TY - JOUR A1 - Frauenrath, Tobias A1 - Hezel, Fabian A1 - Renz, Wolfgang A1 - de Geyer d'Orth, Thibaut A1 - Dieringer, Matthias A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Prothmann, Marcel A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf T1 - Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla JF - Journal of Cardiovascular Magnetic Resonance N2 - Background To demonstrate the applicability of acoustic cardiac triggering (ACT) for imaging of the heart at ultrahigh magnetic fields (7.0 T) by comparing phonocardiogram, conventional vector electrocardiogram (ECG) and traditional pulse oximetry (POX) triggered 2D CINE acquisitions together with (i) a qualitative image quality analysis, (ii) an assessment of the left ventricular function parameter and (iii) an examination of trigger reliability and trigger detection variance derived from the signal waveforms. Results ECG was susceptible to severe distortions at 7.0 T. POX and ACT provided waveforms free of interferences from electromagnetic fields or from magneto-hydrodynamic effects. Frequent R-wave mis-registration occurred in ECG-triggered acquisitions with a failure rate of up to 30% resulting in cardiac motion induced artifacts. ACT and POX triggering produced images free of cardiac motion artefacts. ECG showed a severe jitter in the R-wave detection. POX also showed a trigger jitter of approximately Δt = 72 ms which is equivalent to two cardiac phases. ACT showed a jitter of approximately Δt = 5 ms only. ECG waveforms revealed a standard deviation for the cardiac trigger offset larger than that observed for ACT or POX waveforms. Image quality assessment showed that ACT substantially improved image quality as compared to ECG (image quality score at end-diastole: ECG = 1.7 ± 0.5, ACT = 2.4 ± 0.5, p = 0.04) while the comparison between ECG vs. POX gated acquisitions showed no significant differences in image quality (image quality score: ECG = 1.7 ± 0.5, POX = 2.0 ± 0.5, p = 0.34). Conclusions The applicability of acoustic triggering for cardiac CINE imaging at 7.0 T was demonstrated. ACT's trigger reliability and fidelity are superior to that of ECG and POX. ACT promises to be beneficial for cardiovascular magnetic resonance at ultra-high field strengths including 7.0 T. KW - Interval Time Series KW - Image Quality Score KW - Image Quality Assessment KW - Sound Pressure Level KW - Cardiovascular Magnetic Resonance Y1 - 2010 U6 - https://doi.org/10.1186/1532-429X-12-67 SN - 1532-429X VL - 12 IS - 1 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Fuchs, Katharina A1 - Dieringer, Matthias A. A1 - Özerdem, Celal A1 - Patel, Nishan A1 - Renz, Wolfgang A1 - Greiser, Andreas A1 - Elgeti, Thomas A1 - Niendorf, Thoralf T1 - Detailing the use of magnetohydrodynamic effects for synchronization of MRI with the cardiac cycle: A feasibility study JF - Journal of Magnetic Resonance Imaging N2 - Purpose: To investigate the feasibility of using magnetohydrodynamic (MHD) effects for synchronization of magnetic resonance imaging (MRI) with the cardiac cycle. Materials and Methods: The MHD effect was scrutinized using a pulsatile flow phantom at B0 = 7.0 T. MHD effects were examined in vivo in healthy volunteers (n = 10) for B0 ranging from 0.05–7.0 T. Noncontrast-enhanced MR angiography (MRA) of the carotids was performed using a gated steady-state free-precession (SSFP) imaging technique in conjunction with electrocardiogram (ECG) and MHD synchronization. Results: The MHD potential correlates with flow velocities derived from phase contrast MRI. MHD voltages depend on the orientation between B0 and the flow of a conductive fluid. An increase in the interelectrode spacing along the flow increases the MHD potential. In vivo measurement of the MHD effect provides peak voltages of 1.5 mV for surface areas close to the common carotid artery at B0 = 7.0 T. Synchronization of MRI with the cardiac cycle using MHD triggering is feasible. MHD triggered MRA of the carotids at 3.0 T showed an overall image quality and richness of anatomic detail, which is comparable to ECG-triggered MRAs. Conclusion: This feasibility study demonstrates the use of MHD effects for synchronization of MR acquisitions with the cardiac cycle. J. Magn. Reson. Imaging 2012;36:364–372. © 2012 Wiley Periodicals, Inc. Y1 - 2012 U6 - https://doi.org/10.1002/jmri.23634 SN - 1522-2586 VL - 36 IS - 2 SP - 364 EP - 372 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Franke, Helmut A1 - Streckert, Joachim A1 - Bitz, Andreas A1 - Goeke, Johannes A1 - Hansen, Volkert A1 - Ringelstein, E. Bernd A1 - Nattkämper, Heiner A1 - Galle, Hans-Joachim A1 - Stögbauer, Florian T1 - Effects of Universal Mobile Telecommunications System (UMTS) electromagnetic fields on the blood-brain barrier in vitro JF - Radiation Research Y1 - 2005 U6 - https://doi.org/10.1667/RR3424.1 SN - 1938-5404 VL - 164 IS - 3 SP - 258 EP - 269 ER - TY - JOUR A1 - Foltz, Christian A1 - Schneider, Nicole A1 - Wolf, Martin R. A1 - Kausch, Bernhard A1 - Schlick, Christopher A1 - Luczak, Holger ED - Nagl, Manfred T1 - Usability engineering JF - Collaborative and distributed chemical engineering : from understanding to substantial design process support; results of the IMPROVE Project. - (Lecture notes in computer science ; 4970) Y1 - 2008 SN - 978-3-540-70552-9 U6 - https://doi.org/10.1007/978-3-540-70552-9_21 SP - 527 EP - 554 PB - Springer CY - Berlin ER - TY - CHAP A1 - Finger, Felix A1 - Khalsa, R. A1 - Kreyer, Jörg A1 - Mayntz, Joscha A1 - Braun, Carsten A1 - Dahmann, Peter A1 - Esch, Thomas A1 - Kemper, Hans A1 - Schmitz, O. A1 - Bragard, Michael T1 - An approach to propulsion system modelling for the conceptual design of hybrid-electric general aviation aircraft T2 - Deutscher Luft- und Raumfahrtkongress 2019, 30.9.-2.10.2019, Darmstadt N2 - In this paper, an approach to propulsion system modelling for hybrid-electric general aviation aircraft is presented. Because the focus is on general aviation aircraft, only combinations of electric motors and reciprocating combustion engines are explored. Gas turbine hybrids will not be considered. The level of the component's models is appropriate for the conceptual design stage. They are simple and adaptable, so that a wide range of designs with morphologically different propulsive system architectures can be quickly compared. Modelling strategies for both mass and efficiency of each part of the propulsion system (engine, motor, battery and propeller) will be presented. Y1 - 2019 ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Schmidt, Simon A1 - Stelter, Jonathan K. A1 - Wittrich, Marco A1 - Quick, Harald H. A1 - Bitz, Andreas A1 - Ladd, Mark E. T1 - Performance and safety assessment of an integrated transmit array for body imaging at 7 T under consideration of specificabsorption rate, tissue temperature, and thermal dose JF - NMR in Biomedicine N2 - In this study, the performance of an integrated body-imaging array for 7 T with 32 radiofrequency (RF) channels under consideration of local specific absorption rate (SAR), tissue temperature, and thermal dose limits was evaluated and the imaging performance was compared with a clinical 3 T body coil. Thirty-two transmit elements were placed in three rings between the bore liner and RF shield of the gradient coil. Slice-selective RF pulse optimizations for B1 shimming and spokes were performed for differently oriented slices in the body under consideration of realistic constraints for power and local SAR. To improve the B1+ homogeneity, safety assessments based on temperature and thermal dose were performed to possibly allow for higher input power for the pulse optimization than permissible with SAR limits. The results showed that using two spokes, the 7 T array outperformed the 3 T birdcage in all the considered regions of interest. However, a significantly higher SAR or lower duty cycle at 7 T is necessary in some cases to achieve similar B1+ homogeneity as at 3 T. The homogeneity in up to 50 cm-long coronal slices can particularly benefit from the high RF shim performance provided by the 32 RF channels. The thermal dose approach increases the allowable input power and the corresponding local SAR, in one example up to 100 W/kg, without limiting the exposure time necessary for an MR examination. In conclusion, the integrated antenna array at 7 T enables a clinical workflow for body imaging and comparable imaging performance to a conventional 3 T clinical body coil. KW - body imaging at 7 T MRI KW - thermal dose KW - tissue temperature KW - transmit antenna arrays Y1 - 2022 U6 - https://doi.org/10.1002/nbm.4656 SN - 0952-3480 (Print) SN - 1099-1492 (Online) VL - 35 IS - 5 SP - 1 EP - 17 PB - Wiley ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - Performance analysis of integrated RF microstrip transmit antenna arrays with high channel count for body imaging at 7 T JF - NMR in Biomedicine N2 - The aim of the current study was to investigate the performance of integrated RF transmit arrays with high channel count consisting of meander microstrip antennas for body imaging at 7 T and to optimize the position and number of transmit ele- ments. RF simulations using multiring antenna arrays