TY - JOUR A1 - Fiedler, Thomas M. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus temperature limits JF - Medical Physics Y1 - 2017 U6 - http://dx.doi.org/10.1002/mp.12034 N1 - This article is corrected by: Errata: Erratum: “RF safety assessment of a bilateral four-channel transmit/receive 7 Tesla breast coil: SAR versus tissue temperature limits” [Med. Phys. 44(1), 143–157 (2017)] Volume 44, Issue 2, 772 VL - 44 IS - 1 SP - 143 EP - 157 ER - TY - JOUR A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - A method to approximate maximum local SAR in multichannel transmit MR systems without transmit phase information JF - Magnetic Resonance in Medicine N2 - Purpose To calculate local specific absorption rate (SAR) correctly, both the amplitude and phase of the signal in each transmit channel have to be known. In this work, we propose a method to derive a conservative upper bound for the local SAR, with a reasonable safety margin without knowledge of the transmit phases of the channels. Methods The proposed method uses virtual observation points (VOPs). Correction factors are calculated for each set of VOPs that prevent underestimation of local SAR when the VOPs are applied with the correct amplitudes but fixed phases. Results The proposed method proved to be superior to the worst-case calculation based on the maximum eigenvalue of the VOPs. The mean overestimation for six coil setups could be reduced, whereas no underestimation of the maximum local SAR occurred. In the best investigated case, the overestimation could be reduced from a factor of 3.3 to a factor of 1.7. Conclusion The upper bound for the local SAR calculated with the proposed method allows a fast estimation of the local SAR based on power measurements in the transmit channels and facilitates SAR monitoring in systems that do not have the capability to monitor transmit phases Y1 - 2016 U6 - http://dx.doi.org/10.1002/mrm.26398 SN - 1522-2594 VL - 78 IS - 2 SP - 805 EP - 811 PB - International Society for Magnetic Resonance in Medicine ER - TY - JOUR A1 - Chen, Bixia A1 - Schoemberg, Tobias A1 - Kraff, Oliver A1 - Dammann, Philipp A1 - Bitz, Andreas A1 - Schlamann, Marc A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Sure, Ulrich A1 - Wrede, Karsten H. T1 - Cranial fixation plates in cerebral magnetic resonance imaging: a 3 and 7 Tesla in vivo image quality study JF - Magnetic Resonance Materials in Physics, Biology and Medicine N2 - Objective This study assesses and quantifies impairment of postoperative magnetic resonance imaging (MRI) at 7 Tesla (T) after implantation of titanium cranial fixation plates (CFPs) for neurosurgical bone flap fixation. Materials and methods The study group comprised five patients who were intra-individually examined with 3 and 7 T MRI preoperatively and postoperatively (within 72 h/3 months) after implantation of CFPs. Acquired sequences included T₁-weighted magnetization-prepared rapid-acquisition gradient-echo (MPRAGE), T₂-weighted turbo-spin-echo (TSE) imaging, and susceptibility-weighted imaging (SWI). Two experienced neurosurgeons and a neuroradiologist rated image quality and the presence of artifacts in consensus reading. Results Minor artifacts occurred around the CFPs in MPRAGE and T2 TSE at both field strengths, with no significant differences between 3 and 7 T. In SWI, artifacts were accentuated in the early postoperative scans at both field strengths due to intracranial air and hemorrhagic remnants. After resorption, the brain tissue directly adjacent to skull bone could still be assessed. Image quality after 3 months was equal to the preoperative examinations at 3 and 7 T. Conclusion Image quality after CFP implantation was not significantly impaired in 7 T MRI, and artifacts were comparable to those in 3 T MRI. