accelerative running, ankle joint net mechanical work is positive. Therefore, this study aims to investigate how plantar flexor muscle-tendon behavior is modulated during fast accelerations. Eleven female sprinters performed maximum sprint accelerations from starting blocks, while gastrocnemius muscle fascicle lengths were estimated using ultrasonography. We combined motion analysis and ground reaction force measurements to assess lower limb joint kinematics and kinetics, and to estimate gastrocnemius muscle-tendon unit length during the first two acceleration steps. Outcome variables were resampled to the stance phase and averaged across three to five trials. Relevant scalars were extracted and analyzed using one-sample and two-sample t-tests, and vector trajectories were compared using statistical parametric mapping. We found that an uncoupling of muscle fascicle behavior from muscle-tendon unit behavior is effectively used to produce net positive mechanical work at the joint during maximum sprint acceleration. Muscle fascicles shortened throughout the first and second steps, while shortening occurred earlier during the first step, where negative joint work was lower compared with the second step. Elastic strain energy may be stored during dorsiflexion after touchdown since fascicles did not lengthen at the same time to dissipate energy. Thus, net positive work generation is accommodated by the reuse of elastic strain energy along with positive gastrocnemius fascicle work. Our results show a mechanism of how muscles with high in-series compliance can contribute to net positive joint work. KW - locomotion KW - muscle mechanics KW - running KW - sprint start KW - ultrasonography Y1 - 2021 U6 - http://dx.doi.org/10.1111/sms.13956 SN - 0905-7188 (Druckausgabe) SN - 1600-0838 (Onlineausgabe) VL - 31 IS - 7 SP - 1471 EP - 1480 PB - Wiley-Blackwell CY - Oxford ER -