TY - JOUR A1 - Hackl, Michael A1 - Kahmann, Stephanie Lucina A1 - Wegmann, Kilian A1 - Ries, Christian A1 - Staat, Manfred A1 - Müller, Lars-Peter T1 - Shortening osteotomy of the proximal radius — a treatment option for isolated osteoarthritis of the lateral column of the elbow joint? JF - Knee surgery, sports traumatology, arthroscopy N2 - Treatment of posttraumatic osteoarthritis of the radial column of the elbow joint remains a challenging yet common issue. While partial joint replacement leads to high revision rates, radial head excision has shown to severely increase joint instability. Shortening osteotomy of the radius could be an option to decrease the contact pressure of the radiohumeral joint and thereby pain levels without causing valgus instability. Hence, the aim of this biomechanical study was to evaluate the effects of radial shortening on axial load distribution and valgus stability of the elbow joint. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s00167-016-4080-7 SN - 0942-2056 VL - Volume 24 IS - Supplement 1 SP - 128 EP - 129 PB - Springer CY - Berlin ER - TY - JOUR A1 - Hackl, Michael A1 - Leschinger, T. A1 - Staat, Manfred A1 - Müller, Lars-Peter A1 - Wegmann, Kilian T1 - Reconstruction of the interosseous membrane in the Essex Lopresti lesion — a biomechanical evaluation JF - Knee surgery, sports traumatology, arthroscopy N2 - Surgical reconstruction of the interosseous membrane (IOM) could restore longitudinal forearm stability to avoid persisting disability due to capituloradial and ulnocarpal impingement in Essex Lopresti lesions. This biomechanical study aimed to assess longitudinal forearm stability of intact specimens, after sectioning of the IOM and after reconstruction with a TightRope construct using either a single or double bundle technique. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s00167-016-4080-7 SN - 0942-2056 VL - Volume 24 IS - Supplement 1 SP - 130 EP - 131 PB - Springer CY - Berlin ER - TY - JOUR A1 - Hackl, Michael A1 - Müller, Lars-Peter A1 - Staat, Manfred A1 - Kahmann, Stephanie Lucina A1 - Wegmann, Kilian T1 - Proximal phalangeal neck fractures of the hand — a biomechanical comparison of three fixation techniques JF - Knee surgery, sports traumatology, arthroscopy N2 - Plate osteosynthesis of displaced proximal phalangeal neck fractures of the hand allows early mobilization due to a stable internal fixation. Nevertheless, joint stiffness—because of soft tissue irritation—represents a common complication leading to high complication rates. Del Pinal et al. recently reported promising clinical results for a new, minimally invasive fixation technique with a cannulated headless intramedullary compression screw. Hence, the aim of this study was to compare plate fixation of proximal phalangeal neck fractures to less two less invasive techniques: Crossed k-wire fixation and intramedullary screw fixation. We hypothesized that these fixation techniques provide inferior stability when compared to plate osteosynthesis. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s00167-016-4080-7 SN - 0942-2056 VL - Volume 24 IS - Supplement 1 SP - 148 EP - 149 PB - Springer CY - Berlin ER - TY - CHAP A1 - Hafner, David A1 - Ochs, Peter A1 - Weickert, Joachim A1 - Reißel, Martin ED - Rosenhahn, Bodo T1 - FSI Schemes : Fast Semi-Iterative Solvers for PDEs and Optimisation Methods T2 - Pattern Recognition : 38th German Conference, GCPR 2016, Hannover, Germany, September 12-15, 2016, Proceedings Y1 - 2016 SN - 978-3-319-45886-1 U6 - http://dx.doi.org/10.1007/978-3-319-45886-1_8 N1 - Lecture Notes in Computer Science; Vol. 9796 SP - 91 EP - 102 PB - Springer ER - TY - JOUR A1 - Hamad, E. M. A1 - Bilatto, S. E. R. A1 - Adly, N. Y. A1 - Correa, D. S. A1 - Wolfrum, B. A1 - Schöning, Michael Josef A1 - Offenhäusser, A. A1 - Yakushenko, A. T1 - Inkjet printing of UV-curable adhesive and dielectric inks for microfluidic devices JF - Lab on a Chip N2 - Bonding of polymer-based microfluidics to polymer substrates still poses a challenge for Lab-On-a-Chip applications. Especially, when sensing elements are incorporated, patterned deposition of adhesives with curing at ambient conditions is required. Here, we demonstrate