TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Larue, H. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - A PET system based on data processing of free-running sampled pulses JF - 2001 IEEE Nuclear Science Symposium Conference Record, Vol. 2 N2 - Within the developments for the Crystal Clear small animal PET project (CLEARPET) a dual head PET system has been established. The basic principle is the early digitization of the detector pulses by free running ADCs. The determination of the γ-energy and also the coincidence detection is performed by data processing of the sampled pulses on the host computer. Therefore a time mark is attached to each pulse identifying the current cycle of the 40 MHz sampling clock. In order to refine the time resolution the pulse starting time is interpolated from the samples of the pulse rise. The detector heads consist of multichannel PMTs with a single LSO scintillator crystal coupled to each channel. For each PMT only one ADC is required. The position of an event is obtained separately from trigger signals generated for each single channel. An FPGA is utilized for pulse buffering, generation of the time mark and for the data transfer to the host via a fast I/O-interface. Y1 - 2002 SN - 1082-3654 SP - 693 EP - 694 ER - TY - JOUR A1 - Heinrichs, U. A1 - Pietrzyk, U. A1 - Ziemons, Karl T1 - Design optimization of the PMT-ClearPET prototypes based on simulation studies with GEANT3 JF - 2002 IEEE Nuclear Science Symposium Conference Record, Vol. 3 N2 - Within the Crystal Clear Collaboration four centres are developing 2nd generation high performance small animal PET scanners for different kinds of animals and medical applications. The first prototypes are PMT-based systems including depth of interaction (DOI) detection by using a phoswich layer of LSO and LuYAP. The aim of these simulation studies is to optimize sensitivity and spatial resolution of given designs, which vary in FOVs caused by different detector configurations (ring/octagon) and sizes. For this purpose the simulation tool GEANT3 (CERN) was used. The simulations have shown that all PMT designs with one-to-one coupling of crystals have a very nonlinear axial sensitivity profile. By shifting every other PMT 1/4 of a PMT length in axial direction the sampling of the FOVs became more homogeneous. At an energy threshold of 350keV the regression coefficient increases from 0.818 for the non-shifted to 0.993 for the shifted design. Simulations of a point source centred in the FOV (threshold: 350keV) resulted in sensitivities of 4.2% for a 4×20PMT (LSO/LuYAP a 10mm) and 3.8% for a 4×16PMT (LSO/LuYAP a 8mm) ring design. The 3D-MLEM reconstruction of a point source shows the enormous improvement of resolution using a crystal double layer with DOI (3.1mm at 40mm from CFOV) instead of a 20mm single layer (11.9mm). Y1 - 2003 SN - 1082-3654 SP - 682 EP - 686 ER - TY - JOUR A1 - Heinrichs, U. A1 - Pietrzyk, U. A1 - Ziemons, Karl T1 - Design optimization of the PMT-ClearPET prototypes based on simulation studies with GEANT3 JF - IEEE Transactions on Nuclear Science N2 - Within the Crystal Clear Collaboration (CCC), four centers are developing second generation high performance small animal positron emission tomography (PET) scanners for different kinds of animals and medical applications. The first prototypes are photomultiplier tube (PMT)-based systems including depth of interaction (DOI) detection by using a phoswich layer of lutetium oxyorthosilicate (LSO) and lutetium yttrium aluminum perovskite (LuYAP). The aim of these simulation studies is to optimize sensitivity and spatial resolution of given designs, which vary in fields of view (FOVs) caused by different detector configurations (ring/octagon) and sizes. For this purpose the simulation tool GEANT3 (CERN, Geneva, Switzerland) was used. Y1 - 2003 SN - 0018-9499 VL - 50 IS - 5 SP - 1428 EP - 1432 ER - TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Larue, H. A1 - Saleh, H. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - Pulse shape discrimination of LSO and LuYAP scintillators for depth of interaction detection in PET JF - 2002 IEEE Nuclear Science Symposium Conference Record, Vol. 3 N2 - A feasible way to gain the depth of interaction information in a PET scanner is the use of phoswich detectors. In general the layer of interaction is identified front the pulse shape of the corresponding scintillator material. In this work pulses from LSO and LuYAP crystals were investigated in order to find a practical method of distinguishing. It turned out that such a pulse processing could he kept simple due to an additional slow component in the light decay of the LuYAP pulse. At the same time the short decay time guarantees that the major amount of the light output is still collected within a short pulse recording time. Y1 - 2003 SN - 1082-3654 SP - 1636 EP - 1639 ER - TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Larue, H. A1 - Saleh, H. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - Pulse shape discrimination of LSO and LuYAP scintillators for depth of interaction detection in PET JF - IEEE Transactions on Nuclear Science N2 - A feasible way to gain the depth of interaction information in a positron emission tomography scanner is the use of phoswich detectors. In general, the layer of interaction is identified from the pulse shape of the corresponding scintillator material. In this work, pulses from LSO and LuYAP crystals were investigated in order to find a practical method of distinguishing. It turned out that such a pulse processing could be kept simple because of an additional slow component in the light decay of the LuYAP pulse. At the same time, the short decay time guarantees that the major amount of the light output is still collected within a short pulse recording time. Y1 - 2003 SN - 0018-9499 VL - 50 IS - 3 SP - 344 EP - 347 ER - TY - JOUR A1 - Khodaverdi, M. A1 - Chaziioannou, A. F. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. A1 - Pietrzyk, U. T1 - Investigation of different microCT scanner configurations by GEANT4 simulations JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4 N2 - This study has been performed to design the combination of the new ClearPET TM (ClearPET is a trademark of the Crystal Clear Collaboration), a small animal Positron Emission Tomography (PET) system, with a microComputed Tomography (microCT) scanner. The properties of different microCT systems have been determined by simulations based on GEANT4. We demonstrate the influence of the detector material and the X-ray spectrum on the obtained contrast. Four different detector materials (selenium, cadmium zinc telluride, cesium iodide and gadolinium oxysulfide) and two X-ray spectra (a molybdenum and a tungsten source) have been considered. The spectra have also been modified by aluminum filters of varying thickness. The contrast between different tissue types (water, air, brain, bone and fat) has been simulated by using a suitable phantom. The results indicate the possibility to improve the image contrast in microCT by an optimized combination of the X-ray source and detector material. Y1 - 2004 SN - 1082-3654 SP - 2989 EP - 2993 ER - TY - JOUR A1 - Ziemons, Karl A1 - Auffray, E. A1 - Barbier, R. A1 - Brandenburg, G. T1 - The ClearPET TM LSO/LuYAP phoswich scanner: a high performance small animal PET system JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 3 N2 - A 2nd generation high performance small animal PET scanner, called ClearPET™, has been designed and a first prototype is built by working groups of the Crystal Clear Collaboration (CCC). In order to achieve high sensitivity and maintain good uniform spatial resolution over the field of view in high resolution PET systems, it is necessary to extract the depth of interaction (DOI) information and correct for spatial degradation. The design of the first ClearPET™ Demonstrator based on the use of the multi-anode photomultiplier tube (Hamamatsu R7600-M64) and a LSO/LuYAP phoswich matrix. The two crystal layers of 8*8 crystals (2*2*10 mm3) are stacked on each other and mounted without light guide as one to one on the PMT. A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the crystal ring is 137 mm, the axial detector length is 110 mm. The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the pulse arrival time. Single gamma interactions are recorded and coincidences are found by software. The gantry allows rotation of the detector modules around the field of view. The measurements have been done using the first LSO/LuYAP detector cassettes. Y1 - 2004 SN - 1082-3654 SP - 1728 EP - 1732 ER - TY - JOUR A1 - Streun, M. A1 - Brandenburg, G. A1 - Brökel, M. A1 - Fuss, L. A1 - Larue, H. A1 - Parl, C. A1 - Zimmermann, E. A1 - Ziemons, Karl A1 - Halling, H. T1 - The ClearPET data acquisition JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 5 N2 - Within the Crystal Clear Collaboration a modular system for a small animal PET scanner (ClearPET™) has been developed. The modularity allows the assembly