TY - JOUR A1 - Ciritsis, Alexander A1 - Horbach, Andreas A1 - Staat, Manfred A1 - Kuhl, Christiane K. A1 - Kraemer, Nils Andreas T1 - Porosity and tissue integration of elastic mesh implants evaluated in vitro and in vivo JF - Journal of Biomedical Materials Research: Part B: Applied Biomaterials N2 - Purpose In vivo, a loss of mesh porosity triggers scar tissue formation and restricts functionality. The purpose of this study was to evaluate the properties and configuration changes as mesh deformation and mesh shrinkage of a soft mesh implant compared with a conventional stiff mesh implant in vitro and in a porcine model. Material and Methods Tensile tests and digital image correlation were used to determine the textile porosity for both mesh types in vitro. A group of three pigs each were treated with magnetic resonance imaging (MRI) visible conventional stiff polyvinylidene fluoride meshes (PVDF) or with soft thermoplastic polyurethane meshes (TPU) (FEG Textiltechnik mbH, Aachen, Germany), respectively. MRI was performed with a pneumoperitoneum at a pressure of 0 and 15 mmHg, which resulted in bulging of the abdomen. The mesh-induced signal voids were semiautomatically segmented and the mesh areas were determined. With the deformations assessed in both mesh types at both pressure conditions, the porosity change of the meshes after 8 weeks of ingrowth was calculated as an indicator of preserved elastic properties. The explanted specimens were examined histologically for the maturity of the scar (collagen I/III ratio). Results In TPU, the in vitro porosity increased constantly, in PVDF, a loss of porosity was observed under mild stresses. In vivo, the mean mesh areas of TPU were 206.8 cm2 (± 5.7 cm2) at 0 mmHg pneumoperitoneum and 274.6 cm2 (± 5.2 cm2) at 15 mmHg; for PVDF the mean areas were 205.5 cm2 (± 8.8 cm2) and 221.5 cm2 (± 11.8 cm2), respectively. The pneumoperitoneum-induced pressure increase resulted in a calculated porosity increase of 8.4% for TPU and of 1.2% for PVDF. The mean collagen I/III ratio was 8.7 (± 0.5) for TPU and 4.7 (± 0.7) for PVDF. Conclusion The elastic properties of TPU mesh implants result in improved tissue integration compared to conventional PVDF meshes, and they adapt more efficiently to the abdominal wall. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 827–833, 2018. Y1 - 2018 U6 - https://doi.org/10.1002/jbm.b.33877 SN - 1552-4981 VL - 106 IS - 2 SP - 827 EP - 833 PB - Wiley CY - New York, NY ER - TY - JOUR A1 - Ciobanu, Octavian A1 - Staat, Manfred A1 - Rahimi, Alireza T1 - The use of open source software in biomechanical finite element analysis JF - Buletinul Institutului Politehnic din Iaşi / Universitatea Tehnică Gh. Asachi, Iaşi Secţia 5, Construcţii de maşini = Machine construction = Bulletin of the Polytechnic Institute of Jassy = Izvestija Jasskogo Politechničeskogo Instituta Y1 - 2008 SN - 1011-2855 VL - 54 IS - 7/8 SP - 213 EP - 220 ER - TY - JOUR A1 - Christiaens, P. A1 - Abouzar, Maryam H. A1 - Poghossian, Arshak A1 - Wagner, Torsten A1 - Bijnens, N. A1 - Williams, O. A. A1 - Daenen, M. A1 - Haenen, K. A1 - Schöning, Michael Josef A1 - Wagner, P. T1 - Nanocrystalline diamond-based field-effect capacitive pH sensor JF - Transducers '07 Eurosensors XXI : digest of technical papers ; the14th International Conference on Solid-State Sensors, Actuators and Microsystems, June 10-14, 2007, Lyon, France / Gilles Delapierre (Ed.) Y1 - 2007 SN - 1-4244-0841-5 N1 - Eurosensors 21, 2007, Lyon ; International Conference on Solid-State Sensors, Actuators and Microsystems 14, 2007, Lyon SP - 1891 EP - 1894 PB - IEEE CY - Piscataway ER - TY - JOUR A1 - Christiaens, P. A1 - Abouzar, Maryam H. A1 - Poghossian, Arshak A1 - Wagner, Torsten A1 - Bijnens, N. A1 - Williams, O. A. A1 - Daenen, M. A1 - Haenen, K. A1 - Douthéret, O. A1 - Haen, J. d´ A1 - Mekhalif, Z. A1 - Schöning, Michael Josef A1 - Wagner, P. T1 - pH sensitivity of nanocrystalline diamond films JF - Physica status solidi (A). 