TY - JOUR A1 - Schlamann, Marc A1 - Voigt, Melanie A. A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Kraff, Oliver A1 - Ladd, Susanne C. A1 - Ladd, Mark E. A1 - Forsting, Michael A1 - Wilhelm, Hans T1 - Exposure to high-field MRI does not affect cognitive function JF - Journal of Magnetic Resonance Imaging N2 - Purpose To assess potential cognitive deficits under the influence of static magnetic fields at various field strengths some studies already exist. These studies were not focused on attention as the most vulnerable cognitive function. Additionally, mostly no magnetic resonance imaging (MRI) sequences were performed. Materials and Methods In all, 25 right-handed men were enrolled in this study. All subjects underwent one MRI examination of 63 minutes at 1.5 T and one at 7 T within an interval of 10 to 30 days. The order of the examinations was randomized. Subjects were referred to six standardized neuropsychological tests strictly focused on attention immediately before and after each MRI examination. Differences in neuropsychological variables between the timepoints before and after each MRI examination were assessed and P-values were calculated Results Only six subtests revealed significant differences between pre- and post-MRI. In these tests the subjects achieved better results in post-MRI testing than in pre-MRI testing (P = 0.013–0.032). The other tests revealed no significant results. Conclusion The improvement in post-MRI testing is only explicable as a result of learning effects. MRI examinations, even in ultrahigh-field scanners, do not seem to have any persisting influence on the attention networks of human cognition immediately after exposure. Y1 - 2010 U6 - http://dx.doi.org/10.1002/jmri.22065 SN - 1522-2586 VL - 31 IS - 5 SP - 1061 EP - 1066 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Lagemaat, Miriam W. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Kobus, Thiele A1 - Kraff, Oliver A1 - Maderwald, Stefan A1 - Ladd, Mark E. A1 - Scheenen, Tom W. J. T1 - Feasibility of T₂-weighted turbo spin echo imaging of the human prostate at 7 tesla JF - Magnetic Resonance in Medicine N2 - Purpose To demonstrate that high quality T₂-weighted (T2w) turbo spin-echo (TSE) imaging of the complete prostate can be achieved routinely and within safety limits at 7 T, using an external transceive body array coil only. Methods Nine healthy volunteers and 12 prostate cancer patients were scanned on a 7 T whole-body system. Preparation consisted of B₀ and radiofrequency shimming and localized flip angle calibration. T₁ and T₂ relaxation times were measured and used to define the T2w-TSE protocol. T2w imaging was performed using a TSE sequence (pulse repetition time/echo time 3000–3640/71 ms) with prolonged excitation and refocusing pulses to reduce specific absorption rate. Results High quality T2w TSE imaging was performed in less than 2 min in all subjects. Tumors of patients with gold-standard tumor localization (MR-guided biopsy or prostatectomy) were well visualized on 7 T imaging (n = 3). The number of consecutive slices achievable within a 10-g averaged specific absorption rate limit of 10 W/kg was ≥28 in all subjects, sufficient for full prostate coverage with 3-mm slices in at least one direction. Conclusion High quality T2w TSE prostate imaging can be performed routinely and within specific absorption rate limits at 7 T with an external transceive body array. Y1 - 2014 U6 - http://dx.doi.org/10.1002/mrm.24818 SN - 1522-2594 VL - 71 IS - 5 SP - 1711 EP - 1719 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Umutlu, L. A1 - Maderwald, S. A1 - Kinner, S. A1 - Kraff, O. A1 - Bitz, Andreas A1 - Orzada, S. A1 - Johst, S. A1 - Wrede, K. A1 - Forsting, M. A1 - Ladd, M. E. A1 - Lauenstein, T. C. A1 - Quick, H. H. T1 - First-pass contrast-enhanced renal MRA at 7 Tesla: initial results JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-012-2666-0 SN - 1432-1084 VL - 23 IS - 4 SP - 1059 EP - 1066 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ndoumbè Mbonjo Mbonjo, H. A1 - Streckert, J. A1 - Bitz, Andreas A1 - Hansen, V. A1 - Glasmachers, A. A1 - Gencol, S. A1 - Rozic, D. T1 - Generic UMTS test signal for RF bioelectromagnetic studies JF - Bioelectromagnetics Y1 - 2004 U6 - http://dx.doi.org/10.1002/bem.20007 SN - 1521-186X VL - 25 IS - 6 SP - 415 EP - 425 ER - TY - JOUR A1 - Vos, E. K. A1 - Lagemaat, M. W. A1 - Barentsz, J. O. A1 - Fütterer, J. J. A1 - Zamecnik, P. A1 - Roozen, H. A1 - Orzada, S. A1 - Bitz, Andreas A1 - Maas, M. C. A1 - Scheenen, T. W. J. T1 - Image quality and cancer visibility of T2-weighted Magnetic Resonance Imaging of the prostate at 7 Tesla JF - European Radiology N2 - Objectives To assess the image quality of T2-weighted (T2w) magnetic resonance imaging of the prostate and the visibility of prostate cancer at 7 Tesla (T). Materials & methods Seventeen prostate cancer patients underwent T2w imaging at 7T with only an external transmit/receive array coil. Three radiologists independently scored images for image quality, visibility of anatomical structures, and presence of artefacts. Krippendorff’s alpha and weighted kappa statistics were used to assess inter-observer agreement. Visibility of prostate cancer lesions was assessed by directly linking the T2w images to the confirmed location of prostate cancer on histopathology. Results T2w imaging at 7T was achievable with ‘satisfactory’ (3/5) to ‘good’ (4/5) quality. Visibility of anatomical structures was predominantly scored as ‘satisfactory’ (3/5) and ‘good’ (4/5). If artefacts were present, they were mostly motion artefacts and, to a lesser extent, aliasing artefacts and noise. Krippendorff’s analysis revealed an α = 0.44 between three readers for the overall image quality scores. Clinically significant cancer lesions in both peripheral zone and transition zone were visible at 7T. Conclusion T2w imaging with satisfactory to good quality can be routinely acquired, and cancer lesions were visible in patients with prostate cancer at 7T using only an external transmit/receive body array coil. Y1 - 2014 U6 - http://dx.doi.org/10.1007/s00330-014-3234-6 SN - 1432-1084 VL - 24 IS - 8 SP - 1950 EP - 1958 PB - Springer CY - Cham ER - TY - JOUR A1 - Wissen, M. A1 - Bogdanski, N. A1 - Scheer, H.