of scanners of different sizes and characteristics in order to fit the specific needs of the individual member institutions. Now a first demonstrator is being completed in Julich. The system performs depth of interaction detection by using a phoswich arrangement combining LSO and LuYAP scintillators which are coupled to multi-channel photomultipliers (PMTs). A free-running ADC digitizes the signal from the PMT and the complete scintillation pulses are sampled by an FPGA and sent with 20 MB/S to a PC for preprocessing. The pulse provides information about the gamma energy and the scintillator material which identifies the interaction layer. Furthermore, the exact pulse starting time is obtained from the sampled data. This is important as no hardware coincidence detection is implemented. All single events are recorded and coincidences are identified by software. An advantage of that is that the coincidence window and the dimensions of the field of view can be adjusted easily. The ClearPET™ demonstrator is equipped with 10240 crystals on 80 PMTs. This paper presents an overview of the data acquisition system. Y1 - 2004 SN - 1082-3654 SP - 3097 EP - 3100 ER - TY - JOUR A1 - Christ, D. A1 - Hollendung, A. A1 - Larue, H. A1 - Parl, C. A1 - Streun, M. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. T1 - Homogenization of the MultiChannel PM gain by inserting light attenuating masks JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 4 N2 - MultiChannel Photomultipliers (PM), like the R7600-00-M64 or R5900-00-M64 from Hamamatsu, are often chosen as photodetectors in high-resolution positron emission tomography (PET). A major problem of this PM is the nonuniform channel gain. In order to solve this problem, light attenuating masks were created. The aim of the masks is a homogenization of the output of all 64 channels using different hole sizes at the channel positions. The hole area, which is individually defined for the different channels, is inversely proportional to the channel gain. The measurements by inserting light attenuating masks improved a homogenization to a ratio of 1:1.2. Y1 - 2004 SN - 1082-3654 SP - 2382 EP - 2385 ER - TY - JOUR A1 - Ziemons, Karl A1 - Heinrichs, U. A1 - Streun, M. A1 - Pietrzyk, U. T1 - Validation of GEANT3 simulation studies with a dual-head PMT ClearPET™ prototype JF - 2003 IEEE Nuclear Science Symposium Conference Record, Vol. 5 N2 - The ClearPET™ project is proposed by working groups of the Crystal Clear Collaboration (CCC) to develop a 2nd generation high performance small animal positron emission tomograph (PET). High sensitivity and high spatial resolution is foreseen for the ClearPET™ camera by using a phoswich arrangement combining mixed lutetium yttrium aluminum perovskite (LuYAP:Ce) and lutetium oxyorthosilicate (LSO) scintillating crystals. Design optimizations for the first photomultiplier tube (PMT) based ClearPET camera are done with a Monte-Carlo simulation package implemented on GEANT3 (CERN, Geneva, Switzerland). A dual-head prototype has been built to test the frontend electronics and was used to validate the implementation of the GEANT3 simulation tool. Multiple simulations were performed following the experimental protocols to measure the intrinsic resolution and the sensitivity profile in axial and radial direction. Including a mean energy resolution of about 27.0% the simulated intrinsic resolution is about (1.41±0.11)mm compared to the measured of (1.48±0.06)mm. The simulated sensitivity profiles show a mean square deviation of 12.6% in axial direction and 3.6% in radial direction. Satisfactorily these results are representative for all designs and confirm the scanner geometry. Y1 - 2004 SN - 1082-3654 SP - 3053 EP - 3056 ER - TY - JOUR A1 - Auffray, E. A1 - Bruyndonckx, P. A1 - Devroede, O. A1 - Fedorov, A. A1 - Ziemons, Karl T1 - The ClearPET project JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - The Crystal Clear Collaboration has designed and is building a high-resolution small animal PET scanner. The design is based on the use of the Hamamatsu R7600-M64 multi-anode photomultiplier tube and a LSO/LuYAP phoswich matrix with one to one coupling between the crystals and the photo-detector. The complete system will have 80 PM tubes in four rings with an inner diameter of 137 mm and an axial field of view of 110 mm. The PM pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the pulse arrival time. Single gamma interactions are recorded and coincidences are found by software. The gantry allows