204 (2007), H. 9 Y1 - 2007 SN - 0031-8965 SP - 2925 EP - 2930 ER - TY - JOUR A1 - Christ, D. A1 - Hollendung, A. A1 - Larue, H. A1 - Parl, C. A1 - Streun, M. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. T1 - Homogenization of the MultiChannel PM gain by inserting light attenuating masks JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4 N2 - MultiChannel Photomultipliers (PM), like the R7600-00-M64 or R5900-00-M64 from Hamamatsu, are often chosen as photodetectors in high-resolution positron emission tomography (PET). A major problem of this PM is the nonuniform channel gain. In order to solve this problem, light attenuating masks were created. The aim of the masks is a homogenization of the output of all 64 channels using different hole sizes at the channel positions. The hole area, which is individually defined for the different channels, is inversely proportional to the channel gain. The measurements by inserting light attenuating masks improved a homogenization to a ratio of 1:1.2. Y1 - 2004 SN - 1082-3654 SP - 2382 EP - 2385 ER - TY - JOUR A1 - Choi, Chang-Hoon A1 - Felder, Tim A1 - Felder, Jörg A1 - Tellmann, Lutz A1 - Hong, Suk-Min A1 - Wegener, Hans-Peter A1 - Shah, N Jon A1 - Ziemons, Karl T1 - Design, evaluation and comparison of endorectal coils for hybrid MR-PET imaging of the prostate JF - Physics in Medicine & Biology N2 - Prostate cancer is one of the most common cancers among men and its early detection is critical for its successful treatment. The use of multimodal imaging, such as MR-PET, is most advantageous as it is able to provide detailed information about the prostate. However, as the human prostate is flexible and can move into different positions under external conditions, it is important to localise the focused region-of-interest using both MRI and PET under identical circumstances. In this work, we designed five commonly used linear and quadrature radiofrequency surface coils suitable for hybrid MR-PET use in endorectal applications. Due to the endorectal design and the shielded PET insert, the outer face of the coils investigated was curved and the region to be imaged was outside the volume of the coil. The tilting angles of the coils were varied with respect to the main magnetic field direction. This was done to approximate the various positions from which the prostate could be imaged. The transmit efficiencies and safety excitation efficiencies from simulations, together with the signal-to-noise ratios from the MR images were calculated and analysed. Overall, it was found that the overlapped loops driven in quadrature were superior to the other types of coils we tested. In order to determine the effect of the different coil designs on PET, transmission scans were carried out, and it was observed that the differences between attenuation maps with and without the coils were negligible. The findings of this work can provide useful guidance for the integration of such coil designs into MR-PET hybrid systems in the future. Y1 - 2020 U6 - https://doi.org/10.1088/1361-6560/ab87f8 SN - 0031-9155 VL - 65 IS - 11 PB - IOP CY - Bristol ER - TY - JOUR A1 - Chloé, Radermacher A1 - Malyaran, Hanna A1 - Craveiro, Rogerio Bastos A1 - Peglow, Sarah A1 - Behbahani, Mehdi A1 - Pufe, Thomas A1 - Wolf, Michael A1 - Neuss, Sabine T1 - Mechanical loading on cementoblasts: a mini review JF - Osteologie N2 - Orthodontic treatments are concomitant with mechanical forces and thereby cause teeth movements. The applied forces are transmitted to the tooth root and the periodontal ligaments which is compressed on one side and tensed up on the other side. Indeed, strong forces can lead to tooth root resorption and the crown-to-tooth ratio is reduced with the potential for significant clinical impact. The cementum, which covers the tooth root, is a thin mineralized tissue of the periodontium that connects the periodontal ligament with the tooth and is build up by cementoblasts. The impact of tension and compression on these cells is investigated in several in vivo and in vitro studies demonstrating differences in protein expression and signaling pathways. In