a fabrication method for fully printed microfluidic systems with sensing elements using inkjet and stereolithographic 3D-printing. Y1 - 2016 U6 - http://dx.doi.org/10.1039/C5LC01195G SN - 1473-0189 VL - 16 IS - 1 SP - 70 EP - 74 PB - Royal Society of Chemistry CY - Cambridge ER - TY - JOUR A1 - Jahnke, Siegfried A1 - Roussel, Johanna A1 - Hombach, Thomas A1 - Kochs, Johannes A1 - Fischbach, Andreas A1 - Huber, Gregor A1 - Scharr, Hanno T1 - phenoSeeder - A robot system for automated handling and phenotyping of individual seeds JF - Plant physiology N2 - The enormous diversity of seed traits is an intriguing feature and critical for the overwhelming success of higher plants. In particular, seed mass is generally regarded to be key for seedling development but is mostly approximated by using scanning methods delivering only two-dimensional data, often termed seed size. However, three-dimensional traits, such as the volume or mass of single seeds, are very rarely determined in routine measurements. Here, we introduce a device named phenoSeeder, which enables the handling and phenotyping of individual seeds of very different sizes. The system consists of a pick-and-place robot and a modular setup of sensors that can be versatilely extended. Basic biometric traits detected for individual seeds are two-dimensional data from projections, three-dimensional data from volumetric measures, and mass, from which seed density is also calculated. Each seed is tracked by an identifier and, after phenotyping, can be planted, sorted, or individually stored for further evaluation or processing (e.g. in routine seed-to-plant tracking pipelines). By investigating seeds of Arabidopsis (Arabidopsis thaliana), rapeseed (Brassica napus), and barley (Hordeum vulgare), we observed that, even for apparently round-shaped seeds of rapeseed, correlations between the projected area and the mass of seeds were much weaker than between volume and mass. This indicates that simple projections may not deliver good proxies for seed mass. Although throughput is limited, we expect that automated seed phenotyping on a single-seed basis can contribute valuable information for applications in a wide range of wild or crop species, including seed classification, seed sorting, and assessment of seed quality. Y1 - 2016 U6 - http://dx.doi.org/10.1104/pp.16.01122 SN - 0032-0889 VL - 172 IS - 3 SP - 1358 EP - 1370 PB - Oxford University Press CY - Oxford ER - TY - CHAP A1 - Jung, Alexander A1 - Staat, Manfred ED - Erni, Daniel T1 - Computing olympic gold: Ski jumping as an example T2 - 1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen Y1 - 2016 SN - 978-3-940402-06-6 U6 - http://dx.doi.org/10.17185/duepublico/40821 SP - 54 EP - 55 PB - Universität Duisburg-Essen CY - Duisburg ER - TY - CHAP A1 - Jung, Alexander A1 - Staat, Manfred A1 - Müller, Wolfram T1 - Effect of wind on flight style optimisation in ski jumping T2 - 15th International Symposium on Computer Simulation in Biomechanics ; July 9th-11th 2015, Edinburgh, UK Y1 - 2016 SP - 53 EP - 54 PB - The University of Edinburgh ; Loughborough University CY - Edinburgh ER - TY - CHAP A1 - Kahmann, Stephanie A1 - Hackl, Michael A1 - Wegmann, Kilian A1 - Müller, Lars-Peter A1 - Staat, Manfred ED - Erni, Daniel T1 - Impact of a proximal radial shortening osteotomy on the distribution of forces and the stability of the elbow T2 - 1st YRA MedTech Symposium 2016 : April 8th / 2016 / University of Duisburg-Essen N2 - The human arm consists of the humerus (upper arm), the medial ulna and the lateral radius (forearm). The joint between the humerus and the ulna is called humeroulnar joint and the joint between the humerus and the radius is called humeroradial joint. Lateral and medial collateral ligaments stabilize the elbow. Statistically, 2.5 out of 10,000 people suffer from radial head fractures [1]. In these fractures the cartilage is often affected. Caused by the injured cartilage, degenerative diseases like posttraumatic arthrosis may occur. The resulting pain and reduced range of motion have an impact on the patient’s quality of life. Until now, there has not been a treatment which allows typical loads in daily life activities and offers