placed behind the bore liner were performed for realistic exposure conditions for body imaging. Simulations were performed for arrays with as few as eight elements and for arrays with high channel counts of up to 48 elements. The B1+ field was evaluated regarding the degrees of freedom for RF shimming in the abdomen. Worst-case specific absorption rate (SARwc ), SAR overestimation in the matrix compression, the number of virtual obser- vation points (VOPs) and SAR efficiency were evaluated. Constrained RF shimming was performed in differently oriented regions of interest in the body, and the devia- tion from a target B1+ field was evaluated. Results show that integrated multiring arrays are able to generate homogeneous B1+ field distributions for large FOVs, espe- cially for coronal/sagittal slices, and thus enable body imaging at 7 T with a clinical workflow; however, a low duty cycle or a high SAR is required to achieve homoge- neous B1+ distributions and to exploit the full potential. In conclusion, integrated arrays allow for high element counts that have high degrees of freedom for the pulse optimization but also produce high SARwc , which reduces the SAR accuracy in the VOP compression for low-SAR protocols, leading to a potential reduction in array performance. Smaller SAR overestimations can increase SAR accuracy, but lead to a high number of VOPs, which increases the computational cost for VOP evaluation and makes online SAR monitoring or pulse optimization challenging. Arrays with interleaved rings showed the best results in the study. KW - body imaging at UHF MRI KW - integrated transmit coil arrays KW - VOP compression Y1 - 2021 U6 - https://doi.org/10.1002/nbm.4515 SN - 0952-3480 (ISSN) SN - 1099-1492 (eISSN) VL - 34 IS - 7 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Clemens, Markus A1 - Bitz, Andreas T1 - Safety of subjects during radiofrequency exposure in ultra-high-field magnetic resonance imaging JF - IEEE Letters on Electromagnetic Compatibility Practice and Applications N2 - Magnetic resonance imaging (MRI) is one of the most important medical imaging techniques. Since the introduction of MRI in the mid-1980s, there has been a continuous trend toward higher static magnetic fields to obtain i.a. a higher signal-to-noise ratio. The step toward ultra-high-field (UHF) MRI at 7 Tesla and higher, however, creates several challenges regarding the homogeneity of the spin excitation RF transmit field and the RF exposure of the subject. In UHF MRI systems, the wavelength of the RF field is in the range of the diameter of the human body, which can result in inhomogeneous spin excitation and local SAR hotspots. To optimize the homogeneity in a region of interest, UHF MRI systems use parallel transmit systems with multiple transmit antennas and time-dependent modulation of the RF signal in the individual transmit channels. Furthermore, SAR increases with increasing field strength, while the SAR limits remain unchanged. Two different approaches to generate the RF transmit field in UHF systems using antenna arrays close and remote to the body are investigated in this letter. Achievable imaging performance is evaluated compared to typical clinical RF transmit systems at lower field strength. The evaluation has been performed under consideration of RF exposure based on local SAR and tissue temperature. Furthermore, results for thermal dose as an alternative RF exposure metric are presented. Y1 - 2020 SN - 2637-6423 U6 - https://doi.org/10.1109/LEMCPA.2020.3029747 VL - 2 IS - 3 SP - 1 EP - 8 PB - IEEE CY - New York, NY ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - SAR Simulations & Safety JF - NeuroImage Y1 - 2017 U6 - https://doi.org/10.1016/j.neuroimage.2017.03.035 SN - 1053-8119 IS - Epub ahead of print PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus temperature limits JF - Medical Physics Y1 - 2017 U6 - https://doi.org/10.1002/mp.12034 N1 - This article is corrected by: Errata: Erratum: “RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus tissue temperature limits” [Med. Phys. 44(1), 143–157 (2017)] Volume 44, Issue 2, 772 VL - 44 IS - 1 SP - 143 EP - 157 ER - TY - CHAP A1 - Feuerriegel, Uwe A1 - Pook, Michael A1 - Wersch, Gregor A1 - Wittenhorst, Simon A1 - Becker, Jürgen A1 - Ecker, Markus A1 - Hoffmann, Ulrich A1 - Kunz, Ulrich T1 - Simulation von Wärmeübertragungsprozessen N2 - Mit freundlicher Genehmigung der Autoren und des Oldenbourg Industrieverlags https://www.oldenbourg-industrieverlag.de/de/9783835633223-33223 erschienen als Beitrag im Tagungsband zur AALE-Tagung 2012. 9. Fachkonferenz 4.