summary, osteogenic marker changes indicate that cyclic tensile forces support whereas static tension inhibits cementogenesis. Furthermore, cementogenesis experiences the same protein expression changes in static conditions as static tension, but cyclic compression leads to the exact opposite of cyclic tension. Consistent with marker expression changes, the singaling pathways of Wnt/ß-catenin and RANKL/OPG show that tissue compression leads to cementum degradation and tension forces to cementogenesis. However, the cementum, and in particular its cementoblasts, remain a research area which should be explored in more detail to understand the underlying mechanism of bone resorption and remodeling after orthodontic treatments. KW - Cementoblast KW - Compression KW - Tension KW - Mechanotransduction KW - Forces Y1 - 2022 U6 - https://doi.org/10.1055/a-1826-0777 SN - 1019-1291 VL - 31 IS - 2 SP - 111 EP - 118 PB - Thieme CY - Stuttgart ER - TY - JOUR A1 - Chen, Chao A1 - Jost, Peter A1 - Volker, Hanno A1 - Kaminski, Marvin A1 - Wirtssohn, Matti R. A1 - Engelmann, Ulrich M. A1 - Krüger, K. A1 - Schlich, Franziska F. A1 - Schlockermann, Carl A1 - Lobo, Ricardo P.S.M. A1 - Wuttig, Matthias T1 - Dielectric properties of amorphous phase-change materials JF - Physical Review B Y1 - 2017 U6 - https://doi.org/10.1103/PhysRevB.95.094111 SN - 2469-9950 VL - 95 IS - 9 SP - Article number 094111 ER - TY - JOUR A1 - Capri, Miriam A1 - Morsiani, Cristina A1 - Santoro, Aurelia A1 - Moriggi, Manuela A1 - Conte, Maria A1 - Martucci, Morena A1 - Bellavista, Elena A1 - Fabbri, Cristina A1 - Giampieri, Enrico A1 - Albracht, Kirsten A1 - Flück, Martin A1 - Ruoss, Severin A1 - Brocca, Lorenza A1 - Canepari, Monica A1 - Longa, Emanuela A1 - Giulio, Irene Di A1 - Bottinelli, Roberto A1 - Cerretelli, Paolo A1 - Salvioli, Stefano A1 - Gelfi, Cecilia A1 - Franceschi, Claudio A1 - Narici, Marco A1 - Rittweger, Jörn T1 - Recovery from 6-month spaceflight at the International Space Station: muscle-related stress into a proinflammatory setting JF - The FASEB journal : official publication of the Federation of American Societies for Experimental Biology Y1 - 2019 U6 - https://doi.org/10.1096/fj.201801625R VL - 33 IS - 4 SP - 5168 EP - 5180 ER - TY - JOUR A1 - Bäcker, Matthias A1 - Schusser, Sebastian A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Degradationsverhalten bioabbaubarer Polymere : siliziumbasierte Sensorik zur Überwachung JF - GIT Labor-Fachzeitschrift Y1 - 2013 SN - 0016-3538 SP - 32 EP - 33 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Bäcker, Matthias A1 - Schusser, Sebastian A1 - Poghossian, Arshak A1 - Schöning, Michael Josef T1 - Multi-Parametererfassung mit siliziumbasiertem Sensorchip: Aus Drei mach Eins JF - GIT Labor-Fachzeitschrift Y1 - 2014 SN - 0016-3538 IS - 2 SP - 28 EP - 30 PB - Wiley ER - TY - JOUR A1 - Bäcker, Matthias A1 - Raue, Markus A1 - Schusser, Sebastian A1 - Jeitner, C. A1 - Breuer, L. A1 - Wagner, P. A1 - Poghossian, Arshak A1 - Förster, Arnold A1 - Mang, Thomas A1 - Schöning, Michael Josef T1 - Microfluidic chip with integrated microvalves based on temperature- and pH-responsive hydrogel thin films JF - Physica Status Solidi (a) N2 - Two types of microvalves based on temperature-responsive poly(N-isopropylacrylamide) (PNIPAAm) and pH-responsive poly(sodium acrylate) (PSA) hydrogel films have been developed and tested. The PNIPAAm and PSA hydrogel films were prepared by means of in situ photopolymerization directly inside the fluidic channel of a microfluidic chip fabricated by combining Si and SU-8 technologies. The swelling/shrinking properties and height changes of the PNIPAAm and PSA films inside the fluidic channel were studied at temperatures of deionized water from 14 to 36 °C and different pH values (pH 3–12) of Titrisol buffer, respectively. Additionally, in separate experiments, the lower critical solution temperature (LCST) of the PNIPAAm hydrogel was investigated by means of a differential scanning calorimetry (DSC) and a surface plasmon resonance (SPR) method. Mass-flow