good long-term results. A new surgical approach was developed with the motivation to reduce the progress of the posttraumatic arthrosis. Here, the radius is shortened by 3 mm in the proximal part [2]. By this means, the load of the radius is intended to be reduced due to a load shift to the ulna. Since the radius is the most important stabilizer of the elbow it has to be confirmed that the stability is not affected. In the first test (Fig. 1 left), pressure distributions within the humeroulnar and humeroradial joints a native and a shortened radius were measured using resistive pressure sensors (I5076 and I5027, Tekscan, USA). The humerus was loaded axially in a tension testing machine (Z010, Zwick Roell, Germany) in 50 N steps up to 400 N. From the humerus the load is transmitted through both the radius and the ulna into the hand which is fixed on the ground. In the second test (Fig. 1 right), the joint stability was investigated using a digital image correlation system to measure the displacement of the ulna. Here, the humerus is fixed with a desired flexion angle and the unconstrained forearm lies on the ground. A rope connects the load actuator with a hook fixed in the ulna. A guide roller is used so that the rope pulls the ulna horizontally when a tensile load is applied. This creates a moment about the elbow joint with a maximum value of 7.5 Nm. Measurements were performed with varying flexion angles (0°, 30°, 60°, 90°, 120°). For both tests and each measurement, seven specimens were used. Student ́s t-test was employed to determine whether the mean values of the measurements in native specimen and operated specimens differ significantly. Y1 - 2016 U6 - http://dx.doi.org/10.17185/duepublico/40821 SP - 7 EP - 8 PB - Universität Duisburg-Essen CY - Duisburg ER - TY - JOUR A1 - Knobe, M. A1 - Giesen, M. A1 - Plate, S. A1 - Gradl-Dietsch, G. A1 - Buecking, B. A1 - Eschbach, D. A1 - Laack, Walter van A1 - Pape, H.-C. T1 - The Aachen mobility and balance index to measure physiological falls risk: a comparison with the Tinetti POMA scale JF - European Journal Of Trauma And Emergency Surgery N2 - Purpose The most commonly used mobility assessments for screening risk of falls among older adults are rating scales such as the Tinetti performance oriented mobility assessment (POMA). However, its correlation with falls is not always predictable and disadvantages of the scale include difficulty to assess many of the items on a 3-point scale and poor specificity. The purpose of this study was to describe the ability of the new Aachen Mobility and Balance Index (AMBI) to discriminate between subjects with a fall history and subjects without such events in comparison to the Tinetti POMA Scale. Methods For this prospective cohort study, 24 participants in the study group and 10 in the control group were selected from a population of patients in our hospital who had met the stringent inclusion criteria. Both groups completed the Tinetti POMA Scale (gait and balance component) and the AMBI (tandem stance, tandem walk, ten-meter-walk-test, sit-to-stand with five repetitions, 360° turns, timed-up-and-go-test and measurement of the dominant hand grip strength). A history of falls and hospitalization in the past year were evaluated retrospectively. The relationships among the mobility tests were examined with Bland–Altmananalysis. Receiver-operated characteristics curves, sensitivity and specificity were calculated. Results The study showed a strong negative correlation between the AMBI (17 points max., highest fall risk) and Tinetti POMA Scale (28 points max., lowest fall risk; r = −0.78, p < 0.001) with an excellent discrimination between community-dwelling older people and a younger control group. However, there were no differences in any of the mobility and balance measurements between participants with and without a fall history with equal characteristics in test comparison (AMBI vs. Tinetti POMA Scale: AUC 0.570 vs. 0.598; p = 0.762). The Tinetti POMA Scale (cut-off <20 points) showed a sensitivity of 0.45 and a specificity of 0.69, the AMBI a sensitivity of 0.64 and a specificity of 0.46 (cut-off >5 points). Conclusion The AMBI comprises mobility and