TY - JOUR A1 - Ciritsis, Alexander A1 - Horbach, Andreas A1 - Staat, Manfred A1 - Kuhl, Christiane K. A1 - Kraemer, Nils Andreas T1 - Porosity and tissue integration of elastic mesh implants evaluated in vitro and in vivo JF - Journal of Biomedical Materials Research: Part B: Applied Biomaterials N2 - Purpose In vivo, a loss of mesh porosity triggers scar tissue formation and restricts functionality. The purpose of this study was to evaluate the properties and configuration changes as mesh deformation and mesh shrinkage of a soft mesh implant compared with a conventional stiff mesh implant in vitro and in a porcine model. Material and Methods Tensile tests and digital image correlation were used to determine the textile porosity for both mesh types in vitro. A group of three pigs each were treated with magnetic resonance imaging (MRI) visible conventional stiff polyvinylidene fluoride meshes (PVDF) or with soft thermoplastic polyurethane meshes (TPU) (FEG Textiltechnik mbH, Aachen, Germany), respectively. MRI was performed with a pneumoperitoneum at a pressure of 0 and 15 mmHg, which resulted in bulging of the abdomen. The mesh-induced signal voids were semiautomatically segmented and the mesh areas were determined. With the deformations assessed in both mesh types at both pressure conditions, the porosity change of the meshes after 8 weeks of ingrowth was calculated as an indicator of preserved elastic properties. The explanted specimens were examined histologically for the maturity of the scar (collagen I/III ratio). Results In TPU, the in vitro porosity increased constantly, in PVDF, a loss of porosity was observed under mild stresses. In vivo, the mean mesh areas of TPU were 206.8 cm2 (± 5.7 cm2) at 0 mmHg pneumoperitoneum and 274.6 cm2 (± 5.2 cm2) at 15 mmHg; for PVDF the mean areas were 205.5 cm2 (± 8.8 cm2) and 221.5 cm2 (± 11.8 cm2), respectively. The pneumoperitoneum-induced pressure increase resulted in a calculated porosity increase of 8.4% for TPU and of 1.2% for PVDF. The mean collagen I/III ratio was 8.7 (± 0.5) for TPU and 4.7 (± 0.7) for PVDF. Conclusion The elastic properties of TPU mesh implants result in improved tissue integration compared to conventional PVDF meshes, and they adapt more efficiently to the abdominal wall. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 827–833, 2018. Y1 - 2018 U6 - https://doi.org/10.1002/jbm.b.33877 SN - 1552-4981 VL - 106 IS - 2 SP - 827 EP - 833 PB - Wiley CY - New York, NY ER - TY - JOUR A1 - Kohler, Annette A1 - Kirschner-Hermanns, Ruth A1 - Staat, Manfred A1 - Brehmer, Bernhard T1 - Pathogenese, funktionelle und anatomische Aspekte der weiblichen Belastungsinkontinenz T1 - Female stress incontinence: aspects of pathogenesis and functional anatomy JF - Aktuelle Urologie N2 - Der vorliegende Artikel fokussiert sich auf die weibliche Belastungsinkontinenz als Insuffizienz der Speicherfunktion der Blase, auch wenn im klinischen Alltag die Harninkontinenz der Frau häufig verschiedene Ursachen hat und insbesondere eine Belastungsinkontinenz im Alter und bei neurologischer Komorbidität nur selten isoliert vorkommt. Das kleine Becken der Frau ist sowohl als Funktions- als auch als strukturelle Einheit zu betrachten. Dabei unterliegen bei der Frau Blase, Harnröhre, Gebärmutter und Enddarm sowie die muskulären und ligamentösen Strukturen des kleinen Beckens durch Fertilitätsphase, mögliche Schwangerschaften, Geburten und Menopausen-Phase, über das „normale Altern“ hinaus, gravierenden Veränderungen. This article focuses on female stress incontinence in the form of pelvic floor dysfunction and urethral sphincter deficiency, although isolated stress incontinence accounts for less than half of all incontinence cases. Especially in women of old age and those with neurological comorbidities, the causes of incontinence are mostly multifactorial. Also it has to be considered that the female bladder, urethra, uterus and rectum as well as the muscular and ligamentous structures of the female pelvis minor are affected by phases of fertility, possible pregnancies, births and menopause in addition to the normal ageing process. Y1 - 2018 U6 - https://doi.org/10.1055/s-0043-120616 SN - 1438-8820 VL - 49 IS - 1 SP - 47 EP - 51 PB - Thieme CY - Stuttgart ER - TY - JOUR A1 - Keutmann, Sabine A1 - Staat, Manfred A1 - Laack, Walter van T1 - Untersuchung der thermischen Auswirkung von therapeutischem Ultraschall N2 - Zusammenfassung: In der Orthopädie zählt der therapeutische Ultraschall als Mittel zur Prävention und Therapiebegleitung. Er hat mechanische, thermische und physiko-chemische Auswirkungen auf den menschlichen Körper. Um mehr Erkenntnisse über die thermischen Auswirkungen zu erlangen, wurden Versuche an einem Hydrogel-Phantom und an Probanden durchgeführt. Dabei entstand eine signifikante Erwärmung des Gewebes, welche beim Probandenversuch an der Oberfläche und beim Hydrogelversuch in der Tiefe gemessen wurde. Summary: In orthopaedics, therapeutic ultrasound is a tool of prevention