-C. A1 - Bitz, Andreas A1 - Ahrens, G. A1 - Gruetzner, G. T1 - Implication of the light polarisation for UV curing of pre-patterned resists JF - Microelectronic Engineering Y1 - 2005 U6 - http://dx.doi.org/10.1016/j.mee.2004.12.099 SN - 0167-9317 VL - 78-79 SP - 659 EP - 664 ER - TY - JOUR A1 - Noureddine, Yacine A1 - Kraff, Oliver A1 - Ladd, Mark E. A1 - Wrede, Karsten H. A1 - Chen, Bixia A1 - Quick, Harald H. A1 - Schaefers, Gregor A1 - Bitz, Andreas T1 - In vitro and in silico assessment of RF-induced heating around intracranial aneurysm clips at 7 Tesla JF - Magnetic Resonance in Medicine Y1 - 2017 U6 - http://dx.doi.org/10.1002/mrm.26650 SN - 1522-2594 IS - Early view PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Kobus, Thiele A1 - Bitz, Andreas A1 - Uden, Mark J. van A1 - Lagemaat, Miram W. A1 - Rothgang, Eva A1 - Orzada, Stephan A1 - Heerschap, Arend A1 - Scheenen, Tom W. J. T1 - In vivo 31P MR spectroscopic imaging of the human prostate at 7 T: safety and feasibility JF - Magnetic Resonance in Medicine N2 - 31P MR spectroscopic imaging of the human prostate provides information about phosphorylated metabolites that could be used for prostate cancer characterization. The sensitivity of a magnetic field strength of 7 T might enable 3D 31P MR spectroscopic imaging with relevant spatial resolution in a clinically acceptable measurement time. To this end, a 31P endorectal coil was developed and combined with an eight-channel 1H body-array coil to relate metabolic information to anatomical location. An extensive safety validation was performed to evaluate the specific absorption rate, the radiofrequency field distribution, and the temperature distribution of both coils. This validation consisted of detailed Finite Integration Technique simulations, confirmed by MR thermometry and Burn:x-wiley:07403194:media:MRM24175:tex2gif-stack-1 measurements in a phantom and in vivo temperature measurements. The safety studies demonstrated that the presence of the 31P endorectal coil had no influence on the specific absorption rate levels and temperature distribution of the external eight-channel 1H array coil. To stay within a 10 g averaged local specific absorption rate of 10 W/kg, a maximum time-averaged input power of 33 W for the 1H array coil was allowed. For transmitting with the 31P endorectal coil, our safety limit of less than 1°C temperature increase in vivo during a 15-min MR spectroscopic imaging experiment was reached at a time-averaged input power of 1.9 W. With this power setting, a second in vivo measurement was performed on a healthy volunteer. Using adiabatic excitation, 3D 31P MR spectroscopic imaging produced spectra from the entire prostate in 18 min with a spatial resolution of 4 cm3. The spectral resolution enabled the separate detection of phosphocholine, phosphoethanolamine, inorganic phosphate, and other metabolites that could play an important role in the characterization of prostate cancer. Y1 - 2012 U6 - http://dx.doi.org/10.1002/mrm.24175 SN - 1522-2594 VL - 68 IS - 6 SP - 1683 EP - 1695 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Orzada, Stephan A1 - Fiedler, Thomas M. A1 - Bitz, Andreas A1 - Ladd, Mark E. A1 - Quick, Harald H. T1 - Local SAR compression with overestimation control to reduce maximum relative SAR overestimation and improve multi-channel RF array performance JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective In local SAR compression algorithms, the overestimation is generally not linearly dependent on actual local SAR. This can lead to large relative overestimation at low actual SAR values, unnecessarily constraining transmit array performance. Method Two strategies are proposed to reduce maximum relative overestimation for a given number of VOPs. The first strategy uses an overestimation matrix that roughly approximates actual local SAR; the second strategy uses a small set of pre-calculated VOPs as the overestimation term for the compression. Result Comparison with a previous method shows that for a given maximum relative overestimation the number of VOPs can be reduced by around 20% at the cost of a higher absolute overestimation at high actual local SAR values. Conclusion The proposed strategies outperform a previously published strategy and can improve the SAR compression where maximum relative overestimation constrains the performance of parallel transmission. Y1 - 2020 SN - 1352-8661 U6 - http://dx.doi.org/10.1007/s10334-020-00890-0 IS - 34 (2021) SP - 153 EP - 164 PB - Springer CY - Heidelberg ER - TY - JOUR A1 - Sommer, Angela M. A1 - Bitz, Andreas A1 - Streckert, Joachim A1 - Hansen, Volkert W. A1 - Lerchl, Alexander T1 - Lymphoma development in mice chronically exposed to UMTS-modulated radiofrequency electromagnetic fields JF - Radiation Research Y1 - 2007 U6 - http://dx.doi.org/10.1667/RR0857.1 SN - 1938-5404 VL - 168 IS - 1 SP - 72 EP - 80 ER -