balance tasks with increasing difficulty as well as a measurement of the dominant hand-grip strength. Its ability to identify fallers was comparable to the Tinetti POMA Scale. However, both measurement sets showed shortcomings in discrimination between fallers and non-fallers based on a self-reported retrospective falls-status. KW - Tinetti test KW - Mobility KW - Mobility tests KW - Ground-level falls KW - Elderly KW - Fall prevention KW - Co-managed care KW - Balance Y1 - 2016 U6 - http://dx.doi.org/10.1007/s00068-016-0693-2 SN - 1863-9941 VL - 42 IS - 5 SP - 537 EP - 545 PB - Springer CY - Berlin ER - TY - JOUR A1 - Kolditz, Melanie A1 - Albin, Thivaharan A1 - Abel, Dirk A1 - Fasse, Alessandro A1 - Brüggemann, Gert-Peter A1 - Albracht, Kirsten T1 - Evaluation of foot position and orientation as manipulated variables to control external knee adduction moments in leg extension training JF - Computer methods and programs in biomedicine N2 - Background and Objective Effective leg extension training at a leg press requires high forces, which need to be controlled to avoid training-induced damage. In order to avoid high external knee adduction moments, which are one reason for unphysiological loadings on knee joint structures, both training movements and the whole reaction force vector need to be observed. In this study, the applicability of lateral and medial changes in foot orientation and position as possible manipulated variables to control external knee adduction moments is investigated. As secondary parameters both the medio-lateral position of the center of pressure and the frontal-plane orientation of the reaction force vector are analyzed. Methods Knee adduction moments are estimated using a dynamic model of the musculoskeletal system together with the measured reaction force vector and the motion of the subject by solving the inverse kinematic and dynamic problem. Six different foot conditions with varying positions and orientations of the foot in a static leg press are evaluated and compared to a neutral foot position. Results Both lateral and medial wedges under the foot and medial and lateral shifts of the foot can influence external knee adduction moments in the presented study with six healthy subjects. Different effects are observed with the varying conditions: the pose of the leg is changed and the direction and center of pressure of the reaction force vector is influenced. Each effect results in a different direction or center of pressure of the reaction force vector. Conclusions The results allow the conclusion that foot position and orientation can be used as manipulated variables in a control loop to actively control knee adduction moments in leg extension training. KW - External knee adduction moments KW - Manipulated variables KW - Inverse dynamic problem KW - Inverse kinematic problem KW - Musculoskeletal model Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.cmpb.2016.09.005 SN - 0169-2607 N1 - Part of special issue: "SI: Personalised Models and System Identification" VL - 171 SP - 81 EP - 86 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Kolditz, Melanie A1 - Albin, Thivaharan A1 - Albracht, Kirsten A1 - Brüggemann, Gert-Peter A1 - Abel, Dirk T1 - Isokinematic leg extension training with an industrial robot T2 - 6th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) June 26-29, 2016. UTown, Singapore Y1 - 2016 U6 - http://dx.doi.org/10.1109/BIOROB.2016.7523750 SP - 950 EP - 955 ER - TY - JOUR A1 - Kolditz, Melanie A1 - Albin, Thivaharan A1 - Brüggemann, Gert-Peter A1 - Abel, Dirk A1 - Albracht, Kirsten T1 - Robotergestütztes System für ein verbessertes neuromuskuläres Aufbautraining der Beinstrecker JF - at - Automatisierungstechnik N2 - Neuromuskuläres Aufbautraining der Beinstrecker ist ein wichtiger Bestandteil in der Rehabilitation und Prävention von Muskel-Skelett-Erkrankungen. Effektives Training erfordert hohe Muskelkräfte, die gleichzeitig hohe Belastungen von bereits geschädigten Strukturen bedeuten. Um trainingsinduzierte Schädigungen zu vermeiden, müssen diese Kräfte kontrolliert werden. Mit heutigen Trainingsgeräten können diese Ziele allerdings nicht erreicht werden. Für ein sicheres und effektives