and therapy support. It has mechanical, thermal and physico-chemical effects on the human body. Tests with a hydrogel phantom and with human probands have been performed in order to obtain more knowledge about their thermal effects. Both tests measured temperature increases in cell tissue, on the surface with the human proband test and in depth with the hydrogel phantom test. T2 - Research about the thermal effects of therapeutic ultrasound Y1 - 2018 SN - 2193-5793 SN - 2193-5785 (Druckausgabe) VL - 7 IS - 10 SP - 518 EP - 522 PB - Deutscher Ärzte-Verl. CY - Köln ER - TY - JOUR A1 - Bhattarai, Aroj A1 - Staat, Manfred T1 - Computational comparison of different textile implants to correct apical prolapse in females JF - Current Directions in Biomedical Engineering N2 - Prosthetic textile implants of different shapes, sizes and polymers are used to correct the apical prolapse after hysterectomy (removal of the uterus). The selection of the implant before or during minimally invasive surgery depends on the patient’s anatomical defect, intended function after reconstruction and most importantly the surgeon’s preference. Weakness or damage of the supporting tissues during childbirth, menopause or previous pelvic surgeries may put females in higher risk of prolapse. Numerical simulations of reconstructed pelvic floor with weakened tissues and organ supported by textile product models: DynaMesh®-PRS soft, DynaMesh®-PRP soft and DynaMesh®-CESA from FEG Textiletechnik mbH, Germany are compared. Y1 - 2018 U6 - https://doi.org/10.1515/cdbme-2018-0159 VL - 4 IS - 1 SP - 661 EP - 664 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Horbach, Andreas A1 - Staat, Manfred T1 - Optical strain measurement for the modeling of surgical meshes and their porosity JF - Current Directions in Biomedical Engineering N2 - The porosity of surgical meshes makes them flexible for large elastic deformation and establishes the healing conditions of good tissue in growth. The biomechanic modeling of orthotropic and compressible materials requires new materials models and simulstaneoaus fit of deformation in the load direction as well as trannsversely to to load. This nonlinear modeling can be achieved by an optical deformation measurement. At the same time the full field deformation measurement allows the dermination of the change of porosity with deformation. Also the socalled effective porosity, which has been defined to asses the tisssue interatcion with the mesh implants, can be determined from the global deformation of the surgical meshes. Y1 - 2018 U6 - https://doi.org/10.1515/cdbme-2018-0045 SN - 2364-5504 VL - Band 4 IS - 1 SP - 181 EP - 184 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Peloni, Alessandro A1 - Dachwald, Bernd A1 - Ceriotti, Matteo T1 - Multiple near-earth asteroid rendezvous mission: Solar-sailing options JF - Advances in Space Research Y1 - 2017 U6 - https://doi.org/10.1016/j.asr.2017.10.017 SN - 0273-1177 IS - In Press, Corrected Proof PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Hackl, Michael A1 - Wegmann, Kilian A1 - Kahmann, Stephanie Lucina A1 - Heinze, Nicolai A1 - Staat, Manfred A1 - Neiss, Wolfram F. A1 - Scaal, Martin A1 - Müller, Lars P. T1 - Radial shortening osteotomy reduces radiocapitellar contact pressures while preserving valgus stability of the elbow JF - Knee Surgery, Sports Traumatology, Arthroscopy Y1 - 2017 U6 - https://doi.org/10.1007/s00167-017-4468-z SN - 1433-7347 VL - 25 IS - 7 SP - 2280 EP - 2288 PB - Springer CY - Berlin ER - TY - JOUR A1 - Tran, Linda A1 - Mottaghy, K. A1 - Arlt-Körfer, Sabine A1 - Waluga, Christian A1 - Behbahani, Mehdi T1 - An experimental study of shear-dependent human platelet adhesion and underlying protein-binding mechanisms in a cylindrical Couette system JF - Biomedizinische Technik Y1 - 2017 U6 - https://doi.org/10.1515/bmt-2015-0034 SN - 0013-5585 VL - 62 IS - 4 SP - 383 EP - 392 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Albanna, Walid A1 - Lueke, Jan Niklas A1 - Sjapic, Volha A1 - Kotliar, Konstantin A1 - Hescheler, Jürgen A1 - Clusmann, Hans A1 - Sjapic, Sergej A1 - Alpdogan, Serdan A1 - Schneider, Toni A1 - Schubert, Gerrit Alexander A1 - Neumaier, Felix T1 - Electroretinographic Assessment of Inner Retinal Signaling in the Isolated and Superfused Murine Retina JF - Current Eye Research Y1 - 2017 U6 - https://doi.org/10.1080/02713683.2017.1339807 SN - 1460-2202 IS - Article in press SP - 1 EP - 9 PB - Taylor & Francis CY - London ER - TY - JOUR A1 - Kotliar, Konstantin A1 - Hauser, Christine A1 - Ortner, Marion A1 - Muggenthaler, Claudia A1 - Diehl-Schmid, Janine A1 - Angermann, Susanne A1 - Hapfelmeier, Alexander A1 - Schmaderer, Christoph A1 - Grimmer, Timo T1 - Altered neurovascular coupling as measured by optical imaging: a biomarker for Alzheimer’s disease JF - Scientific Reports Y1 - 2017 U6 - https://doi.org/10.1038/s41598-017-13349-5 SN - 2045-2322 N1 - Article 12906 VL - 7 IS - 1 SP - 1 EP - 11 PB - Springer Nature CY - Cham ER -