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s10334-016-0548-1 SN - 1352-8661 VL - 29 IS - 3 SP - 389 EP - 398 PB - Springer CY - Berlin ER - TY - JOUR A1 - Maas, Marnix C. A1 - Vos, Eline K. A1 - Lagemaat, Miriam W. A1 - Bitz, Andreas A1 - Orzada, Stephan A1 - Kobus, Thiele A1 - Kraff, Oliver A1 - Maderwald, Stefan A1 - Ladd, Mark E. A1 - Scheenen, Tom W. J. T1 - Feasibility of T₂-weighted turbo spin echo imaging of the human prostate at 7 tesla JF - Magnetic Resonance in Medicine N2 - Purpose To demonstrate that high quality T₂-weighted (T2w) turbo spin-echo (TSE) imaging of the complete prostate can be achieved routinely and within safety limits at 7 T, using an external transceive body array coil only. Methods Nine healthy volunteers and 12 prostate cancer patients were scanned on a 7 T whole-body system. Preparation consisted of B₀ and radiofrequency shimming and localized flip angle calibration. T₁ and T₂ relaxation times were measured and used to define the T2w-TSE protocol. T2w imaging was performed using a TSE sequence (pulse repetition time/echo time 3000–3640/71 ms) with prolonged excitation and refocusing pulses to reduce specific absorption rate. Results High quality T2w TSE imaging was performed in less than 2 min in all subjects. Tumors of patients with gold-standard tumor localization (MR-guided biopsy or prostatectomy) were well visualized on 7 T imaging (n = 3). The number of consecutive slices achievable within a 10-g averaged specific absorption rate limit of 10 W/kg was ≥28 in all subjects, sufficient for full prostate coverage with 3-mm slices in at least one direction. Conclusion High quality T2w TSE prostate imaging can be performed routinely and within specific absorption rate limits at 7 T with an external transceive body array. Y1 - 2014 U6 - http://dx.doi.org/10.1002/mrm.24818 SN - 1522-2594 VL - 71 IS - 5 SP - 1711 EP - 1719 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Umutlu, Lale A1 - Kraff, Oliver A1 - Fischer, Anja A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Nassenstein, Kai A1 - Nensa, Felix A1 - Grüneisen, Johannes A1 - Orzada, Stephan A1 - Bitz, Andreas A1 - Forsting, Michael A1 - Ladd, Mark E. A1 - Lauenstein, Thomas C. T1 - Seven-Tesla MRI of the female pelvis JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-013-2868-0 SN - 1432-1084 VL - 23 IS - 9 SP - 2364 EP - 2373 PB - Springer CY - Berlin ER - TY - JOUR A1 - Bitz, Andreas A1 - Felder, Jorg A1 - Wittig, Tilmann T1 - Designing MRI Coils with Aid of Simulation JF - Microwaves & RF Y1 - 2013 SN - 0745-2993 VL - 52 IS - 7 SP - 56 PB - Penton CY - Cleveland, Ohio ER - TY - JOUR A1 - Kraff, Oliver A1 - Wrede, Karsten H. A1 - Schoemberg, Tobias A1 - Dammann, Philipp A1 - Noureddine, Yacine A1 - Orzada, Stephan A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - MR safety assessment of potential RF heating from cranial fixation plates at 7 T JF - Medical Physics Y1 - 2013 U6 - http://dx.doi.org/10.1118/1.4795347 SN - 2473-4209 VL - 40 IS - 4 SP - 042302-1 EP - 042302-10 PB - Wiley CY - Hoboken ER - TY - JOUR A1 - Umutlu, L. A1 - Maderwald, S. A1 - Kinner, S. A1 - Kraff, O. A1 - Bitz, Andreas A1 - Orzada, S. A1 - Johst, S. A1 - Wrede, K. A1 - Forsting, M. A1 - Ladd, M. E. A1 - Lauenstein, T. C. A1 - Quick, H. H. T1 - First-pass contrast-enhanced renal MRA at 7 Tesla: initial results JF - European Radiology Y1 - 2013 U6 - http://dx.doi.org/10.1007/s00330-012-2666-0 SN - 1432-1084 VL - 23 IS - 4 SP - 1059 EP - 1066 PB - Springer CY - Berlin ER - TY - JOUR A1 - Umutlu, Lale A1 - Orzada, Stephan A1 - Kinner, Sonja A1 - Maderwald, Stefan A1 - Bronte, Irina A1 - Bitz, Andreas A1 - Kraff, Oliver A1 - Ladd, Susanne C. A1 - Antoch, Gerald A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Lauenstein, Thomas C. T1 - Renal imaging at 7 Tesla: preliminary results JF - European Radiology N2 - Objective To investigate the feasibility of 7T MR imaging of the kidneys utilising a custom-built 8-channel transmit/receive radiofrequency body coil. Methods In vivo unenhanced MR was performed in 8 healthy volunteers on a 7T whole-body MR system. After B0 shimming the following sequences were obtained: 