rotation of the detector modules around the field of view. Simulations, and measurements a 2×4 module test set-up predict a spatial resolution of 1.5 mm in the centre of the field of view and a sensitivity of 5.9% for a point source in the centre of the field of view. Y1 - 2004 SN - 0168-9002 N1 - Proceedings of the 2nd International Conference on Imaging Technologies in Biomedical Sciences VL - 527 IS - 1-2 SP - 171 EP - 174 ER - TY - JOUR A1 - Streun, M. A1 - Christ, D. A1 - Hellendung, A. A1 - Larue, H. A1 - Ziemons, Karl A1 - Halling, H. T1 - Effects of crosstalk and gain nonuniformity using multichannel PMTs in the Clearpet® scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - The ClearPET® scanners developed by the Crystal Clear Collaboration use multichannel PMTs as photodetectors with scintillator pixels coupled individually to each channel. In order to localize an event each channel anode is connected to a comparator that triggers when the anode signal exceeds a common predefined threshold. Two major difficulties here are crosstalk of light and the gain nonuniformity of the PMT channels. Crosstalk can generate false triggering in channels adjacent to the actual event. On the one hand this can be suppressed by sufficiently increasing the threshold, but on the other hand a threshold too high can already prevent valid events on the lower gain channels from being detected. Finally, both effects restrict the dynamic range of pulse heights that can be processed. The requirements to the dynamic range are not low as the ClearPET® scanners detect the depth of interaction by phoswich pixels consisting of LSO and Lu0.7Y0.3AP, two scintillators with different light yields. We will present a model to estimate the achievable dynamic range and show solutions to increase it. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 402 EP - 405 ER - TY - JOUR A1 - Ziemons, Karl A1 - Auffray, E. A1 - Barbier, R. A1 - Brandenburg, G. A1 - Bruyndonckx, P. T1 - The ClearPET™ project: Development of a 2nd generation high-performance small animal PET scanner JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment N2 - Second generation high-performance PET scanners, called ClearPET™1, have been developed by working groups of the Crystal Clear Collaboration (CCC). High sensitivity and high spatial resolution for the ClearPET camera is achieved by using a phoswich arrangement combining two different types of lutetium-based scintillator materials: LSO from CTI and LuYAP:Ce from the CCC (ISTC project). In a first ClearPET prototype, phoswich arrangements of 8×8 crystals of 2×2×10 mm3 are coupled to multi-channel photomultiplier tubes (Hamamatsu R7600). A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the ring is 120 mm, the axial detector length is 110 mm.The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the exact pulse starting time, which is subsequently used for coincidence detection. The gantry allows rotation of the detector modules around the field of view. Preliminary data shows a correct identification of the crystal layer about (98±1)%. Typically the energy resolution is (23.3±0.5)% for the luyap layer and (15.4±0.4)% for the lso layer. early studies showed the timing resolution of 2 ns FWHM and 4.8 ns FWTM. the intrinsic spatial resolution ranges from 1.37 mm to 1.61 mm full-width of half-maximum (FWHM) with a mean of 1.48 mm FWHM. further improvements in image and energy resolution are expected when the system geometry is fully modeled. Y1 - 2005 SN - 0168-9002 N1 - Proceedings of the 7th International Conference on Inorganic Scintillators and their Use in Scientific and Industrial Applications VL - 537 IS - 1-2 SP - 307 EP - 311 ER - TY - JOUR A1 - Beer, S. A1 - Streun, M. A1 - Hombach, T. A1 - Buehler, J. A1 - Jahnke, S. A1 - Khodaverdi, M. A1 - Larue, H. A1 - Minwuyelet, S. A1 - Parl, C. A1 - Roeb, G. A1 - Schurr, U. A1 - Ziemons, Karl T1 - Design and initial performance of PlanTIS: a high-resolution positron emission tomograph for plants JF - Physics in Medicine and Biology N2 - Positron emitters such as 11C, 13N and 18F and their labelled compounds are widely used in clinical diagnosis and animal studies, but can also be used to study metabolic and physiological functions in plants dynamically and in vivo. A very particular tracer molecule is 11CO2 since it can be applied to a leaf as a gas. We have developed a Plant Tomographic Imaging System (PlanTIS), a high-resolution PET scanner for plant studies. Detectors, front-end electronics and data