Training sollen durch den Einsatz der Robotik, Sensorik, eines Regelkreises sowie Muskel-Skelett-Modellen Belastungen am Zielgewebe direkt berechnet und kontrolliert werden. Auf Basis zweier Vorstudien zu möglichen Stellgrößen wird der Aufbau eines robotischen Systems vorgestellt, das sowohl für Forschungszwecke als auch zur Entwicklung neuartiger Trainingsgeräte verwendet werden kann. Y1 - 2016 U6 - http://dx.doi.org/10.1515/auto-2016-0044 SN - 2196-677X VL - 64 IS - 11 SP - 905 EP - 914 PB - De Gruyter CY - Berlin ER - TY - JOUR A1 - Kotliar, Konstantin A1 - Lanzl, I. M. T1 - Mit Statistik gemeistert: perfekte Augentropfen und idealer Screeningtest : Möglichkeiten und Grenzen statistischer Methoden beim Glaukom JF - Der Ophthalmologe: Zeitschrift Der Deutschen Ophthalmologischen Gesellschaft N2 - Hintergrund Die Anwendung und das Verständnis von Statistik sind sehr wichtig für die biomedizinische Forschung und für die klinische Praxis. Dies gilt insbesondere auch zur Abschätzung der Möglichkeiten unterschiedlichster Diagnostik- und Therapieoptionen beim Glaukom. Die scheinbare Komplexität der Statistik, die zum Teil dem „gesunden Menschenverstand“ zu widersprechen scheint, zusammen mit der nur vorsichtigen Akzeptanz der Statistik bei vielen Medizinern können zu bewussten und unbewussten Manipulationen bei der Datendarstellung und -interpretation führen. Ziel der Arbeit Ziel ist die verständliche Darstellung einiger typischer Fehler in der medizinisch-statistischen Datenbehandlung. Material und Methoden Anhand hypothetischer Beispiele aus der Glaukomdiagnostik erfolgen die Darstellung der Wirkung eines hypotensiven Medikamentes sowie die Beurteilung der Ergebnisse eines diagnostischen Tests. Es werden die typischsten statistischen Einsatzbereiche und Irrtumsquellen ausführlich und verständlich analysiert Ergebnisse Mechanismen von Datenmanipulation und falscher Dateninterpretation werden aufgeklärt. Typische Irrtumsquellen bei der statistischen Auswertung und Datendarstellung werden dabei erläutert. Schlussfolgerungen Die erläuterten praktischen Beispiele zeigen die Notwendigkeit, die Grundlagen der Statistik zu verstehen und korrekt anwenden zu können. Fehlendes Grundlagenwissen und Halbwissen der medizinischen Statistik können zu folgenschweren Missverständnissen und falschen Entscheidungen in der medizinischen Forschung, aber auch in der klinischen Praxis führen. Y1 - 2016 U6 - http://dx.doi.org/10.1007/s00347-016-0312-y SN - 0941-293X SN - 1433-0423 N1 - Englischer Titel: Mastered with statistics: perfect eye drops and ideal screening test : possibilities and limits of statistical methods for glaucoma IS - 113 SP - 838 EP - 843 PB - Springer CY - Berlin ER - TY - JOUR A1 - Kowalski, Julia A1 - Linder, Peter A1 - Zierke, S. A1 - Wulfen, B. van A1 - Clemens, J. A1 - Konstantinidis, K. A1 - Ameres, G. A1 - Hoffmann, R. A1 - Mikucki, J. A1 - Tulaczyk, S. A1 - Funke, O. A1 - Blandfort, D. A1 - Espe, Clemens A1 - Feldmann, Marco A1 - Francke, Gero A1 - Hiecker, S. A1 - Plescher, Engelbert A1 - Schöngarth, Sarah A1 - Dachwald, Bernd A1 - Digel, Ilya A1 - Artmann, Gerhard A1 - Eliseev, D. A1 - Heinen, D. A1 - Scholz, F. A1 - Wiebusch, C. A1 - Macht, S. A1 - Bestmann, U. A1 - Reineking, T. A1 - Zetzsche, C. A1 - Schill, K. A1 - Förstner, R. A1 - Niedermeier, H. A1 - Szumski, A. A1 - Eissfeller, B. A1 - Naumann, U. A1 - Helbing, K. T1 - Navigation technology for exploration of glacier ice with maneuverable melting probes JF - Cold Regions Science and Technology N2 - The Saturnian moon Enceladus with its extensive water bodies underneath a thick ice sheet cover is a potential candidate for extraterrestrial life. Direct exploration of such extraterrestrial aquatic ecosystems requires advanced access and sampling technologies with a high level of autonomy. A new technological approach has been developed as part of the collaborative research project Enceladus Explorer (EnEx). The concept is based upon a minimally invasive melting probe called the IceMole. The force-regulated, heater-controlled IceMole is able to travel along a curved trajectory as well as upwards. Hence, it allows maneuvers which may be necessary for obstacle avoidance or target selection. Maneuverability, however, necessitates a sophisticated on-board navigation system capable of autonomous operations. The development of such a navigational system has been the focal part of the EnEx project. The original IceMole has been further developed to include relative positioning based on in-ice attitude determination, acoustic positioning, ultrasonic obstacle and target detection integrated through a high-level sensor fusion. This paper describes the EnEx technology and discusses implications for an actual extraterrestrial mission concept. Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.coldregions.2015.11.006 SN - 0165-232X IS - 123 SP - 53 EP - 70 PB - Elsevier CY - Amsterdam ER - TY - CHAP A1 - Laack, Walter van T1 - Schnittstelle Tod: Wo stehen wir nach 40 Jahren NTE-Forschung? Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:101:1-201603132912 SN - 978-3-936624-30-4 (Print-Ausgabe) SN - 978-3-936624-32-8 (Online-Ausgabe) N1 - Tagungsbeiträge des 4. Europäischen Seminars am 07. November 2015 in Aachen zum Thema Nahtoderfahrungen mit dem Serientitel: "Schnittstelle Tod" PB - van Laack GmbH CY - Aachen ER - TY - JOUR A1 - Levers, A. A1 - Staat, Manfred A1 - Laack, Walter van T1 - Analysis of the long-term effect of the MBST® nuclear magnetic resonance therapy on gonarthrosis JF - Orthopedic Practice Y1 - 2016 VL - 47 IS - 11 SP - 521 EP - 528 ER - TY - CHAP A1 - Marinova, V. A1 - Kerroumi, I. A1 - Lintermann, A. A1 - Göbbert, J.H. A1 - Moulinec, C. A1 - Rible, S. A1 - Fournier, Y. A1 - Behbahani, Mehdi T1 - Numerical Analysis of the FDA Centrifugal Blood Pump T2 - NIC Symposium 2016 Y1 - 2016 SN - 978-3-95806-109-5 SP - 355 EP - 364 ER - TY - JOUR A1 - Miyamoto, Ko-Ichiro A1 - Sato, Takuya A1 - Abe, Minami A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Light-addressable potentiometric sensor as a sensing element in plug-based microfluidic devices JF - Micromachines N2 - A plug-based microfluidic system based on the principle of the light-addressable potentiometric sensor (LAPS) is proposed. The LAPS is a semiconductor-based chemical sensor, which has a free addressability of the measurement point on the sensing surface. By combining a microfluidic device and LAPS, ion sensing can be performed anywhere inside the microfluidic channel. In this study, the sample solution to be measured was introduced into the channel in a form of a plug with a volume in the range of microliters. Taking advantage of the light-addressability, the position of the plug could be monitored and pneumatically controlled. With the developed system, the pH value of a plug with a volume down to 400 nL could be measured. As an example of plug-based operation, two plugs were merged in the channel, and the pH change was detected by differential measurement. KW - light-addressable potentiometric sensor KW - plug-based microfluidic device KW - chemical sensor Y1 - 2016 U6 - http://dx.doi.org/10.3390/mi7070111 SN - 2072-666X N1 - This article belongs to the Special Issue "Micro/Nano Devices for Chemical Analysis" VL - 7 IS - 7 SP - 111 PB - MDPI CY - Basel ER - TY - JOUR A1 - Miyamoto, Ko-ichiro A1 - Yu, Bing A1 - Isoda, Hiroko A1 - Wagner, Torsten A1 - Schöning, Michael Josef A1 - Yoshinobu, Tatsuo T1 - Visualization of the recovery process of defects in a cultured cell layer by chemical imaging sensor JF - Sensors and Actuators B: Chemical N2 - The chemical imaging sensor is a field-effect sensor which is able to visualize both the distribution of ions (in LAPS mode) and the distribution of impedance (in SPIM mode) in the sample. In this study, a novel cell assay is proposed, in which the chemical imaging sensor operated in SPIM mode is applied to monitor the recovery of defects in a cell layer brought into proximity of the sensing surface. A reduced impedance at a defect formed artificially in a cell layer was successfully visualized in a photocurrent image. The cell layer was cultured over two weeks, during which the temporal change of the photocurrent distribution corresponding to the recovery of the defect was observed. Y1 - 2016 U6 - http://dx.doi.org/10.1016/j.snb.2016.04.018 SN - 0925-4005 VL - 236 SP - 965 EP - 969 PB - Elsevier CY - Amsterdam ER -