1) 2D and 3D spoiled gradient-echo sequences (FLASH, VIBE), 2) T1-weighted 2D in and opposed phase 3) True-FISP imaging and 4) a T2-weighted turbo spin echo (TSE) sequence. Visual evaluation of the overall image quality was performed by two radiologists. Results Renal MRI at 7T was feasible in all eight subjects. Best image quality was found using T1-weighted gradient echo MRI, providing high anatomical details and excellent conspicuity of the non-enhanced vasculature. With successful shimming, B1 signal voids could be effectively reduced and/or shifted out of the region of interest in most sequence types. However, T2-weighted TSE imaging remained challenging and strongly impaired because of signal heterogeneities in three volunteers. Conclusion The results demonstrate the feasibility and diagnostic potential of dedicated 7T renal imaging. Further optimisation of imaging sequences and dedicated RF coil concepts are expected to improve the acquisition quality and ultimately provide high clinical diagnostic value. Y1 - 2011 SN - 1432-1084 VL - 21 IS - 4 SP - 841 EP - 849 PB - Springer CY - Berlin ER - TY - JOUR A1 - Orzada, Stephan A1 - Johst, Sören A1 - Maderwald, Stefan A1 - Bitz, Andreas A1 - Solbach, Klaus A1 - Ladd, Mark E. T1 - Mitigation of B1(+) inhomogeneity on single-channel transmit systems with TIAMO JF - Magnetic Resonance in Medicine Y1 - 2013 U6 - http://dx.doi.org/10.1002/mrm.24453 SN - 1522-2594 VL - 70 IS - 1 SP - 290 EP - 294 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Orzada, S. A1 - Maderwald, S. A1 - Poser, B. A. A1 - Johst, S. A1 - Kannengiesser, S. A1 - Ladd, M. E. A1 - Bitz, Andreas T1 - Time-interleaved acquisition of modes: an analysis of SAR and image contrast implications JF - Magnetic Resonance in Medicine N2 - s the magnetic field strength and therefore the operational frequency in MRI are increased, the radiofrequency wavelength approaches the size of the human head/body, resulting in wave effects which cause signal decreases and dropouts. Especially, whole-body imaging at 7 T and higher is therefore challenging. Recently, an acquisition scheme called time-interleaved acquisition of modes has been proposed to tackle the inhomogeneity problems in high-field MRI. The basic premise is to excite two (or more) different Burn:x-wiley:07403194:media:MRM23081:tex2gif-stack-1 modes using static radiofrequency shimming in an interleaved acquisition, where the complementary radiofrequency patterns of the two modes can be exploited to improve overall signal homogeneity. In this work, the impact of time-interleaved acquisition of mode on image contrast as well as on time-averaged specific absorption rate is addressed in detail. Time-interleaved acquisition of mode is superior in Burn:x-wiley:07403194:media:MRM23081:tex2gif-stack-2 homogeneity compared with conventional radiofrequency shimming while being highly specific absorption rate efficient. Time-interleaved acquisition of modes can enable almost homogeneous high-field imaging throughout the entire field of view in PD, T2, and T2*-weighted imaging and, if a specified homogeneity criterion is met, in T1-weighted imaging as well. Y1 - 2012 U6 - http://dx.doi.org/10.1002/mrm.23081 SN - 1522-2594 VL - 67 IS - 4 SP - 1033 EP - 1041 PB - Wiley-Liss CY - New York ER - TY - CHAP A1 - Granat, Andreas A1 - Höfken, Hans-Wilhelm A1 - Schuba, Marko T1 - Intrusion Detection of the ICS Protocol EtherCAT N2 - Control mechanisms like Industrial Controls Systems (ICS) and its subgroup SCADA (Supervisory Control and Data Acquisition) are a prerequisite to automate industrial processes. While protection of ICS on process management level is relatively straightforward – well known office IT security mechanisms can be used – protection on field bus level is harder to achieve as there are real-time and production requirements like 24x7 to consider. One option to improve security on field bus level is to introduce controls that help to detect and to react