measurements have shown the feasibility of the prepared hydrogel films to work as an on-chip integrated temperature- or pH-responsive microvalve capable to switch the flow channel on/off. Y1 - 2012 U6 - https://doi.org/10.1002/pssa.201100763 SN - 1862-6319 VL - 209 IS - 5 SP - 839 EP - 845 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Bäcker, Matthias A1 - Rakowski, D. A1 - Poghossian, Arshak A1 - Biselli, Manfred A1 - Wagner, Patrick A1 - Schöning, Michael Josef T1 - Chip-based amperometric enzyme sensor system for monitoring of bioprocesses by flow-injection analysis JF - Journal of Biotechnology N2 - A microfluidic chip integrating amperometric enzyme sensors for the detection of glucose, glutamate and glutamine in cell-culture fermentation processes has been developed. The enzymes glucose oxidase, glutamate oxidase and glutaminase were immobilized by means of cross-linking with glutaraldehyde on platinum thin-film electrodes integrated within a microfluidic channel. The biosensor chip was coupled to a flow-injection analysis system for electrochemical characterization of the sensors. The sensors have been characterized in terms of sensitivity, linear working range and detection limit. The sensitivity evaluated from the respective peak areas was 1.47, 3.68 and 0.28 μAs/mM for the glucose, glutamate and glutamine sensor, respectively. The calibration curves were linear up to a concentration of 20 mM glucose and glutamine and up to 10 mM for glutamate. The lower detection limit amounted to be 0.05 mM for the glucose and glutamate sensor, respectively, and 0.1 mM for the glutamine sensor. Experiments in cell-culture medium have demonstrated a good correlation between the glutamate, glutamine and glucose concentrations measured with the chip-based biosensors in a differential-mode and the commercially available instrumentation. The obtained results demonstrate the feasibility of the realized microfluidic biosensor chip for monitoring of bioprocesses. Y1 - 2013 U6 - https://doi.org/10.1016/j.jbiotec.2012.03.014 SN - 0168-1656 VL - 163 IS - 4 SP - 371 EP - 376 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Bäcker, Matthias A1 - Pouyeshman, S. A1 - Schnitzler, Thomas A1 - Poghossian, Arshak A1 - Wagner, Patrick A1 - Biselli, Manfred A1 - Schöning, Michael Josef T1 - A silicon-based multi-sensor chip for monitoring of fermentation processes JF - Physica status solidi (a) : applications and material science. 208 (2011), H. 6 Y1 - 2011 SN - 1862-6319 SP - 1364 EP - 1369 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Bäcker, Matthias A1 - Poghossian, Arshak A1 - Schöning, Michael Josef A1 - Schnitzler, Thomas A1 - Biselli, Manfred A1 - Zang, Werner A1 - Wagner, P. T1 - Entwicklung eines modularen festkörperbasierten Sensorsystems für die Überwachung von Zellkulturfermentationen JF - Sensoren und Messsysteme 2010 [Elektronische Ressource] : Vorträge der 15. ITG/GMA-Fachtagung vom 18. bis 19. Mai 2010 in Nürnberg / Informationstechnische Gesellschaft im VDE (ITG); VDI/VDE-Gesellschaft Mess- und Automatisierungstechnik (GMA) Y1 - 2010 SN - 978-3-8007-3260-9 N1 - Fachtagung Sensoren und Messsysteme 15, 2010, Nürnberg ; Gesellschaft Mess- und Automatisierungstechnik SP - 688 EP - 698 PB - VDE Verlag CY - Berlin ER - TY - JOUR A1 - Bäcker, Matthias A1 - Poghossian, Arshak A1 - Abouzar, Maryam H. A1 - Wenmackers, Sylvia A1 - Janssens, Stoffel D. A1 - Haenen, Ken A1 - Wagner, Patrick A1 - Schöning, Michael Josef T1 - Capacitive field-effect (bio-)chemical sensors based on nanocrystalline diamond films JF - Diamond Electronics and Bioelectronics — Fundamentals to Applications III, edited by P. Bergonzo, [u.a.] Y1 - 2010 N1 - MRS Proceedings Volume 1203 paper 1203-J17-31 ; Mater. Res. Soc. Sympos. Proc. Vol 1203 (2010) ; Materials Research Society SP - 1 EP - 6 ER - TY - JOUR A1 - Bäcker, Matthias A1 - Kramer, F. A1 - Huck, Christina A1 - Poghossian, Arshak A1 - Bratov, A. A1 - Abramova, N. A1 - Schöning, Michael Josef T1 - Planar and 3D interdigitated electrodes for biosensing applications: The