acquisition architecture of the scanner are based on the ClearPET™ system. The detectors consist of LSO and LuYAP crystals in phoswich configuration which are coupled to position-sensitive photomultiplier tubes. Signals are continuously sampled by free running ADCs, and data are stored in a list mode format. The detectors are arranged in a horizontal plane to allow the plants to be measured in the natural upright position. Two groups of four detector modules stand face-to-face and rotate around the field-of-view. This special system geometry requires dedicated image reconstruction and normalization procedures. We present the initial performance of the detector system and first phantom and plant measurements. Y1 - 2010 U6 - https://doi.org/10.1088/0031-9155/55/3/006 SN - 1361-6560 VL - 55 IS - 3 SP - 635 EP - 646 PB - IOP CY - Bristol ER - TY - JOUR A1 - Herzog, Hans A1 - Pietrzyk, Uwe A1 - Shah, N. Jon A1 - Ziemons, Karl T1 - The current state, challenges and perspectives of MR-PET JF - Neuroimage N2 - Following the success of PET/CT during the last decade and the recent increasing proliferation of SPECT/CT, another hybrid imaging instrument has been gaining more and more interest: MR-PET. First combined, simultaneous PET and MR studies carried out in small animals demonstrated the feasibility of the new approach. Concurrently, some prototypes of an MR-PET scanner for simultaneous human brain studies have been built, their performance is being tested and preliminary applications have already been shown. Through this pioneering work, it has become clear that advances in the detector design are necessary for further optimization. Recently, the different issues related to the present state and future prospects of MR-PET were presented and discussed during an international 2-day workshop at the Forschungszentrum Jülich, Germany, held after, and in conjunction with, the 2008 IEEE Nuclear Science Symposium and Medical Imaging Conference in Dresden, Germany on October 27–28, 2008. The topics ranged from small animal MR-PET imaging to human MR-BrainPET imaging, new detector developments, challenges/opportunities for ultra-high field MR-PET imaging and considerations of possible future research and clinical applications. This report presents a critical summary of the contributions made to the workshop. Y1 - 2010 U6 - https://doi.org/10.1016/j.neuroimage.2009.10.036 SN - 1053-8119 VL - 49 IS - 3 SP - 2072 EP - 2082 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Parl, C. A1 - Larue, H. A1 - Streun, M. A1 - Ziemons, Karl T1 - Double-side-readout technique for SiPM-matrices JF - 2010 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) N2 - In our case the double-side-method is used to minimize the complexity of a matrix-readout. Here the number of channels is reduced to 2√N̅. It is also possible to benefit from the method in a single pixel readout system. One signal can be used to measure position and energy of the event, the other one can be applied to a fast trigger-circuit at the same time. In a next step we will investigate timing behavior and electrical crosstalk of the circuit. Y1 - 2011 SN - 1095-7863 SP - 1486 EP - 1487 PB - IEEE CY - New York ER - TY - CHAP A1 - Streun, M. A1 - Al-Kaddoum, R. A1 - Parl, C. A1 - Pietrzyk, U. A1 - Ziemons, Karl A1 - Waasen, S. van T1 - Simulation studies of optical photons in monolithic block scintillators T2 - 2011 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) N2 - The interest in PET detectors with monolithic block scintillators is growing. In order to obtain high spatial resolutions dedicated positioning algorithms are required. But even an ideal algorithm can only deliver information which is provided by the detector. In this simulation study we investigated the light distribution on one surface of cuboid LSO scintillators of different size. Scintillators with a large aspect ratio (small footprint and large height) showed significant position information only for a minimum interaction depth of the gamma particle. The results allow a quantitative estimate for a useful aspect ratio. Y1 - 2012 SN - 978-1-4673-0120-6 (electronic ISBN) SN - 978-1-4673-0118-3 (print ISBN) U6 - https://doi.org/10.1109/NSSMIC.2011.6154621 SP - 1380 EP - 1382 PB - IEEE CY - New York ER - TY - JOUR A1 - Gebhardt, Andreas A1 - Fateri, Miranda T1 - 3D printing and its applications JF - RTejournal - Forum für Rapid Technologie N2 - Eine zunehmende Anzahl von Artikeln in Publikumszeitschriften und Journalen rückt die direkte Herstellung von