on attacks. This paper introduces an initial set of intrusion detection mechanisms for the field bus protocol EtherCAT. To this end existing Ethernet attack vectors including packet injection and man-in-the-middle attacks are tested in an EtherCAT environment, where they could interrupt the EtherCAT network and may even cause physical damage. Based on the signatures of such attacks, a preprocessor and new rule options are defined for the open source intrusion detection system Snort demonstrating the general feasibility of intrusion detection on field bus level. Y1 - 2017 N1 - International Conference on Computer, Network Security and Communication Engineering (CNSCE 2017), March 26-27, 2017, Bangkok, Thailand SP - 1 EP - 5 ER - TY - CHAP A1 - Bitz, Andreas A1 - Klomp, D.W. A1 - Ladd, M.E. T1 - Experimental and numerical determination of SAR and temperature distribution of a human endorectal coil for MR imaging of the prostate at 7T (903.) T2 - 16th annual ISMRM scientific meeting and exhibition 2008 : Toronto, Ontario, Canada, 3 - 9 May 2008 Y1 - 2009 SN - 978-1-61567-196-0 PB - Curran CY - Red Hook, NY ER - TY - CHAP A1 - Bitz, Andreas A1 - El Ouardi, A. A1 - Streckert, J. A1 - Hansen, V. T1 - Efficient calculation of human exposure in front of base station antennas by a combination of the FDTD and Hybrid(2)-method T2 - Proceedings of the 16th International Zurich Symposium on Electromagnetic Compatibility, Topical Meeting on Biomedical EMC, Zurich, Switzerland, February 2005 Y1 - 2005 SN - 3-9521199-9-7 SP - 115 EP - 118 PB - ETH Zentrum CY - Zürich ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V. T1 - Combination of the hybrid/sup [2]/-method and the FDTD-method for solution of boundary value problems with electrically large and high-resolution bodies T2 - IEEE Antennas and Propagation Society International Symposium (IEEE Cat. No.02CH37313) 2002 : San Antonio, TX, USA, 16-21 June Y1 - 2002 SN - 0-7803-7330-8 U6 - http://dx.doi.org/10.1109/APS.2002.1018209 SP - 278 EP - 281 PB - IEEE CY - Piscataway, NJ ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V.W. T1 - Absorption rates inside human body due to radiated electro-magnetic fields of multi-band base station antennas. Boundary value problems with electrically large and high-resolution bodies T2 - 24th BEMS Annual Meeting, Quebec City, Quebec, Canada, 23 - 27 June 2002 Y1 - 2002 SP - 40 EP - 41 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Alaydrus, M. A1 - Streckert, J. A1 - Hansen, V.W. T1 - Base station antennas embedded in complex environments: rf exposure in controlled areas T2 - 23rd BEMS Annual Meeting, St. Paul, Minnesota, USA, 49-50, June 2001 Y1 - 2001 SP - 49 EP - 50 PB - Bioelectromagnetics Society CY - Frederick, MD ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Stogbauer, F. ED - Hietanen, Maila ED - Jokela, Kari ED - Juutilainen, Jukka ED - Työterveyslaitos, T1 - Shielded RF-exposure system with an integrated LF-measuring system for in vitro experiments with cell-layers T2 - Proceedings of the Ebea 2001 : 5th International Congress of the European Bioelectromagnetics Association (Ebea) ; 6-8 September 2001, Marina Congress Center, Helsinki, Finland Y1 - 2001 SN - 9789518024401 SP - 67 EP - 68 PB - Finnish Institute of Occupational Health CY - Helsinki ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J.R. A1 - Hansen, V.W. A1 - Lerchl, A. ED - Danesy, Dorothea T1 - Freely moving or restrained animals in bioelec-tromagnetic experiments - pros and cons T2 - AP 2000 : Millennium Conference on Antennas & Propagation, Davos, Switzerland, 9 - 14 April 2000 Y1 - 2000 IS - Band 1 SP - 489 PB - ESA Publications Division, ESTEC CY - Noordwijk ER - TY - CHAP A1 - Bitz, Andreas A1 - Streckert, J. A1 - Hansen, V.W. A1 - Buschmann, J. T1 - RF exposure of non restrained animals in an overmoded radial waveguide T2 - 22nd BEMS Annual Meeting, Munich, Germany, 2000 Y1 - 2000 SP - 63 PB - Bioelectromagnetics Society CY - Frederick, MD ER -