impact of a dielectric barrier on the sensor properties JF - Physica Status Solidi (A) - Applications and Materials Science N2 - Planar and three-dimensional (3D) interdigitated electrodes (IDE) with electrode digits separated by an insulating barrier of different heights were electrochemically characterized and compared in terms of their sensing properties. Due to the impact of the surface resistance, both types of IDE structures display a non-linear behavior in low-ionic strength solutions. The experimental data were fitted to an electrical equivalent circuit and interpreted taking into account the surface-charge-governed properties. The effect of a charged polyelectrolyte layer electrostatically assembled onto the sensor surface on the surface resistance in solutions with different KCl concentration is studied. In case of the same electrode footprint, 3D-IDEs show a larger cell constant and a higher sensitivity to molecular adsorption than that of planar IDEs. The obtained results demonstrate the potential of 3D-IDEs as a new transducer structure for a direct label-free sensing of charged molecules. Y1 - 2014 U6 - https://doi.org/10.1002/pssa.201330416 SN - 1521-396X (E-Journal); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) VL - 211 IS - 6 SP - 1357 EP - 1363 PB - Wiley-VCH CY - Weinheim ER - TY - JOUR A1 - Bäcker, Matthias A1 - Kowollik, Silvia A1 - Schnitzler, Thomas A1 - Wang, J. A1 - Poghossian, Arshak A1 - Zang, Werner A1 - Biselli, Manfred A1 - Wagner, P. A1 - Schöning, Michael Josef T1 - Entwicklung eines festkörperbasierten Sensorsystems für die Bioprozessautomatisierung JF - 9. Dresdner Sensor-Symposium : Dresden, 07.-09. Dezember 2009 / Gerlach, Gerald ; Hauptmann, Peter [Hrsg.] Y1 - 2009 SN - 978-3-941298-44-6 N1 - Dresdner Sensor-Symposium ; (9, 2009, Dresden) SP - 197 EP - 200 PB - TUDpress CY - Dresden ER - TY - JOUR A1 - Bäcker, Matthias A1 - Koch, Claudia A1 - Eiben, Sabine A1 - Geiger, Fania A1 - Eber, Fabian A1 - Gliemann, Hartmut A1 - Poghossian, Arshak A1 - Wege, Christina A1 - Schöning, Michael Josef T1 - Tobacco mosaic virus as enzyme nanocarrier for electrochemical biosensors JF - Sensors and Actuators B: Chemical N2 - The conjunction of (bio-)chemical recognition elements with nanoscale biological building blocks such as virus particles is considered as a very promising strategy for the creation of biohybrids opening novel opportunities for label-free biosensing. This work presents a new approach for the development of biosensors using tobacco mosaic virus (TMV) nanotubes or coat proteins (CPs) as enzyme nanocarriers. Sensor chips combining an array of Pt electrodes loaded with glucose oxidase (GOD)-modified TMV nanotubes or CP aggregates were used for amperometric detection of glucose as a model system for the first time. The presence of TMV nanotubes or CPs on the sensor surface allows binding of a high amount of precisely positioned enzymes without substantial loss of their activity, and may also ensure accessibility of their active centers for analyte molecules. Specific and efficient immobilization of streptavidin-conjugated GOD ([SA]-GOD) complexes on biotinylated TMV nanotubes or CPs was achieved via bioaffinity binding. These layouts were tested in parallel with glucose sensors with adsorptively immobilized [SA]-GOD, as well as [SA]-GOD crosslinked with glutardialdehyde, and came out to exhibit superior sensor performance. The achieved results underline a great potential of an integration of virus/biomolecule hybrids with electronic transducers for future applications in biosensorics and biochips. Y1 - 2017 U6 - https://doi.org/10.1016/j.snb.2016.07.096 SN - 0925-4005 VL - 238 SP - 716 EP - 722 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Bäcker, Matthias A1 - Delle, L. A1 - Poghossian, Arshak A1 - Biselli, Manfred A1 - Zang, Werner A1 - Wagner, P. A1 - Schöning, Michael Josef T1 - Electrochemical sensor array for bioprocess monitoring JF - Electrochimica Acta (2011) Y1 - 2011 VL - 56 IS - 26 SP - 9673 EP - 9678 PB - Elsevier CY - Amsterdam ER -