-5. Mai 2012, Aachen, Fachhochschule. ISBN 9783835633223 S 5-1 S. 127-135 Es werden Ergebnisse unterschiedlicher Projekte aus dem Bereich der Simulation von Wärmeübertragungsprozessen mit Excel-VBA vorgestellt. - Thermische Behandlung hochviskoser Fruchtzubereitungen, verschiedene Projekte und Kooperationen mit der Zentis GmbH & Co. KG, Aachen (J. Becker, U. Feuerriegel, G. Wersch). - Untersuchung des dynamischen Verhaltens von dampfbeheizten Ethylen-Verdampfern. Projekt mit der TGE Gas Engineering GmbH, Bonn (M. Ecker, U. Feuerriegel, U. Hoffmann, S. Wittenhorst). - Dynamische Simulation des axialen Temperaturverlaufs von elektrisch beheizten Rohrreaktoren. Kooperation mit dem Institut für Chemische Verfahrenstechnik, TU Clausthal (U. Feuerriegel, U. Kunz, M. Pook, S. Wittenhorst). KW - Wärmeübertragung Y1 - 2012 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Strack, Andreas A1 - Lakemeyer, Gerhard T1 - Laser-Based Localization with Sparse Landmarks / Strack, Andreas ; Ferrein, Alexander ; Lakemeyer, Gerhard JF - RoboCup 2005: Robot Soccer World Cup IX Y1 - 2006 SN - 978-3-540-35437-6 N1 - Lecture Notes in Computer Science ; 4020 SP - 569 EP - 576 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - Vassos, Stavros T1 - Action-Based Imperative Programming with YAGI N2 - Many tasks for autonomous agents or robots are best described by a specification of the environment and a specification of the available actions the agent or robot can perform. Combining such a specification with the possibility to imperatively program a robot or agent is what we call the actionbased imperative programming. One of the most successful such approaches is Golog. In this paper, we draft a proposal for a new robot programming language YAGI, which is based on the action-based imperative programming paradigm. Our goal is to design a small, portable stand-alone YAGI interpreter. We combine the benefits of a principled domain specification with a clean, small and simple programming language, which does not exploit any side-effects from the implementation language. We discuss general requirements of action-based programming languages and outline YAGI, our action-based language approach which particularly aims at embeddability. Y1 - 2012 N1 - Cognitive Robotics AAAI Technical Report WS-12-06 SP - 24 EP - 31 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Potgieter, Anet T1 - Establishing the RoboCup Standard League in Africa - applying for the RoboCup Standard League with a German-Austrian-South African Research Project Y1 - 2007 SP - 1 EP - 5 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Potgieter, Anet T1 - RoboCup Standard Platform League - Team Zadeat : an intercontinental research effort Y1 - 2008 SP - 1 EP - 7 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald A1 - McPhillips, Graeme A1 - Niemüller, Tim A1 - Potgieter, Anet T1 - Team Zadeat 2009 : team report Y1 - 2009 VL - 6 SP - 1 ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - 20 Years of RoboCup - A Subjective Retrospection JF - KI - Künstliche Intelligenz N2 - This summer, RoboCup competitions were held for the 20th time in Leipzig, Germany. It was the second time that RoboCup took place in Germany, 10 years after the 2006 RoboCup in Bremen. In this article, we give an overview on the latest developments of RoboCup and what happened in the different leagues over the last decade. With its 20th edition, RoboCup clearly is a success story and a role model for robotics competitions. From our personal view point, we acknowledge this by giving a retrospection about what makes RoboCup such a success. Y1 - 2016 U6 - https://doi.org/10.1007/s13218-016-0449-5 SN - 1610-1987 VL - 30 IS - 3 SP - 225 EP - 232 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - Looking back on 20 Years of RoboCup JF - KI - Künstliche Intelligenz Y1 - 2016 U6 - https://doi.org/10.1007/s13218-016-0443-y SN - 1610-1987 VL - 30 IS - 3-4 SP - 321 EP - 323 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - The Interplay of Aldebaran and RoboCup JF - KI - Künstliche Intelligenz Y1 - 2016 U6 - https://doi.org/10.1007/s13218-016-0440-1 SN - 1610-1987 VL - 30 IS - 3-4 SP - 325 EP - 326 PB - Springer CY - Berlin ER - TY - JOUR A1 - Ferrein, Alexander A1 - Steinbauer, Gerald T1 - On the Way to High-Level Programming for Resource-Limited Embedded Systems with Golog JF - Simulation, Modeling, and Programming for Autonomous Robots Y1 - 2010 N1 - Lecture Notes in Computer Science ; 6472 ; Second International Conference, SIMPAR 2010, Darmstadt, Germany, November 15-18, 2010. Proceedings SP - 229 EP - 240 ER -