TY - JOUR A1 - Hasan, Istabrak A1 - Keil, Ludger A1 - Staat, Manfred A1 - Wahl, Gerhard A1 - Bourauel, Christoph T1 - Determination of the frictional coefficient of the implant-antler interface : experimental approach JF - Biomedical Engineering / Biomedizinische Technik N2 - The similar bone structure of reindeer antler to human bone permits studying the osseointegration of dental implants in the jawbone. As the friction is one of the major factors that have a significant influence on the initial stability of immediately loaded dental implants, it is essential to define the frictional coefficient of the implant-antler interface. In this study, the kinetic frictional forces at the implant-antler interface were measured experimentally using an optomechanical setup and a stepping motor controller under different axial loads and sliding velocities. The corresponding mean values of the static and kinetic frictional coefficients were within the range of 0.5–0.7 and 0.3–0.5, respectively. An increase in the frictional forces with increasing applied axial loads was registered. The measurements showed an evidence of a decrease in the magnitude of the frictional coefficient with increasing sliding velocity. The results of this study provide a considerable assessment to clarify the suitable frictional coefficient to be used in the finite element contact analysis of antler specimens. Y1 - 2012 SN - 1862-278X VL - 57 IS - 5 SP - 359 EP - 363 PB - De Gruyter CY - Berlin ER -