Bauteilen und Figuren immer mehr in das Bewusstsein einer breiten Öffentlichkeit. Leider ergibt sich nur selten ein einigermaßen vollständiges Bild davon, wie und in welchen Lebensbereichen diese Techniken unseren Alltag verändern werden. Das liegt auch daran, dass die meisten Artikel sehr technisch geprägt sind und sich nur punktuell auf Beispiele stützen. Dieser Beitrag geht von den Bedürfnissen der Menschen aus, wie sie z.B. in der Maslow’schen Bedürfnispyramide strukturiert dargestellt sind und unterstreicht dadurch, dass 3D Printing (oder Additive Manufacturing resp. Rapid Prototyping) bereits alle Lebensbereiche erfasst hat und im Begriff ist, viele davon zu revolutionieren. N2 - An increasing amount of popular articles focus on making models and sculptures by 3D Printing thus making more and more even private users aware of this technology. Unfortunately they mostly draw an incomplete picture of how our daily life will be influenced by this new technology. Often this is caused by a very technical point of view based on not very representative examples. This article focuses on the peoples needs as they have been structured by the so-called Maslow pyramid. Doing so, it underlines that 3D Printing (called Additive Manufacturing or Rapid Prototyping as well) already touches all aspects of life and is about to revolutionize most of them. Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:0009-2-35626 SN - 1614-0923 VL - 10 IS - 1 PB - Fachhochschule Aachen CY - Aachen ER - TY - JOUR A1 - Kirchner, Patrick A1 - Oberländer, Jan A1 - Suso, Henri-Pierre A1 - Rysstad, Gunnar A1 - Keusgen, Michael A1 - Schöning, Michael Josef T1 - Towards a wireless sensor system for real-time H2O2 monitoring in aseptic food processes JF - Physica status solidi (a) N2 - A wireless sensor system based on the industrial ZigBee standard for low-rate wireless networking was developed that enables real-time monitoring of gaseous H2O2 during the package sterilization in aseptic food processes. The sensor system consists of a remote unit connected to a calorimetric gas sensor, which was already established in former works, and an external base unit connected to a laptop computer. The remote unit was built up by an XBee radio frequency (RF) module for data communication and a programmable system-on-chip controller to read out the sensor signal and process the sensor data, whereas the base unit is a second XBee RF module. For the rapid H2O2 detection on various locations inside the package that has to be sterilized, a novel read-out strategy of the calorimetric gas sensor was established, wherein the sensor response is measured within the short sterilization time and correlated with the present H2O2 concentration. In an exemplary measurement application in an aseptic filling machinery, the suitability of the new, wireless sensor system was demonstrated, wherein the influence of the gas velocity on the H2O2 distribution inside a package was determined and verified with microbiological tests. KW - calorimetric gas sensor;hydrogen peroxide;wireless sensor system Y1 - 2013 U6 - https://doi.org/10.1002/pssa.201200920 SN - 1862-6319 VL - 210 IS - 5 SP - 877 EP - 883 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Khodaverdi, M. A1 - Chatziioannou, A. F. A1 - Weber, S. A1 - Ziemons, Karl A1 - Halling, H. A1 - Pietrzyk, U. T1 - Investigation of different MicroCT scanner configurations by GEANT4 simulations JF - IEEE Transactions on Nuclear Science N2 - This study has been performed to design the combination of the new ClearPET (ClearPET is a trademark of the Crystal Clear Collaboration), a small animal positron emission tomography (PET) system, with a micro-computed tomography (microCT) scanner. The properties of different microCT systems have been determined by simulations based on GEANT4. We will demonstrate the influence of the detector material and the X-ray spectrum on the obtained contrast. Four different detector materials (selenium, cadmium zinc telluride, cesium iodide and gadolinium oxysulfide) and two X-ray spectra (a molybdenum and a tungsten source) have been considered. The spectra have also been modified by aluminum filters of varying thickness. The contrast between different tissue types (water, air, brain, bone and fat) has been simulated by using a suitable phantom. The results indicate the possibility to improve the image contrast in microCT by an optimized combination of the X-ray source and detector material. Y1 - 2005 SN - 0018-9499 VL - 52 IS